    							OB Table Editor
    LOGIQ 5 Basic User Manual 9-85
    Direction 2300002-100 Rev. 2
    O B  Ta b l e  Te m p l a t e s
    Tool Type:
    • Distance: 2D Caliper
    •  Circumference: 2D Ellipse, 2D Trace, 2D Caliper
    Te m p l a t e  1
    Table 9-4: Template 1 (based on Hadlock)
    Template 1: SD Range Table
    Fetal Age Table Table Format MEAS MEAN SD
    Table Unit mm Week Week
    Table Range 1SD
    Graph Range 1SD
    Measurement Result Value [cm]
    GA [#w#d]
    Min [#w#d]
    Max [#w#d]
    Fetal Growth Table Table Format AGE MEAN SD
    Table Unit Week mm Week
    Others are same as above 
    						
    							OB/GYN
    9-86 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Te m p l a t e  2
    Te m p l a t e  3
    Table 9-5: Template 2 (based on Tokyo)
    Template 2: SD Range Table
    Fetal Age Table Table Format MEAS MEAN SD
    Table Unit mm Day Day
    Table Range 1SD
    Graph Range 1SD
    Measurement Result Value [cm]
    GA [#w#d]
    SD: day(+/-)
    Fetal Growth Table Table Format AGE MEAN SD
    Table Unit Day mm Day
    Others are same as above
    Table 9-6: Template 3 (based on Osaka)
    Template 3: SD Table
    Fetal Age Table Table Format MEAS MEAN SD
    Table Unit mm Day mm
    Table Range 1SD
    Graph Range 1SD
    Measurement Result Value [cm]
    GA [#w#d]
    SD: sd=(mv-pv)/sd
    Fetal Growth Table Table Format AGE MEAN SD
    Table Unit Day mm Day
    Others are same as above 
    						
    							OB Table Editor
    LOGIQ 5 Basic User Manual 9-87
    Direction 2300002-100 Rev. 2
    Te m p l a t e  4
    Table 9-7: Template 4 (based on several European tables)
    Template 4: 5%-95% Table
    Fetal Age Table Table Format MEAS MEAN SD
    Table Unit mm WeekDay mm
    Table Range 5%:95%
    Graph Range 5%:95%
    Measurement Result Value [cm]
    GA [#w#d]
    GP [%]
    GP is calculated by Fetal Growth Table. If you did 
    not edit Growth Table, GP is not calculated by the 
    system,
    Fetal Growth Table Table Format AGE MEAN SD
    Table Unit WeekDay mm Day
    Table Range 5%:95%
    Graph Range 5%:95% 
    						
    							OB/GYN
    9-88 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Te m p l a t e  5
    Table 9-8: Template 5 (based on several European tables)
    Template 5: 5% - 95% Table
    Fetal Age Table Table Format MEAS MEAN SD
    Table Unit mm WeekDay mm
    Table Range 1SD
    Graph Range 5%:95%
    Measurement Result Value [cm]
    GA [#w#d]
    GP [%]
    GP is calculated by Fetal Growth Table. If you did not 
    edit Growth Table, GP is not calculated by the system,
    Fetal Growth Table Table Format AGE MEAN MAX
    Table Unit WeekDay mm mm
    Table Range 1SD
    Graph Range 5%:95% 
    						
    							OB Table Editor
    LOGIQ 5 Basic User Manual 9-89
    Direction 2300002-100 Rev. 2
    Te m p l a t e  6
    Table 9-9: Template 6 (based on several European tables)
    Template 6: 10% - 90% Table
    Fetal Age Table Table Format MEAS MIN MEAN SD
    Table Unit mm WeekDay WeekDay WeekDay
    Table Range 10%-90%
    Graph Range 10%-90%
    Measurement Result Value [cm]
    GA [#w#d]
    GP [%]
    GP is calculated by Fetal Growth Table. If you did not edit Growth 
    Table, GP is not calculated by the system,
    Fetal Growth Table Table Format AGE MIN MEAN SD
    Table Unit WeekDay mm mm mm
    Table Range 10%-90%
    Graph Range 10%-90% 
    						
    							OB/GYN
    9-90 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Te m p l a t e  7
    Table 9-10: Template 7 (Based on several european tables)
    Template 7: 10% - 90% Table
    Fetal Age Table Table Format MEAS MEAN SD
    Table Unit mm WeekDay mm
    Table Range 1SD
    Graph Range 10%:90%
    Measurement Result Value [cm]
    GA [#w#d]
    GP [%]
    GP is calculated by Fetal Growth Table. If you did not 
    edit Growth Table, GP is not calculated by the system,
    Fetal Growth Table Table Format AGE MEAN MAX
    Table Unit WeekDay mm mm
    Table Range 1SD
    Graph Range 10%:90% 
    						
    							OB Table Editor
    LOGIQ 5 Basic User Manual 9-91
    Direction 2300002-100 Rev. 2
    OB Table Edit Menu
    The data you enter in the OB Table Edit Menu depends on 
    whether the table type is Fetal Age or Fetal Growth.
    Fetal Age Table
    If you are creating or editing a Fetal Age table, the OB Table Edit 
    Menu is as follows:
     Figure 9-28. OB Table Edit Menu: Fetal Age Table 
    						
    							OB/GYN
    9-92 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Fetal Age Table (continued)
    Complete the field1.  Input value to Min, Max and Interval of the Parameters field. 
    The system automatically fills in the MEAS column.
    Input value to the columns of MEAN and SD.
    NOTE:  To move between the fields in the table, use the up, down, 
    left, and right arrow keys.
    NOTE:  You must enter a minimum of two rows of data. Any lines 
    with a blank cell are not saved. 
    To save the Table Data, move the Trackball to Exit to Save 
    and press Set. If you want cancel this table, move the 
    Trackball to Cancel and press Set. 
    						
    							OB Table Editor
    LOGIQ 5 Basic User Manual 9-93
    Direction 2300002-100 Rev. 2
    Fetal Growth Table
    If you are creating or editing a Fetal Growth table, the OB Table 
    Edit Menu is as follows:
     Figure 9-29. OB Table Edit Menu: Fetal Growth Table 
    Complete the field1.  Input value to the required columns.
    NOTE:  To move between the fields in the table, use the up, down, 
    left, and right arrow keys.
    NOTE:  You must enter a minimum of two rows of data. Any lines 
    with a blank cell are not saved. 
    2.  To save the Table Data, move the Trackball to Exit to Save 
    and press Set. If you want cancel this table, move the 
    Trackball to Cancel and press Set.
    After you complete the OB table, it is now available for the 
    selected study. To use the measurement, you must assign it to a 
    Touch Panel.   
    						
    							OB/GYN
    9-94 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    GYN Measurements
    Introduction
    The Gynecology exam category includes the following three 
    studies:
    •  Generic. This study is common to all exam categories. See 
    ‘Generic Measurements’ for more information. 
    •  General Gynecology. This study includes uterine, ovarian, 
    ovarian follicle, and endometrium measurements.
    •  OB/GYN Vessel. This study includes the following vessels: 
    uterine, ovarian, umbilical, middle cerebral artery, aorta, 
    placenta, and descending aorta.
    NOTE:  The calculation formulas are listed in the Advanced Reference 