    							Cardiology Measurements
    LOGIQ 5 Basic User Manual 10-51
    Direction 2300002-100 Rev. 2
    Mitral Valve (continued)
    Two distance
    measurement•  Mitral Valve E-Peak to A-Peak Ratio (A-C and D-E) (MV E/A 
    Ratio)
    Making a two
    distance
    measurement1. Select MV E/A Ratio; an active caliper displays.
    2.  Make the first distance measurement:
    a.  To position the active caliper at the start point, move the 
    Trackball.
    b.  To fix the start point, press Set.
    The system fixes the first caliper and displays a second 
    active caliper.
    c.  To position the second active caliper at the end point, 
    move the Trackball.
    A dotted line connects the measurement points.
    d.  To complete the measurement, press Set.
    The system displays the distance value in the Results 
    Window. After the first measurement, the system 
    displays an active caliper.
    3.  To make the second distance measurement, repeat steps 
    a–d above.
    The system displays the measurements and ratio in the 
    Results Window. 
    						
    							Cardiology
    10-52 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Mitral Valve (continued)
    One time interval/
    slope
    measurements•  Mitral Valve Acceleration Time (MV AccT)
    •  Mitral Valve Deceleration Time (MV DecT)
    Making a time
    interval/slope
    measurement1.  Select the measurement; an active caliper displays.
    2.  To position the caliper at the start point, move the Trackball.
    3.  To fix the start point, press Set.
    The system fixes the first caliper and displays a second 
    active caliper.
    4.  To position the second caliper at the end point, move the 
    Trackball.
    A dotted line shows the slope.
    5.  To complete the measurement, press Set. 
    The system displays the time interval and slope 
    measurements in the Results Window. 
    						
    							Cardiology Measurements
    LOGIQ 5 Basic User Manual 10-53
    Direction 2300002-100 Rev. 2
    Mitral Valve (continued)
    One time interval
    measurements•  Mitral Valve Ejection Time ((MV Trace)
    •  Mitral Valve A-Wave Duration (MV A Dur)
    •  Mitral Valve Time to Peak (MV Trace)
    • Time
    Making a time
    interval
    measurement1.  Select the measurement.
    The system displays an active caliper and a vertical dotted 
    line. 
    2.  To position the caliper at the start point, move the Trackball.
    3.  To fix the first caliper, press Set.
    The system displays a second active caliper.
    4.  To position the caliper at the end point, move the Trackball.
    5.  To complete the measurement, press Set.
    The system displays the time interval in the Results 
    Window.
    Tw o  t i m e  i n t e r v a l
    measurementBody surface area and stroke volume measurements:
    •  Stroke Volume Index by Mitral Flow (MV Trace)
    The system calculates body surface area from the patient’s 
    height and weight.
    One distance and two velocity measurement:
    •  Mitral Valve Area from Continuity Equation (MV Vmax) 
    						
    							Cardiology
    10-54 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Pulmonic Valve
    The following are cardiac Doppler Mode Pulmonic Valve 
    measurements:
    One peak velocity
    measurements•  Pulmonic Insufficiency Peak Pressure Gradient (PR Vmax)
    •  Pulmonic Insufficiency End-Diastolic Pressure Gradient (PR 
    Trace)
    •  Pulmonic Valve Peak Pressure Gradient (PV Vmax)
    •  Pulmonic End-Diastolic Pressure Gradient (PR Trace)
    •  Pulmonic Insufficiency Peak Velocity (PR Vmax)
    •  Pulmonic Insufficiency End-Diastolic Velocity (PRend 
    Vmax)
    •  Pulmonic Valve Peak Velocity (PV Vmax)
    •  Pulmonic End-Diastolic Velocity (PV Trace)
    Making a peak
    velocity
    measurement1.  Select the measurement; an active caliper with a vertical 
    dotted line displays.
    2.  To position the caliper at the desired measurement point, 
    move the Trackball.
    3.  To complete the measurement, press Set. 
    The system displays the velocity measurement in the 
    Results Window. 
    						
    							Cardiology Measurements
    LOGIQ 5 Basic User Manual 10-55
    Direction 2300002-100 Rev. 2
    Pulmonic Valve (continued)
    One velocity flow
    trace
    measurements•  Pulmonary Artery Diastolic Pressure (PV Trace)
    •  Pulmonic Insufficiency Mean Pressure Gradient (PR Trace)
    •  Pulmonic Valve Mean Pressure Gradient (PV Trace)
    •  Pulmonic Insufficiency Mean Velocity (PR Trace)
    •  Pulmonic Insufficiency Mean Square Root Velocity(PR 
    Trace)
    •  Pulmonic Insufficiency Velocity Time Integral (PR Trace)
    •  Pulmonic Valve Mean Velocity (PV Trace)
    •  Pulmonic Valve Mean Square Root Velocity (PV Trace)
    •  Pulmonic Valve Velocity Time Integral (PV Trace)
    Making a velocity
    flow trace
    measurement1.  Select the measurement; an active caliper displays.
    2.  To position the caliper at the start point, move the Trackball.
    3.  To fix the trace start point, press Set.
    The system displays a vertical dotted line.
    4.  To trace the envelope, move the Trackball.
    A line shows the traced area.
    5.  To complete the trace, press Set.
    The system displays a second vertical dotted line.
    6.  To position the second dotted line at the start of the next 
    envelope, move the Trackball.
    7.  To complete the measurement, press Set.
    The system displays the measurement in the Results 
    Window. 
    						
