GE Logiq P5 Service Manual
Have a look at the manual GE Logiq P5 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 95 8-2-41 Cup Holder 8-2-41-1 Tools • No need tool 8-2-41-2 Removal procedure 1.) Remove the cup holder. 2.) Perform the following functional tests. If all ar e successful, include the debrief script provided below. 8-2-41-3 Mounting Procedure Install the new parts in the reverse order of removal. Figure 8-131 Cup holder Table 8-42 Functional Tests Service Manual Section Functional Test / Diagnostic TestDebrief Script Section 4-3-1 Power On/Boot Up “Service Manual, Direction 5394141, Rev 1+, Section 8-2-41. Equipment passed all required tests and is ready for use. “ Section 4-3-2 Power Off / Shutdown Section 4-8-1 Cover Parts Function Validation Section 10-5-5Physical Inspection
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 96 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-2-42 ECG Cable Hook 8-2-42-1 Tools • Common pilIips screwdrivers 8-2-42-2 Removal procedure 1.) Unscrew 2 screws (1-2) to remove the ECG Cable Hook. 2.) Perform the following functional tests. If all are successful, include the debrief script provided below. 8-2-42-3 Mounting Procedure Install the new parts in the reverse order of removal. Figure 8-132 Screws of cable hook Table 8-43 Functional Tests Service Manual Section Functional Test / Diagnostic TestDebrief Script Section 4-3-1 Power On/Boot Up “Service Manual, Direction 5394141, Rev 1+, Section 8-2-42. Equipment passed all required tests and is ready for use. “ Section 4-3-2 Power Off / Shutdown Section 4-8-1 Cover Parts Function Validation Section 10-5-5Physical Inspection (1) (2)
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 97 8-2-43 Air Filter Set 8-2-43-1 Tools • No need tool 8-2-43-2 Preparations • Shut down the system and switch off the main breaker. 8-2-43-3 Removal procedure 1.) Pull out the 2 Air filters from the bottom left and right side of the system (1-2). 2.) Perform the following functional tests. If all ar e successful, include the debrief script provided below. 8-2-43-4 Mounting Procedure Install the new parts in the reverse order of removal. Figure 8-133 Air filters Table 8-44 Functional Tests Service Manual Section Functional Test / Diagnostic TestDebrief Script Section 4-3-1 Power On/Boot Up “Service Manual, Direction 5394141, Rev 1+, Section 8-2-43. Equipment passed all required tests and is ready for use. “ Section 4-3-2 Power Off / Shutdown Section 4-8-3 Air filter set Function Validation Procedure Section 10-5-5Physical Inspection (1) (2)
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 98 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-2-44 DVD R/W Drive 8-2-44-1 Tools • Common pilIips screwdrivers 8-2-44-2 Preparations • Shut down the system and switch off the main breaker. 8-2-44-3 Removal procedure 1.) Remove the side left cover. Refer to the 8-2-15 Side Left Cover on page 8-51. 2.) Remove EMI L Cover. Refer to the 8-2-26 EMI Cover L on page 8-68. 3.) .) Remove the HDD assy. Refer to the 8-2-47 SATA HDD Assy on page 8-104. 4.) Disconnect the HDD power cable and unsc rew 4 screws of the HDD fixture bracket. 5.) Remove the Front Cover. Refer to the 8-2-17 Front Cover on page 8-53. Figure 8-134 ODD SATA Power Cable and SATA cable of DVD Drive SATA Cable ODD SATA Power CAble
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 99 8-2-44-3 Removal procedure (cont’d) 6.) Unscrew the 4 screws on the front of the DVD drive and remove the DVD Assy. 7.) Unscrew 8 screws to remove the bracket from the DVD Assy. 8.) Perform the following functional tests. If all ar e successful, include the debrief script provided below. 8-2-44-4 Mounting Procedure Install the new parts in the reverse order of removal.Figure 8-135 Unscrew 4screws and remove the DVD Assy Figure 8-136 Unscrew 8screws and remove the DVD Drive from the bracket Table 8-45 Functional Tests Service Manual Section Functional Test / Diagnostic TestDebrief Script Section 4-3-1 Power On/Boot Up “Service Manual, Direction 5394141, Rev 1+, Section 8-2-44. Equipment passed all required tests and is ready for use. “ Section 4-3-2 Power Off / Shutdown
