GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 77 3-9-16-3 Create a New Dataflow with Worklist and DICOM Storage (for CA-1000 sites) This dataflow is intended for Vivid 7 scanners in a DI COM network, where the data should not be stored locally on the scanner. A site with the integrated EchoPAC PC and CA-1000 is an example of this. To create a new dataflow with Worklist an d DICOM Storage, do the following steps: 1.) Add a new dataflow in Config screen, Conn ectivity, Dataflow tab. Enter a name, e.g. Worklist - DICOM Storage . 2.) Select OK. 3.) Add the Worklist inpu t. Configure (at least) IP address, AE title and Port Number . 4.) Select OK. 5.) Add the Dicom Storage output. 6.) Configure (at least) IP address, AE title and Port Number . If the site has inte grated EchoPAC PC and CA-1000, enable Allow SR and Allow SR Private Data . 7.) Select OK. 8.) If this dataflow should be the default at start-up, enable Default for the dataflow. Figure 3-76 Dicom SR Settings
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 78 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x 3-9-17 Query/Retrieve (Q/R) Setup for Software v7.x/v6.x/v5.x/v4.x 3-9-17-1 Overview The Query/Retrieve function makes it possible to search for and retrieve DICOM data from a DICOM server for further analysis on the Vivid 7. NOTE: You may have to set up Vivid 7 as a destination on the server. 3-9-17-2 Query/Retrieve Setup on the Vivid 7 with Software v7.x/v6.x/v5.x/v4.x 1.) Press CONFIG (F2) and log on as adm, see 4-2-4 Log On to the System as ‘ADM’ on page 4-12 . 2.) Select Connectivity 3.) Select Dataflow 4.) Select Query Retrieve from the Name pull down menu Figure 3-77 Select Query Retrieve NAME PULLDOWN MENU QUERY RETRIEVE
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 79 3-9-17-2 Query/Retrieve Setup on the Vivid 7 with Software v7.x/v6.x/v5.x/v4.x (cont’d) In Figure 3-78 , the selected dataflow, Query/Retrieve, is shown. 5.) Select QueryRetrieve so it is hi ghlighted and then select Propertie s to display the Properties dialog. Figure 3-78 Query/Retrieve workflow Figure 3-79 Select Qu ery/Retrieve’s Properties SELECT QUERY RETRIEVE AND THEN SELECT PROPERTIES QUERY RETRIEVE PROPERTIES DIALOG
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 80 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x 3-9-17-2 Query/Retrieve Setup on the Vivid 7 with Software v7.x/v6.x/v5.x/v4.x (cont’d) 6.) Select the IP-address down-arrow to choose t he DICOM Query/Retrieve server from the pull down menu. In some cases, the server to use is the same as used for DICOM Storage. If the server to use is missing from the list, select from the pull down menu and edit the setup for one of the predefined servers. 7.) Enter the correct AE Title and Port Number for th e DICOM Query/Retrieve server in the respective fields in the Query/Retrieve screen. Figure 3-80 DICOM Query/Retrieve properties IP-ADDRESS DOWN-ARROW AE TITLE PORT NO DO NOT SELECT THE SEARCH CRITERIA BUTTON IF NOT TOLD SO BY THE ONLINE CENTER.
