    DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL
    Section 4-4 - Software Configuration Checks 4 - 25
    4-3-12 ECG Check Out 
    Connect the ECG Harness and check:
    Section 4-4
    Software Configuration Checks
    Table 4-5     ECG Harness Check
    StepTask to doExpected Result(s)
    1.Connect the ECG at the Connector on the Front of the 
    scanner. It will display a curve along the bottom edge of the image 
    sector
    Table 4-6     Software Configuration Checks
    StepTask to doExpected Result(s)
    1.
    Check Date and Time setting Date and Time are correct
    2. Check that Location (Hospital Name) is  correct Location Name is correct
    3. Check Language settings Desired Language is displayed
    4. Check assignment of Printer Keys and User Define Keys Print1, 2 Keys and User define Keys are assigned as 
    desired by the customer
    5. Check that all of the customer’s options ar e set up correct All authorized functions are enabled 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
    4 - 26 Section 4-4 - Software Configuration Checks
    Section 4-5
    Peripheral Checks
    Check that peripherals work as described below:
    Table 4-7    Peripheral Checks
    StepTask to do Expected Result(s)
    1.
    Press ( FREEZE) Stop image acquisition.
    2. Press ( P1
    ) or (P2) on the Keyboard  The image displayed on the screen is printed on B&W or Color printer, 
    depending on the key assignment configuration
    3. Press  RECORD
     on the Keyboard. To start the video counter at a different point:
    4. Press  UPDATE MENU
     in the Trackball area. The Video Set Counter / Search Dialog window  is displayed
    5. Use the 
    alphanumeric keyboard  to enter 
    the counter number in the counter field. Number appears
    6. Press  SET COUNTER
     to save the change. Number is saved
    7. Press  RECORD
     on the KEYBOARD to return to the scanning mode
    8. Press  REC/PAUSE
     icon on the screen to Start Recording
    A red dot is displayed in the VCR status area
     on the Title bar  to indicate 
    that recording has begun
    9. Press  REC/PAUSE
     icon on the screen. To Stop recording
    The video status icon is changed to (Pause)
    10. Press RECORD
     on the KEYBOARD and the 
    assignable play To start, Play back an examination
    11. Use the 
    Assignable keys  on the Control 
    Panel to perform actions on the recorded session, such as stop, pause, rewind 
    or fast forward.
    The video status icon in updated accordingly.
    12. While in playback mode, use the 
    TRACKBALL
     to adjust the video playback 
    speed and scroll through the record. To search on the tape
    13. Press the assignable  PAUSE
    to stop the tape at the desired frame.
    14. When playing back an examination part of it can be stored on the computer’s memory as a cineloop.The 
    cineloop enables the user to perform further operations on the stored 
    section (see for further information on cineloop operation).
    15. Press (
    FREEZE
    ) while playing back a 
    recorded session. To store a recorded sequence as a cineloop. The
     last few seconds are 
    stored as a cineloop. 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL
    Section 4-4 - Software Configuration Checks 4 - 27
    Section 4-6Monitor Function Checks
    4-6-1 15 LCDMON II Assy & LCD Panel with filter Function Validation Procedure
    1.) Turn on the system.
    2.) Select Utility Key” on the Key board and then choose “Test Pattern” and “R esolution” marked in the 
    figure below.
    3.) All the Numbers from 0 to 100 on the fi gure above should be easily distinguished. 
    Figure 4-23   15” LCDMON Assy Function Check 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
    4 - 28 Section 4-4 - Software Configuration Checks
    4-6-2 15 OSD Key Assy Functi on Validation Procedure
    1.) Turn on the system.
    2.) Select “OSD Button” and “Lamp Button” marked below in the figure below.
    3.) OSD Menu should be displayed within 5sec and the Lamp should be turned on as soon as Lamp 
    Button is pushed.
    4.) Select “OSD Button” marked below in the figure below and check that contrast value is changed as  soon as the button pushed. 
    5) Displayed value should be change d as soon as the button pushed.
    Figure 4-24   Select OSD Button & Lamp Button
    Figure 4-25   Select OSD Button
    Lamp ButtonOSD button for 
    menu display
    OSD button for adjusting values 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL
    Section 4-7 - Keyboard Function Checks 4 - 29
    Section 4-7Keyboard Function Checks
    4-7-1 Main Keyboard Assy Function Validation Procedure
    1.) Boot up the system.
    2.) Insert the Service Dongle and go to the maintenance mode.
    3.) Click Start and go to Run.
    4.) Click the Run and then you can get the Run dialog box.
    5.) Write the cmd command into Run dialog box  and press OK button to display command prompt. 
    Figure 4-26   Start and run
    Figure 4-27   Run dialog box
    Figure 4-28   command prompt 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
    4 - 30 Section 4-7 - Keyboard Function Checks
    4-7-1              Main Keyboard Assy Fu nction Validation Procedure (cont’d)
    6.) Write the scfpapitest.exe on the command prompt and press Enter button to pop up the  scfpapitest program.
