    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 8 - Replacement Procedures 8 - 93
    8-14-4      Rear Caster s Removal Procedure (cont’d)
    5.) Activate the brakes on all Casters, except the one you intend to replace.
    6.) Turn the Rear Caster that stands on the Bevel Edged Board in the direction as shown in 
    Figure 8-88 .
    7.) The system is now ne arly balanced between one  Front and one Rear Caster (illustrated in this 
    Procedure; Left Front and Right Rear). 
    8.) Make the System rest on both Front Casters a nd lift the Rear Caster. Put the Wooden Wedge under 
    the AC Transformer. This will stab ilize the System with the Rear Caster free from the floor.
    Figure 8-88   Activate Rear Brakes
    Figure 8-89   System stabilized with Rear Caster free from the floor
    Rear Wheel free 
    from the floor
    Wooden Wedge under 
    AC Transformer 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    8 - 94 Section 8-14 - Rear Casters Replacement Procedure
    8-14-4      Rear Caster s Removal Procedure (cont’d)
    9.) Remove four (4x) screws fastening Caster.
    10.) Remove Caster
    8-14-5 Rear Casters In stallation Procedure
    1.) Replace the Caster and tightly fa sten it with four (4x) screws.
     
    2.) Remove the Wooden Wedge.
    3.) Roll the system off the Bevel Edged Board.
    4.) To replace the other Rear Caster, repeat all the steps from  Rear Casters Removal Procedure on 
    page 8-92,  but now using the Bevel Edged board and t he Wooden Wedge on the other side of the 
    system.
    5.) Install Lower Rear Cover, see  Lower Rear Cover Installation Procedure on page 8-10 .
    6.) Install Filter Cover, see  Upper Rear Cover Installation Procedure on page 8-6 .
    7.) Install Side Covers, see  Side Covers Installation Procedure on page 8-5 .
    Figure 8-90   Rear Caster fastening
    WARNINGWARNINGUse extreme caution in the next two steps. The scanner is unstable.
    Screws 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 8 - Replacement Procedures 8 - 95
    Section 8-15
    Front Casters Replacement Procedure
    8-15-1 Manpower
    1 persons, 30 minutes per Caster
    8-15-2 Tools
    • Torx screwdriver T-20
    • Open end wrenches, 10 mm and 134 mm
    • 5 mm hexagon key
    • Bevel Edged Board and Wooden Wedge (service parts, see  Table 9-77 Service Kits, Overview on 
    page 9-99 .)
    8-15-3 Preparations
    1.) Shut down the system, see  4-2-3 Power Shut Down on page 4-7 .
    2.) Remove the Front Bumper, see  Front Bumper Removal Procedure on page 8-100 .
    3.) Turn Front Casters as shown in  Figure 8-92.
    4.) Activate Direction Lock.
    5.) Put the Bevel Edged Board on the floor. Figure 8-91   Bevel Edged Board and Wooden Wedge.
    Wooden Wedge
    Bevel Edged Board 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    8 - 96 Section 8-15 - Front Casters Replacement Procedure
    8-15-4 Front Casters Removal Procedure
      
    6.) Pull the system backwards up the board incline. 
    7.) Turn the Rear Caster on the board in the direction as shown in  Figure 8-93.
    8.) Activate both Rear Brakes as shown in  Figure 8-93.
    WARNINGWARNINGPrior to elevating scanner, verify that the floating Operator Panel is locked in its lowest 
    position. 
    WARNINGWARNINGUse extreme caution as long as VIVID 7 is un-stable, not resting on all four Casters.
    Figure 8-92   Pull system backwards up the board incline
    Figure 8-93   Activate both rear brakes 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 8 - Replacement Procedures 8 - 97
    8-15-4      Front Caster s Removal Procedure (cont’d)
    9.) Put the Wooden Wedge under the  Front-End Card Cage. This will stabilize the System with the right 
    Front Wheel free from the floor.
    .
    Figure 8-94   System stabilized with front wheel free from the floor 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    8 - 98 Section 8-15 - Front Casters Replacement Procedure
    8-15-4      Front Caster s Removal Procedure (cont’d)
    10.) Remove the lock pin that secures the ball joint.
    11.) Pull balljoint apart.
    .
    12.) Unscrew four (4x) screws.
    13.) Remove the Caster.Figure 8-95   Front Caster Brake Arm and Ball Joint Parts
    Figure 8-96   Caster fastening (both Casters shown)
    Lock pin
    Adjustment nut
    Retaining nut
    Brake lever
    Screws
    Screws
    Screws
    Screws 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 8 - Replacement Procedures 8 - 99
    8-15-5 Front Casters In stallation Procedure
    1.) Move the ball piece of the joint from old Caster to new Caster (See  Figure 8-95).
    2.) Align new Caster on the frame and fasten it tightly with four (4x) hexagon screws.
    3.) Press the ball joint together and install the lock pin (See  Figure 8-95).
    4.) To replace the other Front Cast er, repeat all the steps from Front Casters Removal Procedure on 
    page 8-96,  but now using the Bevel Edged board and t he Wooden Wedge on the other side of the 
    system.
    5.) Install Bumper and Covers, see  Front Bumper Installation Procedure on page 8-100 .
    6.) Roll the system off the Bevel Edged Board. 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    8 - 100 Section 8-16 - Front Bumper Replacement Procedure
    Section 8-16
    Front Bumper Replacement Procedure
    8-16-1 Manpower
    One person, 30 minutes
    8-16-2 Tools
    • Phillips size 2
    • Torx screwdriver T-20
    8-16-3 Preparations
    1.) Shut down the system, see  4-2-3 Power Shut Down on page 4-7 .
    2.)  Remove Front Cover, see  Front Cover Removal Procedure on page 8-7 .
    8-16-4 Front Bumper Removal Procedure
    1.) Unscrew 4 Torx screws under the bumper.
    2.) Lift the bumper off.
    8-16-5 Front Bumper Installation Procedure
    1.) Put the new bumper in position.
    2.) Fasten the 4 screws firmly.
    NOTE: Do NOT overtighten these screws. 3.) Install Front Cover, see  Front Cover Installation Procedure on page 8-7 .
    Figure 8-97   System front and Front Bumper
    Screw holes
    Front Bumper 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 8 - Replacement Procedures 8 - 101
    Section 8-17
    Brake Pedal Replacement Procedure
    8-17-1 Manpower
    One person, 15 minutes + travel
    8-17-2 Tools
    Torx screwdriver T-20
    8-17-3 Brake Pedal Replacement Procedure
    1.) Unscrew the four screws under the pedal.
    2.) Replace the pedal with a new one.
    3.) Fasten the four screws. Figure 8-98   Brake Pedal
    Brake Pedal 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    8 - 102 Section 8-18 - Direction Lock Pedal Replacement Procedure
    Section 8-18
    Direction Lock Pedal Replacement Procedure
    8-18-1 Manpower
    One person, 15 minutes
    8-18-2 Tools
    Torx screwdriver T-20
    8-18-3 Direction Lock Replacement Procedure
    1.) Unscrew 4 screws, 3 under and 1 at the left side of the pedal.
    2.) Replace the pedal with a new one.
    3.) Fasten 4 screws. Figure 8-99   Direction Lock Pedal
    Direction Lock Pedal 
    						