    Section 3-4 - Preparing for Installation 3-11
    Section 3-5Completing the Installation
    3-5-1 Power On/Boot Up
    NOTE: After turning off the system, wait at least ten seconds before turning it on again. The system 
    may not be able to boot if power is recycled too quickly.
    3-5-1-1 Scanner Power On 1.) Connect the Main Power Cable at the rear of the System.
    2.) Connect the Main Power cable to an appropriate mains power outlet.
    3.) Switch ON the Main Circuit Breaker at the rear of the System.
    When power is applied to the Scanner, and the Rear  Circuit breaker is turned ON, Power is distributed 
    to the DC Power supply unit and power control  assy in the AC power assy. When the Power ON/OFF 
    key is pressed once, the DC power supply is enab led and feed each DC power to board assemblies. 
    and system run the software.
    4.) Press the  ON/OFF 
    key at the front of the System once.
    WARNINGWARNINGProtective earth must be taken when connecting AC power cable (200V) without its plug 
    to wall outlet.
    Figure 3-8   Circuit Breaker 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
    3-12 Section 3-4 - Preparing for Installation
    3-5-1-1              Scanner Power On (cont’d)
    Figure 3-9   Power On /Off Switch Location 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL
    Section 3-4 - Preparing for Installation 3-13
    3-5-1-2 Power Up Sequence
    1.) The Start Up Screen will be shown on the Mo nitor display when the system is turned ON.
    2.) After initialization is complete , all lighted buttons on the Control Panel light and the default B-Mode 
    screen or Patient screen (no probes are conn ected) is displayed on the monitor display.
    3-5-1-3 Power Up Sequence  NOTE: To enter the Maintenance Mode, select Maintenance button.
    NOTE: Figure 3-11 on page 3-13 displays when Service Dongle inserted in Rear Panel.
    NOTE: Start is selected autom atically when it time out.
    NOTE: In case the system is locked with task manager, use  arirang as a password to unlock the 
    system. Figure 3-10   Start Up Screen Display
    Figure 3-11   Start Application Window 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
    3-14 Section 3-4 - Preparing for Installation
    3-5-2 Power Off/ Shutdown
    NOTE: After turning off a system, wait at least ten  seconds before turning it on again. The system may 
    not be able to boot if power is recycled too quickly.
    3-5-2-1 Power Shutdown 1.) Press the  ON/OFF
     key at the front of the System. refer to Figure 3-9 on page 3-12 .
    2.) Click the shutdown icon in  the shutdown dialog box and wa it until the system shutdown.
    3.) Switch OFF the Main Circuit Breaker at the rear  of the system refer to Figure 3-8 on page 3-11 .
    4.) Disconnect the Main Power Cable is necessary.  For example : Relocating the scanner. 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL
    Section 3-4 - Preparing for Installation 3-15
    3-5-3 Transducer Connection
    1.) Connect a transducer to one of the three rightmost transducer receptacle as follows:
    A.) Ensure that the transducer twist lo ck lever to the horizontal position.
    B.) Insert the transducer  connector on the receptacle guide pi n until it touches the receptacle 
    mating surface.
    C.) Twist the transducer twis t lock lever to horizontal position to lock it in place. Twist the lever 
    to the vertical position to disconnect the transducer.
    NOTE: It is not necessary to turn OFF po wer to connect or disconnect a transducer. 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
    3-16 Section 3-4 - Preparing for Installation
    Section 3-6
    System Configuration
    3-6-1 System Specifications
    3-6-1-1 System Settings
    1.) Press Insert key on the A/N keyboard to enter  the utility menu. 
    2.) Select Utility > System > General.
    3.) Set the  Hospital name, Department, Date and Time, Language,  and Units.
    4.) Click on  Exit to terminate the  utility function.
    Table 3-1    System Configuration
    Configuration CategoryDescription
    Settings
    Enables the user or service personnel to set the 
    date, time, unit, language, basic information about 
    the organization such as the institution name and 
    department.
    Figure 3-12   Setting Display 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL
    Section 3-4 - Preparing for Installation 3-17
    3-6-1-2 Physical Dimensions
    The physical dimensions of the LOGIQ™ P5/A5/A5Pro  unit are summarized in Table 3-2 on page 3-17 .
    The Size of LOGIQ™ P5/A5/A5Pro, with monitor and peripherals
    Table 3-2    Physical Dimensions of LOGIQ™ P5/A5/A5Pro
    HeightWidthDepthUnit
    135/141 43.064.0 cm
    53.15 - 55.50 16.97 25.2inches
    Figure 3-13   Overall Dimensions
    WEIGHT : 75Kg (165bs)
    including Monitor
    NOTE : Length is in mm Variation +/- 10%
    1350 / 1410
    840 / 900 430
    640 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
    3-18 Section 3-4 - Preparing for Installation
    3-6-1-3 Weight without Monitor and Peripherals
    3-6-1-4 Acoustic Noise Output:
    Less than 70dB (A) according to DIN 45635 - 19 - 01 - KL2.
    3-6-2 Electrical  Specifications
    3-6-3 On-Board Optional Peripherals
    See each option installation instructions  for installation and connection procedures.
    NOTE: There are no external peripheral options for the LOGIQ™ P5/A5/A5Pro. Table 3-3    Weight of LOGIQ™ P5/A5/A5Pr
    o With Monitor and Without Other Peripherals
    ModelWeight [kg]Weight [lbs]
    LOGIQ™ P5/A5/
    A5Pro Approximately 75 Approximately 165
    Table 3-4    Electrical Specifi
    cations for LOGIQ™ P5/A5/A5Pro
    SystemVoltageCurrentFrequency
    1110 ~ 120 VAC 7 ~ 9A 50~60Hz
    2 220 ~ 240 VAC 3.5 ~ 4.5A 50~60Hz
    Table 3-5    List of Recording Devices
    DeviceManufacturerModelVideo Signal
    B/W Video Printer SONYUP-D897MDN/A (* USB Interface)
    Video Cassette Recorder MistubishiHS-MD3000U
    HS-MD3000E NTSC/PAL (USB Interface)
    DVD Video Recorder PanasonicLQ-MD800 NTSC/PAL (USB Interface)
    A6 Color Video PrinterSONYUP-D23MDN/A (* USB Interface) 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL
    Section 3-4 - Preparing for Installation 3-19
    3-6-4 External I/O Connector Panel
    Located on the rear panel are video input and output connectors, audio input and output, USB, 
    footswitch connector power connector and control connections for VCR or DVD Recorder, printer, and 
    service tools. 
    This section indicates the pin  assignment for each connector.
    NOTE: Each outer (case) ground line of peripheral/accessory connectors are protectively grounded.  Signal ground lines are not isolated , except the Service port (3). All of signal lines (include signal 
    GND) of the Service port are isolated. Figure 3-14   Rear Connector Panel 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
    3-20 Section 3-4 - Preparing for Installation
    3-6-4-1 External I/O Pin Outs
    Pin No.SignalPin No.Signal
    1RED9 N /A
    2 GREEN 10SGND
    3B LU E1 1 N /A
    4N /A1 2N /A
    5 GND 13HSYNC
    6 RGND 14VSYNC
    7 GGND 15N/A
    8B GND
    Table 3-6    Pin Assignments of DB15 connector for External VGA
    Pin No.SignalPin No.Signal
    1+ 5 V DC3D ATA  +
    2 DATA - 4GND
    Table 3-7    Pin Assignments of USB
    Pin No.SignalPin No.Signal
    1T X+5 N C
    2 TX- 6RX-
    3R X+7 N C
    4N C8N C
    Table 3-8    Pin Assignments of Ethernet 
    						