Home > GE > Ultrasound System > GE Vivid 7 Service Manual

GE Vivid 7 Service Manual

    Download as PDF Print this page Share this page

    Have a look at the manual GE Vivid 7 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    Page
    of 858
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 10 - Care & Maintenance 10 - 23
    10-7-9-4 Data Sheet for Transducer Source Leakage Current The test passes when all readings  measure less than the values shown in  Table 10-13 and 
    Table 10-14 . Record all data on the PM Inspection Certificate.
    . 
    CAUTIONEquipment damage possibility. Never switch the Polarity and the status of Neutral when the unit 
    is powered ON. Be sure to turn the unit power  OFF before switching them using the POLARITY 
    switch and/or the NEUTRAL switch. Otherwise, the unit may be damaged
    Table 10-21   Typical Data Sheet For Transducer Source Leakage Current
    Transducer Tested:
     Unit Power
    Tester Power Polarity  SwitchTester GROUND or NUETRAL Switch
    Measurement
    ON NORM OPEN
    ON NORM CLOSED
    ON REV OPEN
    ONREVCLOSED
    OFFNORMOPEN
    OFFNORMCLOSED
    OFF REV OPEN
    OFF REV CLOSED 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    10 - 24 Section 10-8 - When Theres Too Much Leakage Current...
    Section 10-8
    When Theres Too Much Leakage Current...
    CHASSIS FAILS
    Check the ground on the power cord and plug for cont inuity. Ensure the ground is not broken, frayed, 
    or intermittent. Replac e any defective part.
    Tighten all grounds. Ensure star  washers are under all ground studs.
    Inspect wiring for bad crimps, poor connections, or damage.
    Test the wall outlet; verify it is grounded and is fr ee of other wiring abnormalities. Notify the user or 
    owner to correct any deviations. As a work around, ch eck the other outlets to see if they could be used 
    instead.
    NOTE: No outlet tester can detect the condition wh ere the white neutral wire and the green grounding 
    wire are reversed. If later tests indicate high leakage currents, this should be suspected as a 
    possible cause and the outlet wiring should be visually inspected.
    PROBE FAILS
    Test the probe in another connector to isolate  if the fault lies with the probe or the scanner.
    NOTE: Each probe will have some am ount of leakage, dependent on its design.  Small variations in 
    probe leakage currents are norm al from probe to probe. Other variations will result from 
    differences in line voltage and test lead placem ent. The maximum allowable leakage current for 
    body surface contact probe differs from inter-cavity probe. Be sure to enter the correct probe 
    type in the appropriate space on the check list.
    If excessive leakage current is slot dependent,  inspect the system connector for bent pins, poor 
    connections, and ground continuity.
    If the problem remains with the probe, replace the probe.
    PERIPHERAL FAILS
    Tighten all grounds. Ensure star  washers are under all ground studs.
    Inspect wiring for bad crimps, poor connections, or damage.
    STILL FAILS
    If all else fails, begin isolation by removing the pr obes, external peripherals, then the on board ones, 
    one at a time while monitoring the leakage current measurement.
    NEW UNIT
    If the leakage current measurement tests fail on a new un it and if situation can not be corrected, submit 
    a Safety Failure Report to  document the system problem.  Remove unit from operation.
    ECG FAILS
    Inspect cables for damage or poor connections 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 10 - Care & Maintenance 10 - 25
    ULTRASOUND INSPECTION CERTIFICATE  
    * Scan Format: Phased Array, Linear Array,  Curved Array, Mechanical Array or Other
    FUNCTIONAL CHECKS PHYSICAL IN SPECTION AND CLEANING 
    COMMENTS:
    ________________________________ ______________________________________________
    ________________________________ ______________________________________________
    ________________________________ ______________________________________________
    Customer Name: System ID: Dispatch Number  / Date Performed:Warranty/Contract/HBS
    System Type Model Number: Serial Number: Manufacture Date:
