GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 23 8-7-11 Record Software Versions - Software Version v7.x/v6.x/v5.x/v4.x 1.) Select the About tab. 2.) Record software and hardware versions in Table 8-8 on page 8-23. Figure 8-20 Information available on the About tab Table 8-8 Record Software versions - Software Version v7.x/v6.x/v5.x/v4.x DESCRIPTIONRECORD VERSIONS APPLICATION SW SYSTEM SW PROCESSOR (BEP) MODEL
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 24 Section 8-7 - Software Loading Procedure - Software Version v7.x, v6.x, v5.x and v4.x 8-7-12 Software Loading - Software Version v7.x/v6.x/v5.x/v4.x 8-7-12-1 Overview This software loading procedure covers: • Re-installation of BT’08 softwa re (System Software v3.0.x and Application Software v7.x) on a Vivid 7 unit with BEP4.3 • Re-installation of BT’08 softwa re (System Software v2.1.1 and Application Software v7.x) on a Vivid 7 unit with BEP3.x or BEP4.2 • Re-installation of BT’06 softwa re (System Software v2.x.x and Application Software v6.x) on a Vivid 7 unit. • Re-installation of BT’05 softwa re (System Software v2.x.x and Application Software v5.x) on a Vivid 7 unit. • Re-installation of BT’04 softwa re (System Software v2.x.x and Application Software v4.x) on a Vivid 7 unit. Refer to Table 8-8 Record Software versions - Software Version v7.x/v6.x/v5.x/v4.x on page 8-23 where you recorded the un it’s software versions. The software for Vivid 7 is divided on two CDs: • one CD with the System Software • one CD with the Application Software For a software CD overview, see 8-7-4 Tools Provided With Unit or After an Upgrade on page 8-13 . 8-7-12-2 Introduction to Software Reloading • Starting with System So ftware version v3.0.0 (and Applicat ion Software v7.0.0), the compressed System and Application software has been stored on the Vivid 7’s hard disk drive, making a software reinstallation easier and faster than with previous software. - To reinstall the Application Software, go to: 8-7-12-4 Loading Application Software v7.x on Vivid 7 with BEP4.3 on page 8-27 . - To reinstall both the System Software and the Application Software, go to: 8-7-12-3 Loading Sy stem Software v3.0.x on Vivid 7 with BEP4.3 on page 8-25 . • For Vivid 7 with System Software version v2.x.x , the software CDs must be used if the software needs to be reinstalled. - To reinstall the Application Software, go to: 8-7-12-6 Application SW Loading - Software Version v7.x/v6.x/v5.x/v4.x on page 8-32 - To reinstall both the System Software and the Application Software, go to: 8-7-12-5 Vivid 7 System Software Loading - Application Software Version v7.x/v6.x/v5.x/ v4.x on page 8-29
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 25 8-7-12-3 Loading System Software v3.0.x on Vivid 7 with BEP4.3 NOTE: This procedure describes how to reload software that already has been stored as compressed files on the BEP4.3’s hard disk drive. A CD is not used in this procedure. When using this procedure, you can choose to reload partition C: only, or to erase all partitions. The selection is done later in this procedure. 1.) Reboot the Vivid 7, and press the ESC-key each second. The boot se quence will be interrupted and a boot-menu appears on the screen. There are two entries in the menu. 1.) Normal startup 2.) Recovery Use the arrow keys to select (highlight) your choice. 2.) Select Recovery. 3.) Press Enter to display the System Recovery Menu on the screen, see next page. Figure 8-21 Select Recovery Normal Startup Recovery GNU GRUB version 8.95 (639k lower / 523112X upper memory Use the ^ and v keys to select which entry is highlighted. Press enter to boot the selected OS or “p” to enter a Password to unlock the next set of features.
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 26 Section 8-7 - Software Loading Procedure - Software Version v7.x, v6.x, v5.x and v4.x 8-7-12-3 Loading System Software v3.0.x on Vivid 7 with BEP4.3 (cont’d) 4.) In this step, you can choose if you want to do a co mplete Vivid 7 software installation that overwrites all data, or if you only want to update or reinstall the software on the C: partition (C: drive). Usually it should be OK to select “ B” to update the C: drive only. This will keep Patient Archive and Presets intact. Continue with either step a (if possible) or step b (if you have to): NOTE: If you select “ A” below, all program data, Patient Archive and Presets will be erased. a.) ONLY select “ A” if a “ B” update cannot be used: Select A to do a complete Vivid 7 software installation. The reload process start automatically. When done, a message tells you to initiate Application software installation. Continue with step 5. b.) Select “ B” to update/re-install the software on the C: partition. (Only data on the C: partition will be erased. This will keep Patient Archive and Presets intact.) The reload process starts automatically. When done, a message tells you to initiate Application software installation. 5.) Press Y to confirm. The Vivid 7 reboots and the Application software inst allation starts automatically. When completed, the Vivid 7 reboots normally. • If you selected A above, the User Presets, the System Configurations a nd the local Patient Archive has been reset to the factory default settings. • If you selected B above, the User Presets, the System Configurations a nd the local Patient Archive remains as before the software reload. Figure 8-22 System Recovery Menu NOTICEIf you select “A” in the next step, ALL existing software and data will be erased. If backup has not been performed as described earlier in this manual; ALL DATA like Patient Database, System Configuration and User Configurations (Customer Presets) will be lost. System Recovery Menu A. Full reload (erase all partitions) B. Reload Partition C: only (Keep Archive) C. Reload Application Software R. Restart System Select (A, B, C,,R) _
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 27 8-7-12-4 Loading Application Software v7.x on Vivid 7 with BEP4.3 NOTE: This procedure describes how to reload software that already has been stored as compressed files on the BEP4.3’s hard disk drive. A CD is not used in this procedure. To reload application softwa re, follow this procedure: 1.) Reboot the Vivid 7, and press the ESC-key each second. The boot se quence will be interrupted and a boot-menu appears on the screen. There are two entries in the menu. 1.) Normal startup 2.) Recovery Use the arrow keys to select (highlight) your choice. 2.) Select Recovery. 3.) Press Enter to display the System Recovery Menu on the screen, see next page. Figure 8-23 Select Recovery Normal Startup Recovery GNU GRUB version 8.95 (639k lower / 523112X upper memory Use the ^ and v keys to select which entry is highlighted. Press enter to boot the selected OS or “p” to enter a Password to unlock the next set of features.
