    DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL
    Section 3-4 - Preparing for Installation 3-21
    3-6-5 Video Specification
    Video specifications may be needed to be able to connect the VCR or DVD Video Recorder to the 
    LOGIQ™ P5/A5/A5Pro.
    Table 3-9    Video Specifications
    Timing ParameterNTSCPAL
    640x480
    60Hz800x60060Hz800x60075Hz1024x768 60Hz
    Horizontal Rate [kHz] 60.02 31.25 31.47 37.88 46.88 48.36
    Horizontal Period [µ s]16.66 32.00 31.78 26.40 21.33 20.68
    Pixel Clock [MHz] 78.75 29.50 24.55 40.00 49.50 65.00
    H Blank Width  [µs] 3.66 5.97 5.706.405.17 4.92
    H Sync Width  [µs] 1.22 2.34 2.363.201.62 2.09
    H Front Porch  [µs] 0.20 0.75 0.731.000.32 0.37
    Active Horizontal Period  [µs] 13.00 26.03 26.07 20.00 16.16 15.75
    Vertical Rate [Hz] 75.03 50.00 59.94 60.32 75.00 60.00
    Vertical Period [ms] 13.33 20.00 16.68 16.58 13.33 16.67
    V Sync Width [lines=ms] 32=0.53 49=1.57 45=1.43 28=0.74 25=0.53 38=0.79
    V Front Porch  [lines=µ s] 3=50.00 5=160.00 6=190.70 4=105.60 3=64.00 6=124.10
    Equalization Gate  [lines= µs] 1=16.66 5=160.0 6=190.7 1=26.4 1=21.3 3=62.00
    Lines: Field/Frame 800 625/625 525/525 628 625806
    Active Lines/Frame 768576 480600600768 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
    3-22 Section 3-4 - Preparing for Installation
    3-6-6 Setting VCR (Mitsubishi MD3000) USB Interface
    1.) Turn off the Power of the VCR.
    2.) Open the rear panel of the VCR.
    3.) Set dip switches. Off 1~6 pins, On 7 pin, and Off 8 pin.
    NOTE: 7 Pin ON = USB Interface, OFF = RS232C Interface.
    Figure 3-15   Setting VCR USB Interface 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL
    Section 3-8 - Software/Option Configuration 3-23
    Section 3-7
    Available Probes
    See Specifications in the LOGIQ™ P5/A5/A5Pro User Manual for probes and intended use.
    See Chapter 9 , for part numbers to be used when ordering new or replacement probes.
    Section 3-8
    Software/Option Configuration
    refer to the LOGIQ™ P5/A5/A5Pro Basic User Manual, Chapter 16, Customizing Your System for 
    information on conf iguring items like Hospital,  Department, Language, Units (of measure), Date, Time 
    and Date Format.
    For information on configuring Software Options,  refer to the LOGIQ™ P5/A5/A5Pro Basic User 
    Manual, Chapter 16, Customizing Your System.
    For information on configuring DI COM Connectivity, refer to the LOGIQ™ P5/A5/A5Pro Basic User 
    Manual, Chapter 16, Customizing Your System. 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
    3-24 Section 3-8 - Software/Option Configuration
    Section 3-9
    Connectivity Installation Worksheet
    Site System Information
    Services (Destination Devices) TCP/IP Settings
    Device Type
    Manufacturer
    Name
    IP Address Port
    AE Title
    1
    2
    3
    4
    5
    6
    7
    8
    9
    10
    11
    12
    Name - AE Title:
    Dept:
    LOGIQ SN:Type:
    Floor:
    Room:
    REV:
    Comments:
    CONTACT INFORMATION
    Name Title
    PhoneE-Mail Address
    Site:
    IP Settings
    IP Address:
    Subnet Mask:
    Default Gateway:
    Remote Archive Setup
    Remote Archive IP:
    Remote Archive Name: 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL
    Section 3-11 - Loading Base System Software 3-25
    Section 3-10
    Insite IP Address Configuration
    This information needed to facilitate the OnLine Center engineers in connecting to the system is found 
    in the InSite Installation Manual.
    Section 3-11
    Loading Base System Software
    3-11-1 Loading Base system Software
    For more information on  loading base system software, refer to Chapter 8.
    Section 3-12
    Warnings
    3-12-1 Warnings
    3-12-1-1 USB HDD
    External USB type Hard drive may be used to back up the data. But not all kinds of USB type can 
    be supported.
