    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 3 - System Setup 3 - 87
    3-9-18-3 Examination List Screen Figure 3-91   Examination List Screen
    Table 3-19    Examination List Screen
    ITEMVIVID 7 FIELD NAMEDICOM TAG NUMBERDICOM TAG NAME
    1 REFERRAL REASON (0010,2180) ADDITIONAL PATIENT’S HISTORY
    2 DIAGNOSIS (0008,103E) SERIES DESCRIPTION
    1
    2 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    3 - 88 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x
    3-9-18-4 Worklist Search Screen Figure 3-92   Worklist Search Screen
    Table 3-20    Worklist Search Screen
    ITEMVIVID 7 FIELD NAMEDICOM TAG NUMBERDICOM TAG NAME
    1 LAST NAME, FIRST NAME (0010,0010) PATIENT’S NAME
    2 SEX (0010,0040) PATIENT’S SEX
    3 ACCESSION (0008,0050) ACCESSION NUMBER
    4BIRTHDATE(0010,0030)PAT I E N T ’ S  B I RT H D AT E
    5PATIENT ID(0010,0020)PATIENT ID
    6STUDY ID(0040,1001)REQUESTED PROCEDURE ID
    7 EXAM BETWEEN (0040,0002) SCEDULED PROCEDURE STEP 
    START DATE
    1
    2
    45
    7
    [ACCESSION  FIELD 
    AVAILABLE ON SOME 
    SOFTWARE VERSIONS]
    6
    [ STUDY ID  FIELD 
    AVAILABLE ON SOME 
    SOFTWARE VERSIONS
    (NOT SHOWN HERE)]
    3 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 3 - System Setup 3 - 89
    Section 3-10
    Connectivity Setup - Software v3.x
    NOTE: If connected to a stand-alone network (Peer-to-Peer network with a Vivid 7 scanner, an  EchoPAC PC Workstation and an optional network printer), you should use default delivery 
    settings. 
    3-10-1 Introduction
    To be able to use the network functions when con nected to a hospital network, the scanner must have 
    a proper network address. 
    • Before you can set up the scanner, you need to collect some information. The  Pre-quote 
    Worksheet for DICOM Network Information on page 2-14  can be used for gathering this 
    information. Typical source for this in formation is the network administrator.
    • Follow the steps below to prepare the scanner for use on the network.
    3-10-2 Contents in this Section
    • 3-10-3 Compatib ility on page 3-89
    • 3-10-4 Select TCP/IP Set-up Screen - Software Version v3.x on page 3-90
    • 3-10-5 Set the Scanner’s Network Information - Software Version v3.x on page 3-91
    • 3-10-6 Set the Remote Archive’s Network Information - Software Version v3.x on page 3-92
    • 3-10-7 Save the New Settings on page 3-92
    • 3-10-8 Connect to a DICOM Server in a Network - Software Version v3.x on page 3-93
    • 3-10-9 HL7 Communication Setup - Software v3.x on page 3-98
    • 3-10-10 Query/Retrieve (Q/R) Setup for Software v3.x on page 3-102
    3-10-3 Compatibility
    Vivid 7 with Software Version v3.3.x (System  Software v1.5.1) can communicate with: 
    • EchoPAC PC with software version v3.3.x
    • EchoPAC PC with software version v4.x
    • EchoPAC PC with software version v5.x
    • EchoPAC PC with software version v6.x
    • EchoPAC PC with software version v7.x
    • Image Vault 3
    • Image Vault 4
    • EchoServer 7 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    3 - 90 Section 3-10 - Connectivity Setup - Software v3.x
    3-10-4 Select TCP/IP Set-up Scr een - Software Version v3.x
    1.) Press CONFIG (F2) and log on as  adm, see  4-2-4 Log On to the System as ‘ADM’ on page 4-12 .
    2.) If not already selected, select  CONNECTIVITY
     from the bottom row of “buttons” on the screen.
    3.) Select the TCP/IP TAB (it is named Tcpip). 
    .
    Figure 3-93   TCP/IP Set-up screen for Vivid 7, Overview (Example)
    Computer Name: 
    For Vivid 7, this name is on the form:  Vivid7-
    00nnnn , where “00nnnn” is a number (nnnn is the 
    scanner’s serial number). 
    IP SETTINGS:
    IP SETTINGS AREA FOR THE Vivid 7 
    SCANNER. 
    DEFAULT SETUP:
    IP-ADDRESS: 10.0.0.3
    SUBNET MASK: 255.255.255.255
    GATEWAY: 0.0.0.0
    Remote Archive Setup area:
    IP settings area for a Remote Archive.
    Example: EchoServer.
    Default Setup:
    Remote Archive IP-Addr: 10.0.0.4
    Remote Archive Name: EchoPAC7-000001
    The AETitles field and the Port field are onl y available on BT’03 software (and beyond). 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 3 - System Setup 3 - 91
    3-10-5 Set the Scanner’s  Network Information - Software Version v3.x
    In the IP settings area of the screen, enter the;
    1.) IP Address for scanner. (Default IP  Address from factory: 10.0.0.3).
    2.) Subnet Mask for scanner. (Default Subn et Masks from factory: 255.255.255.0).
    3.) IP address for Default Gateway. (Default  Gateways from factory IP address: 0.0.0.0).
    4.) In addition, the AE Title for the scanner mu st be entered in the DICOM server’s setup. 
    .
    Figure 3-94   TCP/IP Set-up for Vivid 7
    IP Address
    Subnet Mask
    Default Gateway
    AETitle
    Port Number 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    3 - 92 Section 3-10 - Connectivity Setup - Software v3.x
    3-10-6 Set the Remote Archive’s Networ k Information - Software Version v3.x
    In the Remote Archive Setup area of the screen (see example in  Figure 3-95), enter the;
    1.) Remote Archive IP address. (Default  IP Address from factory: 10.0.0.4).
    2.) Remote Archive Name. (Default Remote Ar chive Name from factory: ECHOPAC7-000001).
    .
    3-10-7 Save the New Settings
    1.) Press SAVE SETTINGS to save the new settings. 
    .
    2.) The new settings are saved to a common settings file. After a restart, the settings are also included 
    in other screens.
    3.) Restart Vivid 7 to activate the changes. Figure 3-95   TCP/IP Set-up for Vivid 7
    Figure 3-96   Save New TCP/IP settings
    IP Address for the remote archive
    Computer Name for the remote archive
    Save settings 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 3 - System Setup 3 - 93
    3-10-8 Connect to a DICOM Server in  a Network - Software Version v3.x
    In this case the Vivid 7 is configured to work with DICOM servers in a network environment. Images are 
    first saved on the local image buffer on the scanner. At the end of the examination the images are sent 
    to the DICOM server via a DICOM spooler and to  the local database, depending on dataflow.
    This scenario requires that the scanner is configur ed to be connected to DICOM servers as described 
    below.
    3-10-8-1 Overview To work against the DICOM server, the following  information has to be entered in the scanner:
    • The DICOM server IP address.
    • The DICOM server port number.
    • The DICOM server AE title (the server application’s name).
    • Name of Device.
    3-10-8-2 DICOM Server IP Address Setting on the Scanner - Software Version v3.x 1.) Press CONFIG
     (F2) and log on as  adm, see  4-2-4 Log On to the System as ‘ADM’ on page 4-12 .
    2.) Select  CONNECTIVITY
     (in the lower part of the window).
    3.) Select the  DATAFLOW
     tab.
    4.) Select the arrow to the right for the Name field to list all dataflows in a pull-down menu, see  Figure 
    3-97 . 
     
