Home > GE > Medical Equipment > GE Logiq P5 Service Manual

GE Logiq P5 Service Manual

    Download as PDF Print this page Share this page

    Have a look at the manual GE Logiq P5 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    Page
    of 520
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL
    Section 4-2 - Required Equipment 4 - 15
    4-3-10 Using the DVD-R
    4-3-10-1 Using the DVD-R
    1.) Go to the Utility->Connectivity->Removable Media.
    2.) Insert CD/DVD media.
    3.) Press Verify 
    4.) The properties will be like below. a.) Capacity, Free space depends on media type.
    b.) Formatted / Finalized : No
    5.) Go to Patient.
    6.) Press Data Transfer (located at the upper-left side of Patient screen)
    7.) Select Export task.
    NOTICENever move the unit with a disk in the DVD-R because the drive actuator will not be locked 
    and the DVD-R could break. 
    Figure 4-7   Removable Media 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
    4 - 16 Section 4-2 - Required Equipment
    4-3-10-1              Using the DVD-R (cont’d)
    8.) Select the type of Removable CD/DVD on the [To] combo box.
    9.) It will do auto -formatting.
    10.) After auto-formatting, select some patients on  the upper list area and then press the [Transfer] 
    button.
    11.) it will do trans ferring to CD/DVD.Figure 4-8   Formatting
    Figure 4-9   Transferring 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL
    Section 4-2 - Required Equipment 4 - 17
    4-3-10-1              Using the DVD-R (cont’d)
    12.) The selected data is transferred correctly.
    13.) Press F3 and select the [Yes and Verify files].
    14.) It will do finalizing. Figure 4-10   Transfer
    Figure 4-11   Finalizing 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
    4 - 18 Section 4-2 - Required Equipment
    4-3-10-1              Using the DVD-R (cont’d)
    15.) It will do Verifying.
    16.) It will eject the CD/DVD and Go  to Utility and reinsert the CD/DVD.
    17.) Go to Utility- >Connectivity->Removable Media.
    18.) Press Verify 
    19.) The properties will be like below.  a.) Capacity is the size of transferred data.
    b.) Free space is 0.
    c.) Formatted / Database  Present / Finalized : Yes
    (DICOMDIR Present : depends on setting) Figure 4-12   Verifying
    Figure 4-13   Removable Media
    NOTICEAvoid mechanical ejection whenever possible . Mechanical ejection leaves the actuator 
    unlocked. If forced to use this method, reboot the system, then insert and eject a known good 
    disk using one of the other methods. 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL
    Section 4-2 - Required Equipment 4 - 19
    4-3-11 Backup and Restore Database, Preset Configurations and Images
    NOTE: When upgrading from R1.0.x or R2.0.x to R3.0.x or R4.0.x ,preset cannot be transferred. Refer 
    to the table below.
    4-3-11-1 Formatting Media 1.) To format the backup media, CD- R or DVD-R, select the UTILITY
     key on the Front Panel.
    2.) Select CONNECTIVITY
    , then RERNOVABLE MEDIA. Properly label and Insert the backup 
    media. 
    3.) Select the media type from the drop down menu.
    4.) Enter the label for the media as shown in Figure  4-14. It is best to use all capital letters with no 
    Table 4-4    LP5 BT11(R4.x.x) compatibility
    Preset
    Compatibility
    R1.x.xR2.x.xR3.x.x
    Patient Archive
    YesYes Yes
    Report Archive YesYes Yes
    Imaging Presets NoNoYes
    Connectivity ConfigurationNoNoYes
    Measurement ConfigurationNoNoYes
    Comment/Body Pattern LibrariesNoNoYes
    Report Templates (Same Software Version Only) NoNoNo
    3D/4D NoNoYes
    Key Macro NoNoYes
    All OthersNoNoYes 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
    4 - 20 Section 4-2 - Required Equipment
    spaces or punctuation marks.
    5.) The Ultrasound system displays a pop-up menu, as shown in Figure 4-15. When the formatting  has been completed, press OK
     to continue.
    6.) If desired, verify that the form at was successful by returning to  Utility > Connectivity > 
    Removable Media  and selecting VERIFY
     as shown in Figure 4-12. 
    Figure 4-14   Format and Verify Media
    Figure 4-15   Format Successful Pop-up Menu
    Verify Format
