GE Logiq P5 User Manual
Have a look at the manual GE Logiq P5 User Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
LOGIQ 5 Basic User Manual 3-1 Direction 2300002-100 Rev. 2 Chapter 3 Preparing the System for Use Describes the site requirements, console overview, system positioning/transporting, powering on the system, adjusting the display monitor, probes and operator controls.
Preparing the System for Use 3-2 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Site Requirements Introduction NOTE: Only qualified physicians or sonographers should perform ultrasound scanning on human subjects for medical diagnostic reasons. Request training, if needed. Do not attempt to install the system alone. General Electric, Affiliate, or Distributor Field Engineers and Application Specialists will install and setup the system. See ‘Contact Information’ for more information. The LOGIQ 5 does not contain any operator serviceable internal components. Ensure that unauthorized personnel do not tamper with the unit. Perform regular preventive maintenance. See ‘System Care and Maintenance’ for more information. Maintain a clean environment. Turn off the system circuit breaker before cleaning the unit. See ‘Cleaning the system’ for more information. Never set liquids on the unit to ensure that liquid does not drip into the control panel or unit.
Site Requirements LOGIQ 5 Basic User Manual 3-3 Direction 2300002-100 Rev. 2 Before the system arrives NOTICE This medical equipment is approved, in terms of the prevention of radio wave interference, to be used in hospitals, clinics and other institutions which are environmentally qualified. The use of this equipment in an inappropriate environment may cause some electronic interference to radios and televisions around the equipment. This equipment can be used in residential areas only under the supervision of physicians or qualified technicians. Ensure that the following is provided for the new system: • A separate power outlet with a 5 amp circuit breaker for 120 VAC for 120V area, 8 amp circuit breaker for 220-240VAC for 220/240V area or 15 amp circuit breaker for 100 VAC for Japan. • Take precautions to ensure that the console is protected from electromagnetic interference. Precautions include: • Operate the console at least 15 feet away from motors, typewriters, elevators, and other sources of strong electromagnetic radiation. • Operation in an enclosed area (wood, plaster or concrete walls, floors and ceilings) helps prevent electromagnetic interference. • Special shielding may be required if the console is to be operated in the vicinity of radio broadcast equipment.
Preparing the System for Use 3-4 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Environmental Requirements The system should be operated, stored, or transported within the parameters outlined below. Either its operational environment must be constantly maintained or the unit must be turned off. Table 3-1: System Environmental Requirements OperationalStorageTransport (
Console Overview LOGIQ 5 Basic User Manual 3-5 Direction 2300002-100 Rev. 2 Console Overview Console graphics The following are illustrations of the console: Figure 3-1. LOGIQ 5 System (right and left side views) 1. Probe and Gel Bottle Holder 2. Storage area of Peripheral Device 3. Storage area of Peripheral Device 4. Release lever to raise and lower Monitor and Control Panel 5. Side Pocket 6. Task Light 100-120V 150VA Max 6 23 4 5 31 CAUTIONDO NOT place probes or the footswitch into either the side pocket or the peripheral device’s storage area.
Preparing the System for Use 3-6 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Console graphics (continued) Figure 3-2. LOGIQ 5 System (front and back views) 1. Optional B/W Video Page Printer 2. Probe Ports with optional CWD Probe Port, when available 3. Footswitch connector 4. Optional Physiological Input Panel 5. MOD 6. CD ROM 7. Rear Panel 8. Task Lamp 8 1 6 5 42 3 7
Console Overview LOGIQ 5 Basic User Manual 3-7 Direction 2300002-100 Rev. 2 External drives (CD-RW and/or MO) The external CD-RW drive (Standard) and MO drive (Option) are located next to the probe port. They can be used to perform software upgrades, image archiving and service diagnostics. CAUTIONMake sure that the CD-RW Disk tray and/or MO Disk are securely placed in each device during system operation. Mechanical damage may occur if other objects hit the tray or MO disk.
Preparing the System for Use 3-8 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Storage areas Several convenient storage areas are provided within the console as shown in the figure below. Use them to store gel, options, probe cables, accessories, etc. Figure 3-3. Storage Areas 1. Storage Areas 100-120V 150VA Max PDIPW L R B Pause Auto DepthMZoom Ellipse Body PatternMeasure M/D Cursor Scan AreaSetClear Comment Patient 21L21 1 1 1 1 1 1 1 CAUTIONDO NOT place probes or the footswitch into either the side pocket or the peripheral device’s storage area.
Console Overview LOGIQ 5 Basic User Manual 3-9 Direction 2300002-100 Rev. 2 Task Lamp The task lamp is used to illuminate the keyboard. Figure 3-4. Task Lamp 1. Task lamp CAUTIONThe task lamp may overheat. Take proper precautions to avoid touching the task lamp. Contact your local Service Representative when the task lamp bulb needs to be replaced. 1.
Preparing the System for Use 3-10 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Peripheral/Accessory Connection Peripheral/Accessory Connector Panel LOGIQ 5 peripherals and accessories can be properly connected using the rear connector panel located behind the rear door. Only the B/W Printer can be connected to the front accessory panel. Located on the rear panel are video input and output, audio input and output, camera rec/expose, Insite, Ethernet, power and control connections for VCR, printer and service tools. See Figure 3-5 for more specifics. CAUTIONEach outer (case) ground line of peripheral/accessory connectors are Earth Grounded. Signal ground lines are Not Isolated. CAUTIONUse only approved probes, peripherals or accessories. Be sure to plug the power supply cable of the peripherals to the connector panel of the system to avoid unexpected electrical shock. DO NOT connect any probes or accessories without approval by GE.