    Never use USB types which requ ire additional power source.
    3-12-1-2 Starting Echoloader in Maintenance mode If Echoloader never started once, starting Echoloader  in the maintenance mode will not work. In 
    other words, in order to be able to start Echoloa der in maintenace mode, Echoloader should have 
    been started at least once. 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
    3-26 Section 3-12 - Warnings
    Section 3-13
    Paperwork
    NOTE: During and after installation, the documentation  (i.e. User Manuals, Installation Manuals...) for the 
    peripheral units must be ke pt as part of the original system do cumentation. This will ensure that all 
    relevant safety and user information is available during the operation and service of the complete 
    system.
    3-13-1 Product Locator Installation
    NOTE: The Product Locator Installation Card shown  may not be same as the provided Product Locator 
    card.
    3-13-2 User Manual(s)
    Check that the correct User Manual(s) for the syst em and software revision, is included with the 
    installation. Specific language versions of the User  Manual may also be available. Check with your GE 
    Sales Representative for availability.
    3-13-2-1 Reference off-board peripherals and options None. Figure 3-16   Product Locator Installation Card 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL
    Section 4-1 - Overview 4 - 1
    Chapter 4
    Functional Checks
    Section 4-1
    Overview
    4-1-1 Purpose of Chapter 4
    This chapter provides procedures for quickly checking major functions of the LOGIQ™ P5/A5/A5Pro 
    scanner diagnostics by using the built-in service software, and power supply adjustments.
    Section 4-2
    Required Equipment
    • An empty (blank) DVD-R or CD-R Disk.
    • At least one transducer. Table 4-1    Contents in Chapter 4
    SectionDescriptionPage Number
    4-1 Overview 4-1
    4-2 Required Equipment 4-1
    4-3 General Procedure 4-2
    4-4 Software Configuration Checks 4-25
    4-5 Peripheral Checks 4-26
    4-6 Monitor Function Checks 4-27
    4-7 Keyboard Function Checks 4-29
    4-8 Mechanical Function Checks 4-34
    4-9 Board Function Checks 4-41
    4-10 Site Log 4-54 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
    4 - 2 Section 4-2 - Required Equipment
    Section 4-3
    General Procedure
    4-3-1 Power On/Boot Up
    NOTE: After turning off the system, wait at least ten seconds before turning it on again. The system 
    may not be able to boot if power is recycled too quickly.
    4-3-1-1 Scanner Power On 1.) Connect the Main Power cable to an appropriate mains power outlet.
    2.) Switch ON the Main Circuit Breaker at the rear of the System.
    When power is applied to the Scanner, and the Rear  Circuit breaker is turned ON, Power is distributed 
    to the DC Power supply unit and power control  assy in the AC power assy. When the Power ON/OFF 
    key is pressed once, the DC power supply is enab led and feed each DC power to board assemblies. 
    and system run the software 
    CAUTIONSYSTEM REQUIRES ALL COVERS
    Operate this unit only when all board covers and frame panels are securely in place. The 
    covers are required for safe operation, good system performance and cooling purposes.
    NOTICELockout/Tagout Requirements (For USA only)
    Follow OSHA Lockout/Tagout requirements by ensuring you are in total control of the Power Cable 
    on the system.
    Figure 4-1   Circuit Breaker
    SignedDate
    TA G 
    & LOCKOUT 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL
    Section 4-2 - Required Equipment 4 - 3
    4-3-1-1              Scanner Power On (cont’d)
    3.) Press the  ON/OFF 
    key at the front of the System once.
    Figure 4-2   Power On/Off  Standby Switch Location 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
    4 - 4 Section 4-2 - Required Equipment
    4-3-2 Power Off / Shutdown
    NOTE: After turning off a system, wait at least ten seco nds before turning it on again. The system may not be 
    able to boot if power is recycled too quickly.
    4-3-2-1 Scanner Shutdown 1.) Press the On/Off  Key at the front of the System once to display the SYSTEM - EXIT menu.
    2.) Select Shutdown from the SYSTEM - EXIT menu.
    4-3-2-2 Switch off the System 1.) Switch OFF the Circuit Breake r at the back of the scanner.
    2.) Unplug the power cord if necessary. For ex ample : Servicing or relocating the scanner.
    Figure 4-3   System Exit Menu for Power Down 
    						