GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL Section 7-5 - Common Diagnostics 7 - 11 7-5-2-2 Hard Drive Tests •Hard Drive Test (All) - The Hard Drive test examines your hard disk drive using a series of tests to ensure that your drive can perform its functions. The following tests are performed on your hard drive to ensure the hard disk controller and the drive mechanism are working correctly. The disk surface itself is also checked. NOTE: The hard drive test can take over ten minutes. •Hard Drive: Linear Seek - Hard disk drive heads are moved from track 0 to the maximum track one track at a time. • Hard Drive: Random Seek - Hard disk drive heads are moved randomly several hundred times. • Hard Drive: Funnel Seek - Hard disk drive heads are moved in a funnel fashion, i.e., from the first track on the drive to the la st, then to the second track, then to the second to last track, then to the third track, and so on. • Hard Drive: Surface Scan - This test scans for surface defects on the hard disk drive. 7-5-2-3 Memory Tests •Memory Test (All) - Eighteen test patterns are used to test memory locations. If errors are detected, the address of the block the error was detected in is shown. During testing, these diagnostics will also look out for pa rity errors and other exceptions. These will be added to the Test Log as errors if encountered. • Memory: Pattern Test - PC-Doctor uses several test patterns to test as much memory as possible. • Memory: Parity Test - This test checks for parity errors on the memory bus during intensive string transfer operations. If an error is detected, the address of the failed block is displayed.
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 7 - 12 Section 7-5 - Common Diagnostics 7-5-2-4 CD/DVD Drive Test NOTE: Although the global aspect of the diagnostics a llow for DVD testing, the LOGIQ™ P5/A5/A5Pro does not have a DVD drive.(Option) • DVD/CD-ROM Test (All) - The CD-ROM/DVD test examines the CD-ROM/DVD drive to make sure it is working properly. You may use either a data CD-ROM disk or a DVD disk for these tests. You cannot use an audio CD, such as a regu lar commercial music CD, with this test. NOTE: The PC is sometimes slow in re cognizing a newly inserted CD-ROM/DVD disk. It is best to wait a few seconds after inserting a disk into the drive before you run any tests. There are two types of CD-ROM/DVD tests: 1.)Seek Tests: Ensures the drive can access all locations on a disk, and 2.) Read Tests: Ensures the drive can read all the data. • DVD/CD-ROM: Linear Seek - Tests that all locations can be accessed on a CD-ROM or DVD disk in a linear fashion, starting from the first to the last. • DVD/CD-ROM: Random Seek - Tests that all locations can be accessed on a CD-ROM or DVD disk randomly. • DVD/CD-ROM: Funnel Seek - Tests that all locations can be accessed on a CD-ROM or DVD disk in a funnel fashion, starting from the first then moving to the last, then, at the next pass starting from the second location then moving to the next to the last, and so on • DVD/CD-ROM: Linear Read - Reads the data on the CD-ROM or DVD in a linear fashion, starting from the start to the end. • DVD/CD-ROM: Random Read - Reads the data on the CD-ROM or DVD randomly. 7-5-2-5 Video Test •Video Test (All) - This diagnostic tests your systems vide o capabilities. This involves testing the video memory with 18 patterns, testing your graphics acceleration, and text output. You will see these tests being performed on your monitor. You can cancel this test at any time by hitting the Escape (Esc) key. The following tests are performed on your monitor: • Video: Memory - Video memory is tested by filling the vi deo buffer with 18 test patterns, one pattern at a time. The tests will fill the en tire screen with a single color. • Video: Data Transfer - This tests the graphics acceleration part of your video controller. These tests will appear on your screen as black and white concentric squares and rectangles of various sizes and colors. If errors are detected, the locations of the problems are displayed. • Video: Text Output - This test prints a text string in random sizes and colors to tests your video device driver and video controller.
