    							Cardiology Measurements
    LOGIQ 5 Basic User Manual 10-21
    Direction 2300002-100 Rev. 2
    Pulmonic Valve
    The following are cardiac B-Mode Pulmonic Valve 
    measurements:
    One distance
    measurement•  Pulmonic Valve Annulus Diameter (PV Ann Diam)
    •  Pulmonic Diameter (Pulmonic Diam)
    Making a one
    distance
    measurement1.  Select the measurement; an active caliper displays.
    2.  To position the active caliper at the start point, move the 
    Trackball.
    3.  To fix the start point, press Set.
    The system fixes the first caliper and displays a second 
    active caliper.
    4.  To position the second active caliper at the end point, move 
    the Trackball.
    A dotted line connects the measurement points.
    5.  To complete the measurement, press Set.
    The system displays the measurement in the Results 
    Window. 
    						
    							Cardiology
    10-22 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Right Atrium
    The following are cardiac B-Mode Right Atrium measurements:
    One distance
    measurement•  Right Atrium Diameter, Length (RA Major)
    •  Right Atrium Diameter, Length (RA Minor)
    Making a one
    distance
    measurement1.  Select the measurement; an active caliper displays.
    2.  To position the active caliper at the start point, move the 
    Trackball.
    3.  To fix the start point, press Set.
    The system fixes the first caliper and displays a second 
    active caliper.
    4.  To position the second active caliper at the end point, move 
    the Trackball.
    A dotted line connects the measurement points.
    5.  To complete the measurement, press Set.
    The system displays the measurement in the Results 
    Window. 
    						
    							Cardiology Measurements
    LOGIQ 5 Basic User Manual 10-23
    Direction 2300002-100 Rev. 2
    Right Atrium (continued)
    One trace
    measurements•  Right Atrium Area (RA Area)
    •  Right Atrium Volume, SIngle Plane, Method of Disk (RAAd)
    •  Right Atrium Volume, Systolic, SIngle Plane, Method of Disk 
    (RAAs)
    Making a trace
    measurement1.  Select the measurement; an active caliper displays.
    2.  To position the caliper at the start point, move the Trackball.
    3.  To fix the trace start point, press Set.
    The system displays a vertical dotted line.
    4.  To trace the measurement area, move the Trackball.
    A line shows the traced area.
    5.  To complete the measurement, press Set.
    The system displays the measurement in the Results 
    Window. 
    						
    							Cardiology
    10-24 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Right Ventricle
    The following are cardiac B-Mode Right Ventricle 
    measurements:
    One trace
    measurement•  Left Pulmonary Artery Area (LPA Area)
    •  Right Pulmonary Artery Area (RPA Area)
    Making a trace
    measurement1.  Select the measurement; an active caliper displays.
    2.  To position the caliper at the start point, move the Trackball.
    3.  To fix the trace start point, press Set.
    The system displays a vertical dotted line.
    4.  To trace the measurement area, move the Trackball.
    A line shows the traced area.
    5.  To complete the measurement, press Set.
    The system displays the measurement in the Results 
    Window. 
    						
    							Cardiology Measurements
    LOGIQ 5 Basic User Manual 10-25
    Direction 2300002-100 Rev. 2
    Right Ventricle (continued)
    One distance
    measurements•  Right Ventricle Internal Diameter
    • Diastolic (RVIDd)
    • Systolic (RVIDs)
    •  Right Ventricle Wall Thickness
    • Diastolic (RVAWd)
    • Systolic (RVAWs)
    •  Right Ventricle Outflow Tract Diameter (RVOT Diam)
    •  Left Pulmonary Artery (LPA)
    •  Main Pulmonary Artery (MPA)
    •  Right Pulmonary Artery (RPA)
    Making a one
    distance
    measurement1.  Select the measurement; an active caliper displays.
    2.  To position the active caliper at the start point, move the 
    Trackball.
    3.  To fix the start point, press Set.
    The system fixes the first caliper and displays a second 
    active caliper.
    4.  To position the second active caliper at the end point, move 
    the Trackball.
    A dotted line connects the measurement points.
    5.  To complete the measurement, press Set.
    The system displays the measurement in the Results 
    Window. 
    						
