Home > GE > Medical Equipment > GE Logiq P5 Service Manual

GE Logiq P5 Service Manual

    Download as PDF Print this page Share this page

    Have a look at the manual GE Logiq P5 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    Page
    of 520
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL
    Section 5-4 - Main Board Detail 5 - 59
    Section 5-13Fast Polling
    5-13-1 General 
    This describes Fast Polling feature that will change the polling rate to 15 seconds for 10 minutes.
    5-13-2 Enabling Fast Polling
    1.) Right Click on “Con nect to GE” icon.
    2.) Pressing “OK” will change the polling rate to 15  seconds for 10 minutes.
    Figure 5-63   
    Figure 5-64   Agent Configuration 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL
    5 - 60Section 5-4 - Main Board Detail
    This page was intentionally left blank. 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
    Section 6-1 - Overview 6 - 1
    Chapter 6
    Service Adjustments
    Section 6-1
    Overview
    6-1-1 Purpose of this chapter 6
    This section describes how to test and adjust the scanner. These tests are optional. You may use them 
    to check the system for errors.
    Section 6-2
    Regulatory
    Verify, where applicable, that any regulatory information or tests required by national law are present 
    and accounted for, and any regulatory tests required by national law are performed  and documented.
    Table 6-1    Contents in chapter 
    SectionDescriptionPage Number
    6-1 Overview
    6-1
    6-2 Regulatory 6-1
    6-3 Monitor Adjustments 6-2 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    6 - 2 Section 6-3 - Monitor Adjustments
    Section 6-3 
    Monitor Adjustments
    6-3-1 Rotate and ti lt the monitor
    The monitor position can be adjusted for easy viewing.
    - The monitor can be rotated around its central pivot point.
    - The monitor can be tilted for the optimum viewing angle.
    - the monitor can be flipped up and down.
    NOTE:  When moving the system, f lip down the monitor to its lowest possible position to improve 
    stability.
    Figure 6-1 LCD monitor Flip up and down 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
    Section 6-3 - Monitor Adjustments 6 - 3
    6-3-2 Brightness and Contrast 
    Adjusting the monitors contrast and brightness is  one of the most important factors for proper image 
    quality. If these controls are set incorrectly, th e Gain, TGC, Dynamic Range and even Power Output 
    may have to be changed more often than necessary to compensate.
    The proper setup displays a complete gray scale. T he lowest level of black should just disappear into 
    the background and the highest white should be bright, but not saturated.
    To adjust the contrast:
    1.) Press the Toggle button on the display monitor once.
    2.) Press the Adjustment (+) button  to increase contrast. Press the Adjustment (-) button to decrease 
    contrast.
    3.) The amount of contrast is sh own on a slide bar on the screen.
    To adjust the brightness:
    4.) Press the Toggle button on  the display monitor twice.
    5.) Press the Adjustment (+) button to increase bri ghtness. Press the Adjustment (-) button to decrease 
    brightness.
    6.) The amount of brightness is  shown on a slide bar on the screen.
    Record any changes to the final brightness and contrast settings and leave this information with the 
    system. Generally speaking, do not change the controls  once they have been set. Once set, the display 
    then becomes the reference for the hard copy device(s).
    NOTE:  After readjusting the monitors Contrast and  Brightness, readjust all preset and peripheral 
    settings. Figure 6-2   Brightness and Contrast
    1. Contrast Indicator
    2. Brightness Indicator
    3.  Dim Brightness Indicator
    4. Light Button
    5.  Adjustment (-) Button
    6. Toggle Button
    7.  Adjustment (+) Button 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    6 - 4 Section 6-3 - Monitor Adjustments
    6-3-3 LCD Plastic filter
    6-3-3-1 Tools • Common Phillips screwdrivers
    • Allen/Unbraco wrench
    • Stubby screwdriver (Flat tip and Cross tip)
    6-3-3-2 Removal procedure 1.) Remove the Monitor cover set. Refer to the  8-2-2 Monitor cover set on page 8-5 and the figure 
    below.
    2.) Remove the LCD Plastic filter.
    3.) Put the LCD plastic filter on the LCD monitor  after detaching the tape of the LCD plastic filter.
    Figure 6-3   Removing the LCD plastic filter
    Figure 6-4   Putting the LCD plastic filter on the LCD monitor 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
    Section 6-3 - Monitor Adjustments 6 - 5
    6-3-3-2              Removal procedure  (cont’d)
    4.) Perform the following functional tests. If all ar e successful, include the debrief script provided below.
    6-3-3-3 Mounting Procedure 1.) Install the LCD monitor cover parts in the reve rse order of removal procedure. Refer to the 8-2-2 
    Monitor cover set on page 8-5 .
    2.) Put the label on the rear of the LCD monitor. Table 6-2    Functional Tests
    Service Manual Section
    Functional Test / Diagnostic TestDebrief Script
    4-3-1 Power On/Boot Up
    “Service Manual, Direction
    5394141, Rev 2+, 6-3-3. Equipment passed all  required tests and is ready for use. “
    4-3-2
    Power Off / Shutdown
    4-8-2 LCD plastic filter Function Validation 
    Procedure
    10-5-5 Physical Inspection
    Figure 6-5   Putting the label on the rear of the LCD monitor 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    6 - 6 Section 6-3 - Monitor Adjustments
    6-3-4 Flexible LCD arm - locking mechanism
    Flexible LCD arm’s rotation movement can be adjusted by using the LCD arm’s locking mechanism.
    To unlock the LCD arm:
    1.) Pull the LCD arm toggle knob and turn clockwise to unlock the LCD arm.
    To lock the LCD arm :
    1.) Turn counter clockwise and push the LCD arm toggle knob and to lock the LCD arm. Figure 6-6   Unlocking the Flexible LCD arm
    Figure 6-7   Locking the Flexible LCD arm 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
    Section 6-4 - Trackball Cleaning 6 - 7
    Section 6-4
    Trackball Cleaning
    The trackball on the keyboard need to be cleaning periodically.
    1.) Open the trackball 
    Figure 6-8   Open Trackball
    Figure 6-9   Trackball Hole
    Remove dust inside of bucket 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    6 - 8 Section 6-4 - Trackball Cleaning
    This page was intentionally left blank. 
    						