    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 6 - Service Adjustments 6 - 11
    6-8-2-4      Direction Lock Adjustment (cont’d) 5.) Turn adjustment nuts un til the balljoints matches.
    6.) Press the ba lljoint together.
    7.) Install the lock pins, see  Figure 6-5.
    8.) Fasten the retaining nuts.
    9.) Test Direction Lock. Figure 6-6   Balljoint with Lock Pin
    Adjustment nut
    Retaining nut
    Lock Pin 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    6 - 12 Section 6-8 - Direction Lock and Brake Adjustment
    6-8-3 Rear Brakes Adjustment
    6-8-3-1 Manpower 2 persons, 1 hour
    6-8-3-2 Tools • Hexagon key 3/32 inch
    • Phillips screw driver size 2
    6-8-3-3 Preparations 1.) Remove Filter Cover , see  Upper Rear Cover Removal Procedure on page 8-6 .
    2.) Remove Lower Rear Cover , see  Lower Rear Cover Removal Procedure on page 8-10 .
    3.) Remove the Caster , see  Rear Casters Removal Procedure on page 8-92
    6-8-3-4 Rear Brakes Adjustment 1.) Use a 3/32 inch hexagon key to adjust the br ake until the space between brakedogs and tires are 
    approximate 2 mm.
    1.) Replace the Caster and tightl y fasten it with four screws.
    2.) Install Lower Rear Cover , see  Lower Rear Cover Installation Procedure on page 8-10 .
    3.) Install Filter Cover , see  Upper Rear Cover Installation Procedure on page 8-6 .
     
