    Section 7-5 - Common Diagnostics 7 - 11
    7-5-2-2 Hard Drive Tests
    •Hard Drive Test (All)  - The Hard Drive test examines your hard disk drive using a series of tests to 
    ensure that your drive can perform its functions.
    The following tests are performed on your hard drive to ensure the hard disk controller and the drive 
    mechanism are working correctly. The disk surface itself is also checked.
    NOTE: The hard drive test can take over ten minutes. •Hard Drive: Linear Seek  - Hard disk drive heads are moved from track 0 to the maximum track one 
    track at a time.
    • Hard Drive: Random Seek  - Hard disk drive heads are moved randomly several hundred times.
    • Hard Drive: Funnel Seek  - Hard disk drive heads are moved in a funnel fashion, i.e., from the first 
    track on the drive to the la st, then to the second track, then to the second to last track, then to the 
    third track, and so on.
    • Hard Drive: Surface Scan  - This test scans for surface defects on the hard disk drive.
    7-5-2-3 Memory Tests •Memory Test (All)  - Eighteen test patterns are used to  test memory locations. If errors are 
    detected, the address of the block the error was  detected in is shown. During testing, these 
    diagnostics will also look out for pa rity errors and other exceptions. These will  be added to the Test 
    Log as errors if encountered. 
    • Memory: Pattern Test  - PC-Doctor uses several test patterns  to test as much memory as possible.
    • Memory: Parity Test  - This test checks for parity errors on  the memory bus during intensive string 
    transfer operations. If an error is detected,  the address of the failed block is displayed. 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    7 - 12 Section 7-5 - Common Diagnostics
    7-5-2-4 CD/DVD Drive Test NOTE: Although the global aspect of the diagnostics a llow for DVD testing, the LOGIQ™ P5/A5/A5Pro does 
    not have a DVD drive.(Option)
    • DVD/CD-ROM Test (All)  - The CD-ROM/DVD test examines  the CD-ROM/DVD drive to make sure 
    it is working properly. You may use either a data  CD-ROM disk or a DVD disk for these tests. You 
    cannot use an audio CD, such as a regu lar commercial music CD, with this test.
    NOTE: The PC is sometimes slow in re cognizing a newly inserted CD-ROM/DVD disk. It is best to wait a few 
    seconds after inserting a disk into the drive before you run any tests. 
    There are two types of CD-ROM/DVD tests: 
    1.)Seek Tests:  Ensures the drive can access all locations on a disk, and 
    2.) Read Tests:  Ensures the drive can read all the data.
    • DVD/CD-ROM: Linear Seek - Tests that all locations can be  accessed on a CD-ROM or DVD disk 
    in a linear fashion, starting from the first to the last.
    • DVD/CD-ROM: Random Seek  - Tests that all locations can be accessed on a CD-ROM or DVD 
    disk randomly.
    • DVD/CD-ROM: Funnel Seek  - Tests that all locations can be  accessed on a CD-ROM or DVD disk 
    in a funnel fashion, starting from  the first then moving to the last, then, at the next pass starting from 
    the second location then moving to the next to the last, and so on
    • DVD/CD-ROM: Linear Read  - Reads the data on the CD-ROM or  DVD in a linear fashion, starting 
    from the start to the end.
    • DVD/CD-ROM: Random Read  - Reads the data on the CD-ROM or DVD randomly.
    7-5-2-5 Video Test •Video Test (All)  - This diagnostic tests your systems vide o capabilities. This involves testing the 
    video memory with 18 patterns, testing your graphics acceleration, and text output. You will see 
    these tests being performed on your monitor.
    You can cancel this test at any time by hitting the Escape (Esc) key.
    The following tests are performed on your monitor:
    • Video: Memory  - Video memory is tested by filling the vi deo buffer with 18 test patterns, one pattern 
    at a time. The tests will fill the en tire screen with a single color.
