    							Cardiology Measurements
    LOGIQ 5 Basic User Manual 10-61
    Direction 2300002-100 Rev. 2
    System
    The following are cardiac Doppler Mode System measurements:
    One peak velocity
    measurements•  Pulmonary Artery Peak Velocity (PV Vmax)
    •  Pulmonary Vein Velocity Peak A (reverse) (P Vein A)
    •  Pulmonary Vein Peak Velocity 
    • End-Diastolic (P Vein D)
    • Systolic (P Vein S)
    •  Systemic Vein Peak Velocity 
    • End-Diastolic (PDA Diastolic)
    • Systolic (PDA Systolic)
    •  Ventricular Septal Defect Peak Velocity (VSD Vmax)
    Making a peak
    velocity
    measurement1.  Select the measurement; an active caliper with a vertical 
    dotted line displays.
    2.  To position the caliper at the desired measurement point, 
    move the Trackball.
    3.  To complete the measurement, press Set. 
    The system displays the velocity measurement in the 
    Results Window. 
    						
    							Cardiology
    10-62 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    System (continued)
    One velocity flow
    trace
    measurements•  Pulmonary Artery Velocity Time Integral (Pulmonic VTI)
    •  Systemic Vein Velocity Time Integral (Systemic VTI)
    Making a velocity
    flow trace
    measurement1.  Select the measurement; an active caliper displays.
    2.  To position the caliper at the start point, move the Trackball.
    3.  To fix the trace start point, press Set.
    The system displays a vertical dotted line.
    4.  To trace the envelope, move the Trackball.
    A line shows the traced area.
    5.  To complete the trace, press Set.
    The system displays a second vertical dotted line.
    6.  To position the second dotted line at the start of the next 
    envelope, move the Trackball.
    7.  To complete the measurement, press Set.
    The system displays the measurement in the Results 
    Window.
    One time interval
    measurements•  Pulmonary Vein A-Wave Duration (P Vein A Dur)
    • Time
    •  IsoVolumetric Relaxation Time (IVRT)
    •  IsoVolumetric Contraction Time (IVCT) 
    						
    							Cardiology Measurements
    LOGIQ 5 Basic User Manual 10-63
    Direction 2300002-100 Rev. 2
    System (continued)
    Making a time
    interval
    measurement1.  Select the measurement.
    The system displays an active caliper and a vertical dotted 
    line. 
    2.  To position the caliper at the start point, move the Trackball.
    3.  To fix the first caliper, press Set.
    The system displays a second active caliper.
    4.  To position the caliper at the end point, move the Trackball.
    5.  To complete the measurement, press Set.
    The system displays the time interval in the Results 
    Window.
    Two peak velocity
    measurements•  Pulmonary Vein S/D Ratio (P Vein D, P Vein S)
    •  Ventricular Septal Defect Peak Pressure Gradient (VSD 
    Vmax)
    Two velocity flow trace measurements:
    •  Pulmonic-to-Systemic Flow Ratio (Qp/Qs) 
    						
    							Cardiology
    10-64 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Tricuspid Valve
    The following are cardiac Doppler Mode Tricuspid Valve 
    measurements:
    One peak velocity
    measurements•  Tricuspid Regurgitant Peak Pressure Gradient (TR Vmax)
    •  Tricuspid Valve Peak Pressure Gradient (TV Vmax)
    •  Tricuspid Regurgitant Peak Velocity (TR Vmax)
    •  Tricuspid Valve Peak Velocity (TV Vmax)
    •  Tricuspid Valve Velocity Peak A (TV A Velocity)
    •  Tricuspid Valve Velocity Peak E (TV E Velocity)
    Making a peak
    velocity
    measurement1.  Select the measurement; an active caliper with a vertical 
    dotted line displays.
    2.  To position the caliper at the desired measurement point, 
    move the Trackball.
    3.  To complete the measurement, press Set. 
    The system displays the velocity measurement in the 
    Results Window. 
    						
    							Cardiology Measurements
    LOGIQ 5 Basic User Manual 10-65
    Direction 2300002-100 Rev. 2
    Tricuspid Valve (continued)
    One velocity flow
    trace
    measurements•  Tricuspid Regurgitant Mean Pressure Gradient (TR Trace)
    •  Tricuspid Regurgitant Mean Velocity (TR Trace)
    •  Tricuspid Regurgitant Mean Square Root Velocity (TR 
    Trace)
    •  Tricuspid Regurgitant Velocity Time Integral (TR Trace)
    •  Tricuspid Valve Mean Pressure Gradient (TV Trace)
    •  Tricuspid Valve Mean Velocity (TV Trace)
    •  Tricuspid Valve Mean Square Root Velocity (TV Trace)
    •  Tricuspid Valve Velocity Time Integral (TV Trace)
    Making a velocity
    flow trace
    measurement1.  Select the measurement; an active caliper displays.
    2.  To position the caliper at the start point, move the Trackball.
    3.  To fix the trace start point, press Set.
    The system displays a vertical dotted line.
    4.  To trace the envelope, move the Trackball.
    A line shows the traced area.
    5.  To complete the trace, press Set.
    The system displays a second vertical dotted line.
    6.  To position the second dotted line at the start of the next 
    envelope, move the Trackball.
    7.  To complete the measurement, press Set.
    The system displays the measurement in the Results 
    Window. 
    						