    Probe 1: Frequency: Scan Format*: Model Number: Serial Number:
    Probe 2: Frequency: Scan Format*: Model Number: Serial Number:
    Probe 3: Frequency: Scan Format*: Model Number: Serial Number:
    Probe 4: Frequency: Scan Format*: Model Number: Serial Number:
    Probe 5: Frequency: Scan Format*: Model Number: Serial Number:
    Probe 6: Frequency: Scan Format*: Model Number: Serial Number:
    Probe 7: Frequency: Scan Format*: Model Number: Serial Number:
    Probe 8: Frequency: Scan Format*: Model Number: Serial Number:
    Probe 9: Frequency: Scan Format*: Model Number: Serial Number:
    Functional Check  (if applicable) OK? or 
    N/A Physical Inspection and Cleaning 
    (if applicable) Inspect Clean
    B-Mode Function Console
    Doppler Modes Function Monitor
    CF-Mode Function Touch Panel
    M-Mode Function Air Filter
    Applicable Software Options Probe Holders
    Applicable Hardware Options External I/O
    Operator Panel Wheels, Br akes & Swivel Locks
    Monitor Cables and Connectors
    Touch Panel GE Approved Peripherals (VCR, DVR, CD/DVD, MOD, 
    Printers)
    Measurement Accuracy
    GE Approved Peripherals 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    10 - 26 Section 10-8 - When Theres Too Much Leakage Current...
    ELECTRICAL SAFETY
    Final Check. All system covers are in plac e. System scans with all probes as expected.
    Accepted by: ______________________________________________________________________
    Electrical Test Performed Max Value 
    Allowed Value 
    Measured OK? Comments
    Outlet (correct ground &wiring config.)
    System Ground Continuity
    Chassis Source Leakage Current - Probe
    Chassis Source Leakage Current - Wheel
    Chassis Source Leakage Current - CRT
    Patient Lead Source Leakage
    (Lead to Ground)
    Patient Lead Source Leakage
    (Lead to Lead)
    Patient Lead Source Leakage
    (Isolation)
    Peripheral 1 Leakage Current
    Peripheral 1 Ground Continuity
    Peripheral 2 Leakage Current
    Peripheral 2 Ground Continuity
    Peripheral 3 Leakage Current
    Peripheral 3 Ground Continuity
    PROBES
    Probe Number 
    (from previous page) Max Value 
    Allowed Max Value 
    Measured OK? Comments
    Probe 1:
    Probe 2:
    Probe 3:
    Probe 4:
    Probe 5:
    Probe 6:
    Probe 7:
    Probe 8:
    Probe 9: 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 10 - Care & Maintenance 10 - 27
    Section 10-9
    Site Log
    Table 10-22   Site Log 
    DATESERVICE PERSONPROBLEMCOMMENTS 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    10 - 28 Section 10-9 - Site Log
    Table 10-22   Site Log (cont’d)
    DATESERVICE PERSONPROBLEMCOMMENTS 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 10 - Care & Maintenance 10 - 29
    Table 10-22   Site Log (cont’d)
    DATESERVICE PERSONPROBLEMCOMMENTS 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    10 - 30 Section 10-9 - Site Log
    This page was intentionally left blank. 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Index Index - 1
    Numerics
    2D Mode Checks 4-32
    4D Volume Rendering Test 4-55
    A
    Abbrevations
    Definitions 9-3
    AC Power Replacement Procedure 8-82
    AC Power Parts 9-61
    Accessory Boxes, Vivid 7 / Vivid 7 PRO 9-105
    Acclimate Time 2-2, 3-2
    Acoustic Noise Output 3-14
    Add Printer 3-35
    Adjust LCDs Contrast  and Backlight Intensity 3-34
    Advanced Settings 3-51
    AE Title 3-126
    Air Filter Cleaning 10-10
    Application Turnover Check List 4-82
    Archiving and Loading Presets 4-19
    Archiving Presets to an MO Disk 4-19
    Audio Doppler 5-51
    Available Probes 3-38
    Average Installation Time 3-2
    B
    Back End Processor5-73
    Theory 5-73
    Back Plane (Motherboard) 5-72
    Back-End Processor 9-46
    Back-End Processor Version 1 Parts 9-60
    Back-End Processor Version 2 Parts 9-59
    Back-End Processor Version 2.2 Parts 9-58
    Back-End Processor Version 3.0 Parts 9-57
    Back-End Processor Version 3.2 Parts 9-55
    Back-End Processor Version 4.2
    Parts 9-47, 9-49
    Backup - Software Version v7.x/v6.x/v5.x/v4.x 4-23
    Beamformer Board(s) 5-33
    Beamformer Calibration 6-6
    Beamformer Calibration --  see DC Offset Calibration 7-34
    BF-64 5-33 B-Mode Checks
    4-32
    Brake Assembly 9-37