    • Video: Data Transfer  - This tests the graphics acceleration part of your video controller. These 
    tests will appear on your screen as black and white concentric squares and rectangles of various 
    sizes and colors. If errors are detected, the locations of the problems are displayed.
    • Video: Text Output  - This test prints a text string in random sizes and colors to tests your video 
    device driver and video controller. 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
    Section 7-5 - Common Diagnostics 7 - 13
    7-5-2-6 USB Test (Not yet supported)
    • USB Test (All)
    • USB Configuration Test
    • USB Devices Test
    7-5-2-7 PCI Board Configuration Test A bus is a set of circuits, wires or connectors whic h connect the various components of your system. A 
    PCI (Peripheral Component Interconnect) bus is a fast standard bus which is common in Pentium 
    systems. Various tests are run on the PCI bus in  order to ensure it is working up to speed. 
    The motherboard is scanned to verify the configur ation space and to make sure the diagnostics can 
    communicate with the board.
    7-5-2-8 Network Test 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    7 - 14 Section 7-5 - Common Diagnostics
    7-5-3 PC Diagnostics  (Interactive Tests)
    7-5-3-1 Keyboard Test
    1.) Click Keyboard Test  from the PC Diagnostics interactive tests menu.
    2.) Click the Test button. A keyboard layout is di splayed on which you can see the keys of your 
    keyboard.
    3.) Press each key on your keyboard  once and make sure that the corresponding keys on-screen are 
    removed from view. If a key on-screen is not re moved, that key may be damaged and you may have 
    to have your keyboard repaired or replaced.
    4.) To reset the keyboard, click the Reset Keys button. 
    5.) To exit the test, click Close.
    7-5-3-2 Audio Test 1.) Click Audio Test  from the PC Diagnostic interactive tests menu.
    2.) To test the WAV sound reproduction click Left Chan nel, Right Channel, or Both Channels to test 
    your speakers. You should hear a guitar chord.
    3.) To test MIDI sound reproduction click the butt on under MIDI. You should hear a few seconds of 
    recorded music.
    4.) Click on the Beep button. You should  hear a low “beep” from your PC.
    If no sound is produced in these tests, choose the More Info button in the Audio Test dialog box for 
    information about possible causes and solutions. 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
    Section 7-5 - Common Diagnostics 7 - 15
    7-5-3-3 Mouse Test
    1.) Click Mouse Test  from the PC Diagnostic interactive tests menu.
    The Mouse test dialog box presents you like Figure 7-8.
    2.) The Mouse coordinates are displayed when the trackball is moved.
    3.) Press the right Button (Set B/Pause Button) on the front panel, the Mouse wh ich is turned to Yellow 
    on the right side is shown on the monitor.
    4.) Left Button is the same as above. Figure 7-8   Mouse Test
    Figure 7-9    
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    7 - 16 Section 7-5 - Common Diagnostics
    7-5-4 FRU Test
    7-5-4-1 SYSCONPM/SYSCONCM These programs  are provided for testing  the SYSCONPM/SYSCONCM board.
    Table 7-2    FRU Test, SYSCONPM/SYSCONCM Diagnostics Menu
    MenuDescriptionsUser 
    Level
    Tools
    Full Test
    Performs all of the L1 SYSCON board tests listed below All None
    Assy Revision Test
    Test EEPROM.
    Send test pattern to write to EEPROM and read
    Displays the board revision of the SYSCONPM/SYSCONCM board All
    None
    USC I/F Test
    The CL1TRX, P3RLY, optional ACWD boards are connected to the SYSCONPM/
    SYSCONCM via the USC bus. 
