    Section 7-7 - Trouble Shooting Tree 7 - 31
    7-7-7 Unable to Record to VCR
    Figure 7-18   VCR Troubleshooting
    Unable to record to VCR!
    Check that all signal, and power 
    supply cable connections to the VCR  unit are OK.YES NO
    Is the VCR properly 
    connected?
    Is the VCR tape put into  the device?
    Put the tape into  the device and 
    rewind it.
    NOYES
    Replace VCR.
    Is recording unable even 
    by manual operation?
    YES- Check the signal cable connections  between VCR and Internal I/O.
    - Check the cables between BEP and  Internal I/O
    NO 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    7 - 32 Section 7-7 - Trouble Shooting Tree
    7-7-8 Printer Troubleshooting
    Figure 7-19   Printer Troubleshooting
    Unable to use Printer(s)!
    Check that all signal, and power 
    supply cable connections between  Printer and Scanner are OK.YES NO
    Is the printer(s) properly 
    connected?
    Are there print papers left?
    Insert the printer papers.
    NOYES
    Is proper configuration for 
    the Print key(s) set?
    Configure the Print key(s).
    NO
    See next page. 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
    Section 7-7 - Trouble Shooting Tree 7 - 33
    7-7-8              Printer Troubleshooting (cont’d)
    Figure 7-20   Printer Troubleshooting (contd.)
    Still unable to use the Printer(s)!
    Check the USB cables Between  backplane and usb port.YES NO
    Is printing unavailable 
    even by manual operation?
    Is there an error message  on the monitor display?
    Replace SOM module.
    NO
    YES
    Replace the Printer. 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    7 - 34 Section 7-7 - Trouble Shooting Tree
    7-7-9 DVD-RW Troubleshooting
    Figure 7-21   CD-RW Troubleshooting
    Unable to record to CD-R or DVD -R!
    Insert a new, un-used CD-R disc.YES NO
    Is a new, unused CD-R 
    disc inserted to the drive?
    Replace SOM Module.
    Are CD-Rs rated at the proper write  speed?
    YES NOAcquire CD-Rs rated at the proper write 
    speed. 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
    Section 7-7 - Trouble Shooting Tree 7 - 35
    7-7-10 Network Troubleshooting
    7-7-10-1 No Connection to the Network at All
    1.) Check that the network cable between the sca nner and the wall network is connected and well 
    seated in both ends.
    2.) Try a network cable that is known to be OK.
    3.) Check the cable between the SYSC ONPM(SYSCOMCM) and Rear Panel.
    4.) Connect a Crossed network cable between the Scanner and your PC. Try to ping from the scanner  to the IP address on the PC. If OK, the hardware connection inside the scanner is OK.
    7-7-10-2 No Contact With Server, But Connection to Network is OK Check Network setup on the scanner. 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    7 - 36 Section 7-7 - Trouble Shooting Tree
    7-7-10               Network Troubleshooting (cont’d)
    Figure 7-22   Network Troubleshooting
    No Connection to the Network
    Connect the cable and verify that it is  well seated in both ends.YES NO
    Is the 
    TCP/IP Cable between 
    LOGIQ™ P5/A5/A5Pro  and network outlet 
    connected and well 
    Replace the AEXP cable between  SYSCON and Rear panel.
    YES
    Is the connection OK now?
    NO
    Connect a Crossed Network Cable 
    between the Scanner and your PC/ Laptop. 
    Try to ping from the PC/Laptop to  LOGIQ™ P5/A5/A5Pro.
     If OK, the hardware connection 
    inside LOGIQ™ P5/A5/A5Pro is OK.
    Try a network cable that is known to  be OK
    NO
    Is the 
    cable between SYSCON  and Rear Panel OK?
    YESReplace the Rear Panel. 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
    Section 8-1 - Overview 8 - 1
    Chapter 8
    Replacement Procedures 
    Section 8-1
    Overview
    8-1-1 Purpose of Chapter 8
    This chapter describes replacement procedures for the following modules and subsystems.
    SectionDescriptionPage Number
    8-1 Overview 8-1
    8-2 DISASSEMBLY/RE-ASSEMBLY 8-2
    8-3 Software Loading Procedure 8-117
    8-4 Software Installation 8-129
    8-5 Full Backup/Restore Procedure 8-159
    8-6 Software Reloading 8-163
    8-7 Mechanical Option Installation instruction 8-173
    Table 8-1    Contents in Chapter 8 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    8 - 2 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
    Section 8-2 DISASS EMBLY/RE-ASSEMBLY 
    WARNINGWARNINGONLY QUALIFIED SERVICE PERSONNEL  SHOULD REMOVE ANY COVERS OR 
    PANELS. ELECTRICAL HAZARDS EXISTS AT SEVERAL POINTS INSIDE. BECOME 
    THOROUGHLY FAMILIAR WITH ALL HAZARDOUS VOLTAGES AND HIGH CURRENT 
    LEVELS TO AVOID ACCIDENTAL CONTACT
    CAUTIONDo not wear the ESD wrist strap when you remo ve a part of power supply unit. Turn OFF power 
    and unplug the power cord before removing a part of power supply unit. However be sure to turn 
    off power and wear the strap before you remove a circuit boards.
    WARNINGWARNINGDO NOT SERVICE OR DISASSEMBLE PART S UNDER FRU UNIT LEVEL AT ANY 
    CIRCUMSTANCES. 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
    Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 3
    8-2-1 LCD Plastic filter
    8-2-1-1 Tools • Common Phillips screwdrivers
    • Allen/Unbraco wrench
    • Stubby screwdriver (Flat tip and Cross tip)
    8-2-1-2 Removal procedure 1.) Remove the LP5 Monitor cover set. Refer to the  8-2-2 Monitor cover set on page 8-5.
    2.) Remove the LCD Plastic filter.
    3.) Put the LCD plastic filter on the LCD monitor  after detaching the tape of the LCD plastic filter.
    Figure 8-1   Removing the LCD plastic filter
    Figure 8-2   Putting the LCD plastic filter on the LCD monitor 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    8 - 4 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
    8-2-1-2              Removal procedure  (cont’d)
    4.) Perform the following functional tests. If all are successful, include the debrief script provided below.
    8-2-1-3 Mounting Procedure 1.) Install the LCD monitor cover parts in the reve rse order of removal procedure. Refer to the 8-2-2 
    Monitor cover set on page 8-5 .
    2.) Put the label on the rear of the LCD monitor. Table 8-2    Functional Tests
    Service Manual Section
    Functional Test / Diagnostic TestDebrief Script
    Section 4-3-1 Power On/Boot Up
    “Service Manual, Direction
    5394141, Rev 1+, Section 8-2-1. Equipment 
    passed all required tests and is ready for use. “
    Section 4-3-2
    Power Off / Shutdown
    Section 4-8-2 LCD plastic filter Function Validation 
    Procedure
    Section 10-5-5 Physical Inspection
    Figure 8-3   Putting the label on the rear of the LCD monitor 
    						