GE Logiq P5 Service Manual
Have a look at the manual GE Logiq P5 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 10-6 - Electrical Safety Tests 10 - 23 10-6-8-1 Data Sheet for ECG Leakage Current The test passes when all readings measure less th an the value shown in the table below. Record all data on the PM Inspection Certificate. Table 10-20 Maximum Allowance Limit for ECG Leakage Current AC Power Source Maximum Allowance Limit GROUND OPEN GROUND CLOSED Patient Lead to Ground Leakage Current Test and Patient Lead to Lead Leakage Current Test 115V 10uA10uA 220/240V 500uA10uA Table 10-21 Maximum Allowance Limit for ECG Leakage Current AC Power SourceMaximum Allowance Limit Patient Lead Isolation Current Test 115V 20uA 220/240V 5mA Table 10-22 Typical Data Sheet for ECG Leakage Current ECG Power Tester Polarity SwitchTester Ground SwitchTester Lead Selector RLRALALLC ON NORM CLOSED ON REVERSE CLOSED ON NORM OPEN ONREVERSEOPEN OFFNORMCLOSED OFFREVERSECLOSED OFF NORM OPEN OFF REVERSE OPEN
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL 10 - 24 Section 10-6 - Electrical Safety Tests 10-6-9 Probe Leakage Current Test 10-6-9-1 Definition This test measures the current that would flow to ground from any of the probes through a patient who is being scanned and becomes grounded by touching some other grounded surface. 10-6-9-2 Generic Procedure Measurements should be made with the ground open and closed, with power line polarity normal and reversed, and with the unit Off and On. For each combi nation, the probe must be active to find the worst case condition. NOTE: Each probe will have some amount of leakage current, dependent on its design. Small variations in probe leakage currents are normal from probe to probe. Other variations will resu lt from differences in line voltage and test lead placement. 10-6-9-3 Meter Procedure Using Probe Adapter The Dale 600/600E provides a method for testing probes independently from the system. The meter utilizes a probe adapter to apply a test potential commonly to all connector pins. The probes imaging area is immersed in a saline solution along with a grounding probe from the meter to complete the current path. Saline solution is a mixture of water and salt. The salt adds a free ion to the water, making it conductive. Normal saline solution is 0.9% sa lt or 1/2 gram salt per 1 liter of water. If saline is not available, a mixture of 1 quart water with one or more grams of table salt, mixed thoroughly, will substitute. Follow these steps to test each transducer for leakage current. 1.) Turn the LOGIQ™ P5 unit OFF. 2.) Plug the unit into the test meter, and the meter into the tested AC wall outlet. 3.) Connect the probe for test with the meters appropriate adapter. 4.) Plug the saline probe into t he meters “CHASSIS” connector. 5.) Plug the probe adapter into the meters connector marked “EXTERNAL”. Figure 10-9 Set Up for Probe Leakage Current CAUTIONTo avoid probe damage and possible electric shock, do not immerse probes into any liquid beyond the level indicated in the probe users manual. Do not touch the probe, conductive liquid or any part of the unit under test while the ISO TEST switch is depressed. POWER OUTLET H (BLACK) POLARITY REVERSING SWITCH MOMENTARY SWITCH N (WHITE) G (GREEN)LEAKAGE TEST METER CONSOLE PROBE
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 10-6 - Electrical Safety Tests 10 - 25 10-6-9-3Meter Procedure Using Probe Adapter (cont’d) 6.) Set the meters “FUNCTION” switch to EXTERNAL position. 7.) Add the saline probe and the imaging area of the probe into the saline bath. 8.) Have unit power ON for the first par t; turn it OFF for the second half. 9.) Depress the ISO TEST rocker switch and record the highest current reading. 10.) Follow the test co nditions described in Table 10-23 for every transducer. 11.) Keep a record of the results with other hand copies of PM data. Figure 10-10 Transducer Source Leakage Current Test
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL 10 - 26 Section 10-6 - Electrical Safety Tests 10-6-9-4 No Meter Probe Adapter Procedure Follow these steps to test each transducer for leakage current. 1.) Turn the LOGIQ™ P5 unit OFF. 2.) Plug the unit into the test meter, and the meter into the tested AC wall outlet. 3.) Plug the external probe into the meters (Dale 600) “EXTERNAL” connector. 4.) Set the meters “FUNCTION” switch to EXTERNAL position. 5.) Connect the probe for test with the connector of the console. 6.) Add the saline probe and the imaging area of the probe into the saline bath. 7.) Have unit power ON for the first par t; turn it OFF for the second half. 8.) Depress the ISO TEST rocker switch and record the highest current reading. 9.) Follow the test co nditions described in Table 10-23 for every transducer. 10.) Keep a record of the results with other hand copies of PM data. Figure 10-11 Check Without Probe Adapter
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 10-6 - Electrical Safety Tests 10 - 27 10-6-9-5 Data Sheet for Transducer Source Leakage Current The test passes when all readings measure less than the values shown in Table 10-15 and Table 10- 16 . Record all data on the PM Inspection Certificate. . CAUTIONEquipment damage possibility. Never switch the Polarity and the status of Neutral when the unit is powered ON. Be sure to turn the unit power OFF before switching them using the POLARITY switch and/or the NEUTRAL switch. Otherwise, the unit may be damaged Table 10-23 Typical Data Sheet For Transducer Source Leakage Current Transducer Tested: Unit Power Tester Power Polarity SwitchTester GROUND or NEUTRAL Switch Measurement ON NORM OPEN ON NORM CLOSED ON REV OPEN ONREVCLOSED OFFNORMOPEN OFFNORMCLOSED OFFREV OPEN OFF REV CLOSED
