GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 25 8-2-6-4 Assembly procedure (cont’d) 3.) Insert two of the flexible arm stoppers (A+B ) to the Power Cable and DVI cable in the Order illustrated on the figure be low. Make sure that chamfered shape of the flex ible Arm stopper should face inside. 4.) Insert oilless bush into the flexible arm neck pipe and insert two cables into the flexible arm neck pipe. 5.) Insert the flexible arm into the flexible arm neck pipe. Figure 8-40 Inserting cables to the flexible arm stopper Figure 8-41 Inserting the oilless bushFigure 8-42 Inserting the flexible arm Flexible Arm Stopper
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 26 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-2-6-4 Assembly procedure (cont’d) 6.) Screw 3 screws (2373562, FH M4X10 YEL) to fix the flexible arm stopper after adjusting the holes. 7.) Screw 1 screw (2159625, PH M4X8 W/SP) to install OP bracket. 8.) Screw 3 screws (5327646, HSH M6X16 WHT) to install the arm neck pipe. Make sure place DVI cable be on the right side, power cable on the left side. Figure 8-43 Fixing the flexible arm stopper Figure 8-44 Installing OP Bracket Figure 8-45 Assembling the arm neck pipe
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 27 8-2-6-4 Assembly procedure (cont’d) 9.) Screw 2 screws (5177684, HSH M5X20 WHT) to assemble the LCD monitor with flexible arm Top part. 10.) Insert the two cables inside the Flexible arm and Screw 4 screws (2159625, PH M4X8 W/SP) to assemble the Lower arm cable bracket. 11.) Put in the two cables. Make sure place DVI cabl e be on the right side, power cable on the left side. Figure 8-46 Screwing 2 screws Figure 8-47 Screwing 4 screws Figure 8-48 Putting in the cables
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 28 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-2-6-4 Assembly procedure (cont’d) 12.) Screw 2 screws (2159626, PH M4X20) to assemble the Arm stopper. 13.) Screw 2 screws (2159625, PH M4X8 W/SP) to assemble the Lower arm guide bracket. 14.) Tie the Cables with Cable tie. Figure 8-49 Installing the Arm stopper Figure 8-50 Assembling the Lower arm guide bracket Figure 8-51 Tying the cables
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 29 8-2-6-4 Assembly procedure (cont’d) 15.) Place the cables inside the slot and connect power cable and DVI cable. 16.) Assemble the Cable bracket with 1 Screw (2159625, PH M4X8 W/SP) 17.) Screw 2 screws (2159634, BH M4X1 0 WHT) to assemble the Cable clamp. Figure 8-52 Placing the Cables Figure 8-53 Assembling the Cable bracket Figure 8-54 Assembling the Cable clamp
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 30 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-2-6-4 Assembly procedure (cont’d) 18.) Place the cables like the figure below. 19.) Screw 1 screw (2306565, BH M4X16 WHT) to fix the Power cable. Figure 8-55 Placing the cables Figure 8-56 Assembling the power cord bracket
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 31 8-2-6-4 Assembly procedure (cont’d) 20.) Assemble the LCD cable cover. Figure 8-57 Assembling the LCD cable cover
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 32 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-2-6-4 Assembly procedure (cont’d) 21.) Screw 3 screws (2159634, BH M4x10 WHT) after placing the neck front cover and neck Rear cover. 22.) Screw 1 screw (2159625, PH M4X8 W/SP) to assemble the Lock Knob. Figure 8-58 Assembling the neck front cover and the neck rear cover Figure 8-59 Assembling the Lock Knob
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 33 8-2-6-4 Assembly procedure (cont’d) 23.) Screw 1 screw (2329677, TAP M4X16) to assemble the Lock Cover. Figure 8-60 Assembling the Lock Cover
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 34 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-2-6-4 Assembly procedure (cont’d) 24.) Screw 2 screws (2159632, BH M4X6 WHT) to assemble the Cam Cover L & R. 25.) Screw 1 screw (2329677, TAP M4X16) to assemble the Fold Cover. Figure 8-61 Assembling the Cam Cover L & R Figure 8-62 Assembling the Fold Cover