    4.) Remove the Wooden Wedge.
    5.) Roll the system off the Bevel Edged Board.Figure 6-7   Rear Brakes adjustment
    WARNINGWARNINGUse extreme caution in the next two steps. The scanner will be unstable.  
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 7 - Diagnostics/Troubleshooting 7 - 1
    Chapter 7
    Diagnostics/Troubleshooting
    Section 7-1
    Overview
    7-1-1 Purpose of this Chapter
    This section describes how to setup and run the tool s and software that help maintain image quality and 
    system operation. Very basic host, system and boa rd level diagnostics are run whenever power is 
    applied. Some Service Tools may  be run at the application level. 
    NOTE: Diagnostics must be run from start-up.
    NOTE: After Diagnostics are performed/run, the syste m must be rebooted. If not done, this could cause 
    subsequent false failures due to the ma chine being left in an unknown state.
    7-1-2 Contents in this Chapter
    Table 7-1    Contents in Chapter 7
    SectionDescriptionPage Number
    7-1 Overview 7-1
    7-2 Service Tools Overview 7-2
    7-3 Frequently Asked Questions - FAQ 7-3
    7-4Service Safety Considerations7-4
    7-5Gathering Troubleshooting Data7-5
    7-6Diagnostics Tools7-7
    7-7 Common Diagnostics 7-48
    7-8 Acquisition Diagnostics 7-63 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    7 - 2 Section 7-2 - Service Tools Overview
    Section 7-2
    Service Tools Overview
    7-2-1 Service Software Tools
    Two major Service Software  tools are available: 
    • Common Service Diagnostic Interface (GSUI)  
    This software tool makes  it possible for you to do tests on  various functions on the Back-End 
    Processor. It also includes an acquisition test that  tests most of the PCBs in the Front-End Rack. If 
    this test fails, detailed Acquisit ion Diagnostics tests are availabl e via Vivid 7’s software menus.
    • Acquisition Diagnostics
    This software tool includes detaile d Acquisition Diagnostics tests available via Vivid 7’s software 
    menus. It is running on the Back-End Processor too. It  includes both a Test All function that is the 
    same test as available from the GSUI, but also tests on PCB level.
    7-2-2 Special Service Tool
    Vivid 7 Test Attenuator, Part Number: AA200215 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 7 - Diagnostics/Troubleshooting 7 - 3
    Section 7-3
    Frequently Asked  Questions - FAQ
    7-3-1 Error Message(s) During Start Up
    Table 7-2    Error Message(s) During Start Up
    ERROR MESSAGEDESCRIPTIONWHAT TO DO
    This message, displayed when starting Vivid 7, 
    indicates that one or more of the installed options 
    soon will expire. Typical,  this will be the case when 
    a demo option is about to expire. If you want to keep the option(s) active, it is time to 
    order the option(s) now. Contact your GE sales 
    representative for more details. 
    If you don’t need the option(s), you can ignore this 
    message.
    You must select 
    OK
     to continue the start up of 
    Vivid 7.
    This message indicates that all of the installed 
    options have expired. Typical, this will be the case 
    when all demo options have expired. Contact your GE sales representative for more 
    details. 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    7 - 4 Section 7-4 - Service Safety Considerations
    Section 7-4
    Service Safety Considerations
    DANGERDANGEROUS VOLTAGES, CAPABLE OF  CAUSING DEATH, ARE PRESENT IN 
    THIS EQUIPMENT. USE  EXTREME CAUTION WHEN HANDLING, TESTING AND 
    ADJUSTING.
    WARNINGWARNINGIF THE COVERS ARE REMOVED FROM AN OPERATING Vivid 7, SOME METAL 
    SURFACES MAY BE WARM ENOUGH TO POSE A POTENTIAL HEAT HAZARD IF 
    TOUCHED, EVEN WHILE IN SHUT DOWN MODE.
    WARNINGWARNINGEXPLOSION WARNING
    DO NOT OPERATE THE EQUIPMENT IN AN  EXPLOSIVE ATMOSPHERE. OPERATION 
    OF ANY ELECTRICAL EQUIPMENT IN SUCH AN ENVIRONMENT CONSTITUTES A 
    DEFINITE SAFETY HAZARD.
    CAUTIONUse all Personal Protection Equi pment (PPE) such as gloves, safety shoes, safety glasses, and 
    kneeling pad, to reduc e the risk of injury. 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 7 - Diagnostics/Troubleshooting 7 - 5
    Section 7-5
    Gathering Troubleshooting Data
    7-5-1 Overview
    There may be a time when it would be advantageous to  capture trouble images and system data (logs) 
    for acquisition through remote diagnos tics (InSite) or to be sent back to the manufacturer for analysis.
    There are different options to acquire this data that would give different results.
    7-5-2 Collect Vital  System Information
    The following information is necessary in order to properly analyze data or images being reported as a 
    malfunction or being returned to the manufacturer:
    Product Name = VIVID 7
    From the Config (F2
    )>About  screen:
    • Applications Software
    - Software Version
    - Software Part Number
    • System Software
    - System Software Revision
    - System Software CD Part Number
    • Back-End Processor (BEP) Model
    NOTE: The contents on the About tab will va ry depending on the software version
    Figure 7-1   The About Screen 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    7 - 6 Section 7-5 - Gathering Troubleshooting Data
    7-5-3 Collect a Trouble Image with Logs
    If the system should malfunction,  press the Alt-D keys simultaneously . This will collect a screen capture 
    of the monitor, system presets and several lo g files in a date and time stamped zip file.
    NOTE: Power Supply and Temperature logs are  not currently being updated by the Vivid 7.
    This Alt-D function is available at all times.
    When Alt-D is pressed, a menu  box appears that allows for;
    • a place to enter a description of the issue
    • a check box to indi cate a System lockup
    • a choice to Export to a pre-formatted MOD, CD-RW or save to the Export directory D: drive (for  remote viewing through InSite).
    NOTE: NOTE: You  MUST select one of the available devices as the destination device if it is to be 
    different than the default Expor t directory on the hard drive.
    The screen captur e is a bitmap which eliminates the po ssibility of artifacts from compression.
    Figure 7-2   ALT-D Dialog Box 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 7 - Diagnostics/Troubleshooting 7 - 7
    Section 7-6
    Diagnostics Tools
    7-6-1 Diagnostic Procedure Summary
    7-6-1-1 Purpose of this Sub-Section These software based diagnostics tools are available on Vivid 7:
    •Common Service Diagnostic Interface (GSUI)  and
    • Vivid 7’s Front-End System Test
    To be able to use these Service Tools, the Back-End  Processor must be up and running to some extent. 
    If the Back-End Processor doesn’t  starts, it’s not possible to use the software service tools! 
    A typical approach would be to first log on to the  GSUI and run the tests there. If the Acquisition test 
    fails, it is time to switch to Vivid 7’s Front-End System Test.
    7-6-1-2 List of Service Tools Available via th e Common Service Diagnostic Interface (GSUI)
     
    Table 7-3    List of Service Tools Available via GSUI
    Software ToolDescriptionOperator
    Admin-
    istratorExternal 
    ServicePage 
    Number
    Error logs Vivid 7 Error Logs x x x
    7-11
    Diagnostics Vivid 7 Diagnostics x x x
    7-26
    Image Quality [Not Populated] - - -
    7-32
    CalibrationDC Offset Calibration-xx7-33
    Configuration[Not Populated]---7-35
    UtilitiesVivid 7 Service Utilitiesxxx7-36
    Replacement [Not Populated] - - -
    7-46
    PM [Not Populated] - - -
    7-47 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    7 - 8 Section 7-6 - Diagnostics Tools
    7-6-2 Common Service Diagnostic Interface (GSUI)
    NOTE: It is not necessary to connect  a PC to Vivid 7 to use the GSUI.
    7-6-2-1 Log-in to GSUI Double-click the button with the wren ch icon (  ) in the status bar at the bottom of the video screen. 
    This button links the user or the Field En gineer (FE) to the Service Login Screen.
     
    Log on as External Service  with password: gogems (all lowercase letters).
    Figure 7-3   Service Login Screen 
    						