GE Logiq P5 User Manual
Have a look at the manual GE Logiq P5 User Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
OB-Multigestational LOGIQ 5 Basic User Manual 9-75 Direction 2300002-100 Rev. 2 Using other OB studies (continued) Figure 9-19. OB Touch Panel: ASUM studies NOTE: ASUM studies include the following measurements: • ASUM: AC, BPD, and CRL • ASUM 2001: AC, BPD, CRL, FL, HC, HL, and OFD
OB/GYN 9-76 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Multiple Fetus LOGIQ 5 allows you to measure and report multiple fetus development. The system can report a maximum of four fetuses. To enter the number of fetuses If more than one fetus is imaged during the exam, enter the number of fetuses in the Patient Data Entry Menu. Figure 9-20. Fetus Number When you start an OB exam, the system automatically fills in the Fetus # field with 1. To change the number: 1. Move the cursor to the fetus number and press Set twice. The number is highlighted. 2. Type the correct number and press Set. The system displays a message to confirm that you want to change the fetus number. 3. Select Yes.
OB-Multigestational LOGIQ 5 Basic User Manual 9-77 Direction 2300002-100 Rev. 2 To identify each fetus For measurements, calculations, and worksheet displays, the system labels each fetus A, B, C, or D. Each fetus is identified by a letter and the total number of fetuses. For example, fetus A/3 is fetus A from a total of 3. When scanning, you can enter information about the fetus position and placenta location. You can enter the information in the Patient Data section of the worksheets and the graphs. You can type up to 23 characters in the FetusPos and PLAC fields. Figure 9-21. Patient Data section of the OB Worksheet To select a fetus During measurements and calculations, to change between fetuses, do one of the following: • Adjust the Fetus knob. • Move the Trackball to the Summary Window and select the fetus. Figure 9-22. Summary Window: Multiple fetus You can change between fetuses at any time during the exam. NOTE: After you change to the next fetus, any measurements you make are recorded and reported to that fetus. If you have any active measurement or calculation that is not completed when you change the fetus, the system cancels the measurement or calculation.
OB/GYN 9-78 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 To view multiple fetuses data on graphs You can view multiple gestation data on fetal growth curve graphs. After you have made measurements for each fetus, select Graph Display. • To view the graph for each fetus, do one of the following: • Adjust the Fetus knob. • In the Patient Data section, move the Trackball to highlight the FetusNo field. In the list of fetuses, move the Trackball to select the fetus you want, and press Set. • To display data for multiple fetuses on the same graph, select Fetus Compare. Figure 9-23. Fetal Growth Curve Graph: Fetus Compare The legend at the bottom of the graph shows the symbols and colors that represent each fetus.
OB-Multigestational LOGIQ 5 Basic User Manual 9-79 Direction 2300002-100 Rev. 2 To compare multiple fetus data on a worksheet With multiple fetuses, you can list and compare measurements of the fetuses on the worksheet. Select Worksheet Display, then select Fetus Compare. Figure 9-24. OB Worksheet Display Touch Panel
OB/GYN 9-80 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 To compare multiple fetus data on a worksheet (continued) When you select Fetus Compare, the system lists the measurement results for each fetus on the Worksheet. Figure 9-25. Worksheet Display with Fetus Compare
OB-Multigestational LOGIQ 5 Basic User Manual 9-81 Direction 2300002-100 Rev. 2 To Show Fetal Trending for Multiple Fetuses When you have data for more than one exam, you can show fetal trending and compare fetuses on one graph. To view fetal trending for multiple fetuses: 1. Select Graph Display. 2. Select Fetus Compare. 3. Select Plot Both. NOTE: You can only view fetal trending for multiple fetuses in single graph display. Figure 9-26. Multigestation Fetal Growth Curve Graph The symbol key for fetal trending and multiple fetuses is shown at the bottom of the graph.
OB/GYN 9-82 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 OB Table Editor You can add user programmable OB tables to the system. OB Table Settings Menu You add OB Tables in the Measurement & Analysis menu. To open the menu: 1. On the Touch Panel, select Utility, then select M&A. 2. Check the Exam Category on the far left of the monitor screen. Make sure that Obstetrics is selected. If it is not selected, select Obstetrics and continue selecting the folders until the appropriate area is selected as to where this new OB Table will be entered. For example, select Obstetrics, then select OB-2/3. If there are further folders within OB-2/3, select that appropriate folder.
OB Table Editor LOGIQ 5 Basic User Manual 9-83 Direction 2300002-100 Rev. 2 OB Table Settings Menu (continued) 3. On the monitor display, select the OB Table tab. The system displays the OB Table settings menu. Figure 9-27. OB Table settings
OB/GYN 9-84 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 OB Table Settings Menu (continued) 4. The OB Table settings menu lists OB Table parameters. Specify the following parameter values as appropriate: • Study: Shows the study to which this measurement table belongs. • New/Edit: To create a new OB table, select New Table. To edit an existing user-programmable OB table, select Edit Table. NOTE: You cannot edit the system’s OB Tables. • O B Ta b l e Te m p l a t e: To create a new OB table, select the Template (1 - 7) which you want to use as the basis of the user programmable OB Table. See ‘OB Table Templates’ for more information. To edit an existing user OB table, select the desired OB table to edit. • Tool Type: Select the type of measurement: Distance or Circumference. • Measure Name: Type the name of measurement that will display on the Touch Panel. • Author Name: Type the author’s name. • Ta b l e Ty p e: If necessary, select the Table Type: Fetal Age or Fetal Growth. • Measure Type: Select a measurement type that can be used to calculate EFW, for example BPD. NOTE: Measure Type is used only when calculating EFW. NOTE: The following items are display only: Table Format, Table Unit, SD/GP Range, and Graph Range. The system determines these values automatically, based on the type of OB table you are creating. 5. After specifying all parameter values, move the Trackball to Edit Table and press Set. The system displays the Edit Menu. NOTE: If any of the OB table parameters are not correct, the Edit Menu is not displayed.