Home > GE > Medical Equipment > GE Logiq P5 Service Manual

GE Logiq P5 Service Manual

    Download as PDF Print this page Share this page

    Have a look at the manual GE Logiq P5 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    Page
    of 520
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
    Section 10-6 - Electrical Safety Tests 10 - 23
    10-6-8-1 Data Sheet for ECG Leakage Current The test passes when all readings measure less th an the value shown in the table below. Record all 
    data on the PM Inspection Certificate.
    Table 10-20 Maximum Allowance Limit for ECG Leakage Current
    AC Power 
    Source
    Maximum
    Allowance Limit
    GROUND OPEN
    GROUND  CLOSED
    Patient Lead to Ground Leakage Current Test
    and
    Patient Lead to Lead Leakage Current Test 115V
    10uA10uA
    220/240V 500uA10uA
    Table 10-21   Maximum Allowance Limit for ECG Leakage Current
    AC Power 
    SourceMaximum
    Allowance Limit
    Patient Lead Isolation Current Test 115V
    20uA
    220/240V 5mA
    Table 10-22 Typical Data Sheet for ECG Leakage Current
    ECG 
    Power
    Tester 
    Polarity  SwitchTester 
    Ground  SwitchTester Lead Selector
    RLRALALLC
    ON NORM CLOSED
    ON REVERSE CLOSED
    ON NORM OPEN
    ONREVERSEOPEN
    OFFNORMCLOSED
    OFFREVERSECLOSED
    OFF NORM OPEN
    OFF REVERSE OPEN 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL  
    10 - 24 Section 10-6 - Electrical Safety Tests
    10-6-9 Probe Leakage Current Test
    10-6-9-1 Definition
    This test measures the current that would flow to  ground from any of the probes through a patient who 
    is being scanned and becomes grounded by touching some other grounded surface.
    10-6-9-2 Generic Procedure Measurements should be made with the ground open and closed, with power line polarity normal and 
    reversed, and with the unit Off and On. For each combi nation, the probe must be active to find the worst 
    case condition.
    NOTE: Each probe will have some amount  of leakage current, dependent on its design.  Small variations in 
    probe leakage currents are  normal from probe to probe. Other variations will resu lt from differences in 
    line voltage and test lead placement.
    10-6-9-3 Meter Procedure Using Probe Adapter The Dale 600/600E provides a method for testing probes independently from the system. The meter 
    utilizes a probe adapter to apply a test  potential commonly to all connector pins.
    The probes imaging area is  immersed in a saline solution  along with a grounding probe from the 
    meter to complete the current path. Saline solution is a mixture of water and salt. The salt adds a free 
    ion to the water, making it conductive. Normal saline solution is 0.9% sa lt or 1/2 gram salt per 1 liter of 
    water. If saline is not available, a mixture of 1 quart water with one or more grams of table salt, mixed 
    thoroughly, will substitute.
    Follow these steps to test each transducer for leakage current.
    1.) Turn the LOGIQ™ P5 unit OFF.
    2.) Plug the unit into the test meter, and the meter into the tested AC wall outlet.
    3.) Connect the probe for test with the meters appropriate adapter.
    4.) Plug the saline probe into t he meters “CHASSIS” connector.
    5.) Plug the probe adapter into the meters connector marked “EXTERNAL”. Figure 10-9   Set Up for Probe Leakage Current
    CAUTIONTo avoid probe damage and possible electric shock, do not immerse probes into any liquid 
    beyond the level indicated in the probe users manual. Do not touch the probe, conductive liquid 
    or any part of the unit under test while the ISO TEST switch is depressed.
    POWER
    OUTLET
    H   (BLACK)
    POLARITY REVERSING SWITCH
    MOMENTARY
    SWITCH
    N   (WHITE)
    G   (GREEN)LEAKAGE TEST
    METER
    CONSOLE
    PROBE 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
    Section 10-6 - Electrical Safety Tests 10 - 25
    10-6-9-3Meter Procedure Using Probe Adapter  (cont’d)
    6.) Set the meters “FUNCTION”  switch to EXTERNAL position.