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 100 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-2-45 Nest Fan Assy 8-2-45-1 Tools • Common pilIips screwdrivers 8-2-45-2 Preparations • Shut down the system and switch off the main breaker. 8-2-45-3 Removal procedure 1.) Remove the Left Cover. Refer to the 8-2-15 Side Left Cover on page 8-51. 2.) Remove the EMI Cover Left. Refer to the 8-2-26 EMI Cover L on page 8-68. 3.) Unplug the Connection. 4.) Remove the Right Cover. Refer to the 8-2-16 Side Right Cover on page 8-52. 5.) Remove the EMI Cover Right. Refer to the 8-2-27 EMI Cover R on page 8-69 . 6.) Unscrews 2 screws. Figure 8-137 Power connector of Nest fan Figure 8-138 2 screws on the Nest Fan
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 101 8-2-45-3 Removal procedure (cont’d) 7.) Remove the Nest fan assy. 8.) Perform the following functional tests. If all ar e successful, include the debrief script provided below. 8-2-45-4 Mounting Procedure Install the new parts in the reverse order of removal. Figure 8-139 Pull out the nest fan Table 8-46 Functional Tests Service Manual Section Functional Test / Diagnostic TestDebrief Script Section 4-3-1 Power On/Boot Up “Service Manual, Direction 5394141, Rev 1+, Section 8-2-45. Equipment passed all required tests and is ready for use. “ Section 4-3-2 Power Off / Shutdown Section 4-9-10 Nest fan Function Validation Procedure Section 10-5-5Physical Inspection
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 102 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-2-46 ARP Assy 110V/220V 8-2-46-1 Tools • Common pilIips screwdrivers 8-2-46-2 Preparations • Shut down the system and switch off the main breaker. 8-2-46-3 Removal procedure 1.) Remove side right cover and EMI cover. Refer to the 8-2-16 Side Right Cover on page 8-52 and 8-2-27 EMI Cover R on page 8-69 . 2.) Remove the Rear Cover. Refer to the 8-2-18 Rear Cover on page 8-55. 3.) Unscrew 4 screws (1-4). 4.) Disconnect J15 connector. Figure 8-140 Screws of the rear panel Figure 8-141 AEXP Ca ble of rear panel (1) (3) (4)(2)
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 103 8-2-46-3 Removal procedure (cont’d) 5.) Disconnect three (3) USB connectors or 5 USB connectors from the rear panel and 2 AC power connector. The 4D controller USB and ECG USB co nnector also connect to this rear panel PWA. 6.) Perform the following functional tests. If all ar e successful, include the debrief script provided below. 8-2-46-4 Mounting Procedure Install the new parts in the reverse order of removal. Figure 8-142 Open Rear Panel Table 8-47 Functional Tests Service Manual Section Functional Test / Diagnostic TestDebrief Script Section 4-3-1 Power On/Boot Up “Service Manual, Direction 5394141, Rev 1+, Section 8-2-46. Equipment passed all required tests and is ready for use. “ Section 4-3-2 Power Off / Shutdown Section 4-9-15 ARP Assy Function Validation Procedure Section 4-5 Peripheral Checks Section 10-5-2-2 Peripheral/Option Checks Section 10-5-5 Physical Inspection Section 10-6-3 Outlet Test - Wiring Arrangement - USA & Canada
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 104 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-2-47 SATA HDD Assy 8-2-47-1 Tools • Common pilIips screwdrivers 8-2-47-2 Preparations • Shut down the system and switch off the main breaker. 8-2-47-3 Removal procedure 1.) Remove the side left cover. Refer to the 8-2-15 Side Left Cover on page 8-51. 2.) Remove EMI L Cover. Refer to the 8-2-26 EMI Cover L on page 8-68. 3.) Disconnect the HDD power cable and unsc rew 4 screws of the HDD fixture bracket. Figure 8-143 HDD in System HDD Power Cable