    							Measurement and Calculation Setup
    LOGIQ 5 Basic User Manual 7-41
    Direction 2300002-100 Rev. 2
    Changing or adding measurement parameters (continued)
    Adding
    measurement
    parametersTo add a measurement parameter:
    1.  In the Selection menu, select the measurement.
    2.  To change the tool used to make a measurement:
    In the Measurement section of the Measurement & Analysis 
    screen, select the desired tool from the Tool list. Select the 
    arrow to display the drop-down list.
    NOTE:  If the Tool field is gray, it cannot be changed.
    3.  If necessary, check Fetus (OB only), Location (Loc), or Side:
    •  Fetus: If this is an OB measurement, check this box. 
    (Default ON).
    •  Location: If this measurement includes a Prox, Mid, or 
    Dist location, check this box.
    •  Side: If this measurement includes a Left or Right side, 
    check this box. 
    						
    							General Measurements and Calculations
    7-42 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Changing or adding measurement parameters (continued)
    Adding
    measurement
    parameters
    (continued)4.  In the Measurement section, move the Trackball to an 
    empty line at the bottom of the Parameter list. Press Set.
    The system adds a parameter with a name of (Name). 
     Figure 7-25. Adding a Parameter 
    						
    							Measurement and Calculation Setup
    LOGIQ 5 Basic User Manual 7-43
    Direction 2300002-100 Rev. 2
    Changing or adding measurement parameters (continued)
    Adding
    measurement
    parameters
    (continued)5.  To change the name of the Parameter, move the Trackball 
    to the (Name) and press Set twice. Type a name for the 
    parameter.
    6. Move the Trackball to the Tool result field, and press Set 
    once.
    The Edit Formula window is displayed. 
     Figure 7-26. Edit Formula
    7.  To create a formula:
    a.  In the Value Type field, select a value.
    b.  Do one of the following:
    •  Type a formula in the Formula field.
    •  Select formula components from the Operators, 
    Parameters, and Functions drop-down lists. When you 
    select a component, the system displays it in the 
    Formula field.
    8.  To test the formula, select Check.
    If there are no problems, the system displays “Syntax OK!”.
    If there are any problems with the formula, the system 
    displays an Error message in place of the Formula field 
    label.
    9.  When the formula is correct, select OK to save it.
    The Edit Formula window closes. The formula is displayed 
    in the Tool result field. 
    						
    							General Measurements and Calculations
    7-44 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Changing or adding measurement parameters (continued)
    Formula Unit
    ConversionWhen you create a formula, the system changes the calculation 
    result into an output unit as defined in the following table.
    Table 7-2: Formula Unit Conversion
    Unit
    Conversion 
    (coefficient value)
    Time
    sx1
    ms x1,000
    min x0.0167
    h x0.00027778
    Ratio
    % x100
    Frequency
    bpm or BPM x1.0
    Angle
    rad x1.0
    deg x57.2958
    grad x63.6620
    Distance
    cm x100
    mx1
    dm x10
    mm x1,000
    inch x39.37
    feet x39.37
    pixels x1
    Ve lo c i ty 
    						
    							Measurement and Calculation Setup
    LOGIQ 5 Basic User Manual 7-45
    Direction 2300002-100 Rev. 2
    m/s x1
    dm/s x10
    cm/s x100
    mm/s x1,000
    inch/s x39.37
    Acceleration
    m/s2 x1
    dm/s2 x10
    cm/s2 x100
    mm/s2 x1,000
    inch/s2 x39.37
    Area
    m2 or m^2 x1
    dm2 x100
    cm2 or cm^2 x10,000
    mm2 or mm^2 x1,000,000
    inch2 x1550
    Vo lu m e
    m3 x1
    dm3 x1,000
    cm3 x1,000,000
    l x1,000
    dl x10,000
    cl x100,000
    ml x1,000,000
    gallon x264,178
    quart x1056.71
    Vo lu m e  Fl o w
    Table 7-2: Formula Unit Conversion
    Unit
    Conversion 
    (coefficient value) 
    						
