    							Cardiology Measurements
    LOGIQ 5 Basic User Manual 10-81
    Direction 2300002-100 Rev. 2
    Generic Study
    B-Mode
    The Cardiology M-Mode Generic exam category includes the 
    following measurements: 
    • Caliper
    • Point
    • Area (Trace)
    • Volume
    • Volume (d)
    • Volume (s)
    •  Dist (Distance) Ratio
    • Area Ratio
    • Ellipse
    • Open Trace
    • R-R
    •  VF Diam (Diameter)
     Figure 10-4. Cardiology B-Mode - Generic Touch Panel 
    						
    							Cardiology
    10-82 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    M-Mode
    The Cardiology M-Mode Generic exam category includes the 
    following measurements: 
    • LV Study
    • LA/Ao
    • RV Study
    • D-E Excursion
    • Depth
    • Slope Caliper
    • Caliper
    • Area Ratio
    • Ellipse
    • Time
    • Heart Rate
     Figure 10-5. Cardiology M-Mode - Generic Touch Panel 
    						
    							Cardiology Measurements
    LOGIQ 5 Basic User Manual 10-83
    Direction 2300002-100 Rev. 2
    Doppler Mode
    The Cardiology Doppler Mode Generic exam category includes 
    the following measurements: 
    • Point
    • Caliper
    • Manual Trace
    •  SD Auto Calcs
    • VSD Vmax
    •  MV E/A Ratio
    • PHT
    • Time
    • Heart Rate
     Figure 10-6. Cardiology Doppler Mode - Generic Touch Panel 
    						
    							Cardiology
    10-84 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    ECG Option
    Overview
    A physiological input panel is available for the LOGIQ 5.  This 
    panel has inputs for ECG, PCG and auxiliary signals.
    The physiological module consists of four channels: ECG, PCG, 
    AUX1 and AUX2 (AUX 2 is currently not available). The two 
    channels, AUX1 and AUX2, are both capable of handling 
    external ECG signals from other diagnostic ECG devices. AUX2 
    is also capable of handling a pressure/pulse signal.
    The scanned image that is displayed is synchronized with the 
    ECG and PCG traces. In Doppler or M-Mode, the traces are 
    synchronized to that particular mode’s sweep.
    Approved accessory cables provide the proper signals to the 
    Physiological Panel.
     Figure 10-7. Optional Physiological Input Panel
    WARNING•  Do not use with the Defibrillator.
    This equipment does not have a defibrillator approved 
    applied part.
    •  Do not use the LOGIQ 5 Ultrasound system Physiological 
    traces for diagnosis and monitoring.
    •  Only approved and recommended peripherals and 
    accessories should be used. 
    						
    							ECG Option
    LOGIQ 5 Basic User Manual 10-85
    Direction 2300002-100 Rev. 2
    ECG Cable
    The ECG Cable is a modular cable consisting of two different 
    cable parts:
    •  Single cable with a system connection at one end and a 
    cable splitter at the other.
    •  A triple color-coded electrode cable to be inserted into the 
    splitter device. Each electrode cable hooks up to the 
    appropriate stick-on electrode by a color-coded clip type 
    connector.
    The cable splitter device has a drawing defining the color codes, 
    names and body location. 
     Figure 10-8. ECG Cable
    R: Red
    L: Yellow
    F: Green
    (1)
    (2)
    (3) 
    						
    							Cardiology
    10-86 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Physiological Trace Monitor Display
    The scanned image is synchronized with the ECG and PCG 
    trace. In Doppler or M-Mode, the traces are synchronized with 
    that particular mode’s sweep.
    The user can control the gain, position and sweep speed of the 
    traces using the Touch Panel controls.
     Figure 10-9. Physiological Trace Monitor Display
    1. PCG
    2. AUX
    3. ECG
    4.  Auto Heart Rate Display 
    						
    							ECG Option
    LOGIQ 5 Basic User Manual 10-87
    Direction 2300002-100 Rev. 2
    ECG Touch Panel
    The ECG Touch Panel Menu provides for control of the 
    physiological input signals.
    Without the ECG option, the ECG Touch Panel menu is not 
    displayed.
     Figure 10-10. ECG Touch Panel Menu
    Sweep Speed
    Description Change the speed of the trace. The sweep speed of the physio 
    signal on the B-Mode image can be set independent of the 
    timeline (Doppler and M-Mode) sweep speed.
    Va l u e 1  -  1 6 . 
    						
    							Cardiology
    10-88 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Timer Trigger
    Description Enables intermittent imaging based on a timer.
    NOTE:  If Timer Trigger is turned on, the ECG Trigger is set to None.
    Value On or Off.
    ECG Trigger
    Description Enables intermittent imaging based on the ECG. The trigger 
    location(s) relative to the R trigger are set with the Delay Time 
    key.
    NOTE:  If other than None is selected, Timer Trigger is turned off.
    Value None, Trig1, Trig2, and Both.
    AUX Display
    Description Provides the ability to turn the AUX trace on for display on the 
    monitor.
    Adjusting When the key is selected, the AUX trace toggles between on 
    and off.
    Value On or Off
    PCG Display
    Description Provides the ability to turn the PCG trace on for display on the 
    monitor.
    Adjusting When the key is selected, the PCG trace toggles between on 
    and off.
    Value On or Off 
    						
    							ECG Option
    LOGIQ 5 Basic User Manual 10-89
    Direction 2300002-100 Rev. 2
    ECG Display
    Description Provides the ability to turn on the ECG trace and Auto Heart 
    Rate for display on the monitor.
    Adjusting When the key is selected, the ECG trace and Auto Heart Rate 
    toggle between on and off.
    Value On or Off
    ECG Trigger Period
    Description The control specifies the number of heart cycles (R-waves) that 
    are skipped between ECG triggers. 
    The default is 1 or no skipping.
    Adjusting Adjust the corresponding control.
    Va l u e 1  -  3 0
    Delay Time
    Description In ECG Trigger Mode: If only ECG Trig1 or ECG Trig2 is 
    selected via the ECG Trigger key, the Delay Time key controls 
    the R-Delay time of the active trigger. If both triggers are 
    selected, the primary control is ECG Trig1, but adjusting this 
    control switches to ECG Trig2. Once the trigger is set, the snap 
    shot image is displayed each time the update line passes the 
    active trigger(s).
    In Timer Trigger Mode: Rotating the knob changes the delay 
    time between images.
    Adjusting Adjust the corresponding control.
    Value 0.10 - 10.00 
    						
    							Cardiology
    10-90 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    AUX Gain/Position
    Description Allows for the amplitude control of the AUX trace or allows for 
    the vertical positioning of the AUX trace on the image display.
    Adjusting Press the knob to toggle between Gain and Position. The default 
    is Gain.
    Value Gain: 0.01 - 4.00
    Position: 0.00 - 100.00
    PCG Gain/Position
    Description Allows for the amplitude control of the PCG trace or allows for 
    the vertical positioning of the PCG trace on the image display.
    Adjusting Press the knob to toggle between Gain and Position. The default 
    is Gain.
    Value Gain: 0.01 - 2.00
    Position: 0.00 - 100.00
    ECG Gain/Position
    Description Allows for the amplitude control of the ECG trace or allows for 
    the vertical positioning of the ECG trace on the image display.
    Adjusting Press the knob to toggle between Gain and Position. The default 
    is Gain.
    Value Gain: 0.00 - 0.23
    Position: 0.00 - 100.00 
    						