    Section 8-3 - Software Loading Procedure 8 - 145
    8-4-4-2 Setting Computer Name 1.) Click on Utility, then  select Connectivity.
    2.) Click on TCP/IP tab. 
    3.) Type  Computer Name
    4.) Verify that  Enable DHCP  has no check mark. If checked, remove the mark.
    NOTE: If Enable DHCP mark is checked,  the system cannot checkout by VOLC.
    5.) Check if  Computer Name, IP Address, Subnet Mask, and Default Gateway are proper ones 
    which you wrote down in section 8-3-5-1Saving Connectivity. 
    NOTE: If the “Save” is NOT performed, you can NOT enter Patient screen and NOT perform  Export/Import 
    6.) Click on Save button 
    7.) Click OK for confirmation dialog box 
    8.) The system will powe r OFF automatically.
    9.) Turn on the system for functional checks Figure 8-203   TCPIP tab
    NOTICEThe Computer name is unique. Check the scanner se
    rial number printed on the label located at the 
    rear lower side of the scanner. For example, wh en 12345SU1 is printed, P5-12345SU1 must be 
    entered as a computer name.
    NOTICEIf  Enable DHCP  checkbox is checked and LAN cable is  disconnected from the system, the system 
    bootup time might slow down. 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    8 - 146 Section 8-3 - Software Loading Procedure
    8-4-4-3 Setting Printer Preferences NOTE: If the 1-3-5 printer setting back-up is NOT performed, i gnore this section.
    1.) Insert removable disk or media that the printer setting was saved.
    2.) Go to Utility -> System ->  Peripherals click the Printers.
    3.) Press the left button of the trackball on the pr inter which need to be restore from the saved printer 
    setting -> Click “properties” ->  Click “Printing Preferences”.
    4.) If Sony D897 is used, click “Read”. Figure 8-204   Printer set restore
    Figure 8-205   Printer set restore 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
    Section 8-3 - Software Loading Procedure 8 - 147
    8-4-4-3              Setting Printer Preferences  (cont’d)
    5.) Select removable disk or media that the  printer setting was saved -> Click “Open”.
    6.) Click the “OK” until all  printer windows are closed.
    7.) If Sony D895 is used, repeat the above step 1 an d 3, then set the parameters below in the layout 
    tab and advanced tab. Figure 8-206   Printer set restore
    ParametersDescriptions
    Orientation
    Portrait or Landscape
    Paper size Paper size
    Scaling Rate of length and width
    Table 8-56 Parameters 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    8 - 148 Section 8-3 - Software Loading Procedure
    8-4-4-3              Setting Printer Preferences  (cont’d)
    8.) Click the properties  of the Density Adjust”.
    9.) Click “Load”.
    10.) Select removable disk or media which the  printer setting was saved -> click “Open”.
    11.) Click the “OK” until all  printer windows are closed.
    Figure 8-207   Printer set restore
    Figure 8-208   Printer set restore 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
    Section 8-3 - Software Loading Procedure 8 - 149
    8-4-4-3              Setting Printer Preferences  (cont’d)
    12.) If the printer is not support restore option, restore the printer set from the recorded parameters.
    NOTE: The parameters depend on printers, so the  above parameters are shown to another names and 
    more parameters setting could be needed.
    ParametersDescriptions
    Orientation Portrait or Landscape
    Paper size Paper size
    Scaling Rate of length and width
    Density AdjustGamma curve
    Table 8-57 Parameters 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    8 - 150 Section 8-3 - Software Loading Procedure
    8-4-4-4 Functional Checks for Service Platform (Diagnostic Program) The Service platform uses a web-based user interface (UI) to provide access to common service 
    components.
    NOTE: After perform diagnostics,  shut down and restart to make the system working properly. 
    1.) Go to Utility and press Service to launch Service  Platform. It will take some seconds for activating.
    2.) Make sure that  CAPS is not selected (should be dimmed)
     for password entry performed later.
    3.) The Service Login window for Service Pla tform will be shown on the monitor display.
    Figure 8-209   Utility -> Service
    Figure 8-210   Log in screen
    CAUTIONIf the Login window for Service Platform is not  displayed on the monitor, the installation of the 
    Service Platform has failed. Reload the application software or Base System Software (OS) + 
    application software. Contact a Technical Support for details. 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
    Section 8-3 - Software Loading Procedure 8 - 151
    8-4-4-4              Functional Checks for Service Platform (Diagnostic Program)  (cont’d)
    4.) Select GE Service  at the “Select User Level” field. 
    5.) Enter the password for the Service Platform.
    6.) Click on  Okay.
    7.) Verify that the following screen (Servi ce Platform) is displayed on the monitor.
    8.) Click on  x located at the upper right corner of the  service platform screen to close the Service 
    Platform and return  to the scan panel.Figure 8-211   Select User Level
    NOTICEWhen the service platform is
     NOT displayed, check if CAPS lock is selected. The  CAPS should not be 
    selected.
    Figure 8-212   Service Platform screen 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    8 - 152 Section 8-3 - Software Loading Procedure
    8-4-4-5 Setting the VCR (if necessary) 1.) Select Utility > System > Peripheral tab.
    2.) Under Video Settings, PAL or NTSC.
    3.) Under VCR, select Composite or S-Video.
    4.) Under VCR, select Panasonic LQ-MD800  or Mitsubishi MD3000, as appropriate.
    5.) Click on Save.
    6.) Shut down the scanner, then restart the scanner.
    7.) Verify that the VCR c an be properly operated.
    8.) Perform the following functional tests. If all are successful, include the debrief script provided below. Figure 8-213   Utility
     > System>Peripherals 
    NOTICEBe sure to connect the converter and cable before  booting up the system. Otherwise the LOGIQ™ P5 
    system does not recognize the peripheral devices
    Table 8-58    Functional Tests
    Service Manual Section
    Functional Test / Diagnostic TestDebrief Script
    Section 4-3-1 Power On/Boot Up
    “Service Manual, Direction
    5394141, Rev 1+, Section 8-4-4. Equipment 
    passed all required tests and is ready for use. “
    Section 4-3-2
    Power Off / Shutdown
    Section 4-5 Peripheral Checks 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
    Section 8-3 - Software Loading Procedure 8 - 153
    8-4-4-6 Verifying the VCR (if necessary) 1.) Press Record button, VCR menu shows up at the bottom of the screen.
    2.) Click Counter Reset, then Press  Record menu to start recording.
    Displays a red dot in the VCR status area on the Title bar to indicate that recording has begun.
    3.) Press Stop to stop recording the tape at the desired moment.
    4.) Press Search to adjust the st arting time for video playback.
    5.) Press Play to start VCR playback. Figure 8-214   VCR Menu
    Figure 8-215   Record Menu
    Figure 8-216   VCR Status 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    8 - 154 Section 8-3 - Software Loading Procedure
    8-4-4-7 Setting Printer Properties in Application Software (if necessary) 1.) Select Utility -> Connectivity -> Service tab
    2.) Select Standard Printer, and Add
    3.) Under Properties, select desired Printer, and Save. Figure 8-217   Utility > Connectivity > Service