GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 55 8-2-18 Rear Cover 8-2-18-1 Tools • Common pilIips screwdrivers 8-2-18-2 Preparations • Shut down the system and switch off the main breaker. 8-2-18-3 Removal procedure 1.) Remove the Left, and Right side covers. Refer to the 8-2-15 Side Left Cover on page 8-51 and 8- 2-16 Side Right Cover on page 8-52 . 2.) Remove 4 rubber caps from the rear cover of the system. 3.) Unscrew 4 screws (1-4) from the rear cover. 4.) Remove the Rear Cover. Figure 8-91 4 rubber caps on Rear Cover Figure 8-92 4 screws on rear cover (2) (3) (4) (1)
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 56 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-2-18-3 Removal procedure (cont’d) 5.) Perform the following functional tests. If all are successful, include the debrief script provided below. 8-2-18-4 Mounting Procedure Install the new parts in the reverse order of removal. Table 8-19 Functional Tests Service ManualSection Functional Test / Diagnostic TestDebrief Script Section 4-3-1 Power On/Boot Up “Service Manual, Direction 5394141, Rev 1+, Section 8-2-18. Equipment passed all required tests and is ready for use. “ Section 4-3-2 Power Off / Shutdown Section 4-8-1 Cover Parts Function Validation Section 10-5-5 Physical Inspection
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 57 8-2-19 Middle Cover 8-2-19-1 Tools • Common pilIips screwdrivers • Stubby screwdriver (Flat tip and Cross tip) 8-2-19-2 Preparations • Shut down the system and switch off the main breaker. 8-2-19-3 Removal procedure 1.) Remove the Left, and Right side covers. Refer to the 8-2-15 Side Left Cover on page 8-51 and 8- 2-16 Side Right Cover on page 8-52 . And Remove the Rear Cover. Refer to the 8-2-18 Rear Cover on page 8-55 . 2.) Remove 2 rubber caps (1-2) from the middle cover of the system. 3.) Unscrew 2 screws (1-2) from the middle cover. Figure 8-93 2 rubber cap of middle cover Figure 8-94 2 screws of middle cover 1 2 12
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 58 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-2-19-3 Removal procedure (cont’d) 4.) Perform the following functional tests. If all are successful, include the debrief script provided below. 8-2-19-4 Mounting Procedure Install the new parts in the reverse order of removal. Table 8-20 Functional Tests Service ManualSection Functional Test / Diagnostic TestDebrief Script Section 4-3-1 Power On/Boot Up “Service Manual, Direction 5394141, Rev 1+, Section 8-2-19. Equipment passed all required tests and is ready for use. “ Section 4-3-2 Power Off / Shutdown Section 4-8-1 Cover Parts Function Validation Section 10-5-5 Physical Inspection
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 59 8-2-20 Top Cover 8-2-20-1 Tools • Common pilIips screwdrivers 8-2-20-2 Preparations • Shut down the system and switch off the main breaker. 8-2-20-3 Removal procedure 1.) Remove the Dummy Cover L/R. Refer to the 8-2-25 Dummy Cover L / R on page 8-66. 2.) Remove the Top Bottom Cover. Refer to the 8-2-21 Top Bottom Cover on page 8-60. 3.) Remove 2 rubber caps (1-2) from the top cover of the system and unscrew 2 screws (1-2). 4.) Unscrew 5 screws (1-5) from the bottom of the top cover. 5.) Remove the Top Cover. 6.) Perform the following functional tests. If all ar e successful, include the debrief script provided below. 8-2-20-4 Mounting Procedure Install the new parts in the reverse order of removal. Figure 8-95 2 rubber caps of top cover Table 8-21 Functional Tests Service Manual Section Functional Test / Diagnostic TestDebrief Script Section 4-3-1 Power On/Boot Up “Service Manual, Direction 5394141, Rev 1+, Section 8-2-20. Equipment passed all required tests and is ready for use. “ Section 4-3-2 Power Off / Shutdown Section 4-8-1 Cover Parts Function Validation Section 10-5-5 Physical Inspection (1)(2)
