GE Logiq P5 User Manual
Have a look at the manual GE Logiq P5 User Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
Measurement and Calculation Setup LOGIQ 5 Basic User Manual 7-41 Direction 2300002-100 Rev. 2 Changing or adding measurement parameters (continued) Adding measurement parametersTo add a measurement parameter: 1. In the Selection menu, select the measurement. 2. To change the tool used to make a measurement: In the Measurement section of the Measurement & Analysis screen, select the desired tool from the Tool list. Select the arrow to display the drop-down list. NOTE: If the Tool field is gray, it cannot be changed. 3. If necessary, check Fetus (OB only), Location (Loc), or Side: • Fetus: If this is an OB measurement, check this box. (Default ON). • Location: If this measurement includes a Prox, Mid, or Dist location, check this box. • Side: If this measurement includes a Left or Right side, check this box.
General Measurements and Calculations 7-42 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Changing or adding measurement parameters (continued) Adding measurement parameters (continued)4. In the Measurement section, move the Trackball to an empty line at the bottom of the Parameter list. Press Set. The system adds a parameter with a name of (Name). Figure 7-25. Adding a Parameter
Measurement and Calculation Setup LOGIQ 5 Basic User Manual 7-43 Direction 2300002-100 Rev. 2 Changing or adding measurement parameters (continued) Adding measurement parameters (continued)5. To change the name of the Parameter, move the Trackball to the (Name) and press Set twice. Type a name for the parameter. 6. Move the Trackball to the Tool result field, and press Set once. The Edit Formula window is displayed. Figure 7-26. Edit Formula 7. To create a formula: a. In the Value Type field, select a value. b. Do one of the following: • Type a formula in the Formula field. • Select formula components from the Operators, Parameters, and Functions drop-down lists. When you select a component, the system displays it in the Formula field. 8. To test the formula, select Check. If there are no problems, the system displays “Syntax OK!”. If there are any problems with the formula, the system displays an Error message in place of the Formula field label. 9. When the formula is correct, select OK to save it. The Edit Formula window closes. The formula is displayed in the Tool result field.
General Measurements and Calculations 7-44 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Changing or adding measurement parameters (continued) Formula Unit ConversionWhen you create a formula, the system changes the calculation result into an output unit as defined in the following table. Table 7-2: Formula Unit Conversion Unit Conversion (coefficient value) Time sx1 ms x1,000 min x0.0167 h x0.00027778 Ratio % x100 Frequency bpm or BPM x1.0 Angle rad x1.0 deg x57.2958 grad x63.6620 Distance cm x100 mx1 dm x10 mm x1,000 inch x39.37 feet x39.37 pixels x1 Ve lo c i ty
Measurement and Calculation Setup LOGIQ 5 Basic User Manual 7-45 Direction 2300002-100 Rev. 2 m/s x1 dm/s x10 cm/s x100 mm/s x1,000 inch/s x39.37 Acceleration m/s2 x1 dm/s2 x10 cm/s2 x100 mm/s2 x1,000 inch/s2 x39.37 Area m2 or m^2 x1 dm2 x100 cm2 or cm^2 x10,000 mm2 or mm^2 x1,000,000 inch2 x1550 Vo lu m e m3 x1 dm3 x1,000 cm3 x1,000,000 l x1,000 dl x10,000 cl x100,000 ml x1,000,000 gallon x264,178 quart x1056.71 Vo lu m e Fl o w Table 7-2: Formula Unit Conversion Unit Conversion (coefficient value)
General Measurements and Calculations 7-46 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 m3/s x1 dm3/s x1,000 cm3/s x1,000,000 l/s x1,000 dl/s x10,000 cl/s x100,000 ml/s x1,000,000 m3/min x60 dm3/min x60,000 cm3/min x60,000,000 l/min or L/min x60,000 dl/min x600,000 cl/min 6,000,000 ml/min x60,000,000 ml/m2 x1,000,000 Pressure mmHg x1 Pa x133.322 kPa x0.133322 bar x0.00133322 Pressure/Time mmHg/s x1 Mass kg x1 g x1,000 ounce x35.273962 pound x2.2046226 Others mmHG x100 Table 7-2: Formula Unit Conversion Unit Conversion (coefficient value)
Measurement and Calculation Setup LOGIQ 5 Basic User Manual 7-47 Direction 2300002-100 Rev. 2 Changing or adding measurement parameters (continued) Formula Unit Conversion (continued)For example, when a Volume formula is created: Vol [ml or cm3] = 0.523598*{D1}*{D2}*{D3} (D1, D2, and D3 indicate a measurement result.) In this case, the measurement (D1, D2, and D3) is a distance measurement, so the measured data is a meter [m] unit according to the above table. To change into a milliliter, the system multiplies each measurement value by 100. As a result, it multiplies a formula by 1,000,000. The standard unit of volume is a cube meter, so the system multiplies the result by 1,000,000. The system multiplies the calculation result by the coefficient and converts it. To get a correct result, when you define the formula, you must convert the coefficient itself, such as the coefficient of 10^. For example, if you want to define the following formula: efg[g] = 10^(1.5662- 0.0108*{P1}+0.0468*{P2}+0.171*{D1}+0.00034*{P1}*{P1}- 0.003685*{P2}*{D1}) D1[cm]: Distance P1[cm]: Perimeter P2[cm]: Perimeter
General Measurements and Calculations 7-48 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Changing or adding measurement parameters (continued) Formula Unit Conversion (continued)The system defines the standard value of each measurement as a meter [m]. If the unit of each measurement value of this formula is defined as centimeter [cm], you must define the formula as follows: efw[g] = 10^(1.5662-0.0108*{P1}*100 + 0.0468*{P2}*100 + 0.171*{D1}*100 + 0.00034*{P1}*{P1}*100*100 - 0.003685*{P2}*{D1}*100*100) (This converts each measurement value to a centimeter [cm], since the system standard unit is a meter [m].) The output unit of this formula is a gram. Since the standard unit of the system is defined as a kilogram [kg], the system multiplies the output by 1,000. Because the output of this formula is defined as a gram, it is necessary to define the formula as follows. efw[g] = 10^(1.5662-0.0108*{P1}*100 + 0.0468*{P2}*100 + 0.171*{D1}*100 + 0.00034*{P1}*{P1}*100*100 - 0.003685*{P2}*{D1}*100*100)/1,000 As shown, you can obtain an exact calculation result.
Measurement and Calculation Setup LOGIQ 5 Basic User Manual 7-49 Direction 2300002-100 Rev. 2 Editing Calculations To modify user-defined calculations: In the Measurement menu section, selection Edit Calc. Figure 7-27. Edit Calc
General Measurements and Calculations 7-50 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Editing Calculations (continued) The Modify User CALC window is displayed. Figure 7-28. Modify User CALC window In the User Defined list, select the calculation that you want modified, then select OK.