GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL Section 3-1 - Overview 3-1 Chapter 3 Installation Section 3-1 Overview 3-1-1 Purpose of Chapter 3 This chapter contains information needed to install the unit. Included are references to a procedure that describes how to receive and unpack the equipment and how to file a damage or loss claim. How to prepare the facility and unit of the actual installation, and how to check a nd test the unit, probes, and external peripherals for electrical safety are incl uded in this procedure. Also included in this section are guidelines for transporting the unit to a new site. Table 3-1 Contents in Chapter 3 SectionDescriptionPage Number 3-1 Overview 3-1 3-2 Installation Reminders 3-2 3-3 Receiving and Unpacking the Equipment 3-4 3-4 Preparing for Installation 3-10 3-5 Completing the Installation 3-11 3-6 System Configuration 3-16 3-7 Available Probes 3-23 3-8 Software/Option Configuration 3-23 3-9 Connectivity Installation Worksheet 3-24 3-10 Insite IP Address Configuration 3-25 3-11 Loading Base System Software 3-25 3-12 Warnings 3-25 3-13 Paperwork 3-26
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 3-2 Section 3-1 - Overview Section 3-2 Installation Reminders 3-2-1 Average Installation Time The LOGIQ™ P5/A5/A5Pro has been designed to be installed and checked out by an experienced service technician in approximately four hours. LOGIQ™ P5/A5/A5Pro consoles with optional equipment may take slightly longer. 3-2-2 Installation Warnings 1.) Since the LOGIQ™ P5/A5/A5Pro weighs approximat ely 75 kg.(165 lb.) without options, preferably two people should unpack it. Two people are also pref erable for installing any additional bulky items. 2.) There are no operator serviceable components. To prevent shock, do not remove any covers or panels. Should problems or malfunctions occur, unplug the power cord. Only qualified service personnel should carry out servicing and troubleshooting. NOTE: For information regarding packing labels, refer to LABELS ON PACKAGE. 3.) After being transported, the unit may be very co ld or hot. If this is the case, allow the unit to acclimate before you turn it on. It requires one hour for each 2.5 °C increment its temperature is below 10 °C or above 30 °C. Table 3-2 Average Installation Time DescriptionAverage Installation TimeComments Unpacking the scanner 0.5 hour Scanner wo/options 0.5 hour Dependant on the configuration that is required DICOM Option 0.5 hour Dependant on the amount of configuration InSite Option0.5 hour CAUTIONEquipment damage possibility. Turning the system on without acclimation after arriving at site may cause the system to be damaged. Table 3-3 Acclimation Time ° C60 55 50 45 40 35 30 25 20 15 10 5 0 -5 -10 -15 -20 -25 -30 -35 -40 °F140 131 122 113 104 96 86 77 68 59 50 41 32 23 14 5 -4 -13 -22 -31 -40 hrs864200000002468101214161820
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL Section 3-1 - Overview 3-3 3-2-3 Safety Reminders DANGERWHEN USING ANY TEST INSTRUMENT THAT IS CAPABLE OF OPENING THE AC GROUND LINE (I.E., METER’S GROU N D S W I T C H I S OPEN), DON’T TOUCH THE UNIT! CAUTIONTwo people should unpack the unit because of its weight. Two people are required whenever a part weighing 19kg (35 lb.) or more must be lifted. CAUTIONIf the unit is very cold or hot, do not turn on its power until it has had a chance to acclimate to its operating environment. CAUTIONTo prevent electrical shock, connect the unit to a properly grounded power outlet. Do not use a three to two prong adapter. This defeats safety grounding. CAUTIONDo NOT wear the ESD wrist strap when you work on live circuits and more than 30 V peak is present. CAUTIONDo not use a 20 Amp to 15 Amp adapter on the 120 Vac unit’s power cord. This unit requires a dedicated 20 A circuit and can have a 15A plug if the on board peripherals do not cause the unit to draw more than 14.0 amps. CAUTIONDo not operate this unit unless all board cove rs and frame panels are securely in place. System performance and cooling require this. CAUTIONOPERATOR MANUAL(S) The User Manual(s) should be fully read and understood before operating the LOGIQ™ P5/ A5/A5Pro and kept near the unit for quick reference. CAUTIONACOUSTIC OUTPUT HAZARD Although the ultrasound energy transmitted from the LOGIQ™ P5/A5/A5Pro probe is within FDA limits, avoid unnecessary exposure. Ultrasound energy can produce heat and mechanical damage. Figure 3-1 Environmental Labels
