GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 67 3-9-14-6 DICOM Storage Properties Adjustments (cont’d) DICOM SR If SR is enabled, all M&A that is supported by DICO M is sent to the DICOM server (in a separate job). Allow SR Private Data (introduced with software versions v6.0, v5.2 and v4.2.) By default, this function is disabled. Allow SR Private Data is intended for receivers that can handle “Private Data”, like the CA-1000. In the examples in Figure 3-67 and Figure 3-68 (below), Allow SR Private Data has been enabled. Figure 3-67 Dicom Storage properties Figure 3-68 Dicom Media properties
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 68 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x 3-9-14-6 DICOM Storage Properties Adjustments (cont’d) Retry • Max # - The number of times the software will try to connect if the connection doesn’t succeed the first time(s). • Interval - The interval in seconds between each attempt to connect. Smiley Select the Smiley button to ping th e server. If the network connection is OK, the Smiley will smile, if the connection fail, the Smiley will turn sad. NOTE: The smiley checks if the remote server is accessible (ping). It is not a DICOM Echo (DICOM ping), so it does not check AE title or port number. Reopen per Image If Reopen per image is enabled, Vi vid 7 will create a new connection (asso ciation) for each image. This may be useful for DICOM servers that do not acce pt different image types in the same association. Storage Commitment If you want to enable Storage Co mmitment, check the check-box an d select the button “Storage Commitment”. In the dialog, enter IP address, Name, AE Title and Port Number. Figure 3-69 Storage Commitment
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 69 3-9-14-6 DICOM Storage Properties Adjustments (cont’d) MPPS If you want to enable MPPS, check the check-box and select the button “MPPS”. In the dialog, enter IP address, Name, AE Title and Port Number. Time-out Adjust Time-out setting Introduced with software versions v6.0, v5.2 and v4.2: The retry settings can be used to make jobs retry on bad networks. There is no need to set retries for mobile (off-line) use when these software versions are used. Software v5.1.x/v4.1.x and lower: The retry settings for DICOM Storage are by default for mobile (off- line) use. They ensure that jobs in the spooler will be sent automatically if the user reconnects within the total retry period. If the customer does not have the mobile case, you can reduce the number of retries and the interval. This will make the jo b fail faster if there is a network or server problem (and let the user forward the job to another server if desired). OK/Cancel •Select OK to save all changes. The changes will take place after the scanner has been restarted. •Select Cancel to return without saving any changes. Figure 3-70 MPPS
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 70 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x 3-9-15 Export Configuration The destination for Export of patient records to Exce l and MPEG must be configured prior to use. This is done from the Dataflow screen. 3-9-15-1 Setup on the Remote Share Required setup on the remote share (remote server): 1.) Add user and password.Predefined user: E1c2h3o4C5l6i7e8n9t Predefined password: u1l3t5r7a 2.) Set Share permissions. 3.) Set security permissions.
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 71 3-9-15-2 Display the Dataflow Screen Follow these steps to display the Dataflow Screen: 1.) Press CONFIG (F2) and log on as administrator. 2.) Select Connectivity > Dataflow . The Dataflow sheet is displayed, see Figure 3-71 The Dataflow sheet on page 3-71 . 3.) Select the dataflow Misc. Export in the Name pull-down menu. Figure 3-71 The Dataflow sheet Software version v4.x/v5.x Software version v6.x: A new check-box, Show Only Imp/Export has been added.
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 72 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x 3-9-15-3 Export to Excel Configuration 1.) Select the Excel storage device in the Selected devices pane and press Properties . The Excel properties window is displayed. 2.) Select a removable media or a network volume re mote path as the destination in the Destination pull-down menu. NOTE: Remote paths for network volumes must be entered once in the Remote path field before they can be selected from the Destination Pull-down menu. 3.) Select OK. 4.)Press CONFIG (F2) if you want to quit Config. Figure 3-72 The Excel properties window EXCEL STORAGE PROPERTIES
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 73 3-9-15-4 Export to MPEGVue Configuration 1.) Select the MPEGVue device in the Selected devices pane and press Properties . The MPEGVue properties window is displayed. 2.) Select a removable media or a network volu me remote path as the destination in the Destination Pull-down menu . NOTE: Remote paths of network volumes must be enter ed once in the Remote path field before they can be selected from the Destination Pull-down menu. 3.) Check the options as required. 4.) Select OK. 5.) Press CONFIG (F2) if you want to exit Configuration. Figure 3-73 The MPEGVue properties window
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 74 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x 3-9-15-5 eVue Configuration 1.) Press CONFIG (F2) and log on as administrator. 2.) Select the Connectivity category and Dataflow subgroup. The Dataflow sheet is displayed. 3.) Select the dataflow Local Archive - Int. HD/eVue in the Name pull-down menu. 4.) Select the eVue device in the Selected devices pane and press Properties. The eVue properties window is displayed. 5.) Select a removable media or a network volu me remote path as the destination in the Destination pull-down menu. NOTE: Remote paths of network volumes must be enter ed once in the Remote path field before they can be selected from the Destination Pull-down menu. 6.) Check the options as required. 7.) Select OK and press CONFIG (F2) . Figure 3-74 The eVue dataflow Figure 3-75 The eVue Properties
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 75 3-9-16 Create a New Dataflow 3-9-16-1 Overview It is possible to make new dataflows by combining the predefined settings. The table below describes the legal combination of inputs and outputs in a dataflow. 3-9-16-2 Create a New Dataflow with Worklist, Database and DICOM Storage (Introduced for software v4.0) To create a new dataflow with Work list, Database and DICOM Storage, do the following in exactly these steps: 1.) Add the Worklist Input, highlight it, select proper ties, and name it (i.e.: DR PACS Worklist or just Worklist). Go to the IP Address field, scro ll down and select . This will allow you to edit or add IP Addresses (New Destinations/devices). If the correct IP Address is already in the list, just select it. 2.) Add the Database Input/Output, highlight it, select properties, and name it (i.e.: Local Archive - Int HD). Go to the IP Address field and (if the Database is to be the Local Archive) verify that My Computer is selected. The IP Address should then be correct and the Database Name and Archive File Location entries should auto-populate correctly. Table 3-16 Allowed combinations of inputs and outputs in a dataflow NOTICEFailure to add destinations to the dataflow in the correct order may or will caus e intermittent errors. Output Input No Output Database Remote Database DICOM MOD5.25DICOM MOD3.5 DICOM StorageDICOM Print Database + DICOM MOD 5.25/3.5Database + DICOM Storage Remote Database + DICOM MOD 5.25/3.5Remote Database + DICOM Storage Database XXX Remote DatabaseX XX DICOM WorklistX DICOM CDX DICOM MOD 5.25X DICOM MOD 3.5X Query/RetrieveX EchoPAC MacX Worklist/databaseXX Worklist/rem databaseXX No input deviceXX
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 76 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x 3-9-16-2 Create a New Dataflow with Worklist, Database and DICOM Storage (cont’d) 3.) Add the DICOM Storage Output, highlight it, select properties, and name it (i.e.: DR PACS Store).Go to the IP Address field, scroll down and select . This will allow you to edit or add IP Addresses (New Destinations/devices). If the correct IP Address is alre ady in the list, just select it. If you create the Dataflow in any other order or remove and add an entry later, you may or will receive “you are not connected to the Data base” errors when trying to create a patient locally. The Worklist will work, however they can NOT save patients “created ” on the Vivid 7 to the Archive while using that dataflow. You can change the information on each entry in a Dataflow as long as you do NOT remove or add to the Dataflow out of sequence. You should also set Direct Search to “All patients” for the new dataflow.