    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
    Section 7-6 - LED Descriptions 7 - 21
    7-6-4 APS/APS Pro Assy
    NOTE: HVL+ and HVL- are  not available for APS PRO.
    The APS/APS Pro Assy is located at the No. 4 slot of the NEST Assy.
    Table 7-11     LEDs on APS/APS Pro Assy
    LEDDescriptionsNormal State
    DP5 (Top) Digital 5V output indicator Normally On
    LP5 Live Power 5V output indicator Normally On
    AP6 Analog 6V output indicator Normally On
    DP33 Digital 3.3V output indicator Normally On
    SHV+ SHV+ output indicator status indicator Normally On
    HVL+ HVL+ level indicator Normally On in CFM mode or PW mode
    HVH+ (Bottom) HVH+ level indicator Normally On in B mode 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    7 - 22 Section 7-7 - Trouble Shooting Tree
    Section 7-7Trouble Shooting Tree
    7-7-1 System Does Not  Power On / Boot Up
    Figure 7-10   Power On / Boot Up Flowchart
    Yes
    Is the scanner 
    plugged in?
    Is the AC at the  wall outlet?
    Make sure the scanner is not  plugged into 
    emergency power  (red outlet);
    Is the circuit 
    breaker ON?
    Yes
    Press the on/off switch to  boot the system. Is the 
    on/off switch illuminated  with an green color and 
    can you listen fan sound  of inside?
    Yes
    Yes
    Does the scanner  boot up?
    Yes
    Scanner is ON and ready for use.
    NO
    NO
    NONO
    NO Check if wall power is alive
    Check if the standby 5V od APS is alive 
    Make sure AC power connector and con-
    trol signal connector is plugged properly
    If HDD indicator is blinking after power 
    on, check the monitor cable
    Check BIOS or replace PM SOM module  
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
    Section 7-7 - Trouble Shooting Tree 7 - 23
    7-7-2 System Does Not Boot
    Figure 7-11   System does not Boot
    System doesn’t boot!
    Connect the AC power cable 
    between the Scanner and the wall  power outlet in right way.YES
    YES NO
    Turn ON the main breaker of 
    the Scanner.
    Go on to the next page.
    NO
    Is the AC power cable 
    seated to the Scanner and  to the wall power outlet?
    Is the main breaker turned  ON?
    Press the ON/OFF button on the Control Panel.
    NO
    Has the ON/OFF
     key on the Control Panel  been pressed once?
    YES 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    7 - 24 Section 7-7 - Trouble Shooting Tree
    7-7-2               System Does Not Boot (cont’d)
    Figure 7-12   System does not Boot (contd.)
    System doesn’t boot! (cont’d)
    YES
    Check;
    1.) AC Power
    2.) AC Transformer
    Go on to the next page.
    NODo the fans start to run?
    Check/replace the PM SOM/CM  SOM module.
    YES NO
    Is the SOM module starting 
    to display start-up screens  on the monitor?
    Is the LOGIQ starting  display shown on the  monitor?
    Turn off the power, wait a few 
    seconds, and try to turn on the  power again.
    NO 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
    Section 7-7 - Trouble Shooting Tree 7 - 25
    7-7-2              System Does Not Boot (cont’d)
    Figure 7-13   System does not Boot (contd.)
    System doesn’t boot! (cont’d)
    1.) Check the Control  Panel cable 
    connection.
    2.) Check/replace  Control Panel.YES
    YES NO
    Are the Control Panel
    keys lit? 
    Is regular scan display 
    shown on the monitor?
    Press and hold the main switch for  five (5) seconds to reboot the  system.
    NO
    Go to Scan Procedure section.
    END 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    7 - 26 Section 7-7 - Trouble Shooting Tree
    7-7-3 Noise in Image
    Figure 7-14   Noise in Image
    Noise is shown in Image.
    YES
    NOIs there any electrical 
    devices that affect the  Scanner?
    Keep the Scanner away from those  electrical devices.
    Check probes. 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
    Section 7-7 - Trouble Shooting Tree 7 - 27
    7-7-4 Unable to Scan
    Figure 7-15   Unable to Scan
    Unable to scan!
    Check the probe connection.YESNO
    Are probes connected in 
    right way?
    Restart the scanner.
    Are other operations possible?YES
    Check the HV power supply unit.
    NO
    END 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    7 - 28 Section 7-7 - Trouble Shooting Tree
    7-7-5 Trackball
    Figure 7-16   Trackball Troubleshooting
    The trackball is low sensitive!
    Remove the ball from  the Trackball, 
    and clean it.
    YES
    NODoes the trackball move at  all?
    Replace the trackball unit. 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
    Section 7-7 - Trouble Shooting Tree 7 - 29
    7-7-6 Monitor Troubleshooting
    NOTE: Reset the monitor settings to the factory defaults prior to troubleshooting the monitor.
    NOTICEComments to remember about the 15” monitor.
    The monitor is the diagnostic monitor and is connected to AC 110/220 VAC.
    The monitor powers down with the console.
    Power output from AC Po wer assy via AC distributor in AC Transbox assy:
    When the monitor is not connected to the AC power there is a leakage current which 
    when measured with a DVM can indicate high voltage.
    When a load is applied, the voltage collapses to low voltage if the monitor is not 
    switched on. 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    7 - 30 Section 7-7 - Trouble Shooting Tree
    7-7-6               Monitor Troubleshooting (cont’d)
    • Monitor does not work  - Check AC circuit breaker
    - Verify power is present at monitor outlet
    - Check if video is present at backplane
    • Prints do not match monitor - Verify factory default settings. Chapter 18 in  the User Manual calls out suggested settings for 
    various exam and lighting conditions.
    • Video test patterns are not clear, bright, parallel or square - Replace the monitor Figure 7-17   Monitor Troubleshooting
    Yes
    Is the shutdown dialog 
    box displayed?
    Did the scanner  Shut off?
    Yes Depress and hold the ON/OFF switch on 
    the scanner for approximately 5 
    seconds.
    If the scanner fails to Shut Off, follow the 
    step below.
    Check cable connection refer 7-10-5
    If the scanner still does not Shut off, 
    Switch the circuit breaker to the OFF 
    position. 
    NONO
    Press once the on/off switch.
    Shutdown the scanner from a software 
    screen or depress the ON/OFF switch
    Switch the breaker to the 
    OFF position.to 
    completely Power Down 
    the scanner. Press and hold on/off switch until 
    system shut down. 
    						