    DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL
    Section 5-4 - Main Board Detail 5 - 49
    5-10-9 Utilities
    The Utilities  page contains several miscellaneous tools.
    Figure 5-48   GE Service Utilities PageFigure 5-49   Customer Utilities Page 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL
    5 - 50Section 5-4 - Main Board Detail
    5-10-10 Replacement
    The Replacement  page is intended to contain the tools used to track replacement parts used in the 
    system.
    5-10-11 PM
    The  PM page is intended to contain the tools used in periodic maintenance of the system.  Figure 5-50   Part Replacement Page
    Figure 5-51   Planned Maintenance Page 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL
    Section 5-4 - Main Board Detail 5 - 51
    5-10-12 Remote Software/Option Installation and Updates
    The Service Platform provides for remote software downloads. Any file transfer red to the scanner will 
    be in the form of an installation package. The actual  content of the installation package is dependent on 
    each product. 
    Software installation is simple and automated with a  Graphic User Interface (GUI), where applicable.
    Supports software upload via remote transmission
    Supports software upload  via Service Platform
    Options can be activated permanently
    Options can be activated for a timed period
    Options can be limited to a number of uses
    Options that expire must give notice to  the user that the option has expired
    5-10-12-1 Remote System Shutdown and Restart The Service Platform provides th e ability to restart the system as per standard shutdown/startup 
    procedures. Remote shutdo wn will only be provided if a softwa re-controlled shutdown exists on the 
    scanner. Scanner system restart is  always accessible from the OLC.
    5-10-12-2 File and/or Image Transfer The Service Platform supports file transfers and ar chived image transfers to the OnLine Center. The 
    Service Platform provides interfaces of th e displayed images for transfer to the OLC. 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL
    5 - 52Section 5-4 - Main Board Detail
    Section 5-11RFS (Service For Request)
    5-11-1 General 
    This describes general RFS functionality.
    1.) RFS function is a call for service to be used by cu stomer or FE directly from the system instead of 
    calling GE Cares.
    5-11-2 System Checkout Procedure
    If the system is checked out already, skip  5-8-2) System Checkout Procedure and proceed from 5-8-
    3) Sending RFS Procedure .
    1.) Check if the system is connected to the network.
    2.) Go to “Utility”
    3.) Press “Service”
    4.) Log on to the Service Platform with a password.
    5.) Press “Configuration”
    6.) Press “InSite ExC Agent Configuration”
    Figure 5-52   Agent Configuration 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL
    Section 5-4 - Main Board Detail 5 - 53
    5-11-2              System Checkout Procedure (cont’d)
    All bold fields should be filled
    7.) Device Name and CRM No are pre-populated and can not be changed.
    8.) Verify that “Send”  button is enabled.
    9.) Select Continent and Country
    10.) Put City, State,  and Institution
    11.) Select “Product” for the Enterprise.
    12.) If Proxy needs to be configured, select “Enable”  for the Proxy filed and input IP address and Port 
    number.
    13.) Press “Submit Changes”. “Please Wait” message  is displayed while the system is checking out. 
    Figure 5-53
    14.) Wait until following screen displays. (Figure 5-54)
    15.) Configuration is complete.
    16.) Close the Service Platform. Figure 5-53   Please Wait 
    Figure 5-54   Configuration done 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL
    5 - 54Section 5-4 - Main Board Detail
    5-11-3 Sending RFS Procedure
    Service Platform should be turned off in order to open RFS window.
    System must be connected to network, configured and checked out to be able to send the RFS
    1.) Right Click on “Connect to GE” icon.
    All the fields that are marked with asterisk * are mandatory and should be filled in order to send 
    the RFS request to On Line Center .
    2.) Fill out the Last name, First name,  and Phone number. (Figure 5-56)
    3.) Extension, E-mail, and Other system ID fields are optional.
    4.) Select Problem Type.
    If Problem is about “No Boot”, “No Image”, “Error message”, “Lock up”, “Probe”, “Peripherals”, select 
    “ Service ” for the Problem Type.
    If Problem is about “Presets”, “Reports”, “M easurements”, or “Probe not recognized”, select 
    “ Applications ” for the Problem Type.
    5.) Select the Problem Area.
    6.) Write the detailed problem description in the Problem Description field. Figure 5-55   Contact GE Icon
    Figure 5-56   RFS window 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL
    Section 5-4 - Main Board Detail 5 - 55
    5-11-3              Sending RFS Procedure (cont’d)
    7.) Then the “Send” button is enabled.
    8.) Click “Send” button and wait  for confirmation window with reference number on the screen.
    9.) Verify that “confirmation  message” pops up in the screen.
    10.) RFS request is completed.
    11.) On Line Center will contact  the customer shortly by phone.
    Figure 5-57   RFS send enabled 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL
    5 - 56Section 5-4 - Main Board Detail
    Section 5-12Machine RFS
    5-12-1 General 
    This describes general MRFS (Machine RFS) functionality.
    1.) MRFS function is an automatic call for service  from the system to the OLC when system detects 
    high temperature or high or low voltage.
    5-12-2 Enabling MRFS
    NOTE: In order to enable MRFS, System must be conn ected to network and should be checked out to 
    a product server. Default MRFS user should also be set.
    1.) If a system is not already checked out, the syst em should be checked out to a product server. Refer 
    to  5-11-2 - System Checkout Procedure.
    2.) MRFS function is an automatic call for service  from the system to the OLC when system detects 
    high temperature or high or low voltage.
    3.) Right Click on “Con nect to GE” icon.
    4.) When RFS window pops up, click on “Users” tab. Figure 5-58   Contact GE Icon
    Figure 5-59   User tab 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL
    Section 5-4 - Main Board Detail 5 - 57
    5-12-2              Enabling MRFS (cont’d)
    5.) Press “Add User”.
    All the fields that are marked with asterisk * are mandatory and should be filled
    6.) When fields with asterisk are fill ed, “Add User” button is enabled.
    7.) Press “Add User” button. Figure 5-60   Add User
    Figure 5-61   Add User 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL
    5 - 58Section 5-4 - Main Board Detail
    5-12-2              Enabling MRFS (cont’d)
    8.) check the checkbox next to the name to set to t he default contact. Then Press “Set Default Machine 
    Contact” button,
    9.) Now the MRFS is enabled.
    10.) When system detects high te mperature or abnormal voltage, the system will send the RFS to the 
    OLC automatically. Figure 5-62   Set Default Machine Contact 
    						