    							General Measurements and Calculations
    7-46 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    m3/s x1
    dm3/s x1,000
    cm3/s x1,000,000
    l/s x1,000
    dl/s x10,000
    cl/s x100,000
    ml/s x1,000,000
    m3/min x60
    dm3/min x60,000
    cm3/min x60,000,000
    l/min or L/min x60,000
    dl/min x600,000
    cl/min 6,000,000
    ml/min x60,000,000
    ml/m2 x1,000,000
    Pressure
    mmHg x1
    Pa x133.322
    kPa x0.133322
    bar x0.00133322
    Pressure/Time
    mmHg/s x1
    Mass
    kg x1
    g x1,000
    ounce x35.273962
    pound x2.2046226
    Others
    mmHG x100
    Table 7-2: Formula Unit Conversion
    Unit
    Conversion 
    (coefficient value) 
    						
    							Measurement and Calculation Setup
    LOGIQ 5 Basic User Manual 7-47
    Direction 2300002-100 Rev. 2
    Changing or adding measurement parameters (continued)
    Formula Unit
    Conversion
    (continued)For example, when a Volume formula is created:
    Vol [ml or cm3] = 0.523598*{D1}*{D2}*{D3}
    (D1, D2, and D3 indicate a measurement result.)
    In this case, the measurement (D1, D2, and D3) is a distance 
    measurement, so the measured data is a meter [m] unit 
    according to the above table. 
    To change into a milliliter, the system multiplies each 
    measurement value by 100. As a result, it multiplies a formula 
    by 1,000,000.
    The standard unit of volume is a cube meter, so the system 
    multiplies the result by 1,000,000.
    The system multiplies the calculation result by the coefficient 
    and converts it. To get a correct result, when you define the 
    formula, you must convert the coefficient itself, such as the 
    coefficient of 10^.
    For example, if you want to define the following formula: 
    efg[g] = 10^(1.5662-
    0.0108*{P1}+0.0468*{P2}+0.171*{D1}+0.00034*{P1}*{P1}-
    0.003685*{P2}*{D1})
    D1[cm]: Distance
    P1[cm]: Perimeter
    P2[cm]: Perimeter 
    						
    							General Measurements and Calculations
    7-48 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Changing or adding measurement parameters (continued)
    Formula Unit
    Conversion
    (continued)The system defines the standard value of each measurement as 
    a meter [m]. If the unit of each measurement value of this 
    formula is defined as centimeter [cm], you must define the 
    formula as follows:
    efw[g] = 10^(1.5662-0.0108*{P1}*100 + 0.0468*{P2}*100 + 
    0.171*{D1}*100 + 0.00034*{P1}*{P1}*100*100 - 
    0.003685*{P2}*{D1}*100*100)
    (This converts each measurement value to a centimeter [cm], 
    since the system standard unit is a meter [m].)
    The output unit of this formula is a gram. Since the standard unit 
    of the system is defined as a kilogram [kg], the system multiplies 
    the output by 1,000. 
    Because the output of this formula is defined as a gram, it is 
    necessary to define the formula as follows.
    efw[g] = 10^(1.5662-0.0108*{P1}*100 + 0.0468*{P2}*100 + 
    0.171*{D1}*100 + 0.00034*{P1}*{P1}*100*100 - 
    0.003685*{P2}*{D1}*100*100)/1,000
    As shown, you can obtain an exact calculation result. 
    						
    							Measurement and Calculation Setup
    LOGIQ 5 Basic User Manual 7-49
    Direction 2300002-100 Rev. 2
    Editing Calculations
    To modify user-defined calculations:
    In the Measurement menu section, selection Edit Calc.
     Figure 7-27. Edit Calc 
    						
    							General Measurements and Calculations
    7-50 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Editing Calculations (continued)
    The Modify User CALC window is displayed.
     Figure 7-28. Modify User CALC window
    In the User Defined list, select the calculation that you want 
    modified, then select OK. 
    						