Home > GE > Medical Equipment > GE Logiq P5 Service Manual

GE Logiq P5 Service Manual

    Download as PDF Print this page Share this page

    Have a look at the manual GE Logiq P5 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    Page
    of 520
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
    Section 9-9 - Cables 9 - 17
    Section 9-9      Cables (cont’d)
    ItemPart NameFRU NumberDescriptionQtyFRUR1.0.XR2.0.XR3.0.X~R4.0.X~
    700 Multi Cable Assy 1 5169853DVI Cable + Speaker Cable 
    + KBD Control Cable
    1. DVI Cable 
    Connect one side 
    (Label:DVI) to LCDMON II 
    Assy and another side 
    (Label:J206) to PCB Silk 
    J206 of Backplane Assy
    2. Speaker Cable
    Connect one side (Label: L 
    & R) to Label L,R of 
    Speaker Assy and another 
    side (Label:J207) to PCB 
    Silk J207 of Backplane 
    Assy
    3. Keyboard Control Cable 
    Connect one side 
    (Label:CON1) to Keyboard 
    Assy and another side 
    (Label:J209) to PCB Silk 
    J209 of Backplane Assy1 1 Yes Yes Yes Yes
    701 Multi Cable Assy 2 5168859 Mon Pwr Cable + BW PWR 
    Cable + BW USB Cable + 
    KB USB Cable
    1. Monitor Power Cable 
    - Connect one side 
    (Label:CON7) to Label 
    CON 7 of Monitor power 
    Con cable or Monitor 
    power con long cable and 
    another side (Label: CON 
    11) to Lable CON 11 of 
    Internal AC Power cable 
    assy
    2. BW Printer Power Cable 
    - Connect one side 
    (Label:CON10) to Label 
    CON10 of USB CON Cable 
    and another side (Label: 
    CON9) to Lable CON9 of 
    LP5 ARP Assy
    3.Keyboard USB Cable 
    - Connect one side 
    (Label:KBD) to LP5 
    Keyboard Assy 
    and another side (Label: 
    J221) to PCB Silk J221 of 
    LP5 Backplane Assy”
    4. BW Printer USB Cable 
    - Connect one side 
    (Label:PRN) to BW Printer 
    Bracket and another side 
    (Label: J223) to PCB Silk 
    J223 of LP5 Backplane 
    Assy”1 1 Yes Yes Yes Yes
    Table 9-9    Cables 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL  
    9 - 18 Section 9-9 - Cables
    702 Multi Cable Assy 3 5168797 Front USB Cable + MIC 
    Cable
    1. Front USB Cable 
    - Connect one side (Label: 
    F_USB) to Front USB 
    Bracket and another side 
    (Label: J14) to Label J14 of 
    LP5 ARP Assy
    2. MIC Cable
    - Connect one side 
    (Label:MIC) to LP5 Front 
    USB Bracket and another 
    side (Label: J205) to PCB 
    Silk J205 of LP5 Backplane 
    Assy1 1 Yes Yes Yes Yes
    703 Monitor Power Conn Cable 5166286 Connect one side 
    (label:CON7) to Label 
    “CON7” of Monitor power 
    cable and another side 
    (Label:MON) to LCDMON II 
    Assy1 1 Yes Yes Yes Yes
    704 Monitor Power Conn Long 
    Cable 5168941Connect one side 
    (label:CON7) to Label 
    “CON7” of Monitor power 
    cable and another side 
    (Label:MON) to LCDMON II 
    Assy
    1 1 Yes Yes Yes Yes
    705 ACPC Cable 5168846Connect one side (Label: 
    J203) to PCB Silk “J203” of 
    LP5 Backplane and another 
    side (Label:CON2) to Label 
    “CON2” of Internal ACPC 
    ConCable 1 1 Yes Yes Yes Yes
    706 Printer Cable Assy 1 5177395 Connect one side to BW 
    Printer Power cable and 
    another side to External 
    Printer (short cable)
    1 1 Yes Yes Yes Yes
    707 Printer Cable Assy 2 5168842 Connect one side to BW 
    Printer Power cable and 
    another side to External 
    Printer (long cable)1 1 Yes Yes Yes Yes
