GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 5 - Components and Functions (Theory) 5 - 35 5-3-15 Beam Former Board(s), BF-64 (cont’d) 5-3-15-7 Dip-switches and Jumpers None. 5-3-15-8 LEDs on BF64, Part Number FB200900 Figure 5-25 BF64, P/N: FB200900 - LEDs Table 5-27 LEDs on the BF-64 Boards, P/N: FB200900 LED #LED COLORDESCRIPTIONNORMAL STATE LD1 - Green “OK”. Board tested OK by System. Normally ON LD2 - Red “ERROR”. Board not initialized or an error has occurred Normally OFF LD3 - Yellow Not Used Normally OFF LD4 - YellowNot UsedNormally OFF LD5 - Red“POWER ERROR”. +5V and/or +2.5V is missingNormally OFF LD5 LD4 LD3 LD1 LD2 PLUG USED BY MANUFACTORING
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 5 - 36 Section 5-3 - a Front-End Processor (FEP) 5-3-15-9 LEDs on BF64, Part Number: FC200100 Figure 5-26 LEDs on BF64, P/N FC200100 Table 5-28 LEDs on BF-64 Boards, P/N FC200100 LED #LED COLORDESCRIPTIONNORMAL STATE LD2 - Red “ERROR”. Board not initialized or an error has occurred Normally OFF LD1 - Green “OK”. Board tested OK by System. Normally ON LD3 - Yellow Not Used Normally OFF LD4 - YellowNot UsedNormally OFF LD5 - Red“POWER ERROR”. +5V and/or +2.5V is missingNormally OFF LD6 - YellowMerger status indicatorBlinking LD7 - Yellow Merger status indicator Blinking LD5 LD4 LD3 LD1 LD2 PLUG USED BY MANUFACTORING LD7 LD6
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 5 - Components and Functions (Theory) 5 - 37 5-3-16 Front-End Contro ller Board, FEC NOTE: The Front-End Controller Board is only used in Front-End Rack version 1. In units with Front-End Rack vers ion 2, the RFI board replaces the functionality of this board, see: 5-3-21 Radio Frequency Interface Board, RFI on page 5-61 . 5-3-16-1 General Description The Front-End Controller Board controls the transmitter and receiver boards. Most of the control is done through the Frond End Bus. Main tasks done by the FEC board: • System Master Clock generator • Transmitter Control • ATGC Control • Receive Focusing Control • Analog Test Signal Generator • High Voltage Control and supervision • Probe Sensing • Control of Relay Board • Temperature Sensing for TEE probes • Step Motor Control for TEE probes • Temperature Sensing for circulated air (System Temperature Surveillance) 5-3-16-2 System Temperature Surveillance The FEC board monitors the temperature in the airflow at the air intake and in the airflow at the air outlet. • If the temperature inside the system (at the air outlet) increases to 60 oC, a “High System Temperature” error will be displayed on the screen. • If the temperature increase to 65 oC, the system will start an automatic shutdown. The temperature will normally be recorded at startup, and then repeated every hour. This frequency will increase if the temperature exceeds the limit. The temperature measurements make it possible to monitor the environmental conditions (temperatures) for the scanner and prevent overheating. The temperature measurements are available in a te mperature log available via the Service Platform. It is also available at D:\log\logfile-temp.txt.