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL 10 - 28 Section 10-6 - Electrical Safety Tests Section 10-7 When Theres Too Much Leakage Current... CHASSIS FAILS Check the ground on the power cord and plug for cont inuity. Ensure the ground is not broken, frayed, or intermittent. Replac e any defective part. Tighten all grounds. Ensure star washers are under all ground studs. Inspect wiring for bad crimps, poor connections, or damage. Test the wall outlet; verify it is grounded and is free of other wiring abnormalities. Notify the user or owner to correct any deviations. As a work around, ch eck the other outlets to see if they could be used instead. NOTE: No outlet tester can detect the condition w here the white neutral wire and the green grounding wire are reversed. If later tests indicate high leakage currents, this should be suspected as a possible cause and the outlet wiring should be visually inspected. PROBE FAILS Test the probe in another connector to isolate if the fault lies with the probe or the scanner. NOTE: Each probe will have some am ount of leakage, dependent on its design. Small variations in probe leakage currents are normal from probe to probe. Othe r variations will result from differences in line voltage and test lead placement. The maximum allowable leakage current for body surface contact probe differs from inter-cavity probe. Be sure to enter the correct probe type in the appropriate space on the check list. If excessive leakage current is slot dependent, inspect the system connector for bent pins, poor connections, and ground continuity. If the problem remains with the probe, replace the probe. PERIPHERAL FAILS Tighten all grounds. Ensure star washers are under all ground studs. Inspect wiring for bad crimps, poor connections, or damage. STILL FAILS If all else fails, begin isolation by removing the pr obes, external peripherals, then the on board ones, one at a time while monitoring the leakage current measurement. NEW UNIT If the leakage current measurement tests fail on a new unit and if situation can not be corrected, submit a Safety Failure Report to document the system problem. Remove unit from operation. ECG FAILS Inspect cables for damage or poor connections.
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 10-6 - Electrical Safety Tests 10 - 29 ULTRASOUND INSPECTION CERTIFICATE * Scan Format: Phased Array, Linear Array, Curved Array, Mechanical Array or Other FUNCTIONAL CHECKSPHYSICAL INSPECTION AND CLEANING COMMENTS: ______________________________________________________________________________ ______________________________________________________________________________ ______________________________________________________________________________ ______________________________________________________________________________ Customer Name: System ID:Dispatch Number / Date Performed: Warranty/Contract/HBS System Type Model Number:Serial Number:Manufacture Date: Probe 1: Frequency: Scan Format*: Model Number:Serial Number: Probe 2: Frequency: Scan Format*: Model Number:Serial Number: Probe 3: Frequency: Scan Format*: Model Number:Serial Number: Probe 4: Frequency: Scan Format*: Model Number:Serial Number: Probe 5: Frequency: Scan Format*: Model Number:Serial Number: Probe 6: Frequency: Scan Format*: Model Number:Serial Number: Probe 7: Frequency: Scan Format*: Model Number:Serial Number: Probe 8: Frequency: Scan Format*: Model Number:Serial Number: Probe 9: Frequency: Scan Format*: Model Number:Serial Number: Functional Check (if applicable) OK? or N/A Physical Inspection and Cleaning (if applicable) Inspect Clean B-Mode Function Console Doppler Modes Function Monitor CF-Mode Function Touch Panel M-Mode Function Air Filter Applicable Software Options Probe Holders Applicable Hardware Options External I/O Control Panel Wheels, Brakes & Swivel Locks Monitor Cables and Connectors Touch Panel GE Approved Peripherals (VCR, CD-RW, MOD, Printers) Measurement Accuracy GE Approved Peripherals
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL 10 - 30 Section 10-6 - Electrical Safety Tests ELECTRICAL SAFETY Final Check. All system covers are in plac e. System scans with all probes as expected. Accepted by: ______________________________________________________________________ Electrical Test Performed Max Value Allowed Value Measured OK? Comments Outlet (correct ground &wiring config.) System Ground Continuity Chassis Source Leakage Current - Probe Chassis Source Leakage Current - Caster Chassis Source Leakage Current - CRT Patient Lead Source Leakage (Lead to Ground) Patient Lead Source Leakage (Lead to Lead) Patient Lead Source Leakage (Isolation) Peripheral 1 Leakage Current Peripheral 1Ground Continuity Peripheral 2 Leakage Current Peripheral 2Ground Continuity Peripheral 3 Leakage Current Peripheral 3Ground Continuity PROBES Probe Number (from previous page) Max Value Allowed Max Value Measured OK? Comments Probe 1: Probe 2: Probe 3: Probe 4: Probe 5: Probe 6: Probe 7: Probe 8: Probe 9:
© 2004 and 2007, General Electric Company. GE Medical Systems, a General Electric Comp any, going to market as GE Healthcare. 3200 N. Grandview Boulevard Waukesha, Wisconsin 53149 USA www.gehealthcare.com