    							Adjusting the Display Monitor
    LOGIQ 5 Basic User Manual 3-31
    Direction 2300002-100 Rev. 2
    Brightness and Contrast (continued)
    To adjust the brightness:
    1.  Press the Toggle button on the display monitor twice.
    2.  Press the Adjustment (+) button to increase brightness.  
    Press the Adjustment (-) button to decrease brightness.
    The amount of brightness is shown on a slide bar on the 
    screen.
    To adjust the contrast:
    1.  Press the Toggle button on the display monitor once.
    2.  Press the Adjustment (+) button to increase contrast. Press 
    the Adjustment (-) button to decrease contrast.
    The amount of contrast is shown on a slide bar on the 
    screen.
     Figure 3-17. Brightness and Contrast
    1
    2
    345
    1.  Contrast Indicator 4.  Brightness/Contrast Adjustment (-) 
    Button
    2.  Brightness Indicator 5.  Brightness/Contrast Adjustment (+) 
    Button
    3.  Brightness/Contrast Toggle Button 
    						
    							Preparing the System for Use
    3-32 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Brightness and Contrast (continued)
    Record any changes to the final brightness and contrast settings 
    and leave this information with the system. Generally speaking, 
    do not change the controls once they have been set. Once set, 
    the display then becomes the reference for the hard copy 
    device(s).
    NOTE:  After readjusting the monitors Contrast and Brightness, readjust 
    all preset and peripheral settings. 
    						
    							Adjusting the Display Monitor
    LOGIQ 5 Basic User Manual 3-33
    Direction 2300002-100 Rev. 2
    Manual Degauss
    Degaussing refers to the process of removing magnetic-field 
    effects from the monitor. Operation of the monitor within a 
    magnetic field may adversely effect color purity. Degaussing can 
    be used to correct this problem. The monitor automatically 
    degausses at power on.
    To activate the manual degauss:
    1.  Press and hold the monitor adjustment button for a minute. 
    The monitor adjustment menu appears. 
    2.  Move to the Degauss menu and press the monitor 
    adjustment button (+).  The monitor is degaussed.
    NOTE:  Monitor degaussing (demagnetizing) is done automatically when 
    the system is turned on.
    Speakers
    Stereo audio, provided by speakers located on the sides of the 
    display monitor, is standard for:
    •  Audio Doppler operation (left side is blood flow away/right 
    side is blood flow toward)
    •  Audio playback of videotaped scan sessions
    •  Audio error notification. 
    						
    							Preparing the System for Use
    3-34 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Probes
    Introduction
    Only use approved probes. 
    All imaging probes can be plugged into any of the probe ports.
    Refer to the Probes chapter.
    Selecting a probe
    •  Always start out with a probe that provides optimum focal 
    depths and penetration for the patient size and exam.
    •  Begin the scan session using the default Power Output 
    setting for the probe and exam.
    NOTE:  Selecting a new probe unfreezes the image. 
    						
    							Probes
    LOGIQ 5 Basic User Manual 3-35
    Direction 2300002-100 Rev. 2
    Connecting the Probe
    Probes can be connected at any time, regardless of whether the 
    console is powered on or off. To ensure that the ports are not 
    active, place the system in the image freeze condition.
    To connect a probe:
    1.  Place the probes carrying case on a stable surface and 
    open the case.
    2.  Carefully remove the probe and unwrap the probe cord.
    3.  DO NOT allow the probe head to hang free. Impact to the 
    probe head could result in irreparable damage.
    4.  Turn the connector locking handle counterclockwise.
    5.  Align the connector with the probe port and carefully push 
    into place.
    6.  Turn the connector locking handle clockwise to secure the 
    probe connector.
    7.  Carefully position the probe cord so it is free to move and is 
    not resting on the floor.
    CAUTIONFault conditions can result in electric shock hazard. Do not 
    touch the surface of probe connectors which are exposed 
    when the probe is removed. Do not touch the patient when 
    connecting or disconnection a probe. 
    						
    							Preparing the System for Use
    3-36 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Connecting the Pencil Probe (P2D and P6D)
    When using a pencil probe, it is necessary to attach a probe 
    adapter to the system.
    To attach a probe adapter and connect a pencil probe:
    Turn the locking handle counterclockwise to unlock.
    Align the connector of the adapter with the probe port and 
    carefully push into place.
     Figure 3-18. Align the connector with the probe port
    Turn the connector locking handle clockwise to the full vertical 
    position to lock in place. 
    						
    							Probes
    LOGIQ 5 Basic User Manual 3-37
    Direction 2300002-100 Rev. 2
    Connecting the Pencil Probe (P2D and P6D) (continued)
    Insert the probe connector into the adapter until you hear the 
    locking click. Carefully position the probe cord so it is free to 
    move and is not resting on the floor.
     Figure 3-19. Insert the probe connector to the adapter
    WARNINGWhen connecting the probe adapter to the system, first verify 
    the direction of the connector. The connector portion of the 
    adapter must be properly oriented to the system before locking 
    the probe adapter. The probe adapter should only be 
    connected when the connector portion of the adapter is 
    properly aligned mechanically with the system.
    WARNINGSerious injury and equipment damage may result if the probe 
    adapter is improperly connected to the system by applying 
    excessive force. Carefully align the probe connector to the 
    system and push it firmly into place. Turn the locking handle 
    clockwise to secure the probe connector to the system. 
    						
    							Preparing the System for Use
    3-38 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Cable Handling 
    Take the following precautions with probe cables:
    •  Keep free from wheels
    •  Do not bend the cable acutely
    •  Avoid crossing cables between probes. 
    						
    							Probes
    LOGIQ 5 Basic User Manual 3-39
    Direction 2300002-100 Rev. 2
    Activating the Probe
    To activate the probe, select the appropriate probe from the 
    probe indicators on the Touch Panel.
     Figure 3-20. Probe Indicators
    1. Probe Indicators
    The probe activates in the currently-selected operating mode. 
    The probes default settings for the mode and selected exam are 
    used automatically.
    1 
    						
    							Preparing the System for Use
    3-40 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Deactivating the Probe
    When deactivating the probe, the probe is automatically placed 
    in standby mode.
    To deactivate a probe:
    Disconnecting the Probe
    Probes can be disconnected at any time. However, the probe 
    should not be selected as the active probe.
    •  Move the probe locking handle counterclockwise.
    •  Pull the probe and connector straight out of the probe port.
    •  Carefully slide the probe and connector away from the 
    probe port and around the right side of the keyboard.
    •  Ensure the cable is free.
    •  Be sure that the probe head is clean before placing the 
    probe in its storage box.
    Disconnecting the Pencil Probe (P2D and P6D)
    Pull out the probe connector carefully.
    Ensure the probe cable is free.
    Be sure that the probe head is clean before placing the probe in 
    its case.
    Move the adapter locking handle counterclockwise into the 
    horizontal position.
    Carefully slide the adapter away from the probe port.
    Store the adapter in its case.
    Fr e e ze1. Press the Freeze key.
    2.  Gently wipe the excess gel from the face of the probe.
    3.  Carefully slide the probe around the right side of the 
    keyboard, toward the probe holder. Ensure that the probe 
    is placed gently in the probe holder. 
    						