GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 137 3-13-2 Product Locator Installation Card NOTE: The Product Locator Installa tion Card shown may not be the same as the provided Product Locator card. From the factory, a sheet with five Product Locator cards for transportation and one for Installation are included. Figure 3-147 Product Locator Installation Card (Example)
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 138 Section 3-13 - Installation Paperwork 3-13-3 Post Deli very Check List Before shipment from the factory, this product has been thorough ly tested and visually inspected. The product is a fine tuned electronic instrument and should be treated properly during transportation. To learn about any issues that are discovered at the reception of the package or when unpacking and installing the prod uct, the Post Delivery Chec klist has been introduced. • Please complete the form on the back of this remi nder, the Post Delivery Checklist, and mail or fax it to our System Test department as soon as y ou have unpacked, installed and verified the product. • If you prefer to use E-mail, you can report your findings in an E-mail and send it to: [email protected] 3-13-3-1 Contact Information for the Post Delivery Checklist GE Vingmed Ultrasound AS P. O. Box 141 N-3191 Horten, NORWAY Phone: +47 3302 1100 Fax: +47 3302 1354 E-mail: [email protected]
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 139 Figure 3-148 Post Delivery Checklist GE Medical Systems Document Number: FC250559 01 POST DELIVERY CHECK LIST Complete this form and send it to: GE VINGMED ULTRASOUND AS FAX NO.: +47 3302 1354 EMAIL: [email protected] ATTENTION: SYSTEM TEST DEPARTMENT SYSTEM TESTER (USE BLOCK LETTERS): ____________________________________________ POST DELIVERY CHECK LIST FOR VIVID______________ SERIAL NO:_____________________ SUBJECTOKFAILURECOMMENTS PACKING LOOSE SCREWS / LOOSE HARDWARE OVERALL APPEARANCE SYSTEM DOCUMENTATION FUNCTIONAL TEST 2D IMAGE M-MODE DOPPLER SPECTRAL COLOR DOPPLER CONFIGURATION PERIPHERALS ECHOPAC PC PROBES OTHER CORRECTIVE ACTIONS REPLACED BOARDS/ PROBES DOCUMENTED BY S/N MISSING PARTS CONTACT AND SIGNATURE DO YOU WANT THE RESPONSIBLE SYSTEM TESTER TO CONTACT YOU? YES: NO:DATE: ___________________ YOUR NAME (BLOCK LETTERS):SIGNATURE:YOUR PHONE NUMBER: FAX PAGE No: ______ OF _________
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 140 Section 3-13 - Installation Paperwork This page was intentionally left blank.
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 4 - Functional Checks 4 - 1 Chapter 4 Functional Checks Section 4-1 Overview 4-1-1 Purpose of Chapter 4 This chapter provides procedures for quickly checking major functions of the Vivid 7 scanner, diagnostics by using the built-in service software, and power supply adjustments. 4-1-2 Contents in Chapter 4 4-1-3 Special Equipment Required • An empty (blank) MO Disk; - P/N: 066E0519 (5.4 GB) - P/N: 066E0514 (8.6 GB) - P/N: 066E0516 (9.1 GB) • Un-used/un-formatted CD-R discs • Un-used/un-formatted DVD-R discs • ECG Pads • ECG Harness, CABLE ECG MARQ . AHA/AMERICA, P/N:164L0025 - CABLE ECG MARQ. AHA/ AMERICA, P/N:164L0025 - LEADWIRES ECG MARQ. AHA/ AMERICA, P/N: 164L0027 or - CABLE ECG MARQ. IEC / EU + AS, P/N:164L0026 - LEADWIRES ECG MARQ. IEC / EU + AS, P/N:164L0028 • At least one probe (ideally you should check a ll the probes used on the system by the site.) See Probes on page 9-80, for an overview of available probes. Table 4-1 Contents in chapter 4 SECTIONDESCRIPTIONPAGE NUMBER 4-1 Overview 4-1 4-2 General Procedures 4-2 4-3 Functional Checks 4-28 4-4Application Turnover Check List4-82 4-5Power Supply4-82 4-6Site Log4-82
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 4 - 2 Section 4-2 - General Procedures Section 4-2 General Procedures 4-2-1 Overview Some procedures are used more often than other. The intention with this section is to keep the most used procedures in one place. NOTICESYSTEM REQUIRES ALL COVERS Operate this unit only when all board covers and frame panels are securely in place. The covers are required for safe operation, good system performance and cooling purposes. WARNINGWARNINGENERGY CONTROL AND POWER LOCKOUT FOR Vivid 7 When servicing parts of the system where t here is exposure to voltage greater than 30 Volts: Unplug the system Maintain control of the system power plug There are no test points to verify isolation, you must wait for at least 20 seconds for capacitors to discharge Beware that the AC Control Box, Front End Processor and Back End Processor may be energized even if the power is turned off when the cord is still plugged into the AC Outlet. Signed Date TA G & LOCKOUT
