GYN Measurements LOGIQ 5 Basic User Manual 9-95 Direction 2300002-100 Rev. 2 To Start a Gynecology Exam To begin a gynecology exam, you enter patient data or, if the patient data from a previous exam is saved in the system, find the patient information. For details about how to start an exam, See ‘To Start an Obstetrics Exam’ for more information. After you complete the patient information, you can begin the scan. 1. To change from the Patient Data Entry screen to the Scan screen, do one of the following: • On the keyboard, press Esc or Patient. • On the Touch Panel, select Scan. • On the Control Panel, press the B-Mode key. The system displays the Scan screen. 2. To choose the appropriate probe, select the probe icon on the Touch Panel. 3. On the Control Panel, press Measure. The default Gynecology study is displayed on the Touch Panel.
OB/GYN 9-96 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 B-Mode Measurements In B-Mode, you make the measurements in the General Gynecology study. These measurements include: • Uterine length, width, and height • Ovarian length, width, and height • Ovarian follicle • Endometrium thickness Figure 9-30. General Gynecology study
GYN Measurements LOGIQ 5 Basic User Manual 9-97 Direction 2300002-100 Rev. 2 Follicle measurements You can make left and right ovary follicle measurements from one, two, or three distances. One distance1. To select the left or right, adjust the Side knob. 2. Select Follicle; an active caliper displays. 3. To position the active caliper at the start point, move the Trackball. 4. To fix the start point, press Set. The system fixes the first caliper and displays a second active caliper. 5. To position the second active caliper at the end point, move the Trackball. A dotted line connects the measurement points. 6. To complete the measurement, press Set. 7. Press Clear. The system displays the ovary follicle measurement in the Results Window.
OB/GYN 9-98 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Follicle measurements (continued) Two distances1. To select the left or right, adjust the Side knob. 2. Select Follicle; an active caliper displays. 3. Make the first distance measurement: a. To position the active caliper at the start point, move the Trackball. b. To fix the start point, press Set. The system fixes the first caliper and displays a second active caliper. c. To position the second active caliper at the end point, move the Trackball. A dotted line connects the measurement points. d. To complete the measurement, press Set. The system displays the distance value in the Results Window and displays an active caliper. 4. To make the second distance measurement, repeat steps a–d above. 5. Press Clear. The system displays the ovary follicle measurement in the Results Window.
GYN Measurements LOGIQ 5 Basic User Manual 9-99 Direction 2300002-100 Rev. 2 Follicle measurements (continued) Three distances1. To select the left or right, adjust the Side knob. 2. Select Follicle; an active caliper displays. 3. Make the first distance measurement: a. To position the active caliper at the start point, move the Trackball. b. To fix the start point, press Set. The system fixes the first caliper and displays a second active caliper. c. To position the second active caliper at the end point, move the Trackball. A dotted line connects the measurement points. d. To complete the measurement, press Set. The system displays the distance value in the Results Window. After the first and second distance measurement, the system displays an active caliper. 4. To make the second and third distance measurement, repeat steps a–d above. After the third measurement, the system displays the ovary follicle measurement in the Results Window.
OB/GYN 9-100 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Endometrium thickness (Endo) To measure the endometrium thickness, make one distance measurement. 1. Select Endo; an active caliper displays. 2. To position the active caliper at the start point, move the Trackball. 3. To fix the start point, press Set. The system fixes the first caliper and displays a second active caliper. 4. To position the second active caliper at the end point, move the Trackball. A dotted line connects the measurement points. 5. To complete the measurement, press Set. The system displays the endometrium thickness in the Results Window.
GYN Measurements LOGIQ 5 Basic User Manual 9-101 Direction 2300002-100 Rev. 2 Ovary length, width, and height You can measure the length, width, and height of the left and right ovaries. Each measurement is a typical distance measurement made in the appropriate scan plane. Typically, the width and height are measured on the axial plane while the length is measured on the sagittal plane. To measure ovarian length, width, or height: 1. Scan the patients right or left ovary in the appropriate plane. 2. To select left or right, adjust the Side knob. 3. Select the OV folder, then select OV L, OV W, or OV H. 4. Perform a standard distance measurement: a. To position the active caliper at the start point, move the Trackball. b. To fix the start point, press Set. The system fixes the first caliper and displays a second active caliper. c. To position the second active caliper at the end point, move the Trackball. A dotted line connects the measurement points. d. To complete the measurement, press Set. The system displays the distance value in the Results Window. After the first and second measurement, the system displays an active caliper for the next measurement. 5. To make the second and third distance measurements, repeat steps 3–4. After you complete the length, width, and height measurements, the system displays the ovarian volume in the Results Window.
OB/GYN 9-102 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Uterus length, width, and height Each of these is a standard distance measurement. Length is typically measured in the sagittal plane. Width and height are measured in the axial plane. To measure uterus length, width, or height: 1. Scan the patient in the appropriate scan plane. 2. Select the UT folder, then select UT L, UT W, or UT H. An active caliper displays. 3. Perform a standard distance measurement: a. To position the active caliper at the start point, move the Trackball. b. To fix the start point, press Set. The system fixes the first caliper and displays a second active caliper. c. To position the second active caliper at the end point, move the Trackball. A dotted line connects the measurement points. d. To complete the measurement, press Set. The system displays the distance value in the Results Window. After the first and second measurement, the system displays an active caliper for the next measurement. 4. To make the second and third distance measurement, repeat steps 2–3. After you complete the third distance measurement, the system displays the uterine volume in the Results Window.
GYN Measurements LOGIQ 5 Basic User Manual 9-103 Direction 2300002-100 Rev. 2 M-Mode Measurements M-Mode measurements for the Gynecology exam are identical to M-Mode measurements for the Obstetrics exam. These measurements include % stenosis, A/B ratio, and heart rate. For details regarding these measurements, See ‘M-Mode Measurements’ for more information. Figure 9-31. M-Mode Generic study
OB/GYN 9-104 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Doppler Mode Measurements Doppler measurements for the Gynecology exam are identical to Doppler measurements for the Obstetrics exam. These measurements include the following vessels: uterine, ovarian, umbilical, middle cerebral artery, aorta, placenta, and descending aorta. For each vessel, you can make any of the following measurements: peak systole, mid diastole, end diastole, heart rate, TAMAX, pulsatility index, resistive index, acceleration, PS/ED, ED/PS, and acceleration time. For details regarding these measurements, See ‘Doppler Mode Measurements’ for more information. Figure 9-32. Doppler Mode General Gynecology study