    Section 8-3 - Software Loading Procedure 8 - 135
    8-4-2-1              Partition Checks  (cont’d)
    4.) Select Storage -> Disk Management  and check that only File System, Status, Capacity  and 
    Drive Letter  of each partition are same as follows
    NOTE: If 160GB HDD is used, refer to the Figure 8-191.
    NOTE: Repo partition and  CD/DVD-ROM Drives  Drive Letter should be assigned as Z:, G: 
    respectively. Figure 8-191   Disk Management 160G HDD 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    8 - 136 Section 8-3 - Software Loading Procedure
    8-4-2-2 Driver Checks 1.) On the window XP desktop, left click on My Computer then right click on Properties.
    2.) The System Properties window appears. Select  Hardware > Device Manager.
    3.) Verify that ?, X, or ! is not displayed
    • If X or ! is displayed: - Left-click on the device with X mark, then right -click on  Enable
    - Left-click on the device with ! mark, then right click on  Properties to perform the device driver 
    recognition. When some files are required,  designate the file in the directories of 
    C:\WINNT\SYSTEM32 or C: \WINNT\SYSTEM32\DRIVERS . If they are not found, search 
    them using a window search function. Figure 8-192   Device Manager 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
    Section 8-3 - Software Loading Procedure 8 - 137
    8-4-2-3 USB Printer Checks (USB Printer User Only) Without any newly additional user in vention, LOGIQ™ P5 will recognize th e below 19 kinds of USB printer 
    This procedure assumes that the P5 software  is installed and is functioning properly.
    1.) Attach the USB cable from the printer to the USB port on assigned position
    Printer NameComment
    Sony UP-D895/D895MD Digital BW Printer
    Sony UP-D897
    Mitsubishi P93D
    Sony UP-D23MD
    Digital Color Printer
    Sony UP-D25MD
    Mitsubishi CP30DW
    HP 990cxi
    Standard Printer
    LexMark Z65
    HP 5550
    HP 6122
    HP 6540
    HP 5650 
    HP 1000
    HP 1200
    Xerox Phaser 8400
    HP Office jet Pro K550
    HP Office jet Pro K5400
    HP Deskjet D4360
    HP Deskjet 6940
    HP Officejet 6000 Special  Edition
    Table 8-55    USB Port Position
    USB Port Position Peripheral
    Right USB Port on Rear Connector Panel Standard Printer
    Left USB Port on Rear Connector Panel: Digital Color Printer
    USB Port on OP Prin ter Bracket Back Pane lDigital BW Printer
    NOTICEIf you plug in USB printer for the first  time to the system, restart the system. 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    8 - 138 Section 8-3 - Software Loading Procedure
    8-4-2-3              USB Printer Checks (USB Printer User Only)  (cont’d)
    2.) Make sure that printer properties are proper. (I mage orientation and size). Refer to the latest 
    Release Notes.
    NOTICEOnce a USB printer is connected to one of the USB ports, dont take out the USB printer and plug it in 
    another USB port.
    Figure 8-193   USB Port - Rear Panel Connector
    Figure 8-194   USB Port - OP Printer Bracket Back Panel 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
    Section 8-3 - Software Loading Procedure 8 - 139
    8-4-2-4 DMC Checks (DMC User Only) NOTE: DMC Cable is supplied only as an option. 1.) Check DMC Cable (USB to Serial Bridge).
    2.) Plug in the Cable as shown on the right side of the USB port.
    3.) Ensure USB to Serial Bridge port  is recognized by the system.
    4.) Trackball to My computer icon and left click
    5.) Right click on  Properties.
    6.) Right click on the  Hardware tab.
    7.) Right click on the device manager in the middle right of the hardware tab window. Figure 8-195   DMC Cable
    DMC Cable 
    Plug on the right 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    8 - 140 Section 8-3 - Software Loading Procedure
    8-4-2-4              DMC Checks (DMC User Only)  (cont’d)
    8.) Make sure“USB to Serial Bridge”  is displayed as shown the Figure 8-196
    Figure 8-196   Device Manager Windows 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
    Section 8-3 - Software Loading Procedure 8 - 141
    8-4-3 Installing Application Software
    NOTE: First time SW installation with peripherals on can cause the BlueScreen.
    NOTE: When connecting peripherals to the system, make sure to turn off the system first in order to avoid the  BlueScreen.
    1.) Place the “Application Soft ware” CD into the DVD drive.
    2.) Press the Start button on the tool bar at  the bottom of the LCD screen and select “Run”.
    Enter “G:\LoadSoftware” into the dialog box as shown below.
    3.) You will see a screen that warns you that you are about to load software, press the Y key to  continue. Then the program requests confirma tion “Are you sure?”, Press the Y key again.
    4.) Select the application type to install.
    5.) You will see a Comm and (CMD) window open as shown in the Figure 8-198.
    6.) Press the “Y” key twice and copying back up files starts. This takes a few minutes.
    WARNINGWARNINGBase Image software should also be installed prior to the application.
    Figure 8-197   Run Load Software
    Figure 8-198    Command Window 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    8 - 142 Section 8-3 - Software Loading Procedure
    8-4-3               Installing Application Software (cont’d)
    7.) After copying backup files, the following options appear. 
    8.) elect a product to be installed. 1.) LOGIQ P5 [R4.0.1] is only for LOGIQ P5 BT11 product.
    2.) LOGIQ P5 [R3.0.4] is only for LOGIQ P5 BT09 product.
    3.) LOGIQ A5 [R3.0.4] is only for LOGIQ A5 BT09 product.
    4.) LOGIQ A5 Pro [R3.0.4] is only  for LOGIQ A5 PRO BT09 product.
    9.) Then, press the “Y” key to continue. Then the progr am requests confirmation - “Are you sure?” then 
    Press the “Y” key again.
    10.) The installation process will be st arted. it will take about 20 minutes.
    11.) To apply the preset by region, select one of the below options. Figure 8-199   Select system
    Figure 8-200   Region selection 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
    Section 8-3 - Software Loading Procedure 8 - 143
    8-4-3               Installing Application Software (cont’d)
    12.) After the installa tion is completed the system  will power off automatically.
    NOTE: Some of systems doesn’t power off automatica lly. In this case, press and hold power button 
    more than 5 sec. to power off.
    NOTE: Do not operate with using mouse or keyboard during installation procedures. 13.) Turn on the system and immediately eject the bu tton on the DVD drive to eject the DVDROM Verify 
    that the system boots up no  error.(It will take a few minutes)
    NOTE: The application installation is  completed, reboot is required.
    14.) Shut down the system.
    15.) Start the system Figure 8-201   Instal
    lation Complete Message 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    8 - 144 Section 8-3 - Software Loading Procedure
    8-4-4 Standard Confirma tion & Functional Checks
    Configure the system settings and perform functi onal checks after installation of software.
    Items for configuration and functional checks include: - Confirmation of the software version
    - Preset Data (if necessary)
    - Patient Image Data (if necessary)
    - Functional Checks for Service Platform (Diagnostic Program)
    8-4-4-1 Confirming software version 1.) Touch the Utility bu tton on the Touch Panel
    2.) Click on About tab. The software version will be  shown on the monitor as shown. Confirm the 
    following: Refer to the Figure 8-202.
    3.) Make sure that Application Software Version (R4.x.x), Software Part Number (described on the  Application Software CD-ROM) and Image Part Number (described on the Base Image Software 
    CD-ROM). Figure 8-202   Confirmation of software version 
    						