    708 EIDE Cable 5168955-2Connect one side 
    (Label:ODD) to LP5 ODD 
    Assy and another side 
    (Label: J204) to PCB Silk 
    J204 of LP5 Backplane 
    Assy 11 YesYes
    709 APS AC Power Cable 5168913 Connect one side 
    (Label:APS) to Label APS 
    of LP5 APS Assy
    and another side (Label: 
    CON3) to Label CON3 of 
    LP5 AC Out Cable
    1 1 Yes Yes Yes Yes
    ItemPart NameFRU NumberDescriptionQtyFRUR1.0.XR2.0.XR3.0.X~R4.0.X~
    Table 9-9    Cables 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
    Section 9-9 - Cables 9 - 19
    710 Peripheral AC Power Cable 5165244Connect one side (Label 
    CON5) to Label CON5 of 
    LP5 AC Out Cable
    and another side (Label: 
    CON6) to Label CON6 of 
    LP5 ARP Assy1 1 Yes Yes Yes Yes
    711 AEXP Cable 5172066Connect one side (Label 
    J13) to Label PCB Silk J13 
    of LP5 Syscon Assy
    and another side (Label: 
    J15) to Label J15 of LP5 
    ARP Assy 1 1 Yes Yes Yes Yes
    712 DCWD Cable Assy 5174403 Connect one side to DCWD 
    Bracket and another 
    side(Label: J7) to PCB Silk 
    J7 of ACWD Assy1 1 Yes Yes Yes Yes
    713 IEC Type ECG CABLE 5341188 ECG Patient Cable only for 
    Japan
    1 1 Yes Yes Yes Yes
    714 AHA Type ECG CABLE 5341189 ECG Patient Cable for Asia, 
    US, and Europe.1 1 Yes Yes Yes Yes
    ItemPart NameFRU NumberDescriptionQtyFRUR1.0.XR2.0.XR3.0.X~R4.0.X~
    Table 9-9    Cables 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL  
    9 - 20 Section 9-10 - Options
    Section 9-10Options
    Table 9-10    OPTIONS
    ItemPart NameFRU NumberDescriptionQtyFRUR1.0.XR2.0.XR3.0.X~R4.0.X~
    800Endo Probe Holder 51719331 1
    Yes Yes Yes Yes
    801 VCR DVD Fixture Top 5171717-21 1
    Yes Yes Yes Yes
    802 BW Printer Fixture Keyboard 5172473-31 1
    Yes Yes Yes Yes
    803 BW Printer Fixture Middle  5172353-21 1
    Yes Yes Yes Yes
    804 VCR Color Printer DVD Fixture Middle 5172238-3 1 1
    Yes Yes Yes Yes
    805 Paper Holder 5172484-2 1 1
    Yes Yes Yes Yes
    806 LCD Flexible Arm KIT 5264538-21 1
    Yes Yes Yes Yes
    807 Probe Cable Hanger 51721781 1
    Yes Yes Yes Yes
    808 Drawer 5171620 1 1
    Yes Yes Yes Yes
    809 ECG Kit with AHA cable 5172098-21 1
    Yes Yes Yes Yes
    810 ECG kit with IEC cable  53411861 1
    Yes Yes Yes Yes
    811 4D KIT 5171709-5 1 1
    Yes Yes Yes Yes
    812 3 Button Foot Switch
    53271101 1
    Yes Yes Yes Yes
    1 Button Foot Switch 53271111 1
    Yes Yes Yes Yes
    813 Additional Task Lamp KIT 51715821 1
    Yes Yes Yes Yes
    814 Microphone 5171727 1 1
    Yes Yes Yes Yes
    815 Protective Cover 51764171 1
    Yes Yes Yes Yes
    816 Swivel Lock Caster 51846031 1
    Yes Yes Yes Yes
    817 3rd probe Ports kit 5171892-31 1
    Yes Yes Yes Yes
    818 CW Doppler kit 5171900-2LOGIQ P5 
    ONLY 11
    Yes Yes Yes Yes
    819 ARC Switch Set 53271151 1
    Yes Yes
    820 ARC Receiver Dongle 53361861 1
    Yes Yes 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
    Section 9-11 - Software 9 - 21
    Section 9-11Software
    Table 9-11    Software
    ItemPart NameFRU NumberDescriptionQtyFRU
    900 LP5 BT11 CD SET
    5395395R4.0.X1 1
    LP5 A5 BT09 CD SET 5327125R3.0.X1 1
    LP5 A5 BT08 CD SET 5272499-5R2.0.41 1
    LP5 A5 BT06 CD SET 5196353-8R1.0.81 1 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL  
    9 - 22 Section 9-12 - Probes
    Section 9-12Probes
    Table 9-12    Probes
    ItemPart NameFRU NumberQtyFRUR1.0.XR2.0.XR3.0.X~R4.0.X~