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 60 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-2-21 Top Bottom Cover 8-2-21-1 Tools • Common pilIips screwdrivers • Stubby screwdriver (Flat tip and Cross tip) 8-2-21-2 Preparations • Shut down the system and switch off the main breaker. 8-2-21-3 Removal procedure 1.) Remove the Dummy Cover L/R. Refer to the 8-2-25 Dummy Cover L / R on page 8-66. 2.) Unscrew 7 screws (1-7) from the bottom of the top cover. 3.) Remove the Top Bottom Cover. 4.) Perform the following functional tests. If all are successful, include the debrief script provided below. 8-2-21-4 Mounting Procedure Install the new parts in the reverse order of removal. Figure 8-96 Top bottom cover Table 8-22 Functional Tests Service Manual Section Functional Test / Diagnostic TestDebrief Script Section 4-3-1 Power On/Boot Up “Service Manual, Direction 5394141, Rev 1+, Section 8-2-21. Equipment passed all required tests and is ready for use. “ Section 4-3-2 Power Off / Shutdown Section 4-8-1 Cover Parts Function Validation Section 10-5-5 Physical Inspection
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 61 8-2-22 Pole Cover 8-2-22-1 Tools • Common pilIips screwdrivers 8-2-22-2 Preparations • Shut down the system and switch off the main breaker. 8-2-22-3 Removal procedure 1.) Remove the Dummy Cover L/R. Refer to the 8-2-25 Dummy Cover L / R on page 8-66. 2.) Remove the Top Bottom Cover. Refer to the 8-2-21 Top Bottom Cover on page 8-60. 3.) Remove the Top Cover. Refer to the 8-2-20 Top Cover on page 8-59. 4.) Unscrew 2 screws (1-2). 5.) Perform the following functional tests. If all ar e successful, include the debrief script provided below. 8-2-22-4 Mounting Procedure Install the new parts in the reverse order of removal. Figure 8-97 Pole Cover Table 8-23 Functional Tests Service Manual Section Functional Test / Diagnostic TestDebrief Script Section 4-3-1 Power On/Boot Up “Service Manual, Direction 5394141, Rev 1+, Section 8-2-22. Equipment passed all required tests and is ready for use. “ Section 4-3-2 Power Off / Shutdown Section 4-8-1 Cover Parts Function Validation Section 10-5-5 Physical Inspection (1) (2)
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 62 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-2-23 Neck Front 8-2-23-1 Tools • Common pilIips screwdrivers 8-2-23-2 Preparations • Shut down the system and switch off the main breaker. 8-2-23-3 Removal procedure 1.) Remove 1 rubber cap. 2.) Unscrew 1 screw and remove the Neck Front. Figure 8-98 rubber cap of neck front cover Figure 8-99 Open neck front cover
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 63 8-2-23-3 Removal procedure (cont’d) 3.) Perform the following functional tests. If all ar e successful, include the debrief script provided below. 8-2-23-4 Mounting Procedure Install the new parts in the reverse order of removal. Table 8-24 Functional Tests Service ManualSection Functional Test / Diagnostic TestDebrief Script Section 4-3-1 Power On/Boot Up “Service Manual, Direction 5394141, Rev 1+, Section 8-2-23. Equipment passed all required tests and is ready for use. “ Section 4-3-2 Power Off / Shutdown Section 4-8-1 Cover Parts Function Validation Section 10-5-5 Physical Inspection
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 64 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-2-24 Neck Rear 8-2-24-1 Tools • Common pilIips screwdrivers 8-2-24-2 Preparations • Shut down the system and switch off the main breaker. 8-2-24-3 Removal procedure 1.) Remove 2 rubber caps (1-2). 2.) Unscrew 2 screws and remove the Neck Rear. Figure 8-100 rubber cap of neck rear cover Figure 8-101 Open neck rear cover (2) (1)