Home > GE > Ultrasound System > GE Logiq P5 User Manual

GE Logiq P5 User Manual

    Download as PDF Print this page Share this page

    Have a look at the manual GE Logiq P5 User Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    Page
    of 878
    							Quality Assurance
    LOGIQ 5 Basic User Manual 18-25
    Direction 2300002-100 Rev. 2
    Typical Tests to Perform
    Quality assurance measurements provide results relating to 
    system performance. Typically these are:
    •  Axial Measurement Accuracy
    • Lateral Measurement Accuracy
    •  Axial and Lateral Resolution
    • Penetration
    •  Functional & Contrast Resolution
    •  Gray Scale Photography.
    With these tests, a performance baseline can be set at 
    installation with the phantom in your department. Future test 
    results can be compared to the baseline in order to maintain a 
    record of system performance trends.
    Frequency of tests
    Quality assurance tests are used to determine whether a 
    scanner is providing the same level of performance from day to 
    day.
    The frequency of testing varies with the amount of system usage 
    and modes to be tested. It is recommended that the user 
    perform quality assurance tests at least every three months or 
    every 400 patient studies. Tests should also be performed when 
    a question about system performance exists.
    A mobile system may require more frequent tests.
    Image quality should also be tested immediately after the 
    following events:
    • Service calls
    • System upgrades/modifications
    •  Dropped probe, power surge, etc.  
    						
    							User Maintenance
    18-26 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Phantoms
    Quality Assurance Evaluations should be done with phantoms 
    and test objects that are applicable to the parameters being 
    evaluated or to the users clinical practice.
    Typical phantoms are composed of material that acoustically 
    mimic human tissue. Pins, anechoic and echogenic targets are 
    physically positioned to provide information for a variety of tests.
    Doppler phantoms are currently expensive and complicated to 
    deal with on the user level. If a problem with any Doppler 
    parameters or measurement is suspected, contact a local 
    service representative for evaluation.
    The RMI 403GS phantom is still available. Due to the superior 
    penetration and resolution capabilities of GE ultrasound 
    systems, the RMI 405GSX is recommended. It is the most 
    current one available to our field service personnel and will 
    provide the targets and extended life necessary for consistent 
    system testing. 
    						
    							Quality Assurance
    LOGIQ 5 Basic User Manual 18-27
    Direction 2300002-100 Rev. 2
    Phantoms (continued)
     Figure 18-9. Phantoms
    1. Penetration
    2.  Axial Distance Measurement
    3. Functional Resolution
    4. Lateral Resolution
    5.  Lateral Distance Measurement
    6. Axial Resolution
    7.  Contrast Resolution and Gray Scale Photography
    8.  Gray Scale Plane Targets 
    						
    							User Maintenance
    18-28 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Baselines
    An absolute necessity for a quality assurance program is 
    establishing baselines for each test or check. Baselines are 
    established after the system has been verified to be working 
    properly at installation or after a repair. If a probe or major 
    assembly is replaced, new baselines should be generated.
    Baselines can be made by adjusting system parameters to 
    prescribed levels or to the best possible image. The key factor to 
    remember is reproducibility. The same conditions must be 
    reproduced for each periodic check.
    All system parameters not displayed on the monitor should be 
    recorded for the permanent record.
    Periodic Checks
    Periodic checks should be performed in accordance with your 
    facility’s quality assurance requirements. For the data to be 
    valid, periodic checks should mimic the baseline setup 
    parameters.
    The resulting image, when scanning the phantom exactly as 
    before, should be recorded and compared to the baseline. 
    When a matching image is obtained, it can be assumed that the 
    system performance has not degraded from the baseline.
    If a significant difference between the baseline and periodic 
    check is noted, double check the system setup and repeat the 
    test. If the difference between the baseline and periodic check 
    persists, contact a local Service Representative.
    Failing to reproduce the control settings as in the baselines will 
    introduce errors in the data and potentially invalidate the results. 
    						