    13.5C PROBE 2050357 1 1 Yes Yes Yes Yes
    2 3.5CS PROBE 2051858 1 1 Yes Yes Yes Yes
    3 4C PROBE  5123455 1 1 Yes Yes Yes Yes
    4 8C PROBE 2348094 1 1 Yes Yes Yes Yes
    5 E8C PROBE 2294641 1 1 Yes Yes Yes Yes
    6 E8CS PROBE 47236865 1 1 Yes Yes Yes Yes
    7 BE9C PROBE 2389382 1 1 Yes Yes Yes Yes
    8 8L PROBE 5140738 1 1 Yes Yes Yes Yes
    9 9L PROBE 5149427 1 1 Yes Yes Yes Yes
    10 10L PROBE  2294523 1 1 Yes Yes Yes Yes
    11 11L PROBE  5171885 1 1 Yes Yes Yes
    12 12L PROBE  2295377 1 1 Yes Yes Yes Yes
    13 i739 PROBE 2259206 1 1 Yes Yes Yes Yes
    14 t739 PROBE 2259246 1 1 Yes Yes Yes Yes
    15 3S-02 PROBE 2323337 1 1 Yes Yes Yes Yes
    16 5S-02 PROBE 2347469 1 1 Yes Yes Yes Yes
    17 7S-02 PROBE 2347471 1 1 Yes Yes Yes Yes
    18 ERB PROBE 2258309 1 1 Yes Yes Yes Yes
    19 4D3C-L PROBE KTZ195893 1 1 Yes Yes Yes Yes
    20 P2D TE100024 1 1 Yes Yes Yes Yes
    21 P6D TQ100002 1 1 Yes Yes Yes Yes
    22 UG7C Probe 5147187 1 1 Yes Yes Yes Yes
    23 5CS Probe 5143934 1 1 Yes Yes Yes Yes
    24 3CRF Probe 5196216 1 1 Yes Yes
    25 4DE7C Probe KTZ195859 1 1 Yes Yes Yes
    26 3Sp Probe 5327196 1 1 Yes Yes
    27 5Sp Probe 5327197 1 1 Yes Yes
    28 4D8C Probe KTZ156959 1 1 Yes Yes
    29 BE9CS Probe 5266132 1 1 Yes Yes
    30 i12L Probe 2264883 1 1 Yes Yes 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
    Section 9-13 - Peripherals 9 - 23
    Section 9-13Peripherals
    Table 9-13    Peripherals
    ItemPart NameFRU NumberDescriptionQtyFRU
    1LQ-MD800P 5120592Panasonic LQ-MD800P 
    DVR Video Recorder 11
    2 LQ-MD800E 5120593Panasonic LQ-MD800E 
    DVR Video Recorder 11
    3 VCR Mitsubishi HS-MD3000U 066E9513VCR1 1
    4 VCR Mitsubishi HS-MD3000E 066E9514VCR1 1
    5 Sony UP-D895MD 2288502Digital B/W printer 1 1
    6 SONY UP-D897MD 5171608Digital B/W printer 1 1
    7 Mitsubishi P93D Digital B/W printer 1 1
    8 SONY UP-D23MD 2401985Digital Color printer 1 1
    9 Sony UP-D25MD 5389822Digital Color printer 1 1
    10 Mitsubishi CP30DW 5310492Digital Color printer 1 1
    11 HP 990cxi 066E0407Standard printer 1 1
    12 HP 6122 2378741Standard printer 1 1
    13 HP 6540 5121837Standard printer 1 1
    14 HP 5650  5121839Standard printer 1 1
    15 HP Business Inkjet 1200 066E0424Standard printer 1 1
    16 HP Office jet Pro K550 EP100925Standard printer 1 1
    17 HP Office jet Pro K5400 EQ100158Standard printer 1 1
    18 HP Deskjet D4360  5357534Standard printer 1 1
    19 HP Deskjet 6940  5351754Standard printer 1 1
    20 Mitsubishi P93D Digital B/W printer 1 1
    21 Sony UP-D25MD Digital Color printer 1 1
    22 LexMark Z65 Standard printer 1 1
    23 HP 1000 Standard printer 1 1
    24 Xerox Phaser 8400 Standard printer 1 1
    25 HP Officejet 6000 Special Edition Standard printer 1 1
    25 Sony DVO-1000MD VCR1 1 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL  
    9 - 24 Section 9-13 - Peripherals 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
    Section 10-1 - Overview 10 - 1
    Chapter 10
    Care & Maintenance
    Section 10-1
    Overview
    10-1-1 Periodic Maintenance Inspections 
    It has been determined by engineering that your  LOGIQ™ P5 system does not have any high wear 
    components that fail with use, therefore no Peri odic Maintenance Inspections are mandatory. Some 
    Customers Quality Assurance Programs may require additional tasks and or inspections at a different 
    frequency than listed in this manual.