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL Section 7-5 - Common Diagnostics 7 - 13 7-5-2-6 USB Test (Not yet supported) • USB Test (All) • USB Configuration Test • USB Devices Test 7-5-2-7 PCI Board Configuration Test A bus is a set of circuits, wires or connectors whic h connect the various components of your system. A PCI (Peripheral Component Interconnect) bus is a fast standard bus which is common in Pentium systems. Various tests are run on the PCI bus in order to ensure it is working up to speed. The motherboard is scanned to verify the configur ation space and to make sure the diagnostics can communicate with the board. 7-5-2-8 Network Test
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 7 - 14 Section 7-5 - Common Diagnostics 7-5-3 PC Diagnostics (Interactive Tests) 7-5-3-1 Keyboard Test 1.) Click Keyboard Test from the PC Diagnostics interactive tests menu. 2.) Click the Test button. A keyboard layout is di splayed on which you can see the keys of your keyboard. 3.) Press each key on your keyboard once and make sure that the corresponding keys on-screen are removed from view. If a key on-screen is not re moved, that key may be damaged and you may have to have your keyboard repaired or replaced. 4.) To reset the keyboard, click the Reset Keys button. 5.) To exit the test, click Close. 7-5-3-2 Audio Test 1.) Click Audio Test from the PC Diagnostic interactive tests menu. 2.) To test the WAV sound reproduction click Left Chan nel, Right Channel, or Both Channels to test your speakers. You should hear a guitar chord. 3.) To test MIDI sound reproduction click the butt on under MIDI. You should hear a few seconds of recorded music. 4.) Click on the Beep button. You should hear a low “beep” from your PC. If no sound is produced in these tests, choose the More Info button in the Audio Test dialog box for information about possible causes and solutions.
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL Section 7-5 - Common Diagnostics 7 - 15 7-5-3-3 Mouse Test 1.) Click Mouse Test from the PC Diagnostic interactive tests menu. The Mouse test dialog box presents you like Figure 7-8. 2.) The Mouse coordinates are displayed when the trackball is moved. 3.) Press the right Button (Set B/Pause Button) on the front panel, the Mouse wh ich is turned to Yellow on the right side is shown on the monitor. 4.) Left Button is the same as above. Figure 7-8 Mouse Test Figure 7-9
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 7 - 16 Section 7-5 - Common Diagnostics 7-5-4 FRU Test 7-5-4-1 SYSCONPM/SYSCONCM These programs are provided for testing the SYSCONPM/SYSCONCM board. Table 7-2 FRU Test, SYSCONPM/SYSCONCM Diagnostics Menu MenuDescriptionsUser Level Tools Full Test Performs all of the L1 SYSCON board tests listed below All None Assy Revision Test Test EEPROM. Send test pattern to write to EEPROM and read Displays the board revision of the SYSCONPM/SYSCONCM board All None USC I/F Test The CL1TRX, P3RLY, optional ACWD boards are connected to the SYSCONPM/ SYSCONCM via the USC bus. The DSP in SYSCONPM/SYSCONCM accesses the boards above to check if the USC bus is connected properly to each board. All None FEBUS TestChecks if The DSP in SYSCONPM/SYSC ONCM accessed the boards, CL1TRX/ BL1TRX and ACWD, via the FEBUS. All None TRIG TestCheck the trigger max, min and ever interval All None PCI I/F TestSends the command to the DSP to check the SDRAM in SYSCONPM/SYSCONCM All None DSP Memory TestSends the command to the DSP to check the internal DSP memory All None Board TemperatureRead SYSCON Board Temperature. All None CHACOM I/F TestSends the command to read / write data from DSP to CHACOM GE FE Only None
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL Section 7-5 - Common Diagnostics 7 - 17 7-5-4-2 CL1TRX/BL1TRX These programs are provided for testing the CL1TRX /BL1TRZboard. Table 7-3 FRU Test, CL1TRX/BL1TRX Diagnostics Menu MenuDescriptionsUser Level Tools Full Test Performs all of the CL1TRX/BL1TRX board tests listed below All None Assy Revision TestTest EEPROM. Send test pattern to write to EEPROM and read Displays the board revision of the CL1TRX/BL1TRX boardAll None USC I/F TestChecks if the DSP in SYSCONPM/SYSCO NCM can access the CL1TRX/BL1TRX boards via USB bus. All None FEBUS TestChecks if the DSP in SYSCONPM/SYSCO NCM can access the CL1TRX/BL1TRX board via FE bus. All None HV TestRead HV Power All None PGC TestSends the test data High, Mid, Low level to PGC Register, then CL1TRX/BL1TRX read result data and verify it.All None TEST Wave TestSends the test data High, Mid, Low level to Test wave Register, then CL1TRX/ BL1TRX read result data and verify it.All None TRX Data Memory TestSend the command to read / write to TRX data memory. All None OQX2 TestThe DSP in L1SYSCON reads and writes the OQCARD registers in CL1TRX/BL1TRX via FE bus. All None TPG2 TestThe DSP in L1SYSCON reads and writes the TPG2 registers in CL1TRX/BL1TRX via FE bus. RX Channel TestSend test wave pattern to the input stage of preamp on CL1TRX/BL1TRX, perform 64 channel scanning, Read raw data, and analyze it. TX Channel Test Send test wave pattern to the CL1TRX/BL1TRX, read returned small signal, Read the data and analyze it. Note: APS/APS Pro HV test is required prior to the TRX TX channel test.
