    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
    Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 5
    8-2-2 Monitor cover set
    8-2-2-1 Tools • Common pilIips screwdrivers
    • Allen/Unbraco wrench
    • Stubby screwdriver (Flat tip and Cross tip)
    8-2-2-2 Removal procedure 1.) Remove the LCD cable cover.
    2.) Unscrew 1 screw (2306565, BH M4x16 WHT) to remove the power cord bracket assembled under the LCD cable cover. Figure 8-4   Removing the LCD cable cover
    Figure 8-5   Removing the power cord bracket 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    8 - 6 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
    8-2-2-2              Removal procedure  (cont’d)
    3.) Disconnect power cable and DVI cable.
    4.) Unscrew 2 screws(5177684, HSH M5x20 WHT) to separate LCD monitor.
    5.) Unscrew 4 screws to remove the LCD rear cover. Figure 8-6   Disconnecting power cable and DVI cable
    Figure 8-7   Separating the LCD monitor
    Figure 8-8   Removing the LCD rear cover 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
    Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 7
    8-2-2-2              Removal procedure  (cont’d)
    6.) Unscrew 4 screws to remove the OS D key assy and disconnect the connect.
    7.) Separate the LCD tact knob and LCD lamp window from the OSD key assy.
    8.) Unscrew 5 screws to  remove LCD front cover.
    Figure 8-9   Separating the OSD key assy
    Figure 8-10   Separating the LCD tact knob and LCD lamp window
    Figure 8-11   Removing the LCD front cover 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    8 - 8 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
    8-2-2-2              Removal procedure  (cont’d)
    9.) Perform the following functional tests. If all are successful, include the debrief script provided below.
    8-2-2-3 Mounting Procedure Install the each of the new part in the reverse order of removal procedure. Table 8-3    Functional Tests
    Service ManualSection
    Functional Test / Diagnostic TestDebrief Script
    Section 4-3-1 Power On/Boot Up
    “Service Manual, Direction
    5394141, Rev 1+, Section 8-2-2. Equipment 
    passed all required tests and is ready for use. “
    Section 4-3-2
    Power Off / Shutdown
    Section 4-8-2 LCD plastic filter Function Validation 
    Procedure
    Section 10-5-5 Physical Inspection 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
    Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 9
    8-2-3 OSD button and lamp  set, LCD Plastic Filter
    8-2-3-1 Tools
    • Common pilIips screwdrivers
    8-2-3-2 Preparations • Shut down the system and switch off the main breaker.
    • Maneuver control console to a suitable position for removing the monitor.
    8-2-3-3 Removal procedure 1.) Unscrew 4 screws (1-4) from the LCD front cover.
    2.) Perform the following functional tests. If all ar e successful, include the debrief script provided below.
    8-2-3-4 Mounting Procedure Install the new parts in the reverse order of removal. Figure 8-12   Unscrews 4 screws
    Table 8-4    Functional Tests
    Service Manual Section
    Functional Test / Diagnostic TestDebrief Script
    Section 4-3-1 Power On/Boot Up
    “Service Manual, Direction
    5394141, Rev 1+, Section 8-2-3. Equipment 
    passed all required tests and is ready for use. “
    Section 4-3-2
    Power Off / Shutdown
    Section 4-6-2 15 OSD Key Assy Function Validation 
    Procedure
    Section 10-5-5 Physical Inspection
    (1)
    (2) (3) (4) 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    8 - 10 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
    8-2-4 LCD Std Arm with Cover
    8-2-4-1 Tools • Common pilIips screwdrivers
    • Allen/Unbraco wrench
    8-2-4-2 Preparations • Shut down the system and switch off the main breaker.
    • Maneuver control console to a suitable position for removing the monitor.
    8-2-4-3 Removal procedure 1.) Remove the Power Cable, DVI Cable.
    2.) Unscrew 1 screw to re move the LCD arm cover.
    Figure 8-13   Removing the Cable
    Figure 8-14   Unscrews 1 screw
    POWER CABLEDVI CABLE 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
    Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 11
    8-2-4-3              Removal procedure  (cont’d)
    3.) Perform the following functional tests. If all ar e successful, include the debrief script provided below.
    8-2-4-4 Mounting Procedure Install the new parts in the reverse order of removal. Table 8-5    Functional Tests
    Service ManualSection
    Functional Test / Diagnostic TestDebrief Script
    Section 4-3-1 Power On/Boot Up
    “Service Manual, Direction
    5394141, Rev 1+, Section 8-2-4. Equipment 
    passed all required tests and is ready for use. “
    Section 4-3-2
    Power Off / Shutdown
    Section 4-8-8 Flexible LCD Arm Function Validation 
    Procedure
    Section 10-5-5Physical Inspection 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    8 - 12 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
    8-2-5 Monitor Assy
    8-2-5-1 Tools • Common pilIips screwdrivers
    • Allen/Unbraco wrench
    8-2-5-2 Preparations • Shut down the system and switch off the main breaker.
    • Maneuver control console to a suitable position for removing the monitor.
    8-2-5-3 Removal procedure 1.) Detach the monitor assy from arm. - Open monitor cable cover
    - Unscrew a screw of the power connector fixture bracket
    - Unplug the power plug and DVI connector
    - Flip down the monitor
    - Unscrew the 2 hexa screw of the monitor hinge
    While unscrew the hinge hexa screw from the arm, keep hold the monitor to prevent drop down. 
    2.) Perform the following functional tests. If all are successful, include the debrief script provided below.
    8-2-5-4 Mounting Procedure Install the new parts in the reverse order of removal Figure 8-15   Disassemble Monitor from STD arm
    Table 8-6    Functional Tests
    Service Manual Section
    Functional Test / Diagnostic TestDebrief Script
    Section 4-3-1 Power On/Boot Up
    “Service Manual, Direction
    5394141, Rev 1+, Section 8-2-5. Equipment 
    passed all required tests and is ready for use. “
    Section 4-3-2
    Power Off / Shutdown
    Section 4-6-1 15 LCDMON II Assy & LCD Panel with filter 
    Function Validation Procedure
    Section 10-5-2Functional Checks (See Also Chapter 4) 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
    Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 13
    8-2-6 Flexible LCD Arm Removal Procedure
    8-2-6-1 Tools • Common pilIips screwdrivers
    • Allen/Unbraco wrench
    • Stubby screwdriver (Flat tip and Cross tip)
    8-2-6-2 Preparations • Shut down the system and switch off the main breaker.
    • Maneuver control console to a suitable position for removing the monitor.
    8-2-6-3 Removal procedure 1.) Remove the LCD cable cover.
    2.) Unscrew 1 screw (2306565, BH M4x16 WHT) to remove the power cord bracket assembled under the LCD cable cover. Figure 8-16   Removing the LCD cable cover
    Figure 8-17   Removing the power cord bracket 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    8 - 14 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
    8-2-6-3              Removal procedure  (cont’d)
    3.) Disconnect power cable and DVI cable.
    4.) Unscrew 1 screw (2329677, TAP M4X16) to remove the Fold cover.
    5.) Unscrew 2 screws (2159634, BH M4 x10 WHT) to remove the Cable clamp.
    Figure 8-18   Disconnecting power cable and DVI cable
    Figure 8-19   Removing the Fold cover
    Figure 8-20   Removing the Cable clamp 
    						