GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL Section 4-4 - Software Configuration Checks 4 - 25 4-3-12 ECG Check Out Connect the ECG Harness and check: Section 4-4 Software Configuration Checks Table 4-5 ECG Harness Check StepTask to doExpected Result(s) 1.Connect the ECG at the Connector on the Front of the scanner. It will display a curve along the bottom edge of the image sector Table 4-6 Software Configuration Checks StepTask to doExpected Result(s) 1. Check Date and Time setting Date and Time are correct 2. Check that Location (Hospital Name) is correct Location Name is correct 3. Check Language settings Desired Language is displayed 4. Check assignment of Printer Keys and User Define Keys Print1, 2 Keys and User define Keys are assigned as desired by the customer 5. Check that all of the customer’s options ar e set up correct All authorized functions are enabled
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 4 - 26 Section 4-4 - Software Configuration Checks Section 4-5 Peripheral Checks Check that peripherals work as described below: Table 4-7 Peripheral Checks StepTask to do Expected Result(s) 1. Press ( FREEZE) Stop image acquisition. 2. Press ( P1 ) or (P2) on the Keyboard The image displayed on the screen is printed on B&W or Color printer, depending on the key assignment configuration 3. Press RECORD on the Keyboard. To start the video counter at a different point: 4. Press UPDATE MENU in the Trackball area. The Video Set Counter / Search Dialog window is displayed 5. Use the alphanumeric keyboard to enter the counter number in the counter field. Number appears 6. Press SET COUNTER to save the change. Number is saved 7. Press RECORD on the KEYBOARD to return to the scanning mode 8. Press REC/PAUSE icon on the screen to Start Recording A red dot is displayed in the VCR status area on the Title bar to indicate that recording has begun 9. Press REC/PAUSE icon on the screen. To Stop recording The video status icon is changed to (Pause) 10. Press RECORD on the KEYBOARD and the assignable play To start, Play back an examination 11. Use the Assignable keys on the Control Panel to perform actions on the recorded session, such as stop, pause, rewind or fast forward. The video status icon in updated accordingly. 12. While in playback mode, use the TRACKBALL to adjust the video playback speed and scroll through the record. To search on the tape 13. Press the assignable PAUSE to stop the tape at the desired frame. 14. When playing back an examination part of it can be stored on the computer’s memory as a cineloop.The cineloop enables the user to perform further operations on the stored section (see for further information on cineloop operation). 15. Press ( FREEZE ) while playing back a recorded session. To store a recorded sequence as a cineloop. The last few seconds are stored as a cineloop.
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL Section 4-4 - Software Configuration Checks 4 - 27 Section 4-6Monitor Function Checks 4-6-1 15 LCDMON II Assy & LCD Panel with filter Function Validation Procedure 1.) Turn on the system. 2.) Select Utility Key” on the Key board and then choose “Test Pattern” and “R esolution” marked in the figure below. 3.) All the Numbers from 0 to 100 on the fi gure above should be easily distinguished. Figure 4-23 15” LCDMON Assy Function Check
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 4 - 28 Section 4-4 - Software Configuration Checks 4-6-2 15 OSD Key Assy Functi on Validation Procedure 1.) Turn on the system. 2.) Select “OSD Button” and “Lamp Button” marked below in the figure below. 3.) OSD Menu should be displayed within 5sec and the Lamp should be turned on as soon as Lamp Button is pushed. 4.) Select “OSD Button” marked below in the figure below and check that contrast value is changed as soon as the button pushed. 5) Displayed value should be change d as soon as the button pushed. Figure 4-24 Select OSD Button & Lamp Button Figure 4-25 Select OSD Button Lamp ButtonOSD button for menu display OSD button for adjusting values
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL Section 4-7 - Keyboard Function Checks 4 - 29 Section 4-7Keyboard Function Checks 4-7-1 Main Keyboard Assy Function Validation Procedure 1.) Boot up the system. 2.) Insert the Service Dongle and go to the maintenance mode. 3.) Click Start and go to Run. 4.) Click the Run and then you can get the Run dialog box. 5.) Write the cmd command into Run dialog box and press OK button to display command prompt. Figure 4-26 Start and run Figure 4-27 Run dialog box Figure 4-28 command prompt