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 4 - Functional Checks 4 - 3 4-2-2 Power ON/ Boot UP 4-2-2-1 Connect Mains Power to the Scanner NOTE: Do not cycle the Circuit Breake r ON-OFF-ON in less than five -5- seconds. When turning OFF the Circuit Breaker, the system should de-energize completely before turning the circuit breaker ON. NOTE: When turning on a system from standby mode, it ta kes a few seconds before it responds. Do not push the On/off button again during this period. A second push will initiate a full shutdown. NOTE: If the unit has been in the OFF condition for an extended period of time, (3 to 5 days or more), the unit may not boot, or may beep when turned on. Follow These Steps to Connect Mains Power to the Unit Connecting the Vivid 7 ultrasound unit involves pre liminary checks of the power cord, voltage level and compliance with electrical safety requirements. 1.) Ensure that the wall outlet is of appropriate type, and that the pow er Circuit Breaker is turned off. 2.) Uncoil the power cable, allowing sufficient sl ack so that the unit can be moved slightly. 3.) Verify that the power cabl e is without any visible scratches or any sign of damage. 4.) Verify that the on-site mains volt age is within the limits indicated on the rating label near the Circuit Breaker at the rear of the unit. DANGERALWAYS CONNECT THE UNIT TO A FIXE D POWER SOCKET WHICH HAS THE PROTECTIVE GROUND ING CONNECTOR. DANGERNEVER USE A THREE-TO -TWO PRONG ADAPTER; THIS DEFEATS THE SAFETY GROUND. DANGERENSURE THAT THE POWER CORD AND PLUG ARE INTACT AND THAT THE POWER PLUG IS THE PROPER HOSPITA L-GRADE TYPE (WHERE REQUIRED). CAUTIONSYSTEM REQUIRES ALL COVERS OPERATE THIS UNIT ONLY WHEN ALL BOARD COVERS AND FRAME PANELS ARE SECURELY IN PLACE. THE COVERS ARE REQUIR ED FOR SAFE OPERATION, GOOD SYSTEM PERFORMANCE AND COOLING PURPOSES. NOTICEUse only power supply cords, cables and plugs pr ovided by or designated by GE Healthcare.
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 4 - 4 Section 4-2 - General Procedures 4-2-2-1 Connect Mains Power to the Scanner (cont’d) 5.) Connect the Power Cable’s female plug to the Power Inlet at the rear of the unit. 6.) Lock the plug in position with the Retaining Clasp. 7.) Verify that the Mains Power Circuit Breaker is in OFF position, if not, switch it OFF. 8.) Connect the Power Cable’s other end (male plug ) to a hospital grade mains power outlet with the proper rated voltage, and the unit is ready for Power ON/Bootup. . Figure 4-1 Female power plug with retaini ng clasp off (to the left) and on (to the right) Figure 4-2 Mains Power Circuit Breaker is in OFF position Retaining Clasp offRetaining Clasp on OFF Position
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 4 - Functional Checks 4 - 5 4-2-2-2 Turn Unit On 1.) Switch ON the Mains Power Circuit Breaker at the rear of the unit. You should hear a “click” from the relays in the AC Power/AC Controller and the unit is ready to boot. NOTE: If the unit has been in the OFF co ndition for an extended period of time, (3 to 5 days or more), the unit may not boot, or may beep when turned on. 2.) Press once on the ON/STANDBY key on the Operator Panel to boot the unit. During a normal boot, you may observe that: a.) The unit’s ventilation fan star ts at full speed, but slows down af ter a few seconds (listen to the fan noise). b.) Power is distributed to the peripherals, Operator Panel (Console), Monitor, Front-End Processor and Back-End Processor. c.) Back-End Processor an d rest of Scanner starts with the sequence listed in the next steps: d.) Back-End Processor is turned ON and starts to load the software. e.) The Start Screen (Vivid) is displayed on the monitor Figure 4-3 Switch Mains Power Circuit Breaker ON . Figure 4-4 Press the On/Standby key on the Operator Panel once to boot the unit Mains Power Circuit Breaker in ON position
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 4 - 6 Section 4-2 - General Procedures 4-2-2-2 Turn Unit On (cont’d) f.) A start-up bar indicating the time used fo r software loading, is displayed on the monitor. g.) The software initiates and se ts up the Front-End electronics and the rest of the instrument. h.) The backlight in the keyboard is lit. i.) As soon as the software has been loaded, a 2D screen is displayed on the screen. NOTE: Total time used for start-up is typical three minutes or less, depending on software version and Vivid 7 model. Figure 4-5 Start-up bar Figure 4-6 2D Screen on the display. The Start-Up Screen (Splash Screen) Is di fferent, depending on software version and Vivid 7 model. Vivid 7 Dimension with softwar e version v6.1 or higher (upper, left), Vivid 7 Dimens ion with software version v4.0 to v6.1.x, (mi ddle) and Vivid 7 with software versions v3.x.x and lower (lower, right). 2D SCREEN ON UNITS WITH SOFTWARE VERSION v4.x, v5x, v6x and v7.x. 2D SCREEN ON UNITS WITH SOFTWARE VERSIONS UP TO VERSION v3.x.