    Format
    Media 
    Selection
    Media Label 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL
    Section 4-2 - Required Equipment 4 - 21
    4-3-11-2 Backup System Presets and Configurations
    NOTE: Always backup any preset configurations before a so ftware reload. This ensures that if the presets need 
    to be reloaded, afte r the software update, they will be the same ones the cu stomer was using prior to 
    service.
    1.) Insert a formatted CD-R or DVD-R into the drive.
    2.) Press UTILITY
     key 
    3.) Select SYSTEM
    . on the monitor screen
    4.) On the monitor display, select BACKUP/RESTORE
    .
    5.) In the Backup list, select Patient Archive,  Report Archive, User Defined Configuration.
    6.) In the Media field, select DVD-R.
    7.) Select BACKUP
    .
    The system performs the backup. As  it proceeds, status information is displayed on the Backup/Restore 
    screen.
    Figure 4-16   Backup/Restore Menu
    Check here to 
    backup presets 
    and configurations 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
    4 - 22 Section 4-2 - Required Equipment
    4-3-11-3 Restore System Presets and Configurations
    1.) Insert the Backup/Restore CD-R or DVD-R into the drive. 
    2.) Press the UTILITY
    . key
    3.) Select SYSTEM
    .menu on the monitor screen
    4.) On the monitor display,  select BACKUP/RESTORE
    .
    5.) In the Restore list, select Pa tient Archive, Report Archive, User Defined Configuration.
    6.) In the Media field, select the Backup/Restore DVD-R.
    7.) Select RESTORE
    .
    d.) The system performs the restore. As it proceeds, status information is displayed on the  Backup/Restore screen.
    CAUTIONThe restore procedure  overwrites the existing database on the local hard drive. Make sure to insert 
    the correct CD-R or DVD-R.
    Figure 4-17   Backup/Restore Menu 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL
    Section 4-2 - Required Equipment 4 - 23
    4-3-11-4 Archiving Images
    To export an exam(s) to a compatible Ultrasound system:
    1.) Format the removable media (CD-ROM/DVD-R).  Label the removable media. Answer Yes/OK 
    to the messages.Press Patient. Deselect any select ed patient(s) in the search portion of the 
    Patient screen. Press Exam Data Transfer (located at the upper, left-hand corner of the Patient 
    menu).
    2.) Select Export task Button.
    3.) The [From] combo box is not active. It displa ys Local Archive. The [To] combo box is active. 
    Select the type of removable media CD R/DVD- R. Then please wait until the patient list is 
    visible. Figure 4-18   Exam Data Transfer
    Figure 4-19   Task button
    Figure 4-20   To Combo Box 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
    4 - 24 Section 4-2 - Required Equipment
    4-3-11-4              Archiving Images (cont’d)
    4.) In the patient list at the top of the Patient me nu, select the patient(s) you want to export. You 
    can use Windows commands to select more than  one patient. To select a consecutive list of 
    patients, click the cursor on the first name, move  the cursor to the last name, then press and 
    hold down the Shift+right Set ke y to select all the names. To select a non-consecutive list of 
    patients, click the cursor at the first name, move the cursor to the next name, then press and 
    hold down the Ctrl+right Set key, move the cursor to the next name, then press and hold down 
    the Ctrl+right Set key again, etc.  You can also search for patients via the Search key and string. 
    Or, Select All Button  from the Patient Menu:
    NOTE: You need to use your best judgment when movi ng patients images. If there are lots of images 
    or loops, then only move a few patients at a time.
    5.) Once you have selected all of the patients to export, press Transfer Button as shown in Figure 4-18 from the Patient Menu.
    6.) The progress bar appears as the copy is taking place. Once transfer completed, the exported  patient list will be shown as Figure 4-22.
    7.) Press F3 to eject the media. Specify that you want to  finalize the CD-ROM.
    Figure 4-21   Source Section
    Figure 4-22   Destination Section of Patient Screen 
    						
    All GE manuals Comments (0)