    5.) Select the dataflow you want to configure  from the Dataflow Name pull-down list, see Figure 3-97. 
    Figure 3-97   Select Dataflow
    Arrow 
    Select dataflow from 
    Dataflow Name pull-down 
    menu. 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    3 - 94 Section 3-10 - Connectivity Setup - Software v3.x
    3-10-8-2      DICOM Server IP Address Setting on the Scanner - Software Version v3.x (cont’d)  In Figure 3-98 , the selected flow, Worklist/Local Arch ive - DICOM Server/Int. HD, is shown. 
     
    6.) Select Worklist (so it is highlighted) and then se lect Properties to display the Properties dialog. 
     
    Figure 3-98   Worklist/Local Archive - DICOM Server/Int. HD flow
    Figure 3-99   Select Worklist Properties dialog
    Select Worklist 
    and then select 
    Properties Worklist Properties dialog 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 3 - System Setup 3 - 95
    3-10-8-2      DICOM Server IP Address Setting on the Scanner - Software Version v3.x (cont’d)  7.) Select the IP-Address down-arrow to choose the Worklist Server from the pull-down menu.
    NOTE: It is not possible to change the setting in  the IP-Address field by editing it. See description 
    below, starting with step  8, to change the setting.
    8.) Follow the steps below to change the IP-Address settings: a.) From the Ip-Address pull-down menu, select  to display the IPs dialog.
    b.) Select the server you want to modify fr om the IPs dialog and then select Modify.
    Figure 3-100   DICOM Server’s Properties dialog
    Figure 3-101   Select 
    Figure 3-102   Select the unit to modify
    Modify
    Select server 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    3 - 96 Section 3-10 - Connectivity Setup - Software v3.x
    3-10-8-2      DICOM Server IP Address Setting on the Scanner - Software Version v3.x (cont’d)  c.) Edit the name and/or the IP address of the server, see  Figure 3-103. 
    d.) Select OK to save the new  settings and close the dialog.
    e.) Select OK to close the IPs dialog  to return to the Properties dialog.
    f.) Enter the DICOM server AE Title. This entr y is case sensitive and must match exactly.
    g.) Enter the DICOM server port.
    h.) Select OK to close the Worklist properties dialog and save any changes. Figure 3-103   Edit Name and/or IP address
    Figure 3-104   DICOM Server’s Properties dialog
    Name
    IP address
    AE Title
    Port No 
    						