GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 5 8-2-2 Monitor cover set 8-2-2-1 Tools • Common pilIips screwdrivers • Allen/Unbraco wrench • Stubby screwdriver (Flat tip and Cross tip) 8-2-2-2 Removal procedure 1.) Remove the LCD cable cover. 2.) Unscrew 1 screw (2306565, BH M4x16 WHT) to remove the power cord bracket assembled under the LCD cable cover. Figure 8-4 Removing the LCD cable cover Figure 8-5 Removing the power cord bracket
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 6 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-2-2-2 Removal procedure (cont’d) 3.) Disconnect power cable and DVI cable. 4.) Unscrew 2 screws(5177684, HSH M5x20 WHT) to separate LCD monitor. 5.) Unscrew 4 screws to remove the LCD rear cover. Figure 8-6 Disconnecting power cable and DVI cable Figure 8-7 Separating the LCD monitor Figure 8-8 Removing the LCD rear cover
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 7 8-2-2-2 Removal procedure (cont’d) 6.) Unscrew 4 screws to remove the OS D key assy and disconnect the connect. 7.) Separate the LCD tact knob and LCD lamp window from the OSD key assy. 8.) Unscrew 5 screws to remove LCD front cover. Figure 8-9 Separating the OSD key assy Figure 8-10 Separating the LCD tact knob and LCD lamp window Figure 8-11 Removing the LCD front cover
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 8 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-2-2-2 Removal procedure (cont’d) 9.) Perform the following functional tests. If all are successful, include the debrief script provided below. 8-2-2-3 Mounting Procedure Install the each of the new part in the reverse order of removal procedure. Table 8-3 Functional Tests Service ManualSection Functional Test / Diagnostic TestDebrief Script Section 4-3-1 Power On/Boot Up “Service Manual, Direction 5394141, Rev 1+, Section 8-2-2. Equipment passed all required tests and is ready for use. “ Section 4-3-2 Power Off / Shutdown Section 4-8-2 LCD plastic filter Function Validation Procedure Section 10-5-5 Physical Inspection
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 9 8-2-3 OSD button and lamp set, LCD Plastic Filter 8-2-3-1 Tools • Common pilIips screwdrivers 8-2-3-2 Preparations • Shut down the system and switch off the main breaker. • Maneuver control console to a suitable position for removing the monitor. 8-2-3-3 Removal procedure 1.) Unscrew 4 screws (1-4) from the LCD front cover. 2.) Perform the following functional tests. If all ar e successful, include the debrief script provided below. 8-2-3-4 Mounting Procedure Install the new parts in the reverse order of removal. Figure 8-12 Unscrews 4 screws Table 8-4 Functional Tests Service Manual Section Functional Test / Diagnostic TestDebrief Script Section 4-3-1 Power On/Boot Up “Service Manual, Direction 5394141, Rev 1+, Section 8-2-3. Equipment passed all required tests and is ready for use. “ Section 4-3-2 Power Off / Shutdown Section 4-6-2 15 OSD Key Assy Function Validation Procedure Section 10-5-5 Physical Inspection (1) (2) (3) (4)
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 10 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-2-4 LCD Std Arm with Cover 8-2-4-1 Tools • Common pilIips screwdrivers • Allen/Unbraco wrench 8-2-4-2 Preparations • Shut down the system and switch off the main breaker. • Maneuver control console to a suitable position for removing the monitor. 8-2-4-3 Removal procedure 1.) Remove the Power Cable, DVI Cable. 2.) Unscrew 1 screw to re move the LCD arm cover. Figure 8-13 Removing the Cable Figure 8-14 Unscrews 1 screw POWER CABLEDVI CABLE
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 11 8-2-4-3 Removal procedure (cont’d) 3.) Perform the following functional tests. If all ar e successful, include the debrief script provided below. 8-2-4-4 Mounting Procedure Install the new parts in the reverse order of removal. Table 8-5 Functional Tests Service ManualSection Functional Test / Diagnostic TestDebrief Script Section 4-3-1 Power On/Boot Up “Service Manual, Direction 5394141, Rev 1+, Section 8-2-4. Equipment passed all required tests and is ready for use. “ Section 4-3-2 Power Off / Shutdown Section 4-8-8 Flexible LCD Arm Function Validation Procedure Section 10-5-5Physical Inspection
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 12 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-2-5 Monitor Assy 8-2-5-1 Tools • Common pilIips screwdrivers • Allen/Unbraco wrench 8-2-5-2 Preparations • Shut down the system and switch off the main breaker. • Maneuver control console to a suitable position for removing the monitor. 8-2-5-3 Removal procedure 1.) Detach the monitor assy from arm. - Open monitor cable cover - Unscrew a screw of the power connector fixture bracket - Unplug the power plug and DVI connector - Flip down the monitor - Unscrew the 2 hexa screw of the monitor hinge While unscrew the hinge hexa screw from the arm, keep hold the monitor to prevent drop down. 2.) Perform the following functional tests. If all are successful, include the debrief script provided below. 8-2-5-4 Mounting Procedure Install the new parts in the reverse order of removal Figure 8-15 Disassemble Monitor from STD arm Table 8-6 Functional Tests Service Manual Section Functional Test / Diagnostic TestDebrief Script Section 4-3-1 Power On/Boot Up “Service Manual, Direction 5394141, Rev 1+, Section 8-2-5. Equipment passed all required tests and is ready for use. “ Section 4-3-2 Power Off / Shutdown Section 4-6-1 15 LCDMON II Assy & LCD Panel with filter Function Validation Procedure Section 10-5-2Functional Checks (See Also Chapter 4)
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 13 8-2-6 Flexible LCD Arm Removal Procedure 8-2-6-1 Tools • Common pilIips screwdrivers • Allen/Unbraco wrench • Stubby screwdriver (Flat tip and Cross tip) 8-2-6-2 Preparations • Shut down the system and switch off the main breaker. • Maneuver control console to a suitable position for removing the monitor. 8-2-6-3 Removal procedure 1.) Remove the LCD cable cover. 2.) Unscrew 1 screw (2306565, BH M4x16 WHT) to remove the power cord bracket assembled under the LCD cable cover. Figure 8-16 Removing the LCD cable cover Figure 8-17 Removing the power cord bracket
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 14 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-2-6-3 Removal procedure (cont’d) 3.) Disconnect power cable and DVI cable. 4.) Unscrew 1 screw (2329677, TAP M4X16) to remove the Fold cover. 5.) Unscrew 2 screws (2159634, BH M4 x10 WHT) to remove the Cable clamp. Figure 8-18 Disconnecting power cable and DVI cable Figure 8-19 Removing the Fold cover Figure 8-20 Removing the Cable clamp