    The DSP in SYSCONPM/SYSCONCM accesses  the boards above to check if the 
    USC bus is connected properly to each board.  All None
    FEBUS TestChecks if The DSP in SYSCONPM/SYSC
    ONCM accessed the boards, CL1TRX/
    BL1TRX and ACWD, via the FEBUS. All None
    TRIG TestCheck the trigger max, min and ever interval 
    All None
    PCI I/F TestSends the command to the DSP to check the SDRAM in SYSCONPM/SYSCONCM All None
    DSP Memory TestSends the command to the DSP to check the internal DSP memory All None
    Board TemperatureRead SYSCON Board Temperature. All None
    CHACOM I/F TestSends the command to read / write data from DSP to CHACOM GE FE 
    Only
    None 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
    Section 7-5 - Common Diagnostics 7 - 17
    7-5-4-2 CL1TRX/BL1TRX
    These programs are provided for testing the CL1TRX /BL1TRZboard.
    Table 7-3    FRU Test, CL1TRX/BL1TRX Diagnostics Menu
    MenuDescriptionsUser 
    Level
    Tools
    Full Test
    Performs all of the CL1TRX/BL1TRX board tests listed below All None
    Assy Revision 
    TestTest EEPROM.
    Send test pattern to write to EEPROM and read
    Displays the board revision of the CL1TRX/BL1TRX boardAll
    None
    USC I/F TestChecks if the DSP in SYSCONPM/SYSCO NCM can access the CL1TRX/BL1TRX 
    boards via USB bus. All None
    FEBUS TestChecks if the DSP in SYSCONPM/SYSCO
    NCM can access the CL1TRX/BL1TRX 
    board via FE bus. All None
    HV TestRead HV Power
    All None
    PGC TestSends the test data High, Mid, Low level to PGC Register, then CL1TRX/BL1TRX 
    read result data and verify it.All None
    TEST Wave TestSends the test data High, Mid, Low level to Test wave Register, then CL1TRX/
    BL1TRX read result data and verify it.All None
    TRX Data 
    Memory TestSend the command to read / write to TRX data memory.
    All None
    OQX2 TestThe DSP in L1SYSCON reads and writes the OQCARD registers in CL1TRX/BL1TRX 
    via FE bus. All None
    TPG2 TestThe DSP in L1SYSCON reads and writes the 
    TPG2 registers in CL1TRX/BL1TRX via 
    FE bus.
    RX Channel TestSend test wave pattern to the input stage of preamp on CL1TRX/BL1TRX, perform 64 
    channel scanning, Read raw data, and analyze it. 
    TX Channel Test
    Send test wave pattern to the CL1TRX/BL1TRX, read returned small signal, Read the 
    data and analyze it. 
    Note: APS/APS Pro HV test is required prior to the TRX TX channel test. 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    7 - 18 Section 7-5 - Common Diagnostics
    7-5-4-3 P3RLY/P2RLY These programs are provided for testing the P3RLY/P2RLY board.
    NOTE: A Convex Probe is needed for RLY Channel Test.
    7-5-4-4 ACWD These programs are provided for testing the ACWD board. Table 7-4    FRU Test, P3RLY Diagnostics Menu
    MenuDescriptionsUser 
    Level
    Tools
    Full Test
    Performs all of the P3RLY/P2RLY board tests listed below GE FE 
    Only
    None
    Assy Revision 
    TestTest EEPROM.
    Send test pattern to write to EEPROM and readDisplays the board revision of the 
    P3RLY/P2RLY board All
    None
    FEBUS TestChecks if the DSP in SYSCONPM/SYSCONCM  can access the P3RLY/P2RLY board 
    via FE bus. GE FE 
    Only
    None
    SHV TestSend the command to SHV resister in P3RLY/P2RLY board to read SHV status. GE FE 
    Only
    None
    OPEN TestChecks whether probe connected or not on P3RLY/P2RLY. GE FE 
    OnlyProbe
    PROBE CODEChecks if the 8 bits of the Probe Code in the probe connector is properly connected. 
    If connected probe is available, display code and Probe name. 
