    							Vascular Measurements
    LOGIQ 5 Basic User Manual 11-31
    Direction 2300002-100 Rev. 2
    Manual Vascular Calculation
    For each vessel in Doppler mode, you can make any of the 
    following measurements:
    •  Peak Systole (PS)
    •  End Diastole (ED)
    •  Mid Diastole (MD)
    • Heart Rate
    • TAMAX
    • Pulsatility Index (PI)
    •  Resistive Index (RI)
    • S/D Ratio
    • D/S Ratio
    • Acceleration (Accel)
    •  Acceleration Time (AT) 
    						
    							Vascular
    11-32 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    To select vascular measurements
    Your system is set up to show the measurements that you 
    usually make for each vessel. To make a measurement that is 
    not shown for the selected vessel:
    1.  Select the folder for the vessel you want to measure.
    2. Select Show All. See Figure 11-12.
    The system displays all possible vessel measurements. See 
    Figure 11-13.
    3.  Select the desired measurement.
     Figure 11-12. ECA folder 
    						
    							Vascular Measurements
    LOGIQ 5 Basic User Manual 11-33
    Direction 2300002-100 Rev. 2
    To select vascular measurements (continued)
     Figure 11-13. Show All measurements
    NOTE:  The following instructions assume that you first scan the patient 
    and then press Freeze. 
    						
    							Vascular
    11-34 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Vascular Worksheet
    The vascular worksheet is structured to automatically display 
    vascular measurements made at specific anatomical sites. The 
    worksheet can also display an average, last, maximum, or 
    minimum value of the latest three measurements. Calculated 
    ratios are automatically summarized and displayed.
    To view the Vascular Worksheet
    1. Press Measure.
    2. Select Worksheet Display.
    The system displays the worksheet.
     Figure 11-14. Vascular Worksheet Example 
    						
    							Vascular Worksheet
    LOGIQ 5 Basic User Manual 11-35
    Direction 2300002-100 Rev. 2
    To view the Vascular Worksheet (continued)
    Only measured parameters are displayed. Location information 
    is labeled with vessel name first. Measured parameters of the 
    vessel are grouped under the vessel label.
    Selected value by method is highlighted, however, when the 
    average method is selected, the highlighted cursor is removed.
    When an entire vessel measurement does not have sides (left or 
    right), the side label is not displayed in that vessel study 
    worksheet.
    HINTSSome fields on the worksheet are view only, and others you 
    can change or select. To easily see which fields you can 
    change or select, move the Trackball. As the cursor moves 
    over a field that you can change or select, the field is 
    highlighted. 
    						
    							Vascular
    11-36 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Worksheet Display Touch Panel
     Figure 11-15. Worksheet Display Touch Panel
    1. Worksheet Display: To exit the worksheet display and 
    return to scanning.
    2. Vessel Worksheet: Select this key to display the Vessel 
    Worksheet when the Vessel Summary is displayed.
    3. Vessel Summary: Select this key to display the Vessel 
    Summary when the Vessel Worksheet is displayed.
    4. Examiner’s Comment: Select this key to display the 
    Examiner’s comment window. See ‘Examiner’s Comments’ 
    for more information.
    5. Generic Worksheet: Select this key to display the Generic 
    Worksheet. Generic study measurements/calculations, such 
    as volume and velocity, are displayed on this worksheet.
    6. Delete Value: Use to delete a value (each measurement 
    value). See ‘To edit a worksheet’ for more information.
    7. Exclude Value: Use to exclude a value from the result line. 
    See ‘To edit a worksheet’ for more information.
    8. Intravessel Ratio: Select this key to display the Intravessel 
    Ratio Calculation window. See ‘Intravessel ratio’ for more 
    information.
    9. Page Change (knob): If a worksheet has more data, to view 
    the next page, adjust the Page Change knob. 
    						
    							Vascular Worksheet
    LOGIQ 5 Basic User Manual 11-37
    Direction 2300002-100 Rev. 2
    To edit a worksheet
    To change data on a worksheet:
    1. Select Worksheet Display from any page of the Vascular 
    Calculation Touch Panel Menu.
    2.  To position the cursor at the field you want to change, move 
    the Trackball.
    The field is highlighted.
    3. Press Set.
    NOTE:  The resultant value is not editable.
    4.  Type the new data in the field and move the cursor to 
    another place, then the new data is displayed in blue and 
    asterisk are appended to value and resultant value to 
    indicate that it was manual entered. 
    The average measurements, calculations and ratios are 
    automatically updated to reflect the edited values.
     Figure 11-16. Display of the edited value
    NOTE:  If the user move the cursor to the edited value and press Set 
    key once, the value is back again before edit. 
    						
    							Vascular
    11-38 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    To edit a worksheet (continued)
    To delete data: 
    Delete Value key is used to delete value (each measured 
    value). Whenever particular value is deleted, the group (set) of 
    the measurement/study value is also deleted from the 
    worksheet as a group (set). When PS value is deleted, ED and 
    RI is also deleted if this value was measured as the set of RI 
    measurement.
    1. Select Worksheet from any page of the Vascular 
    Calculation Touch Panel Menu.
    2.  To position the cursor at the field you want to delete or 
    exclude, move the Trackball.
    The field is highlighted.
    3. Select Delete Value.
    For Example:
    1.  If the user measured RI 4 times, however, latest 3 set of 
    measurements were displayed in the worksheet.
    1.  Then, the user deleted PS value of #3 from the worksheet.
    2.  The whole set of #3 measurements is deleted from the 
    worksheet and #1 set of measurements is shifted and 
    displayed as below.
    Result Number #2 #3 #4
    PS 0.500 0.600 0.700
    ED 0.100 0.200 0.300
    RI 0.800 0.667 0.571
    Result Number #1 #2 #4
    PS 0.400 0.500 0.700
    ED 0.000 0.100 0.300
    RI 1.000 0.800 0.571 
    						
    							Vascular Worksheet
    LOGIQ 5 Basic User Manual 11-39
    Direction 2300002-100 Rev. 2
    To edit a worksheet (continued)
    To exclude data: 
    When the user selects a particular value on the Worksheet and 
    select Exclude Value, this value is excluded from result line and 
    resultant value is re-calculated without this value and also 
    calculation values using this value is ‘blank’.
    1. Select Worksheet Display from any page of the Vascular 
    Calculation Touch Panel Menu.
    2.  To position the cursor at the field you want to delete or 
    exclude, move the Trackball.
    The field is highlighted.
    3. Select Exclude Value.
    4.  The data in the field is not visible and is not included in 
    worksheet calculations as below.
    5.  To include a value that you previously excluded, select 
    Exclude Value.
     Figure 11-17. Display of the excluded value 
    						
    							Vascular
    11-40 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    To edit a worksheet (continued)
    To select the method:
    The user can select the method for calculating the cumulative 
    value.
    This value is only calculated by using displaying values. If the 
    user takes parameters more than 3 times, latest 3 values are 
    used for this calculation.
    1.  Move cursor to method column and press Set. 
    2.  The pull-down menu is displayed. Move to cursor any one of 
    methods and press Set. The selected method is displayed 
    in the column.
     Figure 11-18. Pop-up menu of methods 
    						