    Section 8-3 - Software Loading Procedure 8 - 205
    8-7-15 Rear Handle installation
    1.) Unscrew 2 screws (1-2) and remove Rear hook.
    2.) Unscrew 2 screws (1-2) and remove the rear handle bracket. Figure 8-318   Rear Handle Installation
    Figure 8-319   Rear Handle Installation 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    8 - 206 Section 8-3 - Software Loading Procedure
    8-7-15               Rear Handle installation (cont’d)
    3.) Screw 2 screws(5172296, HSH M8x25) from the back of the rear handle.
    4.) Screw 1 screw(5178673, FH M4x6 WHT) and Install handle dummy bracket.  Figure 8-320   Rear Handle Installation
    Figure 8-321   Rear Handle Installation
    (1) 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
    Section 8-3 - Software Loading Procedure 8 - 207
    8-7-16 Additional Task Lamp installation
    1.) This is the installation inst ruction of additional task lamp.
    2.) Connect the USB connector to the location of light bulb mark.
    3.) Install plastic holder to the task lamp bracket. Figure 8-322   Additional Task Lamp Installation
    Figure 8-323   Additional Task Lamp Installation
    Figure 8-324   Additional Task Lamp Installation 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    8 - 208 Section 8-3 - Software Loading Procedure
    8-7-16               Additional Task Lamp installation (cont’d)
    4.) Install task lamp bracket to the rear side of  keyboard by screwing 2 screws(5178673, FH M4x6 
    WHT).
    5.) Fix the task lamp light as shown on the figure below. 
    6.) Install BW hide bracket if it is unin stalled during the installation process.
    Figure 8-325   Additional Task Lamp Installation
    Figure 8-326   Additional Task Lamp Installation
    Figure 8-327   Additional Task Lamp Installation 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
    Section 8-3 - Software Loading Procedure 8 - 209
    8-7-17 Protect Cover Installation
    1.) Refer to the figure below the location  of the protect covers to be installed. 
    2.) Install the protect cover using the double stick tape. Figure 8-328   Protect Cover Installation
    Figure 8-329   Protect Cover Installation
    Front Protect Cover
    Rear Protect Cover 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    8 - 210 Section 8-3 - Software Loading Procedure
    8-7-18 Swivel Lock Caster Installation
    NOTE: Both Front Casters and Rear Casters are swivel  type in LOGIQ™ P5 but in LOGIQ™ A5, only 
    Front Casters are swivel type (Rear Casters are “Fixed” type).
    1.) This is the installation instruction of the Protect  Cover. Refer to the figure below for the location of 
    the protect covers to be installed.
    2.) Lift the system up front and fix  it not to move during installation.
    3.) Unscrew 8 screws (1A-8A) to remove the existing  front casters. Refer to the figure below. keep the 
    screws for use.
    4.) Replace them to Swivel Lock  Casters by screwing 8 screws (1A-8A).
    Figure 8-330   Swivel Lock Caster 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
    Section 8-3 - Software Loading Procedure 8 - 211
    8-7-19 LOGIQ A5 Rear Swiv el Caster Installation
    1.) Refer to the figure below for the location  of the Rear Swivel Caster to be installed.
    2.) Lift the system up front and fix  it not to move during installation.
    3.) Unscrew 8 screws (1-8) to remove the existing fr ont casters. Refer to the figure below. keep the 
    screws for use.
    4.) Replace them to Rear Swivel Casters by screwing 8 screws (1-8).
    Figure 8-331   Rear Swivel Caster 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    8 - 212 Section 8-3 - Software Loading Procedure
    8-7-20 BT11 Drawer  Option Installation
    1.) This section describes the installa tion procedure of the drawer option.
    2.) Assemble the BW PRINTER to DRAWER FIX BRKT  with screwing 4 screws (2306562, FH M3X6).
    3.) Assemble the BT11 DRAWER to DRAWER FIX BRKT using 2 hooks. Figure 8-332   BT11 Drawer Option Installation
    Figure 8-333   BT11 Drawer Option Installation 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
    Section 8-3 - Software Loading Procedure 8 - 213
    8-7-20      BT11 Draw er Option Installation (cont’d)
    4.) Screw 2 screws (2159634, BH M4X10) behind the Assembly.
    5.) Screw 2 screws (5176744, BH M4X20) to fix the DRAWER Assy to the system. Figure 8-334   BT11 Drawer Option Installation
    Figure 8-335   BT11 Drawer Option Installation 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    8 - 214 Section 8-3 - Software Loading Procedure
    8-7-20      BT11 Draw er Option Installation (cont’d)
    6.) Connect the BW Printer cable and fix the cables  with CABLE MOUNT. Refer to the pictures below.
    7.) Installation Completion. Figure 8-336   BT11 Drawer Option Installation