    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
    Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 15
    8-2-6-3              Removal procedure  (cont’d)
    6.) Unscrew 2 screws (5177684, HSH M5x20 WH T) to separate the LCD MONITOR and 1 screw 
    (2159625, PH M4x8 W/SP) to remove the Cable bracket.
    NOTE: While unscrewing the hinge hexa screw from the arm, keep hold of the monitor to prevent  dropping down.
    7.) Unscrew 2 screws (2337572, FH M3x6 WHT) to remove the Arm cover L & R.
    8.) Unscrew 2 screws (2159632, BH M4x6 WHT) to remove the Cam cover L & R.Figure 8-21   Separating the LCD monitor & Removing the Cable bracket.
    Figure 8-22   Removing the Arm cover L & Arm cover R
    Figure 8-23   Removing the Cam cover L & Cam cover R 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    8 - 16 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
    8-2-6-3              Removal procedure  (cont’d)
    9.) Unscrew 4 screws (2159634, BH M4x10 WHT) to remove the Arm bottom cover.
    10.) Unscrew 1 screw (2329677, TAP M4X16) to remove the Lock cover.
    11.)  Unscrew 2 screws (2159625, PH M4x8 W/SP) to remove the Lower arm guide bracket. Figure 8-24   Removing the Arm bottom cover 
    Figure 8-25   Removing the Lock cover
    Figure 8-26   Removing the Lower arm guide bracket 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
    Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 17
    8-2-6-3              Removal procedure  (cont’d)
    12.) Unscrew 2 screws (2159626, PH M4X20) to remove the Arm stopper.
    13.)  Pull out power cable & DVI cable. Ma ke sure flexible arm not to rotate.
    14.)  Unscrew 4 screws (2159625, PH M4x8 W/SP) to remove the Lower arm cable bracket and pull out  power cable & DVI cable. Figure 8-27   Removing the Arm stopper
    Figure 8-28   Pulling out power cable & DVI cable
    Figure 8-29   Unscrewing 4 screws 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    8 - 18 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
    8-2-6-3              Removal procedure  (cont’d)
    15.) Unscrew 3 screws (2159634, BH M4x10 WHT) after  removing the rubber caps to remove neck front 
    cover, neck rear cover.
    16.) Unscrew 3 screws(5327646, HSH M6X16 WHT) to separate the flexible arm neck pipe.
    Figure 8-30   Unscrewing 3 screws and removing the neck front cover & neck rear cover
    Figure 8-31   Separating the neck pipe 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
    Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 19
    8-2-6-3              Removal procedure  (cont’d)
    17.) Unscrew 1 screw (2159625, PH M4x8 W/SP) to remove the OP bracket.
    18.)  Unscrew 3 screws (2373562, FH M4x10 YEL)  to separate the Flexible arm stopper (A+B).
    Figure 8-32   Removing OP bracket
    Figure 8-33   Separating the flexible arm stopper 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    8 - 20 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
    8-2-6-3              Removal procedure  (cont’d)
    19.) Pull out power cable & DVI cable and remove th e oilless bush after separating the neck pipe.
    Figure 8-34   Pulling out power cable & DVI cable and removing the stopper 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
    Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 21
    8-2-6-3              Removal procedure  (cont’d)
    20.) Perform the following functional tests. If all ar e successful, include the debrief script provided below.
    Table 8-7    Functional Tests
    Service Manual Section
    Functional Test / Diagnostic TestDebrief Script
    Section 4-3-1 Power On/Boot Up
    “Service Manual, Direction
    5394141, Rev 1+, Section 8-2-6. Equipment 
    passed all required tests and is ready for use. “
    Section 4-3-2
    Power Off / Shutdown
    Section 4-8-8 Flexible LCD Arm Function Validation 
    Procedure
    Section 10-5-5Physical Inspection 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    8 - 22 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
    8-2-6-4 Assembly procedure First, remove LP5/LA5 Main K/ B Assy and LP5/LA5 Sub K/B Assy
    1.) Change the MON PWR CONN LONG cable and remove  the cable tie of the DVI cable for flexible 
    arm.
    Figure 8-35   Changing the MON PWR CONN LONG cable
    MON PWR CONN LONG cable 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
    Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 23
    8-2-6-4              Assembly procedure  (cont’d)
    2.) Assemble the LCD locking mechanism.
    a.) Put the stopper pin inside the  spring and then put them inside the spring holder. Fix it with 
    spring pin.
    b.) Put lock tite 290 on the M16 tap of  the spring holder Assy and screw it.
    Figure 8-36   Assembling the spring holder Assy
    Figure 8-37   Screwing the spring holder Assy 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    8 - 24 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
    8-2-6-4              Assembly procedure  (cont’d)
    b-1) Mold die of AL neck frame modified as seen in the figure below does not need a nut.
    c.) Insert holder grip as seen the picture below and fix it with 2set screws(2327580, M2.6X4). Figure 8-38   Screwing the spring holder Assy
    Figure 8-39   Assembling the holder grip 
    						