    							Cardiology
    10-56 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Pulmonic Valve (continued)
    One slope
    measurements•  Pulmonic Insufficiency Pressure Half Time (PR PHT)
    •  Pulmonic Valve Flow Acceleration (PV AccT)
    Making a slope
    measurement1.  Select the measurement; an active caliper displays.
    2.  To position the caliper at the start point, move the Trackball.
    3.  To fix the start point, press Set.
    The system fixes the first caliper and displays a second 
    active caliper.
    4.  To position the second caliper at the end point, move the 
    Trackball.
    A dotted line shows the slope.
    5.  To complete the measurement, press Set. 
    The system displays the slope measurement in the Results 
    Window.
    One time interval
    measurements•  Pulmonic Valve Acceleration Time (PV AccT)
    •  Pulmonic Valve Ejection Time (PVET)
    •  QRS complex to end of envelope (Q-to-PV close)
    • Time 
    						
    							Cardiology Measurements
    LOGIQ 5 Basic User Manual 10-57
    Direction 2300002-100 Rev. 2
    Pulmonic Valve (continued)
    Making a time
    interval
    measurement1.  Select the measurement.
    The system displays an active caliper and a vertical dotted 
    line. 
    2.  To position the caliper at the start point, move the Trackball.
    3.  To fix the first caliper, press Set.
    The system displays a second active caliper.
    4.  To position the caliper at the end point, move the Trackball.
    5.  To complete the measurement, press Set.
    The system displays the time interval in the Results 
    Window.
    Two time intervals
    measurements•  Pulmonic Valve Acceleration to Ejection TIme Ratio (PV 
    AccT, PVET)
    •  Pulmonic Valve Pre-Ejection to Ejection Time Ratio 
    (PVPEP, PVET) 
    						
    							Cardiology
    10-58 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Right Ventricle
    The following are cardiac Doppler Mode Right Ventricle 
    measurements:
    One peak velocity
    measurements•  Right Ventricle Outflow Tract Peak Pressure Gradient 
    (RVOT Vmax)
    •  Right Ventricle Systolic Pressure (RVOT Vmax)
    •  Right Ventricle Outflow Tract Peak Velocity (RVOT Vmax)
    Making a peak
    velocity
    measurement1.  Select the measurement; an active caliper with a vertical 
    dotted line displays.
    2.  To position the caliper at the desired measurement point, 
    move the Trackball.
    3.  To complete the measurement, press Set. 
    The system displays the velocity measurement in the 
    Results Window.
    One velocity flow
    trace
    measurement•  Right Ventricle Diastolic Pressure (RVOT Trace)
    •  Right Ventricle Outflow Tract Velocity Time Integral (RVOT 
    Trace) 
    						
    							Cardiology Measurements
    LOGIQ 5 Basic User Manual 10-59
    Direction 2300002-100 Rev. 2
    Right Ventricle (continued)
    Making a velocity
    flow trace
    measurement1.  Select the measurement; an active caliper displays.
    2.  To position the caliper at the start point, move the Trackball.
    3.  To fix the trace start point, press Set.
    The system displays a vertical dotted line.
    4.  To trace the envelope, move the Trackball.
    A line shows the traced area.
    5.  To complete the trace, press Set.
    The system displays a second vertical dotted line.
    6.  To position the second dotted line at the start of the next 
    envelope, move the Trackball.
    7.  To complete the measurement, press Set.
    The system displays the measurement in the Results 
    Window.
    One time interval
    measurement•  Right Ventricle Ejection Time (PV Trace) 
    						
    							Cardiology
    10-60 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Right Ventricle (continued)
    Making a time
    interval
    measurement1.  Select the measurement.
    The system displays an active caliper and a vertical dotted 
    line. 
    2.  To position the caliper at the start point, move the Trackball.
    3.  To fix the first caliper, press Set.
    The system displays a second active caliper.
    4.  To position the caliper at the end point, move the Trackball.
    5.  To complete the measurement, press Set.
    The system displays the time interval in the Results 
    Window.
    One velocity flow trace and one area measurements:
    •  Stroke Volume by Pulmonic Flow (RVOT Trace)
    Body surface area and stroke volume measurements:
    •  Right Ventricle Stroke Volume Index by Pulmonic Flow 
    (RVOT Trace)
    The system calculates body surface area from the patient’s 
    height and weight. 
    						