    Chapter 7 - Diagnostics/Troubleshooting 7 - 29
    7-6-5-4 Control Frame Contains the user interface elements used for:
    • Diagnostic control
    • Operator feedback
    a.) Execute ButtonThis button has two modes - each with appropriate text:
    * Execute - to start the diagnostic, and
    * Abort - to stop a diagnostic
    b.) Loop Count This is an editable text field th at will only accept numeric values with 4 digits or less. When the 
    button described in  Execute Button step a (above) is configured as an “execute” button and 
    pressed, the loop count field will be queried to  determine the number of times to execute the 
    diagnostic.
    c.) Progress Indicator Displays a graphical progress indicator for the user.
    d.) Short Text Message Displays either a starting message or aborting  message, as well as the diagnostic completion 
    status.
    e.) Control Frame Background Color Initially gray, the Control Frame background colo r changes upon completion of a diagnostic 
    to indicate completion status.
    * Fail = Red
    *Pass = Green 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    7 - 30 Section 7-6 - Diagnostics Tools
    7-6-5-4      Control Frame (cont’d) Examine the log in the status frame to find  the test(s) that failed. (In the example in Figure 7-33, all the 
    tests failed.)
    Figure 7-33   User Interface Diagnostic Screen,  Test All Failed 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 7 - Diagnostics/Troubleshooting 7 - 31
    7-6-5-4      Control Frame (cont’d) In Figure 7-34 , the test passed. The green background color in the Control Frame indicates that the test 
    passed.
    7-6-5-5 Common Diagnostics  See: Section 7-7 Common Diagnostics on page 7-48.Figure 7-34   User Interface Diagnostic Screen, 
    Test All Passed 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    7 - 32 Section 7-6 - Diagnostics Tools
    7-6-6 Image Quality
    Field is not populated. Figure 7-35   Image Quality Page 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 7 - Diagnostics/Troubleshooting 7 - 33
    7-6-7 Calibration
    This screen gives you access to:
    • DC Offset calibration
    • Reset DC-Offset cal (calibration) Figure 7-36   Calibration 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    7 - 34 Section 7-6 - Diagnostics Tools
    7-6-7-1 DC Offset Calibration DC Offset calibration also called “Beamformer Calibra tion” or “Front-End Alignment” is performed to 
    calibrate each ADC channel of the two Beamformer  (BF) board inputs. The system calibrates the 
    Beamformer ADCs to zero output when there is no signal on the input.
    These bias voltages are stored on  hard disk in Back-End Processor. 
    You can either access  DC Offset Calibration via the screen displayed above, or as described in 
    Section 6-7 Front-End Alignment Procedure (Beamformer Calibration) on page 6-6 . 
    The user interfaces are different, but they use the sa me adjustment software, so the result from the two 
    access methodes are equal.
    1.) Select DC Offset Calibration
    2.) Wait for the calibration to take place.
    3.) When finished, select Show Graph.
    NOTE: Wavy line signifies system has  been calibrated. A Straight line signifies that the system has not 
    been calibrated and can be the cause of image quality issues.
    NOTE: The LEDs displayed in the below individual b oard sections are relevant when the system is in 
    the default setting (2D-mode). Figure 7-37   DC Calibration of the Beamformer AD-converters Passed 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 7 - Diagnostics/Troubleshooting 7 - 35
    7-6-8 Configuration
    Use the Configuration Screen to add or edit Software Options Strings, alphanumeric strings to 
    enable Software Options (the password(s) used for enabling optional functions).  Figure 7-38   Configuration Page 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    7 - 36 Section 7-6 - Diagnostics Tools
    7-6-9 Utilities
    The Utilities page contains  several miscellaneous tools:
    • Windows 2000 Utilities
    - Disk Usage
    - IP Configuration
    - Network Status
    - Windows 2000 Services
    - User Accounts
    - Shared Resources
    - System Shutdown
    - Disk Defragmenter
    • Scanner Utilities
    - System Shutdown
    - Set Debug Utility
    - Reset Database
    - Clean Userdefs
    - Resource File Utility
    • iLink Configuration Utility
    • ProDiags Configuration Utility
    Figure 7-39   Utilities/Tools Page (example) 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 7 - Diagnostics/Troubleshooting 7 - 37
    7-6-9-1 Windows 2000 Utilities
    Disk Usage
    The Disk Usage page displays an ov erview of Vivid 7’s disk usage.
    Figure 7-40   Disk Usage 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    7 - 38 Section 7-6 - Diagnostics Tools
    7-6-9-1      Windows 2000 Utilities (cont’d)
    IP Configuration
    The IP Configuration page gives you an overview of  Vivid 7’s Internet Protocol (IP) configuration.
    Figure 7-41   IP Configuration 
    						