    							Abdomen
    LOGIQ 5 Basic User Manual 8-7
    Direction 2300002-100 Rev. 2
    M-Mode Measurements
    In M-Mode, all Exam Calcs include the following measurements:
    • % Stenosis
    • A/B Ratio
    • Heart Rate
     Figure 8-3. Abdomnal Calculations M-Mode Touch Panel
    See ‘M-Mode Measurements’ for more information.. 
    						
    							Abdomen and Small Parts
    8-8 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Doppler Mode Measurements
    Generic Exam Calcs
    In Doppler Mode, the Generic Exam Calcs tab includes the 
    following measurements:
    •  PI (Pulsatility Index)
    •  RI (Resistive Index)
    • S/D Ratio
    • D/S Ratio
    • A/B Ratio
    •  Max PG (Pressure Gradient)
    •  Mean PG (Pressure Gradient)
    •  SV (Stroke Volume)
    • Heart Rate
     Figure 8-4. Doppler Mode Touch Panel:  Abdominal 
    Calculations
    See ‘Doppler Mode Measurements’ for more information. 
    						
    							Abdomen
    LOGIQ 5 Basic User Manual 8-9
    Direction 2300002-100 Rev. 2
    Abdomen and Renal Artery Exam Calcs
    Your system is set up to show the measurements that you 
    usually make for each region of interest. To make a 
    measurement that is not shown for the selected region:
    1.  Select the folder for the region you want to measure. Refer 
    to Figure 8-5 and Figure 8-6 for the Abdomen and Renal 
    Artery Exam Calcs, respectively.
     Figure 8-5. Abdomen Exam Calcs Folder (Doppler Mode) in 
    Abdomen exam category 
    						
    							Abdomen and Small Parts
    8-10 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Abdomen and Renal Artery Exam Calcs (continued)
     Figure 8-6. Renal Exam Calcs (Doppler Mode)
    2.  Press Show All.
     Figure 8-7. Aorta folder 
    						
    							Abdomen
    LOGIQ 5 Basic User Manual 8-11
    Direction 2300002-100 Rev. 2
    Abdomen and Renal Artery Exam Calcs (continued)
    The system displays all possible measurements.
     Figure 8-8. Show All measurements
    3.  Select the desired measurement.
    NOTE:  The following instructions assume that you first scan the 
    patient and then press Freeze. 
    						
    							Abdomen and Small Parts
    8-12 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Abdomen and Renal Artery Exam Calcs (continued)
    Acceleration1. Select Accel. 
    The system displays an active caliper with vertical and 
    horizontal dotted lines.
    2.  To position the caliper at peak systole, move the Trackball.
    3.  To fix the measure point, press Set.
    The system displays a second active caliper.
    4.  To position the second caliper at end diastole, move the 
    Trackball.
    5.  To complete the measurement, press Set.
    The system displays the peak systole, end diastole, 
    acceleration time, and acceleration in the Results Window. 
    Acceleration Time
    (AT)1. Select AT. 
    The system displays an active caliper and a vertical dotted 
    line. 
    2.  To position the caliper at the start point, move the Trackball.
    3.  To fix the first caliper, press Set.
    The system displays a second active caliper.
    4.  To position the caliper at the end point, move the Trackball.
    5.  To complete the measurement, press Set.
    The system displays the acceleration time in the Results 
    Window. 
    						
    							Abdomen
    LOGIQ 5 Basic User Manual 8-13
    Direction 2300002-100 Rev. 2
    Abdomen and Renal Artery Exam Calcs (continued)
    Heart Rate
    To calculate the heart rate:
    1. Select Heart Rate. 
    The system displays an active caliper and a vertical dotted 
    line. 
    2.  To position the caliper at a recognizable point in the first 
    cycle, move the Trackball.
    3.  To fix the first caliper, press Set.
    The system displays a second active caliper.
    4.  To position the caliper at the identical point in the second or 
    third cycle, move the Trackball.
     Figure 8-9. Two Heart Beat Reference
    5.  To complete the measurement and transfer the calculation 
    to the worksheet, press Set. 
    						
    							Abdomen and Small Parts
    8-14 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Abdomen and Renal Artery Exam Calcs (continued)
    Peak Systole (PS),
    End Diastole (ED),
    or Mid Diastole
    (MD)To calculate the peak systole, end diastole, or mid diastole:
    1. Select PS, ED, or MD. 
    The system displays an active caliper with vertical and 
    horizontal dotted lines.
    2.  To position the caliper at the measurement point, move the 
    Trackball.
    3.  To complete the measurement, press Set.
    The system displays the peak systole, end diastole, or mid 
    diastole in the Results Window. 
    Pulsatility Index
    (PI)1. Select PI. 
    The system displays a trace caliper and a vertical dotted 
    line.
    2.  To position the caliper at the beginning of the waveform 
    (PS), move the Trackball.
    3.  To fix the start point, press Set.
    The system displays a second active caliper.
    4.  To trace the waveform from PS to ED, move the Trackball.
    5.  To complete the measurement, press Set.
    The system displays peak systole, mid diastole, end 
    diastole, TAMAX, and the PI in the Results Window.  
    						
    							Abdomen
    LOGIQ 5 Basic User Manual 8-15
    Direction 2300002-100 Rev. 2
    Abdomen and Renal Artery Exam Calcs (continued)
    PS/ED or ED/PS
    RatioTo calculate the Peak Systole/End Diastole ratio or End 
    Diastole/Peak Systole ratio:
    1. Select PS/ED or ED/PS. 
    The system displays an active caliper with vertical and 
    horizontal dotted lines.
    2.  To position the caliper at peak systole (PS) or end diastole 
    (ED), move the Trackball.
    3.  To fix the measure point, press Set.
    The system displays a second active caliper.
    4.  To position the second caliper at end diastole (ED) or peak 
    systole (PS), move the Trackball.
    5.  To complete the measurement, press Set.
    The system displays the peak systole, end diastole, and PS/
    ED or ED/PS ratio in the Results Window. 
    Resistive Index
    (RI)1. Select RI. 
    The system displays an active caliper with vertical and 
    horizontal dotted lines.
    2.  To position the caliper at the peak systolic velocity, move the 
    Trackball.
    3.  To fix the measure point, press Set.
    The system displays a second active caliper.
    4.  To position the second caliper at the end diastolic velocity, 
    move the Trackball.
    5.  To complete the measurement, press Set.
    The system displays PS
    , ED, and RI in the Results Window.  
    						
    							Abdomen and Small Parts
    8-16 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Abdomen and Renal Artery Exam Calcs (continued)
    TA M A X1. Select TA M A X. 
    The system displays a trace caliper and a vertical dotted 
    line.
    2.  To position the caliper, move the Trackball.
    3.  To fix the start point, press Set.
    The system displays a second active caliper.
    4.  To trace the waveform, move the Trackball.
    5.  To complete the measurement, press Set.
    The system displays VTI, Time, and TAMAX in the Results 
    Window.  
    						