Home > GE > Ultrasound System > GE Vivid 7 Service Manual

GE Vivid 7 Service Manual

    Download as PDF Print this page Share this page

    Have a look at the manual GE Vivid 7 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    Page
    of 858
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 5 - Components and Functions (Theory) 5 - 35
    5-3-15 Beam Former Board(s), BF-64 (cont’d)
    5-3-15-7 Dip-switches and Jumpers None.
    5-3-15-8 LEDs on BF64, Part Number FB200900 
      
     
    Figure 5-25   BF64, P/N: FB200900 - LEDs
    Table 5-27    LEDs on the BF-64 Boards, P/N: FB200900 
    LED #LED COLORDESCRIPTIONNORMAL STATE
    LD1  - Green “OK”. Board tested OK 
    by System. Normally ON
    LD2  - Red “ERROR”. Board not initialized or an 
    error has occurred Normally OFF
    LD3  - Yellow Not Used Normally OFF
    LD4 - YellowNot UsedNormally OFF
    LD5 - Red“POWER ERROR”. 
    +5V and/or +2.5V is missingNormally OFF
    LD5 LD4 LD3 LD1 LD2
    PLUG USED BY MANUFACTORING 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    5 - 36 Section 5-3 - a  Front-End Processor (FEP)
    5-3-15-9 LEDs on BF64, Part Number: FC200100
      
     
    Figure 5-26   LEDs on BF64, P/N FC200100 
    Table 5-28    LEDs on BF-64 Boards, P/N FC200100 
    LED #LED COLORDESCRIPTIONNORMAL STATE
    LD2  - Red “ERROR”. Board not initialized or an 
    error has occurred Normally OFF
    LD1  - Green “OK”. Board tested OK by System. Normally ON
    LD3  - Yellow Not Used Normally OFF
    LD4 - YellowNot UsedNormally OFF
    LD5 - Red“POWER ERROR”. 
    +5V and/or +2.5V is missingNormally OFF
    LD6 - YellowMerger status indicatorBlinking
    LD7  - Yellow Merger status indicator Blinking
    LD5 LD4 LD3 LD1 LD2
    PLUG USED BY MANUFACTORING
    LD7 LD6 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 5 - Components and Functions (Theory) 5 - 37
    5-3-16 Front-End Contro ller Board, FEC
    NOTE: The Front-End Controller Board is only used in Front-End Rack version 1. 
    In units with Front-End Rack vers ion 2, the RFI board replaces the functionality of this board, 
    see:  5-3-21 Radio Frequency Interface Board, RFI on page 5-61 .
    5-3-16-1 General Description The Front-End Controller Board controls  the transmitter and receiver boards.
     Most of the control is done through the Frond End Bus.
    Main tasks done by the FEC board:
    • System Master Clock generator
    • Transmitter Control
    • ATGC Control
    • Receive Focusing Control
    • Analog Test Signal Generator
    • High Voltage Control and supervision
    • Probe Sensing
    • Control of Relay Board
    • Temperature Sensing for TEE probes
    • Step Motor Control for TEE probes
    • Temperature Sensing for circulated  air (System Temperature Surveillance)
    5-3-16-2 System Temperature Surveillance The FEC board monitors the temperature in the airflow at  the air intake and in the airflow at the air outlet.
    • If the temperature inside the system (at the air outlet) increases to 60 
    oC, a “High System 
    Temperature” error will be displayed on the screen.
    • If the temperature increase to 65 
    oC, the system will start  an automatic shutdown.
    The temperature will normally  be recorded at startup, and then repeated every hour. This frequency will 
    increase if the temperature exceeds the limit.
    The temperature measurements make it possible to monitor the environmental conditions 
    (temperatures) for the scanner and prevent overheating.
    The temperature measurements are available in a te mperature log available via the Service Platform. 
    It is also available at D:\log\logfile-temp.txt. 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    5 - 38 Section 5-3 - a  Front-End Processor (FEP)
    5-3-16-3 Location in the Unit Figure 5-27   FEC-2 Board: Location in Unit (Front-End Rack version 1, only)
    FEC-2 BoardFront of 
    Scanner 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 5 - Components and Functions (Theory) 5 - 39
    5-3-16 Front-End C ontroller Board, FEC (cont’d)
    5-3-16-4 Inputs
    5-3-16-4-1 DC Voltages
     