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 3-4 Section 3-1 - Overview Section 3-3 Receiving and Unpacking the Equipment When a new system arrives, check that any components are not damaged and are not in short supply. If shipping damage or shortage occurs, contact the address shown in Chapter 1. 1.) Cut the three PLASTIC BANDs. Figure 3-2 Cutting the Plastic Bands. CAUTIONDo not lift the unit by the Keyboard. Equipment damage may result. CAUTIONThe crate with the LOGIQ™ P5/A5/A5Pro weighs approximately 75 kg. (165 lb.) Be prepared for a sudden shift of weight as the unit is removed from its base (pallet) Plastic Bands
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL Section 3-1 - Overview 3-5 Section 3-3 Receiving and Unpacking the Equipment (cont’d) 2.) Lift the TOP COVER up and off. Figure 3-3 Removing the top Cover Top Cover
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 3-6 Section 3-1 - Overview Section 3-3 Receiving and Unpacking the Equipment (cont’d) 3.) Remove the Top Plate up and off.. Figure 3-4 Removing the Top Plate Top Plate
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL Section 3-1 - Overview 3-7 Section 3-3 Receiving and Unpacking the Equipment (cont’d) 4.) Remove the three PLASTIC JOINTs from the OUTER SLEEVE. 5.) Remove the OUTER SLEEVE. Figure 3-5 Removing Plastic Joints and Sleeves Plastic Joint Outer Sleeve Palette Assy
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 3-8 Section 3-1 - Overview Section 3-3 Receiving and Unpacking the Equipment (cont’d) 6.) Remove the MONITOR SUPPORTER and Option Box. 7.) Remove the PLASTIC BAG. 8.) Unlock the front caster and carefu lly put the console off the PALETTE. Figure 3-6 Moving the LOGIQ™ P5/A5/A5Pro from the crate Monitor Pad
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL Section 3-1 - Overview 3-9 3-3-1 Moving into Position In general, a single adult can move the LOGIQ™ P5/A5/A5Pro along an even surface with no steep grades. At least two people should move the machine when la rge humps, grooves, or grades will be encountered. (It is better to pull from the rear rather than push from the front of the unit). Before moving, store all loose parts in the unit. Wrap transducers in soft cloth or foam to prevent damage. Although LOGIQ™ P5/A5/A5Pro is a compact and mob ile machine, two people should move it over rough surfaces or up and down grades. 3-3-2 Adjusting System Clock Set the system clock for the LOGIQ™ P5/A5/A5Pro to the local time. For procedure of adjusting the system clock, refer to section 3-6-1-1 on page 16. Fill out proper customer Information the Product Locator Installation Card. Mail this Installation Card “Product Locator” to the address corresponding to your pole. 3-3-3 Product Locator Installation Card NOTE: The Product Locator Installation Card shown may not be same as the provided Product Locator card. CAUTIONDo not lift the unit by the Keyboard. Do not tilt the unit more than 5 degrees to avoid tipping it over. To avoid injury by tipping over. Set the monitor to the lowest position before moving. CAUTIONEquipment Damage Possibility. Lifting the cons ole by holding covers may damage the covers. Do not lift the console by holding any covers. Figure 3-7 Product Locator Installation Card
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 3-10 Section 3-4 - Preparing for Installation Section 3-4 Preparing for Installation 3-4-1 Verify Customer Order Compare items received by the customer to that whic h is listed on the delivery order. Report any items that are missing, back ordered or damaged. 3-4-2 Physical Inspection 3-4-2-1 System Voltage Settings Verify that the scanner is set to the correct voltage. The Voltage settings for the LOGIQ™ P5/A5/A5Pro Scanner is found on a label to the right of the Power switch and External I/O, on the rear of the system. 3-4-2-2 Video Formats Check that the video format for VCR playback is set to the locally used video standard, NTSC or PAL. 3-4-3 EMI Protection This Unit has been designed to minimize the effects of Electro Magnetic Interference (EMI). Many of the covers, shields, and screws are provided primarily to protect the system from image artifacts caused by this interference. For this reason, it is imperative that all covers and hardware are installed and secured before the unit is put into operation. WARNINGWARNINGConnecting a LOGIQ™ P5/A5/A5Pro scanner to the wrong voltage level will most likely destroy the scanner.