GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 115 8-2-51-1 How to start serial no read/write program (cont’d) 7.) When asn.exe starts running, Logon windows po ps up. Type in maintenance password to start asn.exe. Figure 8-156 Execute the “ASN.EXE”
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 116 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-2-51-2 How to read Serial No 1.) Check checkboxes of system boards you want to read or System Serial No checkbox. NOTE: You can read System Serial No, L1SYSCON board, CL1TRX board, P3RLY board, and ACWD board information. 8-2-51-3 How to write info on the system 1.) Check checkboxes of system boards you want to write or System Serial No checkbox. 2.) Put the Serial No, Part No, or Rev on the blank box. 3.) Press “Write” Button. Figure 8-157 Reading System Info
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-3 - Software Loading Procedure 8 - 117 Section 8-3 Software Loading Procedure 8-3-1 General This describes a full system soft ware loading (Base Software Load Image + Arirang Application Load Image) procedure. Use this instruction when: - Performing the HDD with new on e (Base Software Load Image + Arirang Application Load Image) - Performing the system revision-up (Application) 8-3-2 Parts Required LOGIQ™ P5 Base Software DVD LOGIQ™ P5 Application Software CD Blank CD-Rs (for patient image backup) or DVDs Service Dongle NOTE: Patient Image Data will be delet ed when all the drives are ghosted. Back up them on the CD-R or DVD before starting software loading NOTE: Insert the service dongle into this service po rt located at the rear panel befor e installing the Base System Software Load image 8-3-3 TimeRequired Approximately 2hours
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 118 Section 8-3 - Software Loading Procedure 8-3-4 Software Loading Steering Guide Figure 8-158 Flow Chart - Installation of Base and Application CD
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-3 - Software Loading Procedure 8 - 119 8-3-5 Before Starti ng Software Loading 8-3-5-1 Saving Connectivity Before starting upgrade, write down the Connectivity Setting for back-up. 1.) Go to Utility > Connectivity and click the Tcpip tab. 2.) Write down the following parameters:. ParametersDescriptions (Entry for Stand-Alone System) Computer Name Enable DHCP IP-Address Subnet Mask Default Gateway Note : For a stand-alone system, preset values of IP address, subnet mask, and default gateway, shown above, must entered. Figure 8-159 Connectivity
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 120 Section 8-3 - Software Loading Procedure 8-3-5-2 Image Management Guide NOTE: Before upgrade to R4.X.X software, perform recommended image backup procedure. Export/Import Data/Images Between Systems - Use this to copy both patient data and images for specified patient(s) from one system to another. To move exams from one Ultrasound system to another, you need to export/import exam information. NOTE: Both database information and images are expo rted. No data is deleted from the local archive when exporting data. Exporting Data To export an exam(s) to a compatible Ultrasound system: 1.) Format the removable media (DVD or CD-ROM). Label the removable media. Answer Yes/OK to the messages.Press Patient. Deselect any select ed patient(s) in the search portion of the Patient screen. Press Data Transfer (located at the upper, right-hand corner of the Patient menu). 2.) Select Export task Button. 3.) The [From] combo box is not active. It displa ys Local Archive. The [To] combo box is active. Select the type of removable media CD/DVD. Then please wait until the patient list is visible. Figure 8-160 Exam Data Transfer Figure 8-161 Task button Figure 8-162 To Combo Box
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-3 - Software Loading Procedure 8 - 121 8-3-5-2 Image Management Guide (cont’d) 4.) In the patient list at the top of the Patient me nu, select the patient(s) you want to export. You can use Windows commands to select more than one patient. To select a consecutive list of patients, click the cursor on the first name, move the cursor to the last name, then press and hold down the Shift+right Set key to select al l the names.To select a non-consecutive list of patients, click the cursor at the first name, move the cursor to the next name, then press and hold down the Ctrl+right Set key, move the cursor to the next name, then press and hold down the Ctrl+right Set key again, etc.You can also s earch for patients via the Search key and string. Or, Select All Button from the Patient Menu: NOTE: You need to use your best judgment when movi ng patients images. If there are lots of images or loops, then only move a few patients at a time. 5.) Once you have selected all of the patients to export, press Transfer Button as shown in Figure 8-163 on page 8-121 from the Patient Menu. 6.) The progress bar appears as the copy is taking place. Once transfer completed, the exported patient list will be shown as Figure 8-164. 7.) Press F3 to eject the media. Specify th at you want to finalize the CD-ROM or DVD. Figure 8-163 Source Section Figure 8-164 Destination Section of Patient Screen
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 122 Section 8-3 - Software Loading Procedure 8-3-5-2 Image Management Guide (cont’d) Importing Data To import an exam(s) to another Ultrasound system: 1.) At the other Ultrasound system, insert the CD-ROM/DVD. 2.) Press Patient. Press Exam Data Transfer (located at the upper, left-hand corner of the Patient menu). 3.) Select Import task Button. 4.) The [TO] combo box is not active. It displays Local Archive. The [FROM] combo box is active. Select the type of removable media CD/DVD Rom. Then please wait until the patient list is visible. 5.) The Patient menu just shows the patients ava ilable for import from the removable media you just loaded onto the system. 6.) Select the patients to be imported. 7.) Once you have selected all of the patients to import, press Transfer Button. 8.) Please wait for the patient information to be copied to this Ultrasound system. Progress bar appears while the import is taking place. Figure 8-165 Exam Data Transfer Figure 8-166 Task buttonFigure 8-167 From
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-3 - Software Loading Procedure 8 - 123 8-3-5-3 Printer setting Back-up Before starting upgrade, write down, print out or save the printe r setting for back-up. NOTE: All the customized image setting of digital prin ter will be lost after software loading. Take the back up and restore all the digital printers setting at software loading. NOTE: This printer setting is for only digital printers, not analog printers. 1.) Insert the CD or removable media and Format. 2.) Go to Utility -> System -> Pe ripherals and click the Printers. 3.) Press the left button of the trackball on the prin ter which need to be backup -> Click “properties”. Figure 8-168 Printer set backup Figure 8-169 Printer set backup
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 124 Section 8-3 - Software Loading Procedure 8-3-5-3 Printer setting Back-up (cont’d) 4.) Select “Printing Preferences”. 5.) If Sony D897 is used, Click “Write”. Figure 8-170 Printer set backup Figure 8-171 Printer set backup