    Brake Pedal Replacement Procedure 8-101
    Brake Pedal and Direction Lock Assembly Replacement Procedures 8-103
    Brakes and Direction Lock Checks 4-80
    Brakes Check 4-80
    Brightness adjustments on monitor 6-5
    C
    Cables
    ECG 9-78
    Cables - Vivid 7 9-69
    Cancel 4-9
    Chain Signal Chain, TX/RLB/RX/BF Test 7-73
    chapters overview 1-1
    Chassis Leakage Current Test 10-16
    Check the Connection to the DICOM Server 3-130
    Cineloop Check 4-69
    Cleaning 10-9
    Color Flow Processing 5-47, 5-62
    Color Inkjet Printer 3-37
    Color Inkjet Printer, EPSON 980N 3-37
    Color Mode Overview 4-49
    Color Mode Checks 4-43
    Color Printer, HP Deskjet 6127 3-37
    Color Printer, HP Inkjet 1100DTN 3-37
    Color Printer, HP Inkjet 1200DTN 3-37
    Color Printer, HP OfficeJet PRO K550 Installation 3-37
    Common Diagnostics 7-48
    Common Service Interface 5-152
    Compatibility 3-43
    Hardware/Softwar e Compatibility9-7
    Complete Power Down 4-11
    Completing the Installation 3-14
    Components and Functions 5-1
    Configuration 3-28
    Connect a Probe 3-21
    Connect ECG 3-20
    Connect Ethernet 3-19
    Connect Footswitch 3-19
    Connect Mains Power to Scanner 4-3
    Connect Mains Power to the Unit 3-23
    Connect Phono 3-20
    INDEX 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Index - 2  - 
    Connect Pulse Pressure Transducer 3-20
    Connect USB Flash Card 3-19
    Connection to a DICOM Server in a Network 3-58
    Connections on the External IO 3-19
    Connections on the Patient I/O Panel 3-20
    Connectivity 5-148
    Setup 3-42, 3-89 , 3-109
    Connectivity Overview 3-41
    Console Environmental Requirements 2-2
    Mechanical Checks 4-78
    Console Lock 9-22
    Vivid 7 PRO 9-24
    VIVID 7/VIVID 7 Dimension 9-23
    Contrast adjustments on monitor 6-4
    Contrast Checks 4-63
    Conventions Conventions Used in Book 1-9
    Cooling 2-2
    Create a New Dataflow 3-75
    CRT (Monitor) Adjustments 6-3
    CRT Monitor 9-35
    Customer Assistance 1-71
    D
    Damage in Transportation 3-8
    Dangerous Procedure Warnings 1-68
    Data Buffer 5-47, 5-62
    Data Sheet for - Chassis Source Leakage Current 10-17
    - ECG Leakage Current 10-20
    - Transducer Source Leakage Current 10-23
    Date and Time Adjustments 3-29
    DC Offset Calibration 7-34, 8-79
    DC Power 5-133
    Replacement Procedure 8-86
    Default DICOM Setup in Configuration Screen 3-129
    Definitions of Left, Right, Front and Back 8-2, 9-2
    Department Name 3-28
    Depth 3-14
    DHCP Configuration 3-52
    Diagnostics 7-7
    Diagnostics/Troubleshooting 7-1
    DICOM 3-126
    DICOM Option Pre-Inst allation Requirements2-13
    DICOM Server’s Set Up 3-126
    Worksheet for DICOM Network Information 2-14
    DICOM Storage Set Up 3-64
    Digital Signal Processor DSP 5-48
    Digital Video Disc Recorder 9-67
    Direction Lock Check
    4-80
    Pedal Replacement Procedure 8-102
    Direction Lock Checks 4-80
    Disassembly Nester 1-62
    Disconnect Probe 3-22
    Disk Management Configuration 4-21, 8-15
    Display Platform Mechanical Checks 4-78
    Doppler Doppler Processing 5-51
    Doppler Mode Checks 4-49
    DSP Digital Signal Processor 5-48
    DVR 9-67
    DVR Checks 4-74
    E
    ECG
    Connection 3-20
    ECG Cables 9-78
    ECG Check 4-68
    Eject and Remove CD-R Disc 4-16
    Eject and Remove MO Disk 4-15
    Electrical - Requirements 2-3
    - Safety 1-19
    - Safety Tests 10-12
    - Specifications 3-15
    Electrical Part - Fan 9-26
    Electrical Specifications for Vivid 7 2-4, 3-16
    Electromagnetic Compatibility (EMC) 1-70
    Electrostatic Discharge Warning 1-70
    EMC 1-70
    What is EMC? 1-70
    EMI 1-70
    - Limitations 2-7
    - Protection 3-14
    Enter Location 3-28
    Environmental Conditions Probes 2-9
    Error Message(s) During Start Up 7-3
    ESD 1-70
    Establish Network Conne ction to Hospital’s
    Network 3-126
    Ethernet - Signal 5-117
    - Switch Connections 3-133
    Connection 3-19
    eVue configuration 3-74
    Examination List Screen 3-87
    Excel Export to Excel Configuration (software version v4.x  and newer) 3-72 
    						
    All GE manuals Comments (0)

    Related Manuals for GE Vivid 7 Service Manual