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 7 - 18 Section 7-5 - Common Diagnostics 7-5-4-3 P3RLY/P2RLY These programs are provided for testing the P3RLY/P2RLY board. NOTE: A Convex Probe is needed for RLY Channel Test. 7-5-4-4 ACWD These programs are provided for testing the ACWD board. Table 7-4 FRU Test, P3RLY Diagnostics Menu MenuDescriptionsUser Level Tools Full Test Performs all of the P3RLY/P2RLY board tests listed below GE FE Only None Assy Revision TestTest EEPROM. Send test pattern to write to EEPROM and readDisplays the board revision of the P3RLY/P2RLY board All None FEBUS TestChecks if the DSP in SYSCONPM/SYSCONCM can access the P3RLY/P2RLY board via FE bus. GE FE Only None SHV TestSend the command to SHV resister in P3RLY/P2RLY board to read SHV status. GE FE Only None OPEN TestChecks whether probe connected or not on P3RLY/P2RLY. GE FE OnlyProbe PROBE CODEChecks if the 8 bits of the Probe Code in the probe connector is properly connected. If connected probe is available, display code and Probe name. Table 7-5 FRU Test, ACWD Diagnostics Menu MenuDescriptionsUser Level Tools Full Test Performs all of the ACWD board test listed below All None Assy Revision TEstTest EEPROM.Send test pattern to write to EEPROM and readDisplays the board revision of the ACWD board All None USC Interface TestChecks if the DSP in SYSCONPM/SYSCNC M can access the ACWD boards via USB bus. All None FEBUS TestAllows SYSCONPM/SYSCNCM to access FE bus interface in the ACWD in order to check if FE bus line is normal. IQ Data ReadSend the command to perform CW scanning, read IQ data from assigned memory, analyze the data with FFT algorithm. Probe ConnectionChecks whether ACWD probe connected or not. Channel TestGenerate test pattern, perform 32 channel scanning in ACWD, read the data and analyze it.
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL Section 7-5 - Common Diagnostics 7 - 19 7-5-4-5 APS/APS Pro These programs are provided for testing the APS board. 7-5-4-6 FRONT PANEL These programs are provided for testing the Front panel. NOTE: For the EchoBack test, it is highly recomme nded to press a button less than two times per a second. NOTE: When Abort button is presse d to finish EchoBack test, you have to wait until “Completed.” message is displayed to proceed other tests. Table 7-6 FRU Test, APS Diagnostics Menu MenuDescriptionsUser Level Tools Full Test Performs all of the APS test listed below All None HV voltage stabilityRead HV voltage and display HV values.All None LV voltage stabilityRead HV voltage and display LV values.All None Table 7-7 FRU Test, ACWD Diagnostics Menu MenuDescriptionsUser Level Tools FRONT PANEL Overall Performs all of the Front Panel tests listed below All None FRONT PANEL RevisionTest EEPROM.Send test pattern to write to EEPROM and read.Displays the board revision of the Front Panel. All None ROMMakes a check of internal ROM check sum. All None RAM Writes / reads data over all address of RAM and also makes a check of control program. All None EEPROM Writes / reads data over all address of EEPROM All None SWITCHTest switch in Front Panel. All None EchobackRotate or depress any key, Verify that it is displayed on the CRT monitor. To finish the test, press “Abort” key and wait until complete message appears. All None
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 7 - 20 Section 7-6 - LED Descriptions Section 7-6 LED Descriptions 7-6-1 CL1TRX/BL1TRX Assy The CL1TRX/BL1TRX Assy is located at the No.1 slot in the NEST Assy. 7-6-2 P3RLY/P2RLY Assy The P3RLY Assy is lo cated at the front of the NEST Assy. 7-6-3 SYSCONPM/SYSCONCM Assy The SYSCONPM/SYSCONCM Assy is located at the No.3 slot of the NEST Assy. Table 7-8 LEDs on CL1TRX/BL1TRX Assy LEDDescriptionsNormal State DS1 FPGA configuration OK status Indicator Normally Blinking Table 7-9 LEDs on P3RLY/P2RLY Assy LEDDescriptionsNormal State D1 MUX embedded Probe Recognition Indicator Normally On D2 Selected probe connector position number bit 0 Depend on live probe position D3 Selected probe connector position number bit 1 Depend on live probe position D4 FPGA Configuration OK status indicator Normally On Table 7-10 LEDs on L1SYSCON Assy LEDDescriptionsNormal State DS1 CPDI FPGA configuration OK indicator Normally On DS2 Frame Start signal. Scan running stat us indicatorNormally blinking on scanning DS3 FEBC FPGA configuration OK indicator Normally On DS4 DSP operation OK indicator Normally blinking