    							Cardiology
    10-66 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Tricuspid Valve (continued)
    One time interval
    measurements•  Tricuspid Valve Time to Peak (TV TTP)
    •  Tricuspid Valve Closure to Opening (TCO)
    •  Tricuspid Valve A-Wave Duration (TV A Dur)
    •  QRS complex to end of envelope (Q-to-TV close)
    Making a time
    interval
    measurement1.  Select the measurement.
    The system displays an active caliper and a vertical dotted 
    line. 
    2.  To position the caliper at the start point, move the Trackball.
    3.  To fix the first caliper, press Set.
    The system displays a second active caliper.
    4.  To position the caliper at the end point, move the Trackball.
    5.  To complete the measurement, press Set.
    The system displays the time interval in the Results 
    Window.
    One slope
    measurements•  Tricuspid Valve Pressure Half Time (TV PHT)
    Making a slope
    measurement1.  Select the measurement; an active caliper displays.
    2.  To position the caliper at the start point, move the Trackball.
    3.  To fix the start point, press Set.
    The system fixes the first caliper and displays a second 
    active caliper.
    4.  To position the second caliper at the end point, move the 
    Trackball.
    A dotted line shows the slope.
    5.  To complete the measurement, press Set. 
    The system displays the slope measurement in the Results 
    Window. 
    						
    							Cardiology Measurements
    LOGIQ 5 Basic User Manual 10-67
    Direction 2300002-100 Rev. 2
    Tricuspid Valve (continued)
    One velocity flow
    trace and one area
    measurement•  Stroke Volume by Tricuspid Flow (TV Trace)
    Two peak velocity
    measurement•  Tricuspid Valve E-Peak to A-Peak Ratio (TV E/A Velocity) 
    						
    							Cardiology
    10-68 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Color Flow Mode
    Aortic Valve
    The following are cardiac Color Flow Mode Aortic Valve 
    measurements:
    One distance
    measurements•  Proximal Isovelocity Surface Area: Regurgitant Orifice Area 
    (PISA AR)
    •  Proximal Isovelocity Surface Area: Radius of Aliased Point 
    (PISA AR)
    Making a one
    distance
    measurement1.  Select the measurement; an active caliper displays.
    2.  To position the active caliper at the start point, move the 
    Trackball.
    3.  To fix the start point, press Set.
    The system fixes the first caliper and displays a second 
    active caliper.
    4.  To position the second active caliper at the end point, move 
    the Trackball.
    A dotted line connects the measurement points.
    5.  To complete the measurement, press Set.
    The system displays the measurement in the Results 
    Window. 
    						
    							Cardiology Measurements
    LOGIQ 5 Basic User Manual 10-69
    Direction 2300002-100 Rev. 2
    Aortic Valve (continued)
    One velocity flow
    trace
    measurements•  Proximal Isovelocity Surface Area: Regurgitant Flow (PISA 
    AR)
    •  Proximal Isovelocity Surface Area: Regurgitant Volume Flow 
    (PISA AR)
    Making a velocity
    flow trace
    measurement1.  Select the measurement; an active caliper displays.
    2.  To position the caliper at the start point, move the Trackball.
    3.  To fix the trace start point, press Set.
    The system displays a vertical dotted line.
    4.  To trace the envelope, move the Trackball.
    A line shows the traced area.
    5.  To complete the trace, press Set.
    The system displays a second vertical dotted line.
    6.  To position the second dotted line at the start of the next 
    envelope, move the Trackball.
    7.  To complete the measurement, press Set.
    The system displays the measurement in the Results 
    Window.
    One peak velocity
    measurement•  Proximal Isovelocity Surface Area: Aliased Velocity (PISA 
    AR)
    Making a peak
    velocity
    measurement1.  Select the measurement; an active caliper with a vertical 
    dotted line displays.
    2.  To position the caliper at the desired measurement point, 
    move the Trackball.
    3.  To complete the measurement, press Set. 
    The system displays the velocity measurement in the 
    Results Window. 
    						
    							Cardiology
    10-70 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Mitral Valve
    The following are cardiac Color Flow Mode Mitral Valve 
    measurements:
    One distance
    measurements•  Proximal Isovelocity Surface Area: Regurgitant Orifice Area 
    (PISA MR)
    •  Proximal Isovelocity Surface Area: Radius of Aliased Point 
    (PISA MR)
    Making a one
    distance
    measurement1.  Select the measurement; an active caliper displays.
    2.  To position the active caliper at the start point, move the 
    Trackball.
    3.  To fix the start point, press Set.
    The system fixes the first caliper and displays a second 
    active caliper.
    4.  To position the second active caliper at the end point, move 
    the Trackball.
    A dotted line connects the measurement points.
    5.  To complete the measurement, press Set.
    The system displays the measurement in the Results 
    Window.
    One velocity flow
    trace
    measurements•  Proximal Isovelocity Surface Area: Regurgitant Flow (PISA 
    MR) 
    						