    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
    Section 8-3 - Software Loading Procedure 8 - 165
    8-6-2-1              Perform So ftware Reloading in the Service Platform  (cont’d)
    9.) Select “Reload System Software”
    10.) Verify that “Local Archive” is available under Dataflow.
    11.) Press “ReLoad”.
    12.) Confirmation message pops up.
    13.) Press “Yes” to proceed.
    14.) Progress bar is displayed Figure 8-233   Reload system Software
    Figure 8-234   Reloading setup in progress 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    8 - 166 Section 8-3 - Software Loading Procedure
    8-6-2-1              Perform Software Reloading in the Service Platform  (cont’d)
    15.) Setup is completed successfully
    a.) If you do s/w reloading from the system  directly, please shut down the system now.
    b.) For remote s/w reloading, please enable Disruptive mode. 
    16.) Press “Disruptive mode” Figure 8-235   setup completed message
    Figure 8-236   Disruptive mode 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
    Section 8-3 - Software Loading Procedure 8 - 167
    8-6-2-1              Perform So ftware Reloading in the Service Platform  (cont’d)
    17.) If “Disruptive Mode Status” is disabl ed, Press “Yes” to enable disruptive mode.
    For the remote reloading, customer has to agree to the diruptive mode. The following window 
    (Figure 8-237) pops up in the system screen and cu stomer has to press “Yes” to enable Disruptive 
    mode.
    18.) Press “System Shutdown” 
    NOTE: If the service dongle is inserted or the system is logged on to the maintenance mode, “System  shutdown” will not work. Please close progra ms and reboot the system manually.
    Figure 8-237   Insite Notification
    Figure 8-238   System Shutdown 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    8 - 168 Section 8-3 - Software Loading Procedure
    8-6-2-1              Perform Software Reloading in the Service Platform  (cont’d)
    19.) Select “Restart System” 
    20.) Press “Submit”
    21.) Press “Confirm” when confirmation message is displayed.
    22.) After a while, the system  restarts automatically. 
    23.) Verify that correct date an d time are displayed correctly.
    Figure 8-239   Restart System 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
    Section 8-3 - Software Loading Procedure 8 - 169
    8-6-2-2 Software Reloading  1.) The Base S/W load starts auto matically immediately after boot up
    2.) After finishing C: partition reload ing, the system restarts automatically.
    3.) When the system restarts, an application installation st arts automatically.
    Figure 8-240   Reloading C: partition from Reloading Ghost Image
    Figure 8-241   Application installation 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    8 - 170 Section 8-3 - Software Loading Procedure
    8-6-2-2              Software Reloading  (cont’d)
    4.) When installation is finished, the system shuts down automatically.
    5.) Press Power Button to start the system.
    6.) When system fully boots up after the softwa re reloading, please shutdown system again.
    NOTE: For the following procedure, disruptive mode should be approved in order to be in done  remotely.
    7.) Start the system.
    8.) Press “Patient” button, and verify under Dataflow  that “Local Archive” is available. If its not 
    available, shut down the unit one more time.
    9.) To put the Local Time Zone ba ck, Go to Utility->System->General
    10.) Click “Date/Time” 
    Figure 8-242   Date/Time 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
    Section 8-3 - Software Loading Procedure 8 - 171
    8-6-2-2              Software Reloading  (cont’d)
    11.) Press Time Zone
    12.) Change the time zone to the local time zone from Table 1.
    13.) Press “Apply”
    14.) Press “Date & Time” tab. Figure 8-243   Time Zone
    Figure 8-244   Date & Time 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    8 - 172 Section 8-3 - Software Loading Procedure
    8-6-2-2              Software Reloading  (cont’d)
    15.) Change the Date and Time correctly if needed.
    16.) Press “Apply”
    17.) Press “OK”
    18.) Turn off the system.
    19.) Start the system.
    20.) Check if the correct date and time are displayed.
    21.) The software reloading is completed. Figure 8-245   Date and Time check 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
    Section 8-3 - Software Loading Procedure 8 - 173
    Section 8-7
    Mechanical Option Ins tallation instruction
    These service in subtractions descr ibe the installation procedures of following mechanical options to 
    LOGIQ™ P5. 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    8 - 174 Section 8-3 - Software Loading Procedure
    8-7-1 Endo Probe Holder Installation
    1.) Check out the 2 screw holes to be used to install Endo Probe Holder.
    2.) Install Endo Probe holder on the bottom  side of the Cable Arm Hook by screwing 2 
    screws(2159634, BH M4x10 WHT). Figure 8-246   Endo Probe Holder Installation
    Figure 8-247   Endo Probe Holder 
    						