GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 7 - Diagnostics/Troubleshooting 7 - 29 7-6-5-4 Control Frame Contains the user interface elements used for: • Diagnostic control • Operator feedback a.) Execute ButtonThis button has two modes - each with appropriate text: * Execute - to start the diagnostic, and * Abort - to stop a diagnostic b.) Loop Count This is an editable text field th at will only accept numeric values with 4 digits or less. When the button described in Execute Button step a (above) is configured as an “execute” button and pressed, the loop count field will be queried to determine the number of times to execute the diagnostic. c.) Progress Indicator Displays a graphical progress indicator for the user. d.) Short Text Message Displays either a starting message or aborting message, as well as the diagnostic completion status. e.) Control Frame Background Color Initially gray, the Control Frame background colo r changes upon completion of a diagnostic to indicate completion status. * Fail = Red *Pass = Green
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 7 - 30 Section 7-6 - Diagnostics Tools 7-6-5-4 Control Frame (cont’d) Examine the log in the status frame to find the test(s) that failed. (In the example in Figure 7-33, all the tests failed.) Figure 7-33 User Interface Diagnostic Screen, Test All Failed
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 7 - Diagnostics/Troubleshooting 7 - 31 7-6-5-4 Control Frame (cont’d) In Figure 7-34 , the test passed. The green background color in the Control Frame indicates that the test passed. 7-6-5-5 Common Diagnostics See: Section 7-7 Common Diagnostics on page 7-48.Figure 7-34 User Interface Diagnostic Screen, Test All Passed
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 7 - 32 Section 7-6 - Diagnostics Tools 7-6-6 Image Quality Field is not populated. Figure 7-35 Image Quality Page
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 7 - Diagnostics/Troubleshooting 7 - 33 7-6-7 Calibration This screen gives you access to: • DC Offset calibration • Reset DC-Offset cal (calibration) Figure 7-36 Calibration
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 7 - 34 Section 7-6 - Diagnostics Tools 7-6-7-1 DC Offset Calibration DC Offset calibration also called “Beamformer Calibra tion” or “Front-End Alignment” is performed to calibrate each ADC channel of the two Beamformer (BF) board inputs. The system calibrates the Beamformer ADCs to zero output when there is no signal on the input. These bias voltages are stored on hard disk in Back-End Processor. You can either access DC Offset Calibration via the screen displayed above, or as described in Section 6-7 Front-End Alignment Procedure (Beamformer Calibration) on page 6-6 . The user interfaces are different, but they use the sa me adjustment software, so the result from the two access methodes are equal. 1.) Select DC Offset Calibration 2.) Wait for the calibration to take place. 3.) When finished, select Show Graph. NOTE: Wavy line signifies system has been calibrated. A Straight line signifies that the system has not been calibrated and can be the cause of image quality issues. NOTE: The LEDs displayed in the below individual b oard sections are relevant when the system is in the default setting (2D-mode). Figure 7-37 DC Calibration of the Beamformer AD-converters Passed
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 7 - Diagnostics/Troubleshooting 7 - 35 7-6-8 Configuration Use the Configuration Screen to add or edit Software Options Strings, alphanumeric strings to enable Software Options (the password(s) used for enabling optional functions). Figure 7-38 Configuration Page
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 7 - 36 Section 7-6 - Diagnostics Tools 7-6-9 Utilities The Utilities page contains several miscellaneous tools: • Windows 2000 Utilities - Disk Usage - IP Configuration - Network Status - Windows 2000 Services - User Accounts - Shared Resources - System Shutdown - Disk Defragmenter • Scanner Utilities - System Shutdown - Set Debug Utility - Reset Database - Clean Userdefs - Resource File Utility • iLink Configuration Utility • ProDiags Configuration Utility Figure 7-39 Utilities/Tools Page (example)
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 7 - Diagnostics/Troubleshooting 7 - 37 7-6-9-1 Windows 2000 Utilities Disk Usage The Disk Usage page displays an ov erview of Vivid 7’s disk usage. Figure 7-40 Disk Usage
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 7 - 38 Section 7-6 - Diagnostics Tools 7-6-9-1 Windows 2000 Utilities (cont’d) IP Configuration The IP Configuration page gives you an overview of Vivid 7’s Internet Protocol (IP) configuration. Figure 7-41 IP Configuration