    							Configuring Connectivity
    LOGIQ 5 Basic User Manual 16-101
    Direction 2300002-100 Rev. 2
    Adding a dataflow
    1.  In the Dataflow section, select Add.
    The system adds a new dataflow, with a default name such 
    as Workflow05.
    2.  To change the name of the dataflow, move the Trackball to 
    the Name field, press Set, and type a new name.
    The system displays a message to confirm that you want to 
    change the name.
    3. Select Yes.
    Removing a dataflow or service
    1.  Select the dataflow or service.
    2.  Select the appropriate Remove button.
    The system displays a warning message to confirm that you 
    want to remove the dataflow or service.
    3. Select OK. 
    						
    							Customizing Your System
    16-102 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Dataflow (continued)
    The following table lists dataflows that are pre-configured with 
    the ultrasound system.
    Table 16-59: Dataflows
    DataflowDescription
    No Archive Performs an exam without storing the data to the 
    archive.
    Local Archive - Int. 
    HDLocal archive internal hard driveUses the local 
    database for patient archiving.Stores images to the 
    internal hard drive.
    Local Archive - Int. 
    HD/MODLocal archive internal hard drive and Magneto-
    Optical DiskUses the local database for patient 
    archiving.Stores images to the internal hard drive 
    (raw data) and to a MOD as DICOM Media.NOTE: 
    LocArch-HD/MOD enables only “pure” DICOM to 
    both destinations, or both DICOM and raw data to 
    both destinations.
    Remote Archive - 
    Remote HDRemote archive remote hard driveUses a remote 
    database (either on EchoPAC workstation or on 
    EchoServer) for patient archiving.Stores images to 
    a network image volume (either internal HD on 
    EchoPAC workstation or EchoServer volume).
    Remote Archive - 
    MODRemote archive Magneto-Optical DiskUses a 
    remote database (either on EchoPAC workstation 
    or on EchoServer) for patient archiving.Stores 
    images to a MOD.
    Worklist/Local 
    Archive - DICOM 
    Server/Int. HDModality Worklist local archive DICOM server and 
    local hard driveSearches the DICOM Modality 
    Worklist. Copies the patient information to the local 
    database. Stores the patient information and the 
    examination results to the local the database.Stores 
    images to a DICOM Server and to an image volume 
    on the local hard drive.NOTE: WL-LA-DServ: Does 
    not search the local database, only the DICOM 
    Modality Worklist. 
    						
    							Configuring Connectivity
    LOGIQ 5 Basic User Manual 16-103
    Direction 2300002-100 Rev. 2
    Screens
    The Screens tab allows you to configure tools related to patient 
    management. You can specify default system functionality such 
    whether patient ID is required when you archive data, and if you 
    want the system to automatically search the archive for a patient 
    when you enter patient data.
     Figure 16-70. Connectivity Screens Preset Menu 
    						
    							Customizing Your System
    16-104 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Screens (continued)
    Table 16-60: Screens
    Preset ParameterDescription
    Columns in 
    examination listingSelect the columns to add to or delete from the 
    Examination list window.
    Use free text 
    addressesIn the Patient information window:When selected, 
    the address information is entered in a single field 
    (free text).When cleared, the address information 
    (e.g. street, city, etc.) is entered in type-specific 
    fields.
    Use birthdate In the Patient information window, enter either the 
    patient age or the birth date:When selected, enter 
    birth date, then the age is calculated.When cleared, 
    enter age (birth date field not available).
    Use extended patient 
    dialogIn the Patient information window:When selected, 
    the entire patient information data is displayed. 
    When cleared, patient information data displayed is 
    restricted to a minimum (for example, Name and 
    Patient ID).NOTE: When cleared, to display the 
    entire patient information data in the Patient 
    information window, press More.
    Use extended search 
    dialogIn the Search/Create Patient window:When 
    selected, all the searching filters are displayed as 
    default.When cleared, the searching criteria are 
    restricted to a minimum.NOTE: When cleared, to 
    display all the searching filters in the Search/Create 
    Patient window, press More.
    Auto search for 
    patientIn the Search/Create Patient window:When 
    selected, the system automatically searches 
    through the selected patient archive, while the user 
    enters patient information.When cleared, the 
    automatic search tool is turned off. If you are trying 
    to keep the past patient data confidential, DO NOT 
    use this feature. 
    Pre-defined text 
    directlyIn the Examination list window:When selected, the 
    Insert text key launches pre-defined text 
    input.When cleared, the Insert text key opens the 
    extended text field.
    Examination List on 
    Archive buttonWhen a patient is selected, and the user selects 
    Archive:When selected, opens the Examination list 
    window for the selected patient.When cleared, 
    opens the Patient Information window for the 
    selected patient. 
    						