    5-3-16-4-2 Signal Inputs
     
    Table 5-29    DC Input Voltages 
    INPUTDESCRIPTIONCONNECTED TO
    + 3.3 Vd DC output voltage, distributed via Motherboard 
    (Backplane). 
    The “d” indicates that this volt age is used for digital circuits.DC Power Supply via Backplane
    + 5 Vd DC output voltage, distributed via Motherboard 
    (Backplane). 
    The “d” indicates that this volt age is used for digital circuits.DC Power Supply via Backplane
    - 5 Va DC output voltage, distributed via Motherboard 
    (Backplane). 
    The “a” indicates that this voltage is used for analog circuits. DC Power Supply via Backplane
    + 15 Va
    DC output voltage, distributed via Motherboard 
    (Backplane). 
    The “a” indicates that this voltage is used for analog circuits.
    DC Power Supply via Backplane
    - 15 Va
    DC output voltage, distributed via Motherboard  (Backplane). 
    The “a” indicates that this voltage is used for analog circuits.
    DC Power Supply via Backplane
    GNDDistributed via Motherboard (Backplane)DC Power Supply via Backplane
    Table 5-30    Signal Inputs 
    INPUTDESCRIPTIONCONNECTION FROM
    PRPRES Probe Present  Relay Board
    High Voltage V & I  Sense Indicates the voltages and currents from the TX Supply TX Power
    VME Bus Used for communication with the rest of the system (not the 
    Front-End Boards).
    Commands and data from the VME Bus can be routed via  the Front-End bus to the other boards in the Front-End. The Back-End Processor  
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    5 - 40 Section 5-3 - a  Front-End Processor (FEP)
    5-3-16 Front-End C ontroller Board, FEC (cont’d)
    5-3-16-5 Output
    5-3-16-5-1 Output Signals
     
    5-3-16-6 LEDs Description, Main Digital Signal Processor (DSP)
     
    Table 5-31    Output Signals 
    OUTPUTDESCRIPTIONCONNECTION TO
    TXTRIG_L Starts the transmitter sequence TX128 boards SYNC_L Signal used to synchronize the Beam Formers BF64 boards TSIG Test Signal RX-128 Board
    AT G C V P / AT G C V NDifferential signals used to control TGC on the RX BoardRX-128 Board
    TX Voltage ControlOne serial data line and several select lines to the TX  SupplyTX Supply
    Table 5-32     FEC-2, Lower Batch of LEDs, Main DSP
    LED #LED COLORDESCRIPTIONNORMAL STATE
    LD1 - Red Board Failure Normally OFF
    LD2  - Green Board OK Normally ON
    LD3  - Yellow System in scanning mode ON when scanning
    OFF in Freeze
    LD4 - YellowProbe ChangeON during probe change
    LD5 - YellowNo-ProbeOFF if no-probe has been selected
    LD6 - YellowTX Overrange
    OFF.
    ON if excessive TX Voltage is output Blinking, surveillance period 
    LD7  - Yellow VME Access ON when communicating with BEP.
    LD8  - Yellow DSP Running BLINKING (1 Hz)
    “heartbeat” 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 5 - Components and Functions (Theory) 5 - 41
    5-3-16 Front-End C ontroller Board, FEC (cont’d)
    5-3-16-7 LEDs Description, Slave DSP
      
    5-3-16-8 LEDsTable 5-33     FEC-2, Upper Batch of LEDs, Slave DSP
    LED #LED COLORDESCRIPTIONNORMAL STATE
    LD27
     - Yellow Rack temperature surveillance Blinking
    LD28  - Yellow TX Power surveillance Blinking
    LD29  - Green Up and running ON
    Blinking
    LD30 - RefFailure or reprogrammingOFF
    Figure 5-28   FEC-2, LEDs
    LD30
    LD29
    LD28
    LD27
    LD1
    LD2
    LD3
    LD4
    LD5
    LD6
    LD7
    LD8 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    5 - 42 Section 5-3 - a  Front-End Processor (FEP)
    5-3-16 Front-End C ontroller Board, FEC (cont’d)
    5-3-16-9 Test points for DC Voltages
    Figure 5-29   FEC-2, Test points for DC Voltages
    TP8TP10
    TP9
    TP7
    TP20
    TP17TP21 +5V
    +15V
    -15V
    -5Va
    +3.3V
    +6Va
    +10Va 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 5 - Components and Functions (Theory) 5 - 43
    5-3-16-10 Test Points for DC Voltages  See Figure 5-29 FEC-2, Test points for DC Voltages on page 5-42 .
    5-3-16-11 Fuses and Dip-switches None Table 5-34    Test points for DC Voltages 
    TEST POINTDC VOLTAGEACCEPTABLE RANGE
    TP 8 +5 V +4.75 V to +5.25 V
    TP 10 +15 V +/- 10%
    TP 9 -15 V +/- 10%
    TP 7-5 Va+/- 10%
    TP 20+3.3 Va+/- 10%
    TP 17+6 Va+/- 10%
    TP 21 +10 Va +/- 10% 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    5 - 44 Section 5-3 - a  Front-End Processor (FEP)
    5-3-16-12 Jumpers Figure 5-30   FEC-2, Jumpers and Connectors
    SW7
    SW4
    SW1
    SW2
    SW10
    12
    910
    12
    1112
    SW3
    12
    3940
    SW5
    12
    1112
    Connector used in 
    factory test
    Connector used in 
    factory test
    Connector used in 
    factory test
    Connector used in 
    factory test 
    						
    All GE manuals Comments (0)