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 5 - 38 Section 5-3 - a Front-End Processor (FEP) 5-3-16-3 Location in the Unit Figure 5-27 FEC-2 Board: Location in Unit (Front-End Rack version 1, only) FEC-2 BoardFront of Scanner
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 5 - Components and Functions (Theory) 5 - 39 5-3-16 Front-End C ontroller Board, FEC (cont’d) 5-3-16-4 Inputs 5-3-16-4-1 DC Voltages 5-3-16-4-2 Signal Inputs Table 5-29 DC Input Voltages INPUTDESCRIPTIONCONNECTED TO + 3.3 Vd DC output voltage, distributed via Motherboard (Backplane). The “d” indicates that this volt age is used for digital circuits.DC Power Supply via Backplane + 5 Vd DC output voltage, distributed via Motherboard (Backplane). The “d” indicates that this volt age is used for digital circuits.DC Power Supply via Backplane - 5 Va DC output voltage, distributed via Motherboard (Backplane). The “a” indicates that this voltage is used for analog circuits. DC Power Supply via Backplane + 15 Va DC output voltage, distributed via Motherboard (Backplane). The “a” indicates that this voltage is used for analog circuits. DC Power Supply via Backplane - 15 Va DC output voltage, distributed via Motherboard (Backplane). The “a” indicates that this voltage is used for analog circuits. DC Power Supply via Backplane GNDDistributed via Motherboard (Backplane)DC Power Supply via Backplane Table 5-30 Signal Inputs INPUTDESCRIPTIONCONNECTION FROM PRPRES Probe Present Relay Board High Voltage V & I Sense Indicates the voltages and currents from the TX Supply TX Power VME Bus Used for communication with the rest of the system (not the Front-End Boards). Commands and data from the VME Bus can be routed via the Front-End bus to the other boards in the Front-End. The Back-End Processor
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 5 - 40 Section 5-3 - a Front-End Processor (FEP) 5-3-16 Front-End C ontroller Board, FEC (cont’d) 5-3-16-5 Output 5-3-16-5-1 Output Signals 5-3-16-6 LEDs Description, Main Digital Signal Processor (DSP) Table 5-31 Output Signals OUTPUTDESCRIPTIONCONNECTION TO TXTRIG_L Starts the transmitter sequence TX128 boards SYNC_L Signal used to synchronize the Beam Formers BF64 boards TSIG Test Signal RX-128 Board AT G C V P / AT G C V NDifferential signals used to control TGC on the RX BoardRX-128 Board TX Voltage ControlOne serial data line and several select lines to the TX SupplyTX Supply Table 5-32 FEC-2, Lower Batch of LEDs, Main DSP LED #LED COLORDESCRIPTIONNORMAL STATE LD1 - Red Board Failure Normally OFF LD2 - Green Board OK Normally ON LD3 - Yellow System in scanning mode ON when scanning OFF in Freeze LD4 - YellowProbe ChangeON during probe change LD5 - YellowNo-ProbeOFF if no-probe has been selected LD6 - YellowTX Overrange OFF. ON if excessive TX Voltage is output Blinking, surveillance period LD7 - Yellow VME Access ON when communicating with BEP. LD8 - Yellow DSP Running BLINKING (1 Hz) “heartbeat”
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 5 - Components and Functions (Theory) 5 - 41 5-3-16 Front-End C ontroller Board, FEC (cont’d) 5-3-16-7 LEDs Description, Slave DSP 5-3-16-8 LEDsTable 5-33 FEC-2, Upper Batch of LEDs, Slave DSP LED #LED COLORDESCRIPTIONNORMAL STATE LD27 - Yellow Rack temperature surveillance Blinking LD28 - Yellow TX Power surveillance Blinking LD29 - Green Up and running ON Blinking LD30 - RefFailure or reprogrammingOFF Figure 5-28 FEC-2, LEDs LD30 LD29 LD28 LD27 LD1 LD2 LD3 LD4 LD5 LD6 LD7 LD8
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 5 - 42 Section 5-3 - a Front-End Processor (FEP) 5-3-16 Front-End C ontroller Board, FEC (cont’d) 5-3-16-9 Test points for DC Voltages Figure 5-29 FEC-2, Test points for DC Voltages TP8TP10 TP9 TP7 TP20 TP17TP21 +5V +15V -15V -5Va +3.3V +6Va +10Va
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 5 - Components and Functions (Theory) 5 - 43 5-3-16-10 Test Points for DC Voltages See Figure 5-29 FEC-2, Test points for DC Voltages on page 5-42 . 5-3-16-11 Fuses and Dip-switches None Table 5-34 Test points for DC Voltages TEST POINTDC VOLTAGEACCEPTABLE RANGE TP 8 +5 V +4.75 V to +5.25 V TP 10 +15 V +/- 10% TP 9 -15 V +/- 10% TP 7-5 Va+/- 10% TP 20+3.3 Va+/- 10% TP 17+6 Va+/- 10% TP 21 +10 Va +/- 10%
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 5 - 44 Section 5-3 - a Front-End Processor (FEP) 5-3-16-12 Jumpers Figure 5-30 FEC-2, Jumpers and Connectors SW7 SW4 SW1 SW2 SW10 12 910 12 1112 SW3 12 3940 SW5 12 1112 Connector used in factory test Connector used in factory test Connector used in factory test Connector used in factory test