    Table 7-5    FRU Test, ACWD Diagnostics Menu
    MenuDescriptionsUser 
    Level
    Tools
    Full Test
    Performs all of the ACWD board test listed below All None
    Assy Revision 
    TEstTest EEPROM.Send test pattern to write to EEPROM and readDisplays the board 
    revision of the ACWD board All None
    USC Interface 
    TestChecks if the DSP in SYSCONPM/SYSCNC
    M can access the ACWD boards via USB 
    bus. All None
    FEBUS TestAllows SYSCONPM/SYSCNCM to access FE bus 
    interface in the ACWD in order to 
    check if FE bus line is normal. 
    IQ Data ReadSend the command to perform CW scanning, read IQ data from assigned memory, 
    analyze the data with FFT algorithm. 
    Probe 
    ConnectionChecks whether ACWD probe connected or not.
    Channel TestGenerate test pattern, perform 32 channel scanning in ACWD, read the data and 
    analyze it.  
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
    Section 7-5 - Common Diagnostics 7 - 19
    7-5-4-5 APS/APS Pro
    These programs  are provided for testing the APS board.
    7-5-4-6 FRONT PANEL These programs are provided for testing the Front panel.
    NOTE: For the EchoBack test, it is highly recomme nded to press a button less than two times per a 
    second.
    NOTE: When Abort button is presse d to finish EchoBack test, you have to wait until “Completed.” 
    message is displayed to proceed other tests. Table 7-6    FRU Test, APS Diagnostics Menu
    MenuDescriptionsUser 
    Level
    Tools
    Full Test
    Performs all of the APS test listed below All None
    HV voltage 
    stabilityRead HV voltage and display HV values.All None
    LV voltage 
    stabilityRead HV voltage and display LV values.All None
    Table 7-7    FRU Test, ACWD Diagnostics Menu
    MenuDescriptionsUser 
    Level
    Tools
    FRONT PANEL 
    Overall
    Performs all of the Front Panel tests listed below All None
    FRONT PANEL 
    RevisionTest EEPROM.Send test pattern to write to EEPROM and read.Displays the board 
    revision of the Front Panel. All None
    ROMMakes a check of internal ROM check sum. 
    All None
    RAM Writes / reads data over all address of RAM and also makes a check of control 
    program. All None
    EEPROM Writes / reads data over all address of EEPROM
    All None
    SWITCHTest switch in Front Panel. All None
    EchobackRotate or depress any key, Verify that it is displayed on the CRT monitor. To finish the 
    test, press “Abort” key and wait until complete message appears.  All None 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
    7 - 20 Section 7-6 - LED Descriptions
    Section 7-6
    LED Descriptions
    7-6-1 CL1TRX/BL1TRX Assy
    The CL1TRX/BL1TRX Assy is located at the No.1 slot in the NEST Assy.
    7-6-2 P3RLY/P2RLY Assy
    The P3RLY Assy is lo cated at the front of the NEST Assy.
    7-6-3 SYSCONPM/SYSCONCM Assy
    The SYSCONPM/SYSCONCM Assy is located  at the No.3 slot of the NEST Assy.
    Table 7-8    LEDs on CL1TRX/BL1TRX Assy
    LEDDescriptionsNormal State
    DS1 FPGA configuration OK status Indicator Normally Blinking
    Table 7-9    LEDs on P3RLY/P2RLY Assy
    LEDDescriptionsNormal State
    D1 MUX embedded Probe Recognition Indicator Normally On
    D2 Selected probe connector position number bit 0 Depend on live probe position
    D3 Selected probe connector position number bit 1 Depend on live probe position
    D4 FPGA Configuration OK status indicator Normally On
    Table 7-10    LEDs on L1SYSCON Assy
    LEDDescriptionsNormal State
    DS1 CPDI FPGA configuration OK indicator Normally On
    DS2 Frame Start signal. Scan running stat us indicatorNormally blinking on scanning
    DS3 FEBC FPGA configuration OK indicator Normally On
    DS4 DSP operation OK indicator Normally blinking 
    						
    All GE manuals Comments (0)

    Related Manuals for GE Logiq P5 Service Manual