    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
    Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 55
    8-2-18 Rear Cover 
    8-2-18-1 Tools • Common pilIips screwdrivers
    8-2-18-2 Preparations • Shut down the system and switch off the main breaker.
    8-2-18-3 Removal procedure 1.) Remove the Left, and Right side covers. Refer to the  8-2-15 Side Left Cover on page 8-51 and 8-
    2-16 Side Right Cover on page 8-52 .
    2.) Remove 4 rubber caps from the rear cover of the system.
    3.) Unscrew 4 screws (1-4) from the rear cover.
    4.) Remove the Rear Cover. Figure 8-91   4 rubber caps on Rear Cover
    Figure 8-92   4 screws on rear cover
    (2)
    (3) (4)
    (1) 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    8 - 56 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
    8-2-18-3              Removal procedure  (cont’d)
    5.) Perform the following functional tests. If all are successful, include the debrief script provided below.
    8-2-18-4 Mounting Procedure Install the new parts in the reverse order of removal. Table 8-19    Functional Tests
    Service ManualSection
    Functional Test / Diagnostic TestDebrief Script
    Section 4-3-1 Power On/Boot Up
    “Service Manual, Direction
    5394141, Rev 1+, Section 8-2-18. Equipment 
    passed all required tests and is ready for use. “
    Section 4-3-2
    Power Off / Shutdown
    Section 4-8-1 Cover Parts Function Validation
    Section 10-5-5 Physical Inspection 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
    Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 57
    8-2-19 Middle Cover
    8-2-19-1 Tools • Common pilIips screwdrivers
    • Stubby screwdriver (Flat tip and Cross tip)
    8-2-19-2 Preparations • Shut down the system and switch off the main breaker.
    8-2-19-3 Removal procedure 1.) Remove the Left, and Right side covers. Refer to the  8-2-15 Side Left Cover on page 8-51 and 8-
    2-16 Side Right Cover on page 8-52 . And Remove the Rear Cover. Refer to the  8-2-18 Rear 
    Cover on page 8-55 .
    2.) Remove 2 rubber caps (1-2) from  the middle cover of the system.
    3.) Unscrew 2 screws (1-2) from the middle cover. Figure 8-93   2 rubber cap of middle cover
    Figure 8-94   2 screws of middle cover
    1 2
    12 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    8 - 58 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
    8-2-19-3              Removal procedure  (cont’d)
    4.) Perform the following functional tests. If all are successful, include the debrief script provided below.
    8-2-19-4 Mounting Procedure Install the new parts in the reverse order of removal. Table 8-20    Functional Tests
    Service ManualSection
    Functional Test / Diagnostic TestDebrief Script
    Section 4-3-1 Power On/Boot Up
    “Service Manual, Direction
    5394141, Rev 1+, Section 8-2-19. Equipment 
    passed all required tests and is ready for use. “
    Section 4-3-2
    Power Off / Shutdown
    Section 4-8-1 Cover Parts Function Validation
    Section 10-5-5 Physical Inspection 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
    Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 59
    8-2-20 Top Cover
    8-2-20-1 Tools • Common pilIips screwdrivers
    8-2-20-2 Preparations • Shut down the system and switch off the main breaker.
    8-2-20-3 Removal procedure 1.) Remove the Dummy Cover L/R. Refer to the  8-2-25 Dummy Cover L / R on page 8-66.
    2.) Remove the Top Bottom Cover. Refer to the  8-2-21 Top Bottom Cover on page 8-60.
    3.) Remove 2 rubber caps (1-2) from the top cover of the system and unscrew 2 screws (1-2).
    4.) Unscrew 5 screws (1-5) from the bottom of the top cover.
    5.) Remove the Top Cover.
    6.) Perform the following functional tests. If all ar e successful, include the debrief script provided below.
    8-2-20-4 Mounting Procedure Install the new parts in the reverse order of removal. Figure 8-95   2 rubber caps of top cover
    Table 8-21    Functional Tests
    Service Manual Section
    Functional Test / Diagnostic TestDebrief Script
    Section 4-3-1 Power On/Boot Up
    “Service Manual, Direction
    5394141, Rev 1+, Section 8-2-20. Equipment 
    passed all required tests and is ready for use. “
    Section 4-3-2
    Power Off / Shutdown
    Section 4-8-1 Cover Parts Function Validation
    Section 10-5-5 Physical Inspection
    (1)(2) 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    8 - 60 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
    8-2-21 Top Bottom Cover
    8-2-21-1 Tools • Common pilIips screwdrivers
    • Stubby screwdriver (Flat tip and Cross tip)
    8-2-21-2 Preparations • Shut down the system and switch off the main breaker.