    							Quality Assurance
    LOGIQ 5 Basic User Manual 18-29
    Direction 2300002-100 Rev. 2
    Results
    Lack of standardization among test instruments, the wide range 
    of acceptance criteria, and incomplete knowledge regarding the 
    significance of certain performance parameters prohibit the 
    establishment of absolute performance criteria for these tests.
    Quality Assurance Evaluation results should be compared to 
    previously-recorded results. 
    Performance trends can then be detected. Unacceptable 
    performance or diminishing trends should be identified for 
    maintenance or repair before a malfunction or inappropriate 
    diagnosis occurs. 
    The user should determine the best method for recording and 
    archiving the baseline and periodic checks. In most cases the 
    choice is hard copy.
    It is important to maintain good consistent records for 
    inspections that may arise, as well as to detect system 
    performance trends. 
    						
    							User Maintenance
    18-30 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    System Setup
    The user should tailor the tests to their particular needs. It is 
    certainly not necessary to make all checks with all probes. A 
    representative example, with the probes used most often by the 
    customer, should be adequate in judging system performance 
    trends.
    Use a gray scale phantom as the scan object for the tests. 
    Commercial phantoms are supplied with its own operator 
    manual. Be familiar with proper phantom operating procedures 
    prior to use for quality assurance evaluations.
    1.  Adjust image monitor. Brightness and contrast should be set 
    to the normal viewing of a good gray scale image.
    2.  Check all recording devices for proper duplication of image 
    monitor. Ensure that what is seen is what is recorded. 
    Check the B/W or Color page printer, VCR, MIC, or Laser 
    camera.
    3.  Annotate non-displayed image processing controls. 
    4.  Set TGC slide pots to center (detent) position. 
    5.  Place focal zone marker(s) in area of interest for an 
    optimum image.
    Test Procedures
    The following are recommended Quality Assurance tests. A brief 
    description of the test, the benefit it provides and steps to 
    accomplish the test are supplied.
    The importance of recording scan parameters and consistent 
    record keeping cannot be stressed enough. Reproducibility to 
    monitor system trends is the key to quality assurance 
    evaluations.
    Using the systems dual image display format is often very 
    convenient and saves recording media. 
    						
    							Quality Assurance
    LOGIQ 5 Basic User Manual 18-31
    Direction 2300002-100 Rev. 2
    Axial distance measurements
    DescriptionAxial measurements are the distance measurements obtained 
    along the sound beam. See Figure 18-9 for more information.
    BenefitThe accurate measurement of the size, depth and volume of a 
    structure is a critical factor in determining a proper diagnosis. 
    Most imaging systems use depth markers and/or electronic 
    calipers for this purpose.
    MethodAxial distance should be measured in the near, mid and far fields 
    as well as in zoom. If necessary, different depths or fields of 
    view can be tested.
    ProcedureTo measure axial distance:
    1.  Scan a test phantom with precisely-spaced vertical pin 
    targets. Adjust all scan controls, as necessary, for the best 
    image of the pin targets to typical depths for the probe being 
    used.
    2. Press Freeze to stop image acquisition and perform a 
    standard distance measurement between the pins at 
    different points in the image. Record all images for 
    archiving.
    3.  Scan the vertical pins in zoom or at different depth/scale 
    factors.
    4. Press Freeze to stop image acquisition; repeat the distance 
    measurements between pins and record the images for 
    archiving.
    5.  Document the measurements for reference and future 
    comparison.
    Contact a Service Engineer if vertical measurements differ by 
    more than 1.50% of the actual distance. 
    						