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 28 Section 8-7 - Software Loading Procedure - Software Version v7.x, v6.x, v5.x and v4.x 8-7-12-4 Loading Application Soft ware v7.x on Vivid 7 with BEP4.3 (cont’d) 4.) Select “ C. Reload Application SW 5.) Confirm the selection by pressing Y. The system reboots and the Application soft ware installation starts automatically. When completed, the Vivid 7 reboots and starts up normally. The User Presets, the System Configurations and the local Patient Arch ive remains as before the software reload. Figure 8-24 System Recovery Menu NOTICEIf you select “A” in the next step, ALL existing software and data will be erased. If backup has not been performed as described earlier in this manual; ALL DATA like Patient Database, System Configuration and User Configurations (Customer Presets) will be lost. System Recovery Menu A. Full reload (erase all partitions) B. Reload Partition C: only (Keep Archive) C. Reload Application Software R. Restart System Select (A, B, C,,R) _
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 29 8-7-12-5 Vivid 7 System Software Loading - App lication Software Version v7.x/v6.x/v5.x/v4.x NOTE: This procedure covers installation of software from CDs. 1.) Insert the applicable Vivid 7 System Software CD into the CD/DVD drive. 2.) Power down the System, see 4-2-3 Power Shut Down on page 4-7 . 3.) Wait until the LCD’s on the Operator Panel are s witched off and the ON/Standby switch has turned amber. 4.) Power ON the System, see 4-2-3 Power Shut Down on page 4-7 . The scanner will now boot from the CD you inserted. Figure 8-25 BIOS loading (example)
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 30 Section 8-7 - Software Loading Procedure - Software Version v7.x, v6.x, v5.x and v4.x 8-7-12-5 Vivid 7 System Software Loading - Application Software Version v7.x/v6.x/v5.x/v4.x (cont’d) 5.) In this step, you can choose if you want to do a co mplete Vivid 7 installation or if you only want to update or reinstall the software on the C: partition (C: drive). Usually it should be OK to select “B” to update the C: drive only. This will keep Patient Archive and Presets intact. a.) Select B to update/re-install the softw are on the C: partition. (Only data on the C: partition will be erased. This will keep Pati ent Archive and Presets intact.) b.) Use “ A” ONLY if a “ B” update cannot be used: Select A to do a complete Vivid 7 software installation. (All data will be erased.) c.) Press C if you want to Exit. NOTICEIf you select “A” in the next step, ALL existing software and data will be erased. If backup has not been performed as described earlier in this manual; All data like Patient Database, System Configuration and User Configurations (Customer Presets) will be lost. Figure 8-26 Info entry page Press A, B or C to choose the wanted option.
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 31 8-7-12-5 Vivid 7 System Software Loading - Application Software Version v7.x/v6.x/v5.x/v4.x (cont’d) 6.) The Software loader program starts and the below Time Bar Info Screen appears. 7.) Wait for the SW Installation to be completed. (Typic al installation time: 5 minutes). Status bar on the screen indicates process. 8.) You will now be prompted to remove the first CD and restart the unit when the software Installation is completed. Remove the CD, but don’t restart the System this time . The next step is to load the application software. Figure 8-27 Software Loading Underway (Picture distorted here due to photo technical limitations, it should look OK on screen) WARNINGWARNINGDo not interrupt the SW loading at any time.
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 32 Section 8-7 - Software Loading Procedure - Software Version v7.x, v6.x, v5.x and v4.x 8-7-12-6 Application SW Loading - Soft ware Version v7.x/v6.x/v5.x/v4.x 1.) Insert the applicable CD labelled Vivid 7 & EchoPAC PC Application SW . 2.) Keep the ON/Standby switch on the operator panel pressed for a few seconds to power down the scanner. (This is called Forced Shut Down). 3.) Wait until the LCD’s on the Oper ator Panel is switched OFF and the ON/Standby Switch on the Operator Panel has turned yellow. 4.) Power ON the System, see 4-2-2 Power ON/ Boot UP on page 4-3 . 5.) A dialogue box, called Start Ap plication, will appear, select Install SW. 6.) A dialogue box, called Star tloader, will appear, select OK to continue. 7.) A dialogue box will appear, al so named Startloader stating “New System SW installation will start now”, select OK to continue. 8.) A Command prompt window will ap pear, hit any key on the alphanumeric keyboard to continue. 9.) The messages: “Unpacking SW . This will take several minutes.” and “Installing new application software. This will take several minu tes” will be displayed on the screen. 10.) Several dialogue boxe s will appear on the screen during the SW load. 11.) Remove the CD from the CD/DVD drive when prompted. 12.) Press any key to continue. 13.) The Scanner will now pow er down and reboot. WARNINGWARNINGDo not interrupt the software loading at any time.