    7.) Add the saline probe and the imaging area of the probe into the saline bath.
    8.) Have unit power ON for the first par t; turn it OFF for the second half.
    9.) Depress the ISO TEST rocker switch and record the highest current reading.
    10.) Follow the test co nditions described in Table 10-23 for every transducer.
    11.) Keep a record of the results with other hand copies of PM data. Figure 10-10   Transducer Source Leakage Current Test 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL  
    10 - 26 Section 10-6 - Electrical Safety Tests
    10-6-9-4 No Meter Probe Adapter Procedure
    Follow these steps to test each transducer for leakage current.
    1.) Turn the LOGIQ™ P5 unit OFF.
    2.) Plug the unit into the test meter, and the meter into the tested AC wall outlet.
    3.) Plug the external probe into the meters (Dale 600) “EXTERNAL” connector.
    4.) Set the meters “FUNCTION”  switch to EXTERNAL position.
    5.) Connect the probe for test with the connector of the console.
    6.) Add the saline probe and the imaging area of the probe into the saline bath.
    7.) Have unit power ON for the first par t; turn it OFF for the second half.
    8.) Depress the ISO TEST rocker switch and record the highest current reading.
    9.) Follow the test co nditions described in Table 10-23 for every transducer.
    10.) Keep a record of the results with other hand copies of PM data. Figure 10-11   Check Without Probe Adapter 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
    Section 10-6 - Electrical Safety Tests 10 - 27
    10-6-9-5 Data Sheet for Transducer Source Leakage Current The test passes when all readings measure less than the values shown in Table 10-15  and Table 10-
    16 . Record all data on the PM Inspection Certificate.
    . 
    CAUTIONEquipment damage possibility. Never switch the Polarity and the status of Neutral when the unit 
    is powered ON. Be sure to turn the unit power OFF before switching them using the POLARITY 
    switch and/or the NEUTRAL switch. Otherwise, the unit may be damaged
    Table 10-23 Typical Data Sheet For Transducer Source Leakage Current
    Transducer Tested:
     Unit Power
    Tester Power Polarity  SwitchTester GROUND or NEUTRAL Switch
    Measurement
    ON NORM OPEN
    ON NORM CLOSED
    ON REV OPEN
    ONREVCLOSED
    OFFNORMOPEN
    OFFNORMCLOSED
    OFFREV OPEN
    OFF REV CLOSED 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL  
    10 - 28 Section 10-6 - Electrical Safety Tests
    Section 10-7
    When Theres Too Much Leakage Current...
    CHASSIS FAILS
    Check the ground on the power cord and plug for cont inuity. Ensure the ground is not broken, frayed, 
    or intermittent. Replac e any defective part.
    Tighten all grounds. Ensure star  washers are under all ground studs.
    Inspect wiring for bad crimps, poor connections, or damage.
    Test the wall outlet; verify it is grounded and is free of other wiring abnormalities. Notify the user or 
    owner to correct any deviations. As a work around, ch eck the other outlets to see if they could be used 
    instead.
    NOTE: No outlet tester can detect the condition w here the white neutral wire and the green grounding 
    wire are reversed. If later tests indicate high leakage currents, this should be suspected as a 
    possible cause and the outlet wiring should be visually inspected.
    PROBE FAILS
    Test the probe in another connector to isolate if the fault lies with the probe or the scanner.
    NOTE: Each probe will have some am ount of leakage, dependent on its design.  Small variations in 
    probe leakage currents are normal from probe to probe. Othe r variations will result from 
    differences in line voltage and test lead placement. The maximum allowable leakage current for 
    body surface contact probe differs from inter-cavity probe. Be sure to enter the correct probe 
    type in the appropriate space on the check list.
    If excessive leakage current is slot dependent,  inspect the system connector for bent pins, poor 
    connections, and ground continuity.