    							Cardiology
    10-26 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    System
    The following are cardiac B-Mode System measurements:
    One distance
    measurements•  Interventricular Septum Thickness
    • Diastolic (IVSd)
    • Systolic (IVSs)
    •  Inferior Vena Cava (IVC)
    •  Main Pulmonary Artery Diameter (MPA)
    •  Systemic Vein Diameter (Systemic Diam)
    •  Patent Ductus Arterosis Diameter (PDA Diam)
    •  Patent Foramen Ovale Diameter (PFO Diam) 
    •  Ventricular Septal Defect Diameter (VSD Diam)
    Making a one
    distance
    measurement1.  Select the measurement; an active caliper displays.
    2.  To position the active caliper at the start point, move the 
    Trackball.
    3.  To fix the start point, press Set.
    The system fixes the first caliper and displays a second 
    active caliper.
    4.  To position the second active caliper at the end point, move 
    the Trackball.
    A dotted line connects the measurement points.
    5.  To complete the measurement, press Set.
    The system displays the measurement in the Results 
    Window.
    Body surface area
    and stroke volume
    measurements•  Interventricular Septum (IVS) Fractional Shortening (IVSs)
    The system calculates body surface area from the patient’s 
    height and weight. 
    						
    							Cardiology Measurements
    LOGIQ 5 Basic User Manual 10-27
    Direction 2300002-100 Rev. 2
    Tricuspid Valve
    The following are cardiac B-Mode Tricuspid Valve 
    measurements:
    One distance
    measurements•  Tricuspid Valve Annulus Diameter (TV Ann Diam)
    •  Tricuspid Valve Area (TV Area)
    Making a one
    distance
    measurement1.  Select the measurement; an active caliper displays.
    2.  To position the active caliper at the start point, move the 
    Trackball.
    3.  To fix the start point, press Set.
    The system fixes the first caliper and displays a second 
    active caliper.
    4.  To position the second active caliper at the end point, move 
    the Trackball.
    A dotted line connects the measurement points.
    5.  To complete the measurement, press Set.
    The system displays the measurement in the Results 
    Window. 
    						
    							Cardiology
    10-28 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    M-Mode Measurements
    Aorta
    The following are cardiac M-Mode aorta measurements:
    One distance
    measurement•  Aortic Root Diameter (Ao Root Diam)
    Making a one
    distance
    measurement1.  Select the measurement; an active caliper displays.
    2.  To position the active caliper at the start point, move the 
    Trackball.
    3.  To fix the start point, press Set.
    The system fixes the first caliper and displays a second 
    active caliper.
    4.  To position the second active caliper at the end point, move 
    the Trackball.
    A dotted line connects the measurement points.
    5.  To complete the measurement, press Set.
    The system displays the measurement in the Results 
    Window. 
    						
    							Cardiology Measurements
    LOGIQ 5 Basic User Manual 10-29
    Direction 2300002-100 Rev. 2
    Aortic Valve
    The following are cardiac M-Mode aortic valve measurements:
    One distance
    measurements•  Aortic Valve Cusp Separation (AV  C u s p)
    Making a one
    distance
    measurement1.  Select the measurement; an active caliper displays.
    2.  To position the active caliper at the start point, move the 
    Trackball.
    3.  To fix the start point, press Set.
    The system fixes the first caliper and displays a second 
    active caliper.
    4.  To position the second active caliper at the end point, move 
    the Trackball.
    A dotted line connects the measurement points.
    5.  To complete the measurement, press Set.
    The system displays the measurement in the Results 
    Window. 
    						
    							Cardiology
    10-30 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Left Atrium
    The following are cardiac M-Mode Left Atrium measurements:
    Two distance
    measurement
    (ratio)•  Left Atrium Diameter to AoRoot Diameter Ratio (LA/Ao)
    Making two
    distance
    measurements1. Select LA/Ao; an active caliper displays.
    2.  Make the first distance measurement:
    a.  To position the active caliper at the start point, move the 
    Trackball.
    b.  To fix the start point, press Set.
    The system fixes the first caliper and displays a second 
    active caliper.
    c.  To position the second active caliper at the end point, 
    move the Trackball.
    A dotted line connects the measurement points.
    d.  To complete the measurement, press Set.
    The system displays the distance value in the Results 
    Window. After the first measurement, the system 
    displays an active caliper.
    3.  To make the second distance measurement, repeat steps 
    a–d above.
    The system displays the measurements and ratio in the 
    Results Window. 
    						