GE Logiq P5 Service Manual
Have a look at the manual GE Logiq P5 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-3 - Software Loading Procedure 8 - 145 8-4-4-2 Setting Computer Name 1.) Click on Utility, then select Connectivity. 2.) Click on TCP/IP tab. 3.) Type Computer Name 4.) Verify that Enable DHCP has no check mark. If checked, remove the mark. NOTE: If Enable DHCP mark is checked, the system cannot checkout by VOLC. 5.) Check if Computer Name, IP Address, Subnet Mask, and Default Gateway are proper ones which you wrote down in section 8-3-5-1Saving Connectivity. NOTE: If the “Save” is NOT performed, you can NOT enter Patient screen and NOT perform Export/Import 6.) Click on Save button 7.) Click OK for confirmation dialog box 8.) The system will powe r OFF automatically. 9.) Turn on the system for functional checks Figure 8-203 TCPIP tab NOTICEThe Computer name is unique. Check the scanner se rial number printed on the label located at the rear lower side of the scanner. For example, wh en 12345SU1 is printed, P5-12345SU1 must be entered as a computer name. NOTICEIf Enable DHCP checkbox is checked and LAN cable is disconnected from the system, the system bootup time might slow down.
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 146 Section 8-3 - Software Loading Procedure 8-4-4-3 Setting Printer Preferences NOTE: If the 1-3-5 printer setting back-up is NOT performed, i gnore this section. 1.) Insert removable disk or media that the printer setting was saved. 2.) Go to Utility -> System -> Peripherals click the Printers. 3.) Press the left button of the trackball on the pr inter which need to be restore from the saved printer setting -> Click “properties” -> Click “Printing Preferences”. 4.) If Sony D897 is used, click “Read”. Figure 8-204 Printer set restore Figure 8-205 Printer set restore
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-3 - Software Loading Procedure 8 - 147 8-4-4-3 Setting Printer Preferences (cont’d) 5.) Select removable disk or media that the printer setting was saved -> Click “Open”. 6.) Click the “OK” until all printer windows are closed. 7.) If Sony D895 is used, repeat the above step 1 an d 3, then set the parameters below in the layout tab and advanced tab. Figure 8-206 Printer set restore ParametersDescriptions Orientation Portrait or Landscape Paper size Paper size Scaling Rate of length and width Table 8-56 Parameters
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 148 Section 8-3 - Software Loading Procedure 8-4-4-3 Setting Printer Preferences (cont’d) 8.) Click the properties of the Density Adjust”. 9.) Click “Load”. 10.) Select removable disk or media which the printer setting was saved -> click “Open”. 11.) Click the “OK” until all printer windows are closed. Figure 8-207 Printer set restore Figure 8-208 Printer set restore
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-3 - Software Loading Procedure 8 - 149 8-4-4-3 Setting Printer Preferences (cont’d) 12.) If the printer is not support restore option, restore the printer set from the recorded parameters. NOTE: The parameters depend on printers, so the above parameters are shown to another names and more parameters setting could be needed. ParametersDescriptions Orientation Portrait or Landscape Paper size Paper size Scaling Rate of length and width Density AdjustGamma curve Table 8-57 Parameters
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 150 Section 8-3 - Software Loading Procedure 8-4-4-4 Functional Checks for Service Platform (Diagnostic Program) The Service platform uses a web-based user interface (UI) to provide access to common service components. NOTE: After perform diagnostics, shut down and restart to make the system working properly. 1.) Go to Utility and press Service to launch Service Platform. It will take some seconds for activating. 2.) Make sure that CAPS is not selected (should be dimmed) for password entry performed later. 3.) The Service Login window for Service Pla tform will be shown on the monitor display. Figure 8-209 Utility -> Service Figure 8-210 Log in screen CAUTIONIf the Login window for Service Platform is not displayed on the monitor, the installation of the Service Platform has failed. Reload the application software or Base System Software (OS) + application software. Contact a Technical Support for details.
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-3 - Software Loading Procedure 8 - 151 8-4-4-4 Functional Checks for Service Platform (Diagnostic Program) (cont’d) 4.) Select GE Service at the “Select User Level” field. 5.) Enter the password for the Service Platform. 6.) Click on Okay. 7.) Verify that the following screen (Servi ce Platform) is displayed on the monitor. 8.) Click on x located at the upper right corner of the service platform screen to close the Service Platform and return to the scan panel.Figure 8-211 Select User Level NOTICEWhen the service platform is NOT displayed, check if CAPS lock is selected. The CAPS should not be selected. Figure 8-212 Service Platform screen
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 152 Section 8-3 - Software Loading Procedure 8-4-4-5 Setting the VCR (if necessary) 1.) Select Utility > System > Peripheral tab. 2.) Under Video Settings, PAL or NTSC. 3.) Under VCR, select Composite or S-Video. 4.) Under VCR, select Panasonic LQ-MD800 or Mitsubishi MD3000, as appropriate. 5.) Click on Save. 6.) Shut down the scanner, then restart the scanner. 7.) Verify that the VCR c an be properly operated. 8.) Perform the following functional tests. If all are successful, include the debrief script provided below. Figure 8-213 Utility > System>Peripherals NOTICEBe sure to connect the converter and cable before booting up the system. Otherwise the LOGIQ™ P5 system does not recognize the peripheral devices Table 8-58 Functional Tests Service Manual Section Functional Test / Diagnostic TestDebrief Script Section 4-3-1 Power On/Boot Up “Service Manual, Direction 5394141, Rev 1+, Section 8-4-4. Equipment passed all required tests and is ready for use. “ Section 4-3-2 Power Off / Shutdown Section 4-5 Peripheral Checks
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-3 - Software Loading Procedure 8 - 153 8-4-4-6 Verifying the VCR (if necessary) 1.) Press Record button, VCR menu shows up at the bottom of the screen. 2.) Click Counter Reset, then Press Record menu to start recording. Displays a red dot in the VCR status area on the Title bar to indicate that recording has begun. 3.) Press Stop to stop recording the tape at the desired moment. 4.) Press Search to adjust the st arting time for video playback. 5.) Press Play to start VCR playback. Figure 8-214 VCR Menu Figure 8-215 Record Menu Figure 8-216 VCR Status
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 154 Section 8-3 - Software Loading Procedure 8-4-4-7 Setting Printer Properties in Application Software (if necessary) 1.) Select Utility -> Connectivity -> Service tab 2.) Select Standard Printer, and Add 3.) Under Properties, select desired Printer, and Save. Figure 8-217 Utility > Connectivity > Service Figure 8-218 Select the Printer