GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-3 - Software Loading Procedure 8 - 135 8-4-2-1 Partition Checks (cont’d) 4.) Select Storage -> Disk Management and check that only File System, Status, Capacity and Drive Letter of each partition are same as follows NOTE: If 160GB HDD is used, refer to the Figure 8-191. NOTE: Repo partition and CD/DVD-ROM Drives Drive Letter should be assigned as Z:, G: respectively. Figure 8-191 Disk Management 160G HDD
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 136 Section 8-3 - Software Loading Procedure 8-4-2-2 Driver Checks 1.) On the window XP desktop, left click on My Computer then right click on Properties. 2.) The System Properties window appears. Select Hardware > Device Manager. 3.) Verify that ?, X, or ! is not displayed • If X or ! is displayed: - Left-click on the device with X mark, then right -click on Enable - Left-click on the device with ! mark, then right click on Properties to perform the device driver recognition. When some files are required, designate the file in the directories of C:\WINNT\SYSTEM32 or C: \WINNT\SYSTEM32\DRIVERS . If they are not found, search them using a window search function. Figure 8-192 Device Manager
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-3 - Software Loading Procedure 8 - 137 8-4-2-3 USB Printer Checks (USB Printer User Only) Without any newly additional user in vention, LOGIQ™ P5 will recognize th e below 19 kinds of USB printer This procedure assumes that the P5 software is installed and is functioning properly. 1.) Attach the USB cable from the printer to the USB port on assigned position Printer NameComment Sony UP-D895/D895MD Digital BW Printer Sony UP-D897 Mitsubishi P93D Sony UP-D23MD Digital Color Printer Sony UP-D25MD Mitsubishi CP30DW HP 990cxi Standard Printer LexMark Z65 HP 5550 HP 6122 HP 6540 HP 5650 HP 1000 HP 1200 Xerox Phaser 8400 HP Office jet Pro K550 HP Office jet Pro K5400 HP Deskjet D4360 HP Deskjet 6940 HP Officejet 6000 Special Edition Table 8-55 USB Port Position USB Port Position Peripheral Right USB Port on Rear Connector Panel Standard Printer Left USB Port on Rear Connector Panel: Digital Color Printer USB Port on OP Prin ter Bracket Back Pane lDigital BW Printer NOTICEIf you plug in USB printer for the first time to the system, restart the system.
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 138 Section 8-3 - Software Loading Procedure 8-4-2-3 USB Printer Checks (USB Printer User Only) (cont’d) 2.) Make sure that printer properties are proper. (I mage orientation and size). Refer to the latest Release Notes. NOTICEOnce a USB printer is connected to one of the USB ports, dont take out the USB printer and plug it in another USB port. Figure 8-193 USB Port - Rear Panel Connector Figure 8-194 USB Port - OP Printer Bracket Back Panel
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-3 - Software Loading Procedure 8 - 139 8-4-2-4 DMC Checks (DMC User Only) NOTE: DMC Cable is supplied only as an option. 1.) Check DMC Cable (USB to Serial Bridge). 2.) Plug in the Cable as shown on the right side of the USB port. 3.) Ensure USB to Serial Bridge port is recognized by the system. 4.) Trackball to My computer icon and left click 5.) Right click on Properties. 6.) Right click on the Hardware tab. 7.) Right click on the device manager in the middle right of the hardware tab window. Figure 8-195 DMC Cable DMC Cable Plug on the right
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 140 Section 8-3 - Software Loading Procedure 8-4-2-4 DMC Checks (DMC User Only) (cont’d) 8.) Make sure“USB to Serial Bridge” is displayed as shown the Figure 8-196 Figure 8-196 Device Manager Windows
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-3 - Software Loading Procedure 8 - 141 8-4-3 Installing Application Software NOTE: First time SW installation with peripherals on can cause the BlueScreen. NOTE: When connecting peripherals to the system, make sure to turn off the system first in order to avoid the BlueScreen. 1.) Place the “Application Soft ware” CD into the DVD drive. 2.) Press the Start button on the tool bar at the bottom of the LCD screen and select “Run”. Enter “G:\LoadSoftware” into the dialog box as shown below. 3.) You will see a screen that warns you that you are about to load software, press the Y key to continue. Then the program requests confirma tion “Are you sure?”, Press the Y key again. 4.) Select the application type to install. 5.) You will see a Comm and (CMD) window open as shown in the Figure 8-198. 6.) Press the “Y” key twice and copying back up files starts. This takes a few minutes. WARNINGWARNINGBase Image software should also be installed prior to the application. Figure 8-197 Run Load Software Figure 8-198 Command Window
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 142 Section 8-3 - Software Loading Procedure 8-4-3 Installing Application Software (cont’d) 7.) After copying backup files, the following options appear. 8.) elect a product to be installed. 1.) LOGIQ P5 [R4.0.1] is only for LOGIQ P5 BT11 product. 2.) LOGIQ P5 [R3.0.4] is only for LOGIQ P5 BT09 product. 3.) LOGIQ A5 [R3.0.4] is only for LOGIQ A5 BT09 product. 4.) LOGIQ A5 Pro [R3.0.4] is only for LOGIQ A5 PRO BT09 product. 9.) Then, press the “Y” key to continue. Then the progr am requests confirmation - “Are you sure?” then Press the “Y” key again. 10.) The installation process will be st arted. it will take about 20 minutes. 11.) To apply the preset by region, select one of the below options. Figure 8-199 Select system Figure 8-200 Region selection
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-3 - Software Loading Procedure 8 - 143 8-4-3 Installing Application Software (cont’d) 12.) After the installa tion is completed the system will power off automatically. NOTE: Some of systems doesn’t power off automatica lly. In this case, press and hold power button more than 5 sec. to power off. NOTE: Do not operate with using mouse or keyboard during installation procedures. 13.) Turn on the system and immediately eject the bu tton on the DVD drive to eject the DVDROM Verify that the system boots up no error.(It will take a few minutes) NOTE: The application installation is completed, reboot is required. 14.) Shut down the system. 15.) Start the system Figure 8-201 Instal lation Complete Message
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 144 Section 8-3 - Software Loading Procedure 8-4-4 Standard Confirma tion & Functional Checks Configure the system settings and perform functi onal checks after installation of software. Items for configuration and functional checks include: - Confirmation of the software version - Preset Data (if necessary) - Patient Image Data (if necessary) - Functional Checks for Service Platform (Diagnostic Program) 8-4-4-1 Confirming software version 1.) Touch the Utility bu tton on the Touch Panel 2.) Click on About tab. The software version will be shown on the monitor as shown. Confirm the following: Refer to the Figure 8-202. 3.) Make sure that Application Software Version (R4.x.x), Software Part Number (described on the Application Software CD-ROM) and Image Part Number (described on the Base Image Software CD-ROM). Figure 8-202 Confirmation of software version