    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 3 - System Setup 3 - 137
    3-13-2 Product Locator  Installation Card
    NOTE: The Product Locator Installa tion Card shown may not be the same as the provided Product 
    Locator card.
    From the factory, a sheet with five Product Locator  cards for transportation and one for Installation are 
    included. 
    Figure 3-147   Product Locator Installation Card (Example) 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    3 - 138 Section 3-13 - Installation Paperwork
    3-13-3 Post Deli very Check List
    Before shipment from the  factory, this product has been thorough ly tested and visually inspected. The 
    product is a fine tuned electronic instrument and  should be treated properly during transportation.
    To learn about any issues that are discovered at the reception of the package or when unpacking and 
    installing the prod uct, the Post Delivery Chec klist has been introduced.
    • Please complete the form on the back of this remi nder, the Post Delivery Checklist, and mail or fax 
    it to our System Test department as soon as y ou have unpacked, installed and verified the product.
    • If you prefer to use E-mail, you can report your findings in an E-mail and send it to:
    [email protected] 
    3-13-3-1 Contact Information for the Post Delivery Checklist GE Vingmed Ultrasound AS
    P. O. Box 141
    N-3191 Horten, NORWAY
    Phone: +47 3302 1100
    Fax: +47 3302 1354
    E-mail: [email protected]  
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 3 - System Setup 3 - 139
    Figure 3-148   Post Delivery Checklist
    GE Medical Systems
    Document Number:  FC250559 01
    POST DELIVERY CHECK LIST
    Complete this form and send it to:
    GE VINGMED ULTRASOUND AS
    FAX NO.: +47 3302 1354
    EMAIL:  [email protected]
    ATTENTION: SYSTEM TEST DEPARTMENT
    SYSTEM TESTER (USE BLOCK LETTERS): ____________________________________________ 
    POST DELIVERY CHECK LIST FOR VIVID______________ SERIAL NO:_____________________
    SUBJECTOKFAILURECOMMENTS
    PACKING
    LOOSE SCREWS /
    LOOSE HARDWARE
    OVERALL APPEARANCE
    SYSTEM 
    DOCUMENTATION
    FUNCTIONAL TEST
    2D IMAGE
    M-MODE
    DOPPLER SPECTRAL
    COLOR DOPPLER
    CONFIGURATION
    PERIPHERALS
    ECHOPAC PC
    PROBES
    OTHER
    CORRECTIVE ACTIONS
    REPLACED BOARDS/
    PROBES
    DOCUMENTED BY S/N
    MISSING PARTS
    CONTACT AND SIGNATURE
    DO YOU WANT THE RESPONSIBLE SYSTEM TESTER TO CONTACT YOU?     YES:     NO:DATE: ___________________ 
    YOUR NAME (BLOCK LETTERS):SIGNATURE:YOUR PHONE NUMBER:
    FAX PAGE No: ______ OF _________ 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    3 - 140 Section 3-13 - Installation Paperwork
    This page was intentionally left blank. 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 4 - Functional Checks 4 - 1
    Chapter 4 
    Functional Checks
    Section 4-1  
    Overview
    4-1-1 Purpose of Chapter 4
    This chapter provides procedures for quickly checking major functions of the Vivid 7 scanner, 
    diagnostics by using the built-in service software, and power supply adjustments. 
    4-1-2 Contents in Chapter 4
    4-1-3 Special Equipment Required
    • An empty (blank) MO Disk; - P/N: 066E0519 (5.4 GB)
    - P/N: 066E0514 (8.6 GB)
    - P/N: 066E0516 (9.1 GB)
    • Un-used/un-formatted CD-R discs
    • Un-used/un-formatted DVD-R discs
    • ECG Pads
    • ECG Harness, CABLE ECG MARQ . AHA/AMERICA, P/N:164L0025
    - CABLE ECG MARQ. AHA/ AMERICA, P/N:164L0025
    - LEADWIRES ECG MARQ. AHA/ AMERICA, P/N: 164L0027
    or
    - CABLE ECG MARQ. IEC /  EU + AS, P/N:164L0026
    - LEADWIRES ECG MARQ. IEC / EU + AS, P/N:164L0028
    • At least one probe (ideally you should check a ll the probes used on the system by the site.)
    See  Probes  on page  9-80, for an overview of available probes.
    Table 4-1    Contents in chapter 4
    SECTIONDESCRIPTIONPAGE NUMBER
    4-1 Overview 4-1
    4-2 General Procedures 4-2
    4-3 Functional Checks 4-28
    4-4Application Turnover Check List4-82
    4-5Power Supply4-82
    4-6Site Log4-82 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    4 - 2 Section 4-2 - General Procedures
    Section 4-2
    General Procedures
    4-2-1 Overview
    Some procedures are used more often than other. The  intention with this section is to keep the most 
    used procedures in one place.
    NOTICESYSTEM REQUIRES ALL COVERS
    Operate this unit only when all board covers and frame panels are securely in place. The covers are 
    required for safe operation, good system performance and cooling purposes.
    WARNINGWARNINGENERGY CONTROL AND POWER LOCKOUT FOR Vivid 7
    When servicing parts of the system where t here is exposure to voltage greater than 30 
    Volts: Unplug the system Maintain control of the system power plug There are no test 
    points to verify isolation, you must wait for at least 20 seconds for capacitors to 
    discharge Beware that the AC Control Box, Front End Processor and Back End 
    Processor may be energized even if the power is turned off when the cord is still 
    plugged into the AC Outlet.