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 4 - 30 Section 4-7 - Keyboard Function Checks 4-7-1 Main Keyboard Assy Fu nction Validation Procedure (cont’d) 6.) Write the scfpapitest.exe on the command prompt and press Enter button to pop up the scfpapitest program. 7.) Press the keyboard button and rotate rotary butt on. When you press or rotary keys, Log dialog box displays the key log. Button name and displaying log name is like the table. Figure 4-29 scfpapitest program Table 4-8 Button and Rotaty Table Button or Rotary NameThe Name displayed in LogButton or Rotary NameThe Name displayed in Log Patient New Patient Button Active ModeActive Mode Button Probe/Preset Probe Button CFCFM PushEncoder Button Report Report Button PDIPDI Button End Exam End Exam Button Dept (Button)Depth PushEncoder Button Record Record button Dept (Rotary)Depth PushEncoder Rotary P1 Print or record button1 PWPWD PushEncoder Button P2 Print or record button2 CWCW Button Reverse Reverse Button MM Mode PushEncoder Userdefine 1 Userdefine 1 Button Scan AreaScanArea Button Userdefine 2 Userdefine 2 Button M/D CursorMD Cursor Button Userdefine 3 Userdefine 3 Button Zoom (Button)Zoom/Size PushEncoder Button Userdefine 4 Userdefine 4 Button Zoom (Rotary) Zoom/Size PushEncoder Rotary Userdefine 5 Userdefine 5 Button MeasureMeasure Button L (Left) Left Button BodyPattern/Ellipse (Button) BodyPattern/Ellipse PushEncoder Button
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL Section 4-7 - Keyboard Function Checks 4 - 31 8.) Check the Fast radio button into Indicator box on scfpapitest program and press Set All button. R (Right)Right Button BodyPattern/Ellipse (Rotary BodyPattern/Ellipse PushEncoder Rotary THI Harmonics Button Commentcomment Button AO Auto Right Button ClearClear Button B (Button) B PushEncoder Button TGC SlideSlider : n=8 s1 ~ s8 B (Rotary) B PushEncoder Rotary NOTICEWhen you can check the name displayed in log as you press or rotate keys, the main keyboard assy was replaced correctly. Figure 4-30 Fast button NOTICEAfter pressing Fast radio button, indicators should be blanked on the frontpanel. Table 4-8 Button and Rotaty Table Button or Rotary NameThe Name displayed in LogButton or Rotary NameThe Name displayed in Log
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 4 - 32 Section 4-7 - Keyboard Function Checks 4-7-2 Sub Keyboard Assy Func tion Validation Procedure 1.) Boot up the system and run the scfpapitest program. 2.) Press and Rotary sub keyboard, and Paddle (up/down, right/left) button. 3.) Check the log dialog box in scfpapitest program 4-7-3 Trackball Assy validation 1.) Boot up the system. 2.) Press the measurement button. 3.) Move the Trackball. When you move the trackball in measurement mode, the cursor should be moved. 4-7-4 A/N Key assy validation 1.) Boot up the system. 2.) Press Comment button. 3.) Press A/N keyboard. Each key should be displayed on the monitor display. 4-7-5 Freeze key validation 1.) Follow up the procedure 4-7-1 Main Keyboard Assy Function Validation Procedure on page 4-29 to run the scfpapitest program. 2.) Press Freeze button and check up the Log dialog box if the freeze key was release. NOTICEFollow up the procedure 4-7-1 Main Keyboard Assy Function Validation Procedure on page 4-29 to run the scfpapitest program. Table 4-9 Button and Rotary Table Button or Rotary NameThe Name displayed in Log 1 (Button/Rotary) PushEncoder 1 of Touchpanel Button/Rotary 2 (Button/Rotary) PushEncoder 2 of Touchpanel Button/Rotary 3 (Button/Rotary) PushEncoder 3 of Touchpanel Button/Rotary 4 (Button/Rotary) PushEncoder 4 of Touchpanel Button/Rotary 5 (Button/Rotary) PushEncoder 5 of Touchpanel Button/Rotary Up Paddle Up Button Down Paddle Down Button Left Paddle Left Button Right Paddle Right Button NOTICEAfter pressing Freeze button, Freeze button message should be displayed into Log dialogbox on scfpapitest program.
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL Section 4-7 - Keyboard Function Checks 4 - 33 4-7-6 TGC key assay / TGC Knob Set validation 1.) Follow up the procedure 4-7-1 Main Keyboard Assy Function Validation Procedure on page 4-29 to run the scfpapitest program. 2.) Slide TGC pot and check up the value. 4-7-7 Sub Keyboard Encoder Knob Set validation Follow up 4-7-2 Sub Keyboard Assy Function Validation Procedure on page 4-32. NOTICEWhile sliding TGC po t, TGC message should be displayed.
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 4 - 34 Section 4-8 - Mechanical Function Checks Section 4-8Mechanical Function Checks 4-8-1 Cover Parts Function Validation • Check if the FRU parts are assembled tightly by naked eye and hands. • Check if there are dents, scratches, or cracks on the FRU parts. • Check if screws are in place. Table 4-10 Cover parts of the FRU NoItem 1 Right and Left cover 2 Front cover 3 Rear cover 4 Middle cover 5 Top cover 6 Top bottom cover 7 Pole cover 8 Neck front cover 9 Neck rear cover 10 Dummy cover L/R 11 EMI cover L/R 12 LCD Std Arm with cover 13 Rear handle 14 Rear hook 15 ECG cable hook 16 Cable Arm hook 17 LP5/LA5 Monitor Cover set