    							Preparing for a Biopsy
    LOGIQ 5 Basic User Manual 17-29
    Direction 2300002-100 Rev. 2
    Displaying the Guidezone (continued)
     Figure 17-8. Biopsy Guidezones for the 10L Probe
    1.  1 cm increments
    2.  5 cm increments
    The biopsy guidezone represents a path of the needle. The dots 
    which make up the guidezones is the depth readout where: 
    •  Yellow represent 1 cm increments. 
    •  Red represents 5 cm increments.
    The display should be carefully monitored during a biopsy for 
    any needle deviation from the center line or guidezone.
    NOTE:  To set up biopsy guidezones, refer to Table 16-8 for more 
    details. 
    						
    							Probes and Biopsy
    17-30 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Displaying the Guidezone (continued)
    The needle may vary from the center line or guidezone for 
    various reasons:
    •  Needle barrel to needle clearance or strength.
    •  Bracket manufacturing tolerance.
    •  Needle deflection due to tissue resistance.
    •  Needle size chosen. Thinner needles may deflect more.
    NOTE:  Although the multi-angle guides are compatible with the Civco 
    Ultrapro and Ultrapro II, it is recommended the multi-angle 
    guides only be used with the Ultrapro II. Table 17-15: LOGIQ 5 Biopsy Guide Availability
    Fixed AngleMulti-Angle
    ProbeMBX1MBX2MBX3
    3C 4.0 6.0 8.0
    3.5C 7.0
    5C 7.0 4.0 5.5 7.0
    E8C 5.0
    7L 4.0 4.0 5.5 7.0
    10L 2.0 2.0 4.0 6.0
    12L 1.5 2.5 3.5
    DANGERFailure to match the guidezone displayed to the guide may 
    cause the needle to track a path outside the zone.
    It is extremely important that when using the adjustable angle 
    biopsy guides, the angle displayed on the screen matches the 
    angle set on the guide, otherwise the needle will not follow the 
    displayed guidezone which could result in repeated biopsies or 
    patient injury. 
    						
    							Preparing for a Biopsy
    LOGIQ 5 Basic User Manual 17-31
    Direction 2300002-100 Rev. 2
    Preparing the Biopsy Guide Attachment
    Convex, Sector and Linear probes have optional biopsy guide 
    attachments for each probe.  The guide consists of a non-
    disposable bracket to attach to the probe, disposable needle clip 
    to attach to the bracket, sheath, gel (sterile gel if necessary) and 
    disposable needle barrels.
    The disposable needle barrels are available for a variety of 
    needle sizes.
    CAUTIONPlease refer to the manufacturers instructions included in the 
    biopsy kit. 
    						
    							Probes and Biopsy
    17-32 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Fixed Needle Guide Assembly
    Identify the appropriate biopsy guide bracket by matching the 
    label on the bracket with the probe to be used.
    Orient the bracket so that the needle clip attachment will be on 
    the same side as the probe orientation mark (ridge).
     Figure 17-9. Probe/Bracket Alignment
    1. Probe Orientation Mark
    2. Bracket
    Attach the biopsy bracket to the probe by sliding the bracket 
    over the end of the probe until it clicks or locks in place.
    Place an adequate amount of coupling gel on the face of the 
    probe.
    Place the proper sanitary sheath over the probe and biopsy 
    bracket.  Use the rubber bands supplied to hold the sheath in 
    place.
     Figure 17-10. Applying Sanitary Sheath
    1
    2 
    						
    							Preparing for a Biopsy
    LOGIQ 5 Basic User Manual 17-33
    Direction 2300002-100 Rev. 2
    Fixed Needle Guide Assembly (continued)
    Snap the fixed or adjustable needle clip onto the biopsy guide 
    bracket.
     Figure 17-11. Fixed Needle Clip Attachment
    1. Sheath
    Choose the desired gauge (size) needle barrel.  Twist it back 
    and forth to remove it from the plastic tree.
     Figure 17-12. Needle Barrels
    1 
    						
    							Probes and Biopsy
    17-34 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Fixed Needle Guide Assembly (continued)
    Place the needle barrel into the needle clip with the desired 
    gauge facing the needle clip and snap into place.
     Figure 17-13. Needle Barrel Installation
    CAUTIONEnsure that all guide parts are seated properly prior to 
    perfoming a biopsy. 
    						
    							Preparing for a Biopsy
    LOGIQ 5 Basic User Manual 17-35
    Direction 2300002-100 Rev. 2
    Multi-Angle Biopsy Guide Assembly
    Scan the patient and identify the target for biopsy. Move the 
    probe to locate the target to the center of the image. Enable  the 
    system biopsy guidezone and try guidezone angles A1 to A3 to 
    decide the best angle setting for needle path.
    Identify the appropriate biopsy guide bracket by matching the 
    label on the bracket with the probe to be used.
     Figure 17-14. 3cb Multi-Angle Biopsy Guide Bracket
    Orient the bracket so that the needle clip attachment will be on 
    the same side as the probe orientation mark (ridge).
     Figure 17-15. Probe/Bracket Alignment
    1. Probe Orientation Mark
    2. Bracket
    Attach the biopsy bracket to the probe by sliding the bracket 
    over the end of the probe until it clicks or locks in place.
    WARNINGDO NOT attempt to use the multi-angle biopsy bracket and 
    needle guide until the manufacturers instructions, provided 
    with the biopsy bracket and needle guide in the kit, have been 
    read and thoroughly understood.
    1 2 
    						
    							Probes and Biopsy
    17-36 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Multi-Angle Biopsy Guide Assembly (continued)
    Pull up on the knob to freely move the needle guide attachment. 
    Align the knob with the selected position of the needle guide 
    attachment from MBX1, MBX2 and MBX3, to match the 
    guidezone display on the ultrasound system.
     Figure 17-16. Select the angle position
    1. Pull up
    Push the knob down into the desired slot to secure the angle 
    position of the needle guide attachment.
     Figure 17-17. Fix the angle position
    1. Push
    1
    1
    CAUTIONHold the bracket in place on the probe when pushing the knob 
    to secure the angle position of the needle guide attachment. 
    Excessive force may cause the bracket to release from the 
    probe. 
    						
    							Preparing for a Biopsy
    LOGIQ 5 Basic User Manual 17-37
    Direction 2300002-100 Rev. 2
    Multi-Angle Biopsy Guide Assembly (continued)
    Place an adequate amount of coupling gel on the face of the 
    probe.
    Place the proper sanitary sheath tightly over the probe and 
    biopsy bracket.  Use the rubber bands supplied to hold the 
    sheath in place.
     Figure 17-18. Applying Sanitary Sheath
    Snap the needle clip onto the biopsy guide bracket.
     Figure 17-19. Fixing the Needle Clip Attachment
    1. Sheath
    Push the locking mechanism towards the bracket to secure the 
    lock. Make sure the needle guide is firmly attached to the 
    bracket.
     Figure 17-20. Locking the Needle Clip
    1
    1 
    						
    							Probes and Biopsy
    17-38 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Multi-Angle Biopsy Guide Assembly (continued)
    Choose the desired gauge (size) needle barrel.  Twist it back 
    and forth to remove it from the plastic tree.
     Figure 17-21. Needle Barrel Selection
    Place the needle barrel into the needle clip with the desired 
    gauge facing the needle clip and snap into place.
     Figure 17-22. Needle Barrel Installation
    CAUTIONEnsure that all guide parts are seated properly prior to 
    perfoming a biopsy. 
    						