    Signed Date
    TA G 
    & 
    LOCKOUT 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 4 - Functional Checks 4 - 3
    4-2-2 Power ON/ Boot UP
    4-2-2-1 Connect Mains Power to the Scanner
        
    NOTE: Do not cycle the Circuit Breake r ON-OFF-ON in less than five -5- seconds.   When turning OFF 
    the Circuit Breaker, the system should de-energize  completely before turning the circuit breaker 
    ON.
    NOTE: When turning on a system from standby mode, it ta kes a few seconds before it responds. Do not push 
    the On/off button again during  this period. A second push will initiate a full shutdown. 
    NOTE: If the unit has been in the OFF condition for an extended period of time, (3 to 5 days or more),  the unit may not boot, or may beep when turned on. 
    Follow These Steps to Connect Mains Power to the Unit
    Connecting the Vivid 7 ultrasound unit involves pre liminary checks of the power cord, voltage level and 
    compliance with electrical safety requirements. 
    1.) Ensure that the wall outlet is of  appropriate type, and that the pow er Circuit Breaker is turned off.
    2.) Uncoil the power cable, allowing sufficient sl ack so that the unit can be moved slightly.
    3.) Verify that the power cabl e is without any visible scratches or any sign of damage.
    4.) Verify that the on-site mains volt age is within the limits indicated on the rating label near the Circuit 
    Breaker at the rear of the unit.
    DANGERALWAYS CONNECT THE UNIT TO A FIXE D POWER SOCKET WHICH HAS THE 
    PROTECTIVE GROUND ING CONNECTOR. 
    DANGERNEVER USE A THREE-TO -TWO PRONG ADAPTER; THIS DEFEATS THE 
    SAFETY GROUND. 
    DANGERENSURE THAT THE POWER CORD AND  PLUG ARE INTACT AND THAT THE 
    POWER PLUG IS THE PROPER HOSPITA L-GRADE TYPE (WHERE REQUIRED). 
    CAUTIONSYSTEM REQUIRES ALL COVERS
    OPERATE THIS UNIT ONLY WHEN ALL BOARD COVERS AND FRAME PANELS ARE 
    SECURELY IN PLACE. THE COVERS ARE REQUIR ED FOR SAFE OPERATION, GOOD SYSTEM 
    PERFORMANCE AND COOLING PURPOSES.
    NOTICEUse only power supply cords, cables and plugs pr ovided by or designated by GE Healthcare. 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    4 - 4 Section 4-2 - General Procedures
    4-2-2-1      Connect Mains Power to the Scanner (cont’d) 5.) Connect the Power Cable’s female plug to the Power Inlet at the rear of the unit.
    6.) Lock the plug in position  with the Retaining Clasp.
    7.) Verify that the Mains Power Circuit Breaker  is in OFF position, if not, switch it OFF.
    8.) Connect the Power Cable’s other end (male plug ) to a hospital grade mains power outlet with the 
    proper rated voltage, and the unit is ready for Power ON/Bootup.
    .
    Figure 4-1   Female power plug with retaini ng clasp off (to the left) and on (to the right)
    Figure 4-2   Mains Power Circuit  Breaker is in OFF position
    Retaining 
    Clasp offRetaining 
    Clasp on
    OFF Position 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 4 - Functional Checks 4 - 5
    4-2-2-2 Turn Unit On 1.) Switch ON the Mains Power Circuit  Breaker at the rear of the unit.
    You should hear a “click” from the relays in the AC  Power/AC Controller and the unit is ready to boot.
    NOTE: If the unit has been in the OFF co ndition for an extended period of time, (3 to 5 days or more), the unit 
    may not boot, or may beep when turned on. 
    2.) Press once on the  ON/STANDBY
     key on the Operator Panel to boot the unit.
    During a normal boot, you may observe that:  a.) The unit’s ventilation fan star ts at full speed, but slows down af ter a few seconds (listen to the 
    fan noise).
    b.) Power is distributed to the peripherals, Operator Panel (Console), Monitor, Front-End  Processor and Back-End Processor.
    c.) Back-End Processor an d rest of Scanner starts with the sequence listed in the next steps:
    d.) Back-End Processor is turned ON  and starts to load the software.
    e.) The Start Screen (Vivid) is displayed on the monitor Figure 4-3   Switch Mains 
    Power Circuit Breaker ON 
    .
    Figure 4-4   Press the On/Standby key on the Operator Panel once to boot the unit
    Mains Power Circuit Breaker in ON position 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    4 - 6 Section 4-2 - General Procedures
    4-2-2-2      Turn Unit On (cont’d) f.) A start-up bar indicating the time used fo r software loading, is displayed on the monitor.
    g.) The software initiates and se ts up the Front-End electronics and the rest of the instrument.
    h.) The backlight in  the keyboard is lit.
    i.) As soon as the software has been loaded, a 2D screen is displayed on the screen.
    NOTE: Total time used for start-up is typical three  minutes or less, depending on software version and 
    Vivid 7 model. Figure 4-5   Start-up bar
    Figure 4-6   2D Screen on the display.
    The Start-Up Screen (Splash Screen) Is di fferent, depending on software version and Vivid 7 model. Vivid 7 Dimension with softwar e version v6.1 or 
    higher (upper, left), Vivid 7 Dimens ion with software version v4.0 to v6.1.x, (mi ddle) and Vivid 7 with software versions v3.x.x and lower (lower, right).
    2D SCREEN ON 
    UNITS WITH 
    SOFTWARE VERSION 
    v4.x, v5x, v6x and v7.x.
    2D SCREEN ON UNITS 
    WITH SOFTWARE 
    VERSIONS UP TO 
    VERSION v3.x. 
    						