    10-1-2 Purpose of Chapter 10
    This chapter describes Care & Maintenance  on the scanner and peripherals. These procedures are 
    intended to maintain the quality of the ultrasound  systems performance. Read this chapter 
    completely and familiarize yourself with the procedures before performing a task.
    Table 10-1    Contents in Chapter 10
    SectionDescriptionPage Number
    10-1 Overview 10-1
    10-2 Why do Maintenance 10-2
    10-3 Maintenance Task Schedule 10-2
    10-4 Tools Required 10-4
    10-5 System Maintenance 10-7
    10-6 Electrical Safety Tests 10-12
    10-7 When Theres Too Much Leakage Current... 10-28
    CAUTIONPractice good ESD prevention. Wear an anti–sta
    tic strap when handling electronic parts and 
    even when disconnecting/connecting cables.
    DANGERTHERE ARE SEVERAL PLACES ON THE BACKPLANE, THE AC DISTRIBUTION, 
    AND DC DISTRIBUTION TH AT ARE DANGEROUS. BE SURE TO DISCONNECT 
    THE SYSTEM POWER PLUG AND OPEN  THE MAIN CIRCUIT BREAKER 
    BEFORE YOU REMOVE ANY PARTS. BE  CAUTIOUS WHENEVER POWER IS 
    STILL ON AND COVERS ARE REMOVED.
    CAUTIONDo not pull out or insert circ uit boards while power is ON.
    CAUTIONDo not operate this unit unless all board covers and frame panels are securely in place. System 
    performance and cooling require this. 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL  
    10 - 2 Section 10-2 - Why do Maintenance
    Section 10-2
    Why do Maintenance
    10-2-1 Keeping Records
    It is good business practice that ultrasound facilities maintain records of qua lity checks and corrective 
    maintenance. The Ultrasound Inspec tion Certificate (provided on page 10-29) provides the customer 
    with documentation that the ultrasound sc anner is maintained on a periodic basis.
    A copy of the Ultrasound Periodic Maintenance Inspec tion Certificate should be kept in the same room 
    or near the scanner.
    10-2-2 Quality Assurance
    In order to gain accreditation from organizations such as the American College of Radiology (USA), it 
    is the customer’s responsibility to  have a quality assurance program in place fo r each scanner. The 
    program must be directed by a medical physicists, t he supervising radiologist/physician or appropriate 
    designee.
    Routine quality control testing must occur regularly. The same tests are performed during each period 
    so that changes can be monitored over time  and effective corrective action can be taken.
    Testing results, corrective action and the effe cts of corrective action must be documented and 
    maintained on the site.
    Your GE service representative can help you with establishing, performing and maintaining records for 
    a quality assurance program. Please contact us for  coverage information and/or price for service.
    Section 10-3
    Maintenance Task Schedule
    10-3-1 How often should care & maintenance tasks be performed?
    The Care & Maintenance Task Schedule (provided on page  10-3) specifies how often your LOGIQ™ 
    P5 should be serviced and outlines items requiring special attention. 
    NOTE: It is the customer’s responsibility to ensure  the LOGIQ™ P5 care & maintenance is performed 
    as scheduled in order to retain its high le vel of safety, dependability and performance.
    Your GE Service Representative has an in-depth  knowledge of your LOGIQ™ P5 ultrasound scanning 
    system and can best provide competent, efficient se rvice. Please contact us for coverage information 
    and/or price for service.
    The service procedures and recommended intervals shown in the Care & Maintenance Task Schedule 
    assumes that you use your LOGIQ™ P5 for an avera ge patient load (10-12 per day) and not use it as 
    a primary mobile unit which is tran sported between diagnostic facilities.
    NOTE: If conditions exist which exceed typical usa ge and patient load, then it is recommended to 
    increase the maintenance frequencies. 
    						
    All GE manuals Comments (0)