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 81 3-9-17-2 Query/Retrieve Setup on the Vivid 7 with Software v7.x/v6.x/v5.x/v4.x (cont’d) Change Search Criterias It is possible to set up special Se arch Criterias for DICOM Query/Retrieve. In most cases you may leave the Search Criterias as is , and skip this adjustment. Follow the steps below to change the Search Criterias parameters: 1.) Select Search Criterias 2.) Select the correct tag from the Select Tag pull-down menu. 3.) If needed, type in the value. NOTE: In this example, tag 00080060 Modality has been added to the list with value “US”. This may be useful if the studies cont ain non-ultrasound images. 4.) Select Add to List 5.) Select OK to close the Search Criterias window. Figure 3-81 Select Search Criterias SEARCH CRITERIAS ADD TO LIST REMOVE SELECT TAG
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 82 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x 3-9-17-3 Query/Retrieve Verification Follow the Steps Below to do a First Test (TCP-IP Ping) of the Connection 1.) Select Query retrieve and then select Properties. 2.) Select the “Smiley” button to Ping the server. NOTE: The smiley checks if the remote server is accessible (ping). It is not a DICOM Echo (DICOM ping), so it does not check AE title or port number. 3.) If the network connection to the server is OK, it will be illustrated by an smiling “Smiley” . A sad Smiley indicates that the network connection is failing. Typical causes: - Network cable not connected. - Configuration error(s). 4.) When ready, select OK to close the DICOM Query/Retrieve properties and save changes. Figure 3-82 DICOM Query/Retrieve properties SMILEY (NETWORK PING) OK
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 83 3-9-17-3 Query/Retrieve Verification (cont’d) Follow the Steps Below to Test th e DICOM Query/Retrieve Workflow NOTE: This check uses both Ping and DICOM Ping. - Ping is used to verify the TCP/IP connection to the server. - DICOM Ping is used to verify that the DICOM application is answering. 1.) Select QueryRetrieve from the Selected devices list. 2.) Select Check. - If the test pass, a pop-up dialog, reporting the results as OK, is displayed, see Figure 3-84. - Select OK to continue. A green check mark to the left of the Selected devices indicates that the test was successfully passed. Figure 3-83 Check the DICOM Query/Retrieve workflow Figure 3-84 Test (Check) Passed Figure 3-85 Test (Check) Passed QUERYRETRIEVECHECK
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 84 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x 3-9-17-3 Query/Retrieve Verification (cont’d) Follow the Steps Below to Test the DICOM Query/Retrieve Workflow (cont’d) - If the test fails, a pop-up dialog reporting the resu lts as “0 of 1 OK” is displayed. In the example in Figure 3-86, the warning message pop up almost at onc e, and it states that for the Server Check (Ping), 1 of 1 is OK but the Device Che ck failed (0 of 1 OK). This indicates that the network connection functions OK, but the application on the remote server is not answering. If the test fails directly, as described in this example, the AE Title is probably wrong. A red “X” to the left of the Selected devices indicates that the test failed, but it doesn’t indicate the cause. - If the Check button fails after a long time (corresponding to the timeout), the Warning will look like the example in Figure 3-88, where both tests failed. In this case, the IP address or Port Number is probably wrong. If Retrieve is Failing ... If Retrieve is failing but Qu ery is functioning, verify if the scanner has been registered as an receiver on the server. Figure 3-86 Test (Check) Fails directly Figure 3-87 Test (Check) Passed Figure 3-88 Both Tests Failed
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 85 3-9-18 Mapping of Parameters from Vivid 7 to DICOM 3-9-18-1 Search Screen Figure 3-89 Search / Create Patient Screen Table 3-17 Search / Create Patient Screen ITEMVIVID 7 FIELD NAMEDICOM TAG NUMBERDICOM TAG NAME 1 LAST NAME, FIRST NAME (0010,0010) PATIENT’S NAME 2 SEX (0010,0040) PATIENT’S SEX 3 ACCESSION (0008,0050) ACCESSSION NUMBER 4PAT I E N T I D(0010,0020)PATIENT ID 5BIRTHDATE(0010,0030)PAT I E N T ’ s B I R T H D AT E 6STUDY ID(0020,0010)STUDY ID THE COLUMN EXAM DATE, WAS INTRODUCED FOR SOFTWARE VERSION V.6.0. 1 2 3 4 5 6
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 86 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x 3-9-18-2 Patient Information Screen Figure 3-90 Patient Information Screen Table 3-18 Patient Information Screen ITEMVIVID 7 FIELD NAMEDICOM TAG NUMBERDICOM TAG NAME 1 OPERATOR (0008,1070) OPERATOR’S NAME 2 REF. DOC (0008,0090) REFFERING PHYSICIAN’S NAME 3 DIAGN.PHYS. (0008,1050) PERFORMING PHYSICIAN’S NAME 4WARD/DEPT(0008,1040)INSTITUTIONAL DEPARTMENT NAME 5WEIGHT(0010,1030)PATIENT’S WEIGHT 6DESCRIPTION(0008, 1030)STUDY DESCRIPTION 7 HEIGHT (0010,1020) PATIENT’S SIZE 8 CONTRAST (0018,0010) CONTRAST/BOLUS AGENT 1 2 3 4 5 6 7 8