    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
    Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 35
    8-2-6-4              Assembly procedure  (cont’d)
    26.) Screw 2 screws (2337572, FH M3X6 WHT) to assemble the Arm cover L & R.
    27.) Screw 4 screws (2159634, BH M4X10 WHT) to assemble the Arm bottom Cover. Figure 8-63   Assembling the Arm Cover L & R
    Figure 8-64   Assembling the Arm bottom cover 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    8 - 36 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
    8-2-6-4              Assembly procedure  (cont’d)
    28.) Attach LCD Arm Caution Label, LCD Arm Rotation Label, LCD Lock Caution Label.
    Figure 8-65   Flexible Arm
    LCD Arm Caution Label LCD Arm Rotation Label
    LCD Lock Caution Label 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
    Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 37
    8-2-7 Main Keyboard Assy
    8-2-7-1 Tools • Common pilIips screwdrivers
    8-2-7-2 Preparations • Shut down the system and switch off the main breaker.
    8-2-7-3 Removal procedure 1.) Remove the 3 Probe holders & Cup holder from the keyboard.
    2.) Unscrew 4 screws from the  bottom front of the keyboard.
    Figure 8-66   Removing holders
    Figure 8-67   Unscrews 4 screws
    (1) (4)
    (2) (3) 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    8 - 38 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
    8-2-7-3              Removal procedure  (cont’d)
    3.) Unscrew 2 screws from the rear side of the keyboard.
    4.) Lift the top cover of the keyboard and  disconnect 4 cables from the keyboard.
    Figure 8-68   Unscrews 2 screws
    Figure 8-69   Lifting the top cover & Disconnecting 5 Cables
    (1) (2) 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
    Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 39
    8-2-7-3              Removal procedure  (cont’d)
    5.) Perform the following functional tests. If all ar e successful, include the debrief script provided below.
    8-2-7-4 Mounting Procedure Install the new parts in the reverse order of removal. Table 8-8    Functional Tests
    Service ManualSection
    Functional Test / Diagnostic TestDebrief Script
    Section 4-3-1 Power On/Boot Up
    “Service Manual, Direction
    5394141, Rev 1+, Section 8-2-7. Equipment 
    passed all required tests and is ready for use. “
    Section 4-3-2
    Power Off / Shutdown
    Section 4-7-1 Main Keyboard Assy Function Validation 
    Procedure
    Section 4-3-4 System B/M-Mode Checks
    Section 4-3-5 System CFM and PWD Checks
    Section 10-5-2 Functional Checks (See Also Chapter 4) 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    8 - 40 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
    8-2-8 Sub Keyboard Assy 
    8-2-8-1 Tools • Common pilIips screwdrivers
    8-2-8-2 Preparations • Shut down the system and switch off the main breaker.
    8-2-8-3 Removal procedure 1.) Remove the Main Keyboard Assy. Refer to the  8-2-7 Main Keyboard Assy on page 8-37.
    2.) Unscrew 4 screws (1-4). 2 screws(1-2) are back side or the speaker and 2 screws(3-4) are bottom  of sub keyboard. 
    3.) Perform the following functional tests. If all are successful, include the debrief script provided below. Figure 8-70   4 Screw Sub Keyboard Assy
    Table 8-9    Functional Tests
    Service Manual Section
    Functional Test / Diagnostic TestDebrief Script
    Section 4-3-1 Power On/Boot Up
    “Service Manual, Direction
    5394141, Rev 1+, Section 8-2-8. Equipment 
    passed all required tests and is ready for use. “
    Section 4-3-2
    Power Off / Shutdown
    Section 4-7-2 Sub Keyboard Assy Function Validation 
    Procedure
    (1) (2)
    (3) (4) 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
    Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 41
    8-2-8-4 Mounting Procedure 1.) The product emblem is attached on the sub keyb oard, so proper elblem should be attached on the 
    sub keyboard when replace the sub keyboard.
    The sub keyboard FRU include 4 kinds of emblem which are LOGIQ P5, LOGIQ P5 Ob, LOGIQ A5, 
    LOGIQ A5 pro.
    2.) Attache proper product emblem as same as in stall base system before assemble the sub keyboard 
    assy.
    3.) Refer Install the new parts in the reverse order of removal. Figure 8-71   Emblems
    Figure 8-72   Emblem on Sub Keyboard 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    8 - 42 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
    8-2-9 Trackball Assy 
    8-2-9-1 Tools • Common pilIips screwdrivers
    8-2-9-2 Preparations • Shut down the system and switch off the main breaker.
    8-2-9-3 Removal procedure 1.) Remove the Main Keyboard Assy. Refer to the  8-2-7 Main Keyboard Assy on page 8-37.
    2.) Unscrew 4 screws (1-4) and disconnect the cable.
    3.) Perform the following functional tests. If all are successful, include the debrief script provided below.
    8-2-9-4 Mounting Procedure Install the new parts in the reverse order of removal. Figure 8-73   Screws and connector for Trackball
    Table 8-10    Functional Tests
    Service Manual Section
    Functional Test / Diagnostic TestDebrief Script
    Section 4-3-1 Power On/Boot Up
    “Service Manual, Direction
    5394141, Rev 1+, Section 8-2-9. Equipment 
    passed all required tests and is ready for use. “
    Section 4-3-2
    Power Off / Shutdown
    Section 4-7-3 Trackball Assy
     validation
    4 screws
    connector 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
    Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 43
    8-2-10 English, Greek, Russian, Swedish, Norweg ian A/N Key assy 
    8-2-10-1 Tools
    • Common pilIips screwdrivers
    8-2-10-2 Preparations • Shut down the system and switch off the main breaker.
    8-2-10-3 Removal procedure 1.) Remove the Main Keyboard Assy. Refer to the  8-2-7 Main Keyboard Assy on page 8-37.
    2.) Remove USB PCB by unscrewing 6 screws  (1-6) and disconnecting the 6 cables.
    Figure 8-74   A/N Key in keyboard
    Figure 8-75   USB PCB
    Connectors
    1 ~ 3
    Connector 3
    4~ 6 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    8 - 44 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
    8-2-10-3              Removal procedure  (cont’d)
    3.) Unscrew 6 screws from the  back side of the A/N key assy.
    4.) Perform the following functional tests. If all are successful, include the debrief script provided below.
    8-2-10-4 Mounting Procedure Install the new parts in the reverse order of removal. Figure 8-76   Screws on the A/N Key
    Table 8-11    Functional Tests
    Service Manual Section
    Functional Test / Diagnostic TestDebrief Script
    Section 4-3-1 Power On/Boot Up
    “Service Manual, Direction
    5394141, Rev 1+, Section 8-2-10. Equipment 
    passed all required tests and is ready for use. “
    Section 4-3-2
    Power Off / Shutdown
    Section 4-7-4 A/N Key assy validation
    12
    3
    4 5 6 
    						