    8-2-21-3 Removal procedure 1.) Remove the Dummy Cover L/R. Refer to the  8-2-25 Dummy Cover L / R on page 8-66.
    2.) Unscrew 7 screws (1-7) from the bottom of the top cover.
    3.) Remove the Top Bottom Cover.
    4.) Perform the following functional tests. If all are successful, include the debrief script provided below.
    8-2-21-4 Mounting Procedure Install the new parts in the reverse order of removal. Figure 8-96   Top bottom cover
    Table 8-22    Functional Tests
    Service Manual Section
    Functional Test / Diagnostic TestDebrief Script
    Section 4-3-1 Power On/Boot Up
    “Service Manual, Direction
    5394141, Rev 1+, Section 8-2-21. Equipment 
    passed all required tests and is ready for use. “
    Section 4-3-2
    Power Off / Shutdown
    Section 4-8-1 Cover Parts Function Validation
    Section 10-5-5 Physical Inspection 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
    Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 61
    8-2-22 Pole Cover 
    8-2-22-1 Tools • Common pilIips screwdrivers
    8-2-22-2 Preparations • Shut down the system and switch off the main breaker.
    8-2-22-3 Removal procedure 1.) Remove the Dummy Cover L/R. Refer to the  8-2-25 Dummy Cover L / R on page 8-66.
    2.) Remove the Top Bottom Cover. Refer to the  8-2-21 Top Bottom Cover on page 8-60.
    3.) Remove the Top Cover. Refer to the  8-2-20 Top Cover on page 8-59.
    4.) Unscrew 2 screws (1-2).
    5.) Perform the following functional tests. If all ar e successful, include the debrief script provided below.
    8-2-22-4 Mounting Procedure Install the new parts in the reverse order of removal. Figure 8-97   Pole Cover
    Table 8-23    Functional Tests
    Service Manual Section
    Functional Test / Diagnostic TestDebrief Script
    Section 4-3-1 Power On/Boot Up
    “Service Manual, Direction
    5394141, Rev 1+, Section 8-2-22. Equipment 
    passed all required tests and is ready for use. “
    Section 4-3-2
    Power Off / Shutdown
    Section 4-8-1 Cover Parts Function Validation
    Section 10-5-5 Physical Inspection
    (1)
    (2) 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    8 - 62 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
    8-2-23 Neck Front 
    8-2-23-1 Tools • Common pilIips screwdrivers
    8-2-23-2 Preparations • Shut down the system and switch off the main breaker.
    8-2-23-3 Removal procedure 1.) Remove 1 rubber cap.
    2.) Unscrew 1 screw and remove the Neck Front. Figure 8-98   rubber cap of neck front cover
    Figure 8-99   Open neck front cover 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
    Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 63
    8-2-23-3              Removal procedure  (cont’d)
    3.) Perform the following functional tests. If all ar e successful, include the debrief script provided below.
    8-2-23-4 Mounting Procedure Install the new parts in the reverse order of removal. Table 8-24    Functional Tests
    Service ManualSection
    Functional Test / Diagnostic TestDebrief Script
    Section 4-3-1 Power On/Boot Up
    “Service Manual, Direction
    5394141, Rev 1+, Section 8-2-23. Equipment 
    passed all required tests and is ready for use. “
    Section 4-3-2
    Power Off / Shutdown
    Section 4-8-1 Cover Parts Function Validation
    Section 10-5-5 Physical Inspection 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    8 - 64 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
    8-2-24 Neck Rear
    8-2-24-1 Tools • Common pilIips screwdrivers
    8-2-24-2 Preparations • Shut down the system and switch off the main breaker.
    8-2-24-3 Removal procedure 1.) Remove 2 rubber caps (1-2).
    2.) Unscrew 2 screws and remove the Neck Rear. Figure 8-100   rubber cap of neck rear cover
    Figure 8-101   Open neck rear cover
    (2)
    (1) 
    						