    							Configuring Connectivity
    LOGIQ 5 Basic User Manual 16-105
    Direction 2300002-100 Rev. 2
    Examination list window column configuration
    You can create new columns, remove columns, and select the 
    information to display in a column.
    1. Move the Trackball to highlight a column.
    2. Press Set.
    The system displays a list of options.
    3.  Do one of the following:
    •  To create a new column to the left of the selected 
    column, select INSERT.
    •  To delete the selected column, select DELETE.
    •  To select the information to display in the selected 
    column, select from the list of data fields.
    4. Press Set.
    Automatic generation 
    of patient IDIn the Search/Create Patient window:When 
    selected, the Patient ID is not required when 
    entering a new patient in the archive. The system 
    automatically generates an ID number.When 
    cleared, the Patient ID is required when entering a 
    new patient in the archive.
    Request 
    acknowledge of End 
    Exam actionWhen selected, the user is asked to confirm action 
    when ending an examination.
    Go directly to 
    scanning from searchWhen selected, the system goes directly to the 
    Scanning screen after the user selects or creates a 
    patient record.When cleared, after you select or 
    create a patient record, the system displays the 
    Patient information window for further information 
    entry. You must select Begin Exam to go to the 
    scanning screen.
    Table 16-60: Screens
    Preset ParameterDescription 
    						
    							Customizing Your System
    16-106 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Tools
    The Tools tab allows you to:
    •  Verify the DICOM directory on removable media
    •  Format removable media (MO disk, rewritable CD, or ZIP 
    disk)
     Figure 16-71. Connectivity Tools Preset Menu 
    						
    							Configuring Connectivity
    LOGIQ 5 Basic User Manual 16-107
    Direction 2300002-100 Rev. 2
    Tools (continued)
    The bottom of the screen lists properties of the selected media.
    Formatting removable media
    1.  Select the removable media from the Media list.
    2.  Type a name for the removable media in the Label field.
    NOTE:  Do not use the following characters for labelling:
     / : ; . , * < > | + = [ ]
    3. Select Format.
    Verifying removable media
    1.  Select the removable media from the Media list.
    2. Select Verify.Table 16-61: Tools
    Preset ParameterDescription
    Media Select the removable media to format or verify.
    Label Type a label for a new removable MO disk (free 
    text).
    Verify Select to verify DICOM directory on removable 
    DICOM disk.
    Format Select to format removable media. 
    						
    							Customizing Your System
    16-108 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Views
    Views shows you an overview of the Ultrasound system’s 
    connectivity architecture:
    •  The currently selected dataflow
    •  All configured dataflows
    •  The network structure tree
    •  The configured buttons dataflows
    Views also allows you to verify service connectivity.
     Figure 16-72. Connectivity Views Preset Menu 
    						
    							Configuring Connectivity
    LOGIQ 5 Basic User Manual 16-109
    Direction 2300002-100 Rev. 2
    Views (continued)
    A dataflow is a set of services. When you select one of the 
    configured dataflows, the ultrasound system is automatically set 
    up according to the services associated with the selected 
    dataflow. For information about editing dataflows, See ‘Dataflow’ 
    for more information.
    Each view includes a structure tree on the left and properties on 
    the right. The properties section lists information about the 
    selected service. The information shown in each view is as 
    follows:Table 16-62: Views
    Preset ParameterDescription
    Current Select to show the current dataflow hierarchy.
    Dataflow Select to show all dataflows.
    Network Select to show network structure tree.
    Buttons Select to show configured buttons dataflows.
    Config Select to go to the Services tab and configure this 
    service.
    Check Select to verify connectivity for a device or service.If 
    the service is a ***, the system performs a TCPIP 
    ping.If the service is a ***, the system performs a 
    DICOM echo.
    Properties Lists information about the selected device or 
    service. 
    						
    							Customizing Your System
    16-110 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Current dataflow
    The current dataflow view displays a three-level structure tree 
    showing the following information:
    •  The name of the current dataflow (Level 1)
    •  The type of services associated to the dataflow (input or 
    output services) (Level 2)
    •  A list of the services configured for the current dataflow 
    (Level 3)
     Figure 16-73. Connectivity Views: Current Dataflow 
    						