    If the problem remains with the probe, replace the probe.
    PERIPHERAL FAILS
    Tighten all grounds. Ensure star  washers are under all ground studs.
    Inspect wiring for bad crimps, poor connections, or damage.
    STILL FAILS
    If all else fails, begin isolation by removing the pr obes, external peripherals, then the on board ones, 
    one at a time while monitoring the leakage current measurement.
    NEW UNIT
    If the leakage current measurement tests fail on a new  unit and if situation can not be corrected, submit 
    a Safety Failure Report to  document the system problem. Remove unit from operation.
    ECG FAILS
    Inspect cables for damage or poor connections. 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
    Section 10-6 - Electrical Safety Tests 10 - 29
    ULTRASOUND INSPECTION CERTIFICATE  
    * Scan Format: Phased Array, Linear Array,  Curved Array, Mechanical Array or Other
    FUNCTIONAL CHECKSPHYSICAL INSPECTION AND CLEANING 
    COMMENTS: 
    ______________________________________________________________________________
    ______________________________________________________________________________
    ______________________________________________________________________________
    ______________________________________________________________________________
    Customer Name: System ID:Dispatch Number / Date Performed: Warranty/Contract/HBS
    System Type Model Number:Serial Number:Manufacture Date:
    Probe 1: Frequency: Scan Format*: Model Number:Serial Number:
    Probe 2: Frequency: Scan Format*: Model Number:Serial Number:
    Probe 3: Frequency: Scan Format*: Model Number:Serial Number:
    Probe 4: Frequency: Scan Format*: Model Number:Serial Number:
    Probe 5: Frequency: Scan Format*: Model Number:Serial Number:
    Probe 6: Frequency: Scan Format*: Model Number:Serial Number:
    Probe 7: Frequency: Scan Format*: Model Number:Serial Number:
    Probe 8: Frequency: Scan Format*: Model Number:Serial Number:
    Probe 9: Frequency: Scan Format*: Model Number:Serial Number:
    Functional Check (if applicable) OK? or 
    N/A Physical Inspection and Cleaning 
    (if applicable) Inspect Clean
    B-Mode Function Console
    Doppler Modes Function Monitor
    CF-Mode Function Touch Panel
    M-Mode Function Air Filter
    Applicable Software Options Probe Holders
    Applicable Hardware Options External I/O
    Control Panel Wheels, Brakes & Swivel Locks
    Monitor Cables and Connectors
    Touch Panel GE Approved Peripherals (VCR, CD-RW, MOD, 
    Printers)
    Measurement Accuracy
    GE Approved Peripherals 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL  
    10 - 30 Section 10-6 - Electrical Safety Tests
    ELECTRICAL SAFETY
    Final Check. All system covers are in plac e. System scans with all probes as expected.
    Accepted by: ______________________________________________________________________
    Electrical Test Performed Max Value 
    Allowed Value 
    Measured OK? Comments
    Outlet (correct ground &wiring config.)
    System Ground Continuity
    Chassis Source Leakage Current - Probe
    Chassis Source Leakage Current - Caster
    Chassis Source Leakage Current - CRT
    Patient Lead Source Leakage
    (Lead to Ground)
    Patient Lead Source Leakage
    (Lead to Lead)
    Patient Lead Source Leakage
    (Isolation)
    Peripheral 1 Leakage Current
    Peripheral 1Ground Continuity
    Peripheral 2 Leakage Current
    Peripheral 2Ground Continuity
    Peripheral 3 Leakage Current
    Peripheral 3Ground Continuity
    PROBES
    Probe Number 
    (from previous page) Max Value 
    Allowed Max Value 
    Measured OK? Comments
    Probe 1:
    Probe 2:
    Probe 3:
    Probe 4:
    Probe 5:
    Probe 6:
    Probe 7:
    Probe 8:
    Probe 9: 
    						
    							© 2004 and 2007, General Electric Company.
    GE Medical Systems, a General Electric Comp any, going to market as GE Healthcare.
    3200 N. Grandview Boulevard
    Waukesha, Wisconsin  53149
    USA
    www.gehealthcare.com 
    						
    All GE manuals Comments (0)

    Related Manuals for GE Logiq P5 Service Manual