GE Logiq P5 User Manual
Have a look at the manual GE Logiq P5 User Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
Powering the System LOGIQ 5 Basic User Manual 3-21 Direction 2300002-100 Rev. 2 Connecting and Using the System (continued) Figure 3-10. Example Plug and Outlet Configurations 1. 100-120 VAC, 1200 VA Plug and Outlet Configuration 2. 220-240 VAC, 1200 VA Plug and Outlet Configuration WARNINGTo avoid risk of fire, the system power must be supplied from a separate, properly rated outlet. See ‘Before the system arrives’ for more information. Under no circumstances should the AC power plug be altered, changed, or adapted to a configuration rated less than specified. Never use an extension cord or adapter plug. To help assure grounding reliability, connect to a “hospital grade” or “hospital only” grounded power outlet. 12 ab ab
Preparing the System for Use 3-22 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Acclimation Time After being transported, the unit requires one hour for each 2.5° increment its temperature is below 10° C or above 40° C. Table 3-3: System Acclimation Time Chart ° C60 55 50 45 40 35 30 25 20 15 10 ° F140 131 122 113 104 95 86 77 68 59 50 hours86420000000 ° C5 0 -5 -10 -15 -20 -25 -30 -35 -40 ° F41 32 23 14 5 -4 -13 -22 -31 -40 hours2 4 6 8 101214161820
Powering the System LOGIQ 5 Basic User Manual 3-23 Direction 2300002-100 Rev. 2 Power On/Standby Figure 3-11. Power On/Standby Switch Location CAUTIONPress the Power On/Standby switch to turn the power on. The circuit breaker must also be in the on position. For circuit breaker location, see Figure 3-16. Composite B/WB/W PrinterPower220-240V 500VA MaxIncluding rear panel
Preparing the System for Use 3-24 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Power Up Sequence The system is initialized. During this time: • System diagnostics run. Its status is reflected on the monitor by the graphic in Figure 3-12. Figure 3-12. Power Up Graphic Sequence • Probes are initialized for immediate operation. NOTE: If no probe is connected, the system goes into standby mode. • Peripheral devices are activated on power up. After initialization is complete, all lighted buttons on the Control Panel light and the default B-Mode screen or Patient screen (no probes are connected) is displayed in the monitor display. HINTSIf problems occur, freeze the image and take a picture for reference. This will help if there is a need to call for service.
Powering the System LOGIQ 5 Basic User Manual 3-25 Direction 2300002-100 Rev. 2 Password Protection LoginPersonal IDs and associated passwords can be preset in the LOGIQ 7. If IDs and passwords have been entered and the Use Auto Logon is off, the Operator Login Window appears, requesting for an ID and password when Power up sequence is completed or when its needed. Figure 3-13. Operator Login Window 1. Operator: Select the Operator. 2. Password: Enter Operator’s password. 3. Dataflow: Select Database (for example, Local Archive, DICOM, etc.) 4. Select type of Logon or Cancel. • Emergency: Data stored only for the duration of the current examination. • Logon: Standard logon • Cancel: Cancel logon If the correct ID and Password are entered, the system displays the Patient screen. 1 4 3 2
Preparing the System for Use 3-26 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Password Protection (continued) LogoffWhen the user enters the system by login, at the end of the exam the system displays the SYSTEM-EXIT window. Select “Logoff” to logoff the system. Figure 3-14. System Exit Window
Powering the System LOGIQ 5 Basic User Manual 3-27 Direction 2300002-100 Rev. 2 Standby Mode To place the system in standby mode: 1. Press the Power On/Standby switch at the front of the system once. 2. The System-Exit window is displayed. 3. Using the Trackball or Tab key, select Shutdown. The shutdown process takes a few seconds and is completed when the control panel illumination is turned off. The shutdown process takes a few seconds to complete. Figure 3-15. System-Exit Window 4. Disconnect the probes. Clean or disinfect all probes as necessary. Store them in their shipping cases to avoid damage. NOTE: If daily maintenance is to be performed, turn off the circuit breaker in the back of the system to power off.
Preparing the System for Use 3-28 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Power Off To power off the system: 1. Press the Power On/Standby switch at the front of the system once. 2. The System-Exit window is displayed. 3. Using the Trackball or Tab key, select Shutdown. The shutdown process takes a few seconds and is completed when the control panel illumination is turned off. 4. Disconnect the probes. Clean or disinfect all probes as necessary. Store them in their shipping cases to avoid damage. 5. Turn off the Main Circuit Breaker at the rear of the system to power off the system. NOTE: When the Uninterruptible Power Supply (UPS) option is installed, battery replacement may be necessary if the circuit breaker is turned off for long periods of time (3 to 6 months). CAUTIONDO NOT turn off the circuit breaker before the Power on/ Standby switch LED is off. Data may be lost or system software damaged if the circuit breaker is turned off before the Power on/Standby LED is off.
Powering the System LOGIQ 5 Basic User Manual 3-29 Direction 2300002-100 Rev. 2 Circuit breaker The Circuit Breaker is located at the rear panel of the system. On supplies main power to all internal systems. Off removes main power from all internal systems. The circuit breaker automatically shuts off power to the system in case of a power overload. If a power overload occurs: 1. Turn off all peripheral devices. 2. Reactivate the Circuit Breaker switch. The Circuit Breaker switch should stay in the On position; DO NOT hold the switch in the On position. If the Circuit Breaker switch remains On, follow the Power On procedure previously described. NOTE: If the Circuit Breaker switch does not remain in the On position or trips again: 1. Disconnect the Power Cable. 2. Call Service immediately. DO NOT attempt to use the system. Figure 3-16. Location of Circuit Breaker 1. Circuit Breaker On Position 2. Circuit Breaker Off Position 3. Circuit Breaker location 1 23
Preparing the System for Use 3-30 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Adjusting the Display Monitor Rotate, tilt, raise and lower the monitor The monitor position can be adjusted for easy viewing. • The monitor can be rotated around its central pivot point. • The monitor can be tilted for the optimum viewing angle. • The monitor and console panel can be raised or lowered for the best viewing height. The monitor position can be adjusted for easy viewing. NOTE: When moving the LOGIQ 5, lower the monitor and console panel to its lowest possible position to improve stability. Brightness and Contrast Adjusting the monitors contrast and brightness is one of the most important factors for proper image quality. If these controls are set incorrectly, the Gain, TGC, Dynamic Range and even Power Output may have to be changed more often than necessary to compensate. The proper setup displays a complete gray scale. The lowest level of black should just disappear into the background and the highest white should be bright, but not saturated. CAUTIONBe careful when lowering the monitor and control panel that it does not collide with the peripheral device sitting on the top of the console. Use caution and slowly lower the monitor and control panel. DO NOT place objects on the monitor.