Console Overview LOGIQ 5 Basic User Manual 3-11 Direction 2300002-100 Rev. 2 Peripheral/Accessory Connector Panel (continued) Figure 3-5. Peirpheral/Accessory Connector Panel 220-240V 500VA MaxIncluding front printer panel220-240V 500VA MaxIncluding rear panel 100~120V 500VA MaxIncluding front printer panel100~120V 500VA MaxIncluding rear panel Japan/USA/Asia Console (100~120V Europe/USA Console (220V) CAUTIONThe connection of equipment or transmission networks other than as specified in these instructions can result in electric shock hazard. Alternate connections will require verification of compatibility and conformity to IEC/EN 60601-1-1 by the installer.
Preparing the System for Use 3-12 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Peripheral/Accessory Connector Panel (continued) Table 3-2: Peripheral/Accessory Connector Panel Descriptions No.ItemType of connector Required Cable 1. S-Video Out S-Terminal, 4-pin 2. Composite Color/ BW 3. Audio In/Out RCA 4. S-Video In 5. Camera Color/ Camera B/W 6. VGA Shrinked D-Sub, 15-pin 7. Service USB(A) 8. VCR 1 USB(A) 9. VCR 2 USB(A) 10. For Future use 11 . Insite RJ-11 Modular, 6-pin 12. Ethernet RJ-45 Modular, 8-pin 13. Power IEC Couplers 243 1 1 6 115 10 15 1 2 3 4 6 1 8 1
Console Overview LOGIQ 5 Basic User Manual 3-13 Direction 2300002-100 Rev. 2 Footswitch (Option) Use only the recommended optional multifunctional footswitch. Figure 3-6. Footswitch The footswitch may be used as select keys, video recording, etc. The three switches have different function assignments depending on the current application. The footswitch connection is located below the operator control panel. CAUTIONDO NOT place or store the footswitch in the side pocket or the Peripheral Device storage area on the system. This is to prevent the possibility of the footswitch falling and damaging itself.
Preparing the System for Use 3-14 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 System Positioning/Transporting Moving the System When moving or transporting the system, follow the precautions below to ensure the maximum safety for personnel, the system, and other equipment. Before moving the system 1. Press the Power on/Standby switch and turn off the circuit breaker to power off. See ‘Power Off’ for more information. 2. Unplug the power cord. 3. All cables from off-board peripheral devices (IIE camera, external printer, VCR, etc.) must be disconnected from the console. 4. Disconnect the footswitch from the console. 5. Ensure that no loose items are left on the console. 6. Wind the power cable around the cable hook below the rear handle or wrap the cord in a bundle and store it behind the rear storage panel. CAUTIONPlease ensure that the peripheral devices, which are installed in the storage areas, are secured with the supplied straps.
System Positioning/Transporting LOGIQ 5 Basic User Manual 3-15 Direction 2300002-100 Rev. 2 Before moving the system (continued) To prevent damage to the Power Cord, DO NOT pull excessively on the cord or make sharp bends while wrapping. Figure 3-7. Location of rear handle a. Handle 7. Connect all probes to be used while off site. Ensure that probe cables are out of the way from the wheels and not protruding beyond the console. NOTE: If more than three (3) probes are intended to be used, store the additional probes securely. 8. Store all other probes in their original cases or in soft cloth or foam to prevent damage. 9. Store sufficient gel and other essential accessories in the provided space. 10. Adjust the monitor and control panel to its lowest position possible by using the release lever located underneath the control panel. 11. Unlock the wheels. a CAUTIONDO NOT place probes or the footswitch into either the side pocket or the peripheral device’s storage area when moving the system. These are not storage spaces for probes or the footswitch.
Preparing the System for Use 3-16 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 When moving the system 1. Always use the rear handle grips to move the system. 2. Take extra care when moving the system long distances and on inclines. Ask for help if necessary. Avoid ramps that are steeper than ten degrees to avoid tipping over the system. NOTE: Wheel chair ramps are usually less than five degrees. Utilize additional care and personnel when moving on steep incline (>5°) or loading into a vehicle for transport. 3. Use the foot brake (pedal), located on the bottom of the system in the front, when necessary. 4. Do not let the system strike walls or door frames. 5. Use extra care when crossing door or elevator thresholds. 6. Once the destination is reached, lock the wheels. CAUTIONDO NOT attempt to move the console using any cables or fixtures, such as the probe connectors. CAUTIONThe system weighs approximately 180 kg (397 lbs). To avoid possible injury and equipment damage: • Be sure the pathway is clear. • Limit movement to a slow careful walk. • Use two or more persons to move the system on inclines or long distances.
System Positioning/Transporting LOGIQ 5 Basic User Manual 3-17 Direction 2300002-100 Rev. 2 Transporting the System Use extra care when transporting the system using vehicles. In addition to the instructions used when moving the system (See ‘Moving the System’ for more information.), also perform the following: 1. Only use vehicles that are designed for transport of the LOGIQ 5 system. 2. Load and unload the system to a vehicle parked on a level surface. 3. Ensure that the transporting vehicle can handle the weight of the system plus the passengers. 4. Ensure that the load capacity of the lift (a minimum of 180 kg [397 lbs] is recommended) is capable of handling the weight of the system. 5. Ensure that the lift is in good working order. 6. Secure the system while it is on the lift so that it cannot roll. Use either wood chocks, restraining straps, or other similar types of constraints. Do not attempt to hold it in place by hand. NOTE: Strap the system below its handle so that the system does not break loose. 7. Employ two to three persons to load and unload safely from a vehicle. WARNINGNever ride on the lift with the system. A persons weight coupled with the weight of the system may exceed the load capacity of the lift.
Preparing the System for Use 3-18 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Transporting the System (continued) 8. Load the unit aboard the vehicle carefully and over its center of gravity. Keep the unit still and upright. NOTE: Do not lay the unit down. 9. Ensure that the system is firmly secured while inside the vehicle. Any movement, coupled with the weight of the system, could cause it to break loose. 10. Secure system with straps or as directed otherwise to prevent motion during transport. 11. Prevent vibration damage by driving cautiously. Avoid unpaved roads, excessive speeds, and erratic stops or starts.
System Positioning/Transporting LOGIQ 5 Basic User Manual 3-19 Direction 2300002-100 Rev. 2 Wheels Examine the wheels frequently for any obvious defects that could cause them to break or bind. Front wheels Swivel and brake lock. Back wheels Swivel, swivel lock and brake lock. Setting the lock The pedal is located in the front wheels of the unit and enables the user to control the movement of the wheels. Figure 3-8. Front Pedal 1. Unlock 2. Lock 1 2
Preparing the System for Use 3-20 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Powering the System Connecting and Using the System To connect the system to the electrical supply: 1. Ensure that the wall outlet is of the appropriate type. 2. Ensure that the power switch is turned off. 3. Unwrap the power cable. Make sure to allow sufficient slack in the cable so that the plug is not pulled out of the wall if the system is moved slightly. 4. Attach the power plug to the system and secure it in place by using the retaining clamp. Figure 3-9. Power Plug a. Retaining clamp for power plug 5. Push the power plug securely into the wall outlet. a CAUTIONEnsure that the retaining clamp for the power plug is fixed firmly. CAUTIONUse caution to ensure that the power cable does not disconnect during system use. If the system is accidentally unplugged, data may be lost.