    							User Maintenance
    18-32 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Lateral distance measurements
    DescriptionLateral measurements are distance measurements obtained 
    perpendicular to the axis of the sound beam. See Figure 18-9 
    for more information.
    BenefitThe purpose is the same as vertical measurements. Precisely-
    spaced horizontal pin targets are scanned and results compared 
    to the known distance in the phantom.
    MethodLateral distance should be measured in the near, mid and far 
    fields as well as in zoom. If necessary, different depths of fields 
    of view can be tested.
    ProcedureTo measure lateral distance:
    1.  Scan a test phantom with precisely-spaced horizontal pin 
    targets. Adjust all scan controls, as necessary, for the best 
    image of the pin targets from side to side.
    2. Press Freeze to stop image acquisition and perform a 
    standard distance measurement between the pins at 
    different points in the image. Record all images for 
    archiving.
    3.  Scan the horizontal pins in zoom or at different depth/scale 
    factors.
    4. Press Freeze to stop image acquisition; repeat the distance 
    measurements between pins and record the images for 
    archiving.
    5.  Document the measurements for reference and future 
    comparison.
    Contact a Service Engineer if horizontal measurements differ by 
    more than 3mm or 3% of that depth, whichever is greater. 
    						
    							Quality Assurance
    LOGIQ 5 Basic User Manual 18-33
    Direction 2300002-100 Rev. 2
    Axial resolution
    DescpritionAxial resolution is the minimum reflector separation between two 
    closely-spaced objects to produce discrete reflections along the 
    axis of the sound beam. It can also be monitored by checking 
    the vertical size of known pin targets. See Figure 18-9 for more 
    information.
    Axial resolution is affected by the transmitting section of the 
    system and the probe.
    BenefitIn clinical imaging, poor axial resolution displays small 
    structures lying close together as a single dot. This may lead to 
    improper interpretation of the ultrasound image.
    ProcedureTo measure Axial resolution:
    1.  Scan a test phantom with precisely-spaced vertical pin 
    targets. 
    2.  Adjust all scan controls, as necessary, for the best image of 
    the pin targets to typical depths for the probe being used.
    3. Press Freeze to stop image acquisition.
    4.  Perform a standard distance measurement of the pin 
    vertical thickness at different points in the image. Record all 
    images for archiving.
    5.  Scan the vertical pins in zoom or at different depth/scale 
    factors.
    6. Press Freeze to stop image acquisition; repeat the vertical 
    thickness measurements of the pins and record the images 
    for archiving.
    7.  Document the measurements for reference and future 
    comparison.
    Axial resolution should remain stable over time. Contact a 
    Service Engineer if any changes are observed. 
    						
    							User Maintenance
    18-34 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Lateral resolution
    DescriptionLateral resolution is the minimum reflector separation between 
    two closely spaced objects to produce discrete reflections 
    perpendicular to the axis of the sound beam. It can also be 
    monitored by checking the horizontal size of known pin targets. 
    See Figure 18-9 for more information.
    Lateral resolution is dependent upon the beam width produced 
    by the probe. The narrower the beam, the better the lateral 
    resolution.
    The beam width is affected by the frequency, degree of focusing, 
    and distance of the object from the face of the probe.
    BenefitClinically, poor lateral resolution will display small structures 
    lying close together as a single dot. This may lead to improper 
    interpretation of the ultrasound image.
    ProcedureTo measure lateral resolution:
    1.  Scan a test phantom with precisely-spaced horizontal pin 
    targets.
    2.  Adjust all scan controls, as necessary, for the best image of 
    the pin targets from side to side.
    3. Press Freeze to stop image acquisition and perform a 
    standard distance measurement of the horizontal thickness 
    of a pin at different points in the image. Record all images 
    for archiving.
    4.  Scan the horizontal pins in zoom or at different depth/scale 
    factors.
    5. Press Freeze to stop image acquisition; repeat the 
    horizontal thickness measurements of the pins and record 
    the images for archiving.
    6.  Document the measurements for reference and future 
    comparison. 
    Pin width should remain relatively constant over time (1mm). 
    Dramatic changes in pin width may indicate beamforming 
    problems. Contact a Service Engineer if beam width changes 
    consistently over 2 to 3 periodic tests.  
    						
    All GE manuals Comments (1)

    Related Manuals for GE Logiq P5 User Manual