    7.) Press the keyboard button and rotate rotary butt on. When you press or rotary keys, Log dialog box 
    displays the key log. Button name and displaying log name is like the table. Figure 4-29   scfpapitest program
    Table 4-8    Button and Rotaty Table
    Button or Rotary NameThe Name displayed in LogButton or Rotary NameThe Name displayed in Log
    Patient New Patient Button
    Active ModeActive Mode Button
    Probe/Preset Probe Button
    CFCFM PushEncoder Button
    Report Report Button
    PDIPDI Button
    End Exam End Exam Button
    Dept (Button)Depth PushEncoder Button
    Record Record button
    Dept (Rotary)Depth PushEncoder Rotary
    P1 Print or record button1
    PWPWD PushEncoder Button
    P2 Print or record button2
    CWCW Button
    Reverse Reverse Button
    MM Mode PushEncoder
    Userdefine 1 Userdefine 1 Button
    Scan AreaScanArea Button
    Userdefine 2 Userdefine 2 Button
    M/D CursorMD Cursor Button
    Userdefine 3 Userdefine 3 Button
    Zoom (Button)Zoom/Size PushEncoder Button
    Userdefine 4 Userdefine 4 Button
    Zoom (Rotary) Zoom/Size PushEncoder Rotary
    Userdefine 5 Userdefine 5 Button
    MeasureMeasure Button
    L (Left) Left Button
    BodyPattern/Ellipse (Button) BodyPattern/Ellipse PushEncoder 
    Button 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL
    Section 4-7 - Keyboard Function Checks 4 - 31
    8.) Check the Fast radio button into Indicator box on scfpapitest program and press Set All button.
    R (Right)Right Button
    BodyPattern/Ellipse (Rotary BodyPattern/Ellipse PushEncoder 
    Rotary
    THI Harmonics Button
    Commentcomment Button
    AO Auto Right Button
    ClearClear Button
    B (Button) B PushEncoder Button
    TGC SlideSlider : n=8 s1 ~ s8
    B (Rotary) B PushEncoder Rotary
    NOTICEWhen you can check the name displayed in log as you press or rotate keys, the main keyboard assy 
    was replaced correctly.
    Figure 4-30   Fast button 
    NOTICEAfter pressing Fast radio button, indicators should be blanked on the frontpanel.
    Table 4-8    Button and Rotaty Table
    Button or Rotary NameThe Name displayed in LogButton or Rotary NameThe Name displayed in Log 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
    4 - 32 Section 4-7 - Keyboard Function Checks
    4-7-2 Sub Keyboard Assy Func tion Validation Procedure
    1.) Boot up the system and run the scfpapitest program.
    2.) Press and Rotary sub keyboard, and Paddle (up/down, right/left) button.
    3.) Check the log dialog box in scfpapitest program
    4-7-3 Trackball Assy validation
    1.) Boot up the system.
    2.) Press the measurement button.
    3.) Move the Trackball.
    When you move the trackball in measurement mode, the cursor should be moved.
    4-7-4 A/N Key assy validation 
    1.) Boot up the system.
    2.) Press Comment button.
    3.) Press A/N keyboard.
    Each key should be displayed on the monitor display.
    4-7-5 Freeze key validation 
    1.) Follow up the procedure 4-7-1 Main Keyboard Assy Function Validation Procedure on page 4-29  
    to run the scfpapitest program.
    2.) Press Freeze button and check up the Log dialog box if the freeze key was release.
    NOTICEFollow up the procedure  4-7-1 Main Keyboard Assy Function Validation Procedure on page 4-29  to 
    run the scfpapitest program.
    Table 4-9    Button and Rotary Table
    Button or Rotary NameThe Name displayed in Log
    1 (Button/Rotary) PushEncoder 1 of Touchpanel Button/Rotary
    2 (Button/Rotary) PushEncoder 2 of Touchpanel Button/Rotary
    3 (Button/Rotary) PushEncoder 3 of Touchpanel Button/Rotary
    4 (Button/Rotary) PushEncoder 4 of Touchpanel Button/Rotary
    5 (Button/Rotary) PushEncoder 5 of Touchpanel Button/Rotary
    Up Paddle Up Button
    Down Paddle Down Button
    Left Paddle Left Button
    Right Paddle Right Button
    NOTICEAfter pressing Freeze button, Freeze button message should be displayed into Log dialogbox on 
    scfpapitest program. 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL
    Section 4-7 - Keyboard Function Checks 4 - 33
    4-7-6 TGC key assay / TGC Knob Set validation
    1.) Follow up the procedure 4-7-1 Main Keyboard Assy Function Validation Procedure on page 4-29 
    to run the scfpapitest program.
    2.) Slide TGC pot and check up the value.
    4-7-7 Sub Keyboard Encoder Knob Set validation
    Follow up  4-7-2 Sub Keyboard Assy Function Validation Procedure on page 4-32.
    NOTICEWhile sliding TGC po t, TGC message should be displayed. 
    						
    							GE HEALTHCARE
    DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
    4 - 34 Section 4-8 - Mechanical Function Checks
    Section 4-8Mechanical Function Checks
    4-8-1 Cover Parts Function Validation
    • Check if the FRU parts are assembled tightly by naked eye and hands. 
    • Check if there are dents, scratches, or cracks on the FRU parts.
    • Check if screws are in place. Table 4-10    Cover parts of the FRU
    NoItem
    1
    Right and Left cover
    2 Front cover
    3 Rear cover
    4 Middle cover
    5 Top cover
    6 Top bottom cover
    7 Pole cover
    8 Neck front cover
    9 Neck rear cover
    10 Dummy cover L/R
    11 EMI cover L/R
    12 LCD Std Arm with cover
    13 Rear handle
    14 Rear hook
    15 ECG cable hook
    16 Cable Arm hook
    